K251124 · GC America, Inc. · KLE · Oct 2, 2025 · Dental
Device Facts
Record ID
K251124
Device Name
G-Bond Universal
Applicant
GC America, Inc.
Product Code
KLE · Dental
Decision Date
Oct 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1. Direct bonding of restorative materials 2. Bonding of dual-cured core build-up composites 3. Adhesive cementation of indirect restorations 4. Priming of adherent surface of indirect restorations 5. Intraoral repair of fractured restorations 6. Treatment of hypersensitive teeth 7. Sealing of tooth preparation (cavity or abutment) for indirect restorations 8. Treatment of exposed root surfaces 9. Micro-invasive treatment and infiltration of non-cavitated non-carious enamel lesions 10. Post cementation
Device Story
G-BOND Universal is a one-component, light-cured resin bonding agent; available in 5 mL liquid bottle or 0.1 mL unit dose. Used by dental professionals in clinical settings. Functions via demineralization of tooth substance and simultaneous penetration of monomer components; photopolymerization creates chemical and mechanical bonds to tooth tissue, metal, zirconia, alumina, porcelain, lithium disilicate, and composite resins. For hypersensitivity, it penetrates and seals dentinal tubules. For non-cavitated enamel lesions, it infiltrates white spot lesions and hardens via photopolymerization. Output is a cured adhesive layer; assists clinicians in restorative procedures, repair, and desensitization; benefits patients by restoring tooth structure, sealing surfaces, and treating early-stage enamel lesions.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated against Japanese certification criteria for dental adhesives, sealants, and desensitizers. Tests included appearance, curing property, hardness, bond strength, application characteristics, dentin tubule sealing, infiltration performance, and viscosity. Biocompatibility assessed per ISO 10993-1:2018 and ISO 7405:2018.
Technological Characteristics
One-component, light-cured resin bonding agent. Contains phosphate ester monomers and silane coupling agents. Form factor: 5 mL liquid bottle or 0.1 mL unit dose. Principle: chemical/mechanical bonding via photopolymerization. Biocompatibility compliant with ISO 10993-1:2018 and ISO 7405:2018.
Indications for Use
Indicated for dental patients requiring restorative bonding, core build-up, indirect restoration cementation, repair of fractured restorations, treatment of dentin hypersensitivity, sealing of tooth preparations, treatment of exposed root surfaces, and micro-invasive infiltration of non-cavitated non-carious enamel lesions.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
Scotchbond Universal Plus Adhesive (ADH19) (K192961)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 2, 2025
GC America, Inc.
Futoshi Fusejima
Director of PE & Regulatory Affairs
3737 W. 127th Street
Alsip, Illinois 60803
Re: K251124
Trade/Device Name: G-Bond Universal
Regulation Number: 21 CFR 872.3200
Regulation Name: Resin Tooth Bonding Agent
Regulatory Class: Class II
Product Code: KLE, EBC
Dated: October 1, 2025
Received: October 1, 2025
Dear Futoshi Fusejima:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251124 - Futoshi Fusejima
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251124 - Futoshi Fusejima
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
G-Bond Universal
Page 7 of 27
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251124 | ? |
| Please provide the device trade name(s). | | ? |
| G-Bond Universal | | |
| Please provide your Indications for Use below. | | ? |
| Indications for Use:
1. Direct bonding of restorative materials
2. Bonding of dual-cured core build up composites
3. Adhesive cementation of indirect restorations
4. Priming of adherent surface of indirect restorations
5. Intraoral repair of fractured restorations
6. Treatment of hypersensitive teeth
7. Sealing of tooth preparation (cavity or abutment) for indirect restorations
8. Treatment of exposed root surfaces
9. Micro-invasive treatment and infiltration of non-cavitated non-carious enamel lesions
10. Post cementation | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
'GC'
K251124
# Section 5 – 510(k) Summary
## 1. Submitter Information:
GC America Inc.
3737 W. 127th Street
Alsip, IL 60803
Contact Person: Futoshi Fusejima
Phone: (708) 926-3050
Alternate Contact: Tamiko Scott
Phone: (708) 926-3261
Fax: (708) 925-0373
Date Prepared: August 29, 2025
## 2. Device Name:
Proprietary Name: G-BOND Universal
Common Name: Resin tooth bonding agent
Classification Name: Agent, Tooth Bonding, Resin, Pit and fissure sealant and conditioner
Device Classification: Class II, 872.3200
Product Code: KLE/EBC
## 3. Predicate Devices:
| Product | Applicant | 510(k) No. | Code No. | Predicate | Decision Date |
| --- | --- | --- | --- | --- | --- |
| G-Premio BOND | GC America Inc. | K143140 | KLE | Primary | 04/20/2015 |
| Icon (INFILTRATION KIT) | DMG USA, Inc | K100062 | EBC | Secondary | 03/26/2010 |
| Scotchbond Universal Plus Adhesive (ADH19) | 3M Deutschland GmbH ESPE Platz | K192961 | KLE | Reference | 10/31/2019 |
## 4. Description of Device:
G-BOND Universal is a one component, light-cured bonding agent, which can also be used for micro-invasive treatment and infiltration of non-cavitated non-carious enamel lesions, available in a 5 mL liquid bottle or 0.1 mL unit dose.
