K251109 · Ray Co., Ltd. · LLZ · May 21, 2025 · Radiology
Device Facts
Record ID
K251109
Device Name
SMARTDent
Applicant
Ray Co., Ltd.
Product Code
LLZ · Radiology
Decision Date
May 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric, 3rd-Party Reviewed
Indications for Use
SMARTDent is intended for use as a software solution for managing dental diagnostic images and providing tools to support diagnosis.
Device Story
SMARTDent is a dental diagnostic image management and processing software. It ingests images from intraoral systems, extraoral systems, intraoral digital video capture devices, and intraoral scanners. The software provides tools for viewing (2D/3D, FMS, MPR, cross-section, panoramic, implant, TMJ), image manipulation (brightness, contrast, gamma, sharpen, inverse), and measurement (line, multi-line, angle, Cobb's angle). It also supports nerve and implant simulation. Used in dental clinics by healthcare professionals (generalists, specialists, dental operator aides) to support clinical diagnosis and treatment planning. The software runs on Windows 11 workstations. By providing enhanced visualization and measurement tools, it assists clinicians in evaluating dental anatomy and pathology, potentially improving diagnostic accuracy and treatment outcomes.
Clinical Evidence
No clinical data. Bench testing only. Software, hardware, and integration verification and validation were performed per FDA guidance, including validation of risk management mitigations.
Technological Characteristics
Software-based medical image management and processing system. Operates on Windows 11. Supports DICOM, JPEG, PNG, STL, and PLY formats. Features include 2D/3D image viewing, MPR, cross-section, panoramic, implant, and TMJ views. Measurement tools include line, multi-line, angle, and Cobb's angle. Connectivity via standard PC hardware interfaces. No specific algorithm class (e.g., ML/DL) is described; functionality is based on standard image processing and measurement tools.
Indications for Use
Indicated for use as a software solution for managing dental diagnostic images and supporting diagnosis in patients over 5 years old who are candidates for X-ray diagnostic radiation exposure. No restrictions regarding ethnic group, sex, weight, or health condition.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
RAY Co., Ltd.
% Dave Yungvirt
CEO
Third Party Review Group, LLC
25 Independence Blvd
Warren, New Jersey 07059
May 21, 2025
Re: K251109
Trade/Device Name: SMARTDent
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: LLZ
Dated: May 20, 2025
Received: May 20, 2025
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251109 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251109 - Dave Yungvirt
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251109 | |
| Device Name SMARTDent | |
| Indications for Use (Describe) SMARTDent is intended for use as a software solution for managing dental diagnostic images and providing tools to support diagnosis. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Submission- SMARTDent
# K 251109 510(k) Summary
## 1. 510(k) Summary
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
## 2. Date:
20 May 2025
## 3. Administrative Information
| Applicant | | Ray Co., Ltd. |
| --- | --- | --- |
| Address | | 1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si, 16882, Korea |
| Manufacturer | Name | Ray Co., Ltd. |
| | Address | 1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si, 16882, Korea |
| | Tel | +82-31-605-1000 |
| | Fax | +82-2-6280-5534 |
| Contact Person | Name | Sooji Huh |
| | Email | sooji.huh@raymedical.co.kr |
## 4. Device Information
| Trade/Proprietary Name | | SMARTDent |
| --- | --- | --- |
| Common Name | | Medical image management and processing system |
| Classification Name | Device | System, Image Processing, Radiological |
| | Regulation Number | 21 CFR 892.2050 |
| | Class | 2 |
| | Product Code | LLZ |
| | Review Panel | Radiology |
Ray Co.,Ltd
{5}
510(k) Submission- SMARTDent
# 5. Predicate device
| Parameter | Predicate Device |
| --- | --- |
| Device Name | CS Imaging |
| Manufacturer | Carestream dental technology topco limited. |
| 510(K) Number | K173622 Traditional 510k |
| Classification name | System, Image Processing, Radiological |
| Regulation number | 21 CFR 892.2050 |
| Primary product code | LLZ |
# 6. Device Description
SMARTDent provides the function to efficiently manage CT, panorama, cephalometric, intraoral sensor images and intraoral camera images acquired using X-ray imaging equipment of Ray Co., Ltd. and performs various image analysis according to the diagnostic purposes.
# 7. Indication for use
SMARTDent is intended for use as a software solution for managing dental diagnostic images and providing tools to support diagnosis.
# 8. Intended Patient Population
- The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure.
- There is no restriction for ethnic group, Sex, weight, health, or condition.
- We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.
# 9. Comparison with predicate device
The following table provides the summary of the technological characteristics of SMARTDent compared to the predicate device
| Parameter | Proposed Device | Predicate Device |
| --- | --- | --- |
| Manufacturer | RAY Co., Ltd. | Carestream dental technology topco limited. |
| Device name | SMARTDent | CS Imaging |
| 510(K) Number | K251109 (Traditional 510K) | K173622 (Traditional 510K) |
Ray Co.,Ltd
{6}
510(k) Submission- SMARTDent
| Common Name | Medical image management and processing system | Medical image management and processing system |
| --- | --- | --- |
| Indications for use | SMARTDent is intended for use as a software solution for managing dental diagnostic images and providing tools to support diagnosis | CS Imaging is digital imaging software intended to be used with Carestream Dental's digital imaging devices by healthcare professionals to display, adjust, make measurement, print, export and store digital or digitized images to support image diagnosis in medical care, predominantly in dentistry |
| Image Input Sources | Intraoral systemExtraoral systemIntraoral digital video capture devicesIntraoral scanners | Intraoral systemExtraoral systemIntraoral digital video capture devicesIntraoral scanners |
| Image format | DICOM, JPEG, PNG, STL, PLY | DICOM, BMP, JPEG, TIFF, PNG, STL, PLY |
| Image Measurement Tools | line measurementmulti-line measurementangle measurementCobb's angle measurement | line measurementmulti-line measurementangle measurementorthogonal measurement |
| Image viewing | single viewmultiple pop-up viewlayout view for 2D imageslayout view for CT imagesfull mouth series(FMS) viewpre-defined FMS | single viewmultiple pop-up viewlayout view for CT imagesfull mouth series(FMS) viewpre-defined FMS |
| Image manipulation | BrightnessContrastGammaSharpenInverse | BrightnessContrastGammaSharpenInverse |
| 3D imagining capability | MPR viewCross-section viewPanoramic view,implant viewTMJ viewNerveImplant simulation | MPR viewCross-section viewPanoramic view,implant viewTMJ viewNerveImplant simulationAirway analysis |
| Image annotation | line, multi-line, free line, circle, ellipse, arrow, rectangle, text | line, multi-line, free line, spline, circle, ellipse, rectangle, landmark, arrow, text |
Ray Co.,Ltd
{7}
510(k) Submission- SMARTDent
| User | The intended users of SMARTDent are health care professionals such as Oral health generalists, specialists and dental operator aides. | Used by health care professionals such as Oral health generalists, specialists and dental operator aides. |
| --- | --- | --- |
| System Requirements | (2D recommend)CPU : Intel i5 2.0GHz (7th Gen.) or higherRAM : 8GB or higherResolution : 1920 x 1080Video Card : 512MB RAMOS : Windows 11(3D View recommend)CPU : Intel Core i7-12700RAM : 32GB of higherResolution : 1920 x 1080Video Card : NVidia GeForce GTX 4060,8GOS : Windows 11 | CPU: Intel Core i7-2600RAM: 8 GB system RAMResolution: 1280 x 1024, 32-bit colorGraphic card: 1GB dedicated memory with open GL v3.2 support(AMD/ATI and Intel graphics chipset