## 5. Indications for Use:
1. Direct bonding of restorative materials
2. Bonding of dual-cured core build-up composites
3. Adhesive cementation of indirect restorations
4. Priming of adherent surface of indirect restorations
5. Intraoral repair of fractured restorations
6. Treatment of hypersensitive teeth
7. Sealing of tooth preparation (cavity or abutment) for indirect restorations
8. Treatment of exposed root surfaces
9. Micro-invasive treatment and infiltration of non-cavitated non-carious enamel lesions
10. Post cementation
GC AMERICA INC.
3737 West 127th Street • Alsip, IL 60803 • TEL: 800.323.3386 • FAX: 708.897.4062
www.gcamerica.com
Section 5
p. 1
{5}
The indications as dental adhesive system is covered with primary predicate and reference device. The indication as a treatment agent of hypersensitive teeth is covered with primary predicate and reference device.
The indication as a micro-invasive treatment and infiltration agent of non-cavitated non-carious enamel lesions is covered with secondary predicate device.
In conclusion, there are no substantial differences between the applicant device and predicate or reference devices in indications for use, and there are no differences in the safety and effectiveness of the device when used as labeled.
## 6. Comparison of Technology:
The technological principles of the applicant device as a dental adhesive system are similar to those of predicate devices G-Premio BOND and Scotchbond Universal Plus Adhesive, as follows:
1. The product demineralizes the tooth substance and simultaneously penetrates the monomer component, which then photopolymerizes through photopolymerization and bonds to the tooth tissue.
2. The phosphate ester monomer bonds with the tooth structure metal, zirconia, and alumina surfaces, thereby bonding to metal, zirconia, and alumina.
3. The silane coupling agent bonds with the surfaces of porcelain, lithium disilicate glass, and composite resins, thereby bonding to porcelain, lithium disilicate glass, and composite resins.
4. The product demineralizes the tooth substance and monomer component simultaneously penetrates the tooth substance. The product is then bonded and hardened using dental adhesive resin cement, thereby bonding to the tooth tissue.
5. The product penetrates the dentinal tubules of the formed cavity and abutment teeth, etc., and photopolymerizes, then it seals them.
The technological principle of the applicant device as a treatment agent of hypersensitive teeth is similar to the predicate device G-Premio BOND, as follows:
1. The product penetrates the dentinal tubules exposed by hypersensitivity, and photopolymerizes, then it suppresses hypersensitivity by sealing the dentinal tubules.
The technological principle of the applicant device as a micro-invasive treatment and infiltration agent of non-cavitated non-carious enamel lesions is similar to that of the predicate device Icon, as follows:
1. The infiltration mechanism is applying it to white spot lesions (WSL) of enamel substance, then penetrating the tooth structure and adhering chemically and mechanically by photopolymerization of uncured resin monomers.
In conclusion, the applicant device is substantially equivalent to the predicate devices in technological principle.
## 7. Performance Bench Tests:
Since there is no international standard, such as ISO, the physical properties of this product were established based on Japanese certification criteria of "Dental dentin adhesive", "Dental metal adhesive material", "Dental ceramics adhesive", "Dental hypersensitive dentine desensitizer", "Dental sealing/coating material" and "Tooth surface coating material." Additionally, the degree of material penetration when used for infiltrant applications and the viscosity were also tested to confirm the substantial equivalence of the applicant device with predicate devices in terms of safety and effectiveness.
Performance testing includes:
- Appearance
- Curing property
- Hardness
Section 5
p. 2
{6}
- Bond strength
- Application characteristics
- Sealing property of dentin tubules
- Infiltration performance
- Viscosity
Section 5
p. 2
{7}
Section 5
p. 3
# 8. Biocompatibility:
A biocompatibility assessment was completed according to ISO 10993-1:2018 and ISO 7405:2018.
G-BOND Universal is a universal adhesive, so medical device categorization by ISO 10993 for biological evaluation of medical devices is as follows.
Category : Externally communicating medical device
Contact : Tissue / bone / dentin
Contact duration : Long term (> 30 d)
Therefore, biological evaluation tests are planned taking account of the requirements of ISO 10993-1: 2018 ANNEX A and the tests in Table 5.2 are selected and discussed as the results of consideration for the device and its contact to body parts or ingredients.
Table 5.2. Biocompatibility evaluation on ISO 10993-1: 2018 ANNEX A
| | Evaluation tests for consideration |
| --- | --- |
| 1 | Cytotoxicity |
| 2 | Sensitization |
| 3 | Irritation or intracutaneous reactivity |
| 4 | Material mediated pyrogenicity |
| 5 | Acute systemic toxicity |
| 6 | Subacute toxicity |
| 7 | Subchronic toxicity |
| 8 | Chronic toxicity |
| 9 | Implantation effects |
| 10 | Genotoxicity |
| 11 | Carcinogenicity |
Furthermore, in addition to the above test, according to ISO 7405:2018 ANNEX A, the test to be considered for evaluation of biocompatibility shown in Table 5.3 is listed.
Table 5.3. Biocompatibility evaluation on ISO 7405:2018 ANNEX A
| | Evaluation tests for consideration |
| --- | --- |
| 12 | Pulp and dentine usage test |
In conclusion, the biocompatibility of G-BOND Universal is acceptable from the biological evaluation result.
# 9. Conclusion:
Based on similarities in indications for use, technology, safety and effectiveness, the applicant device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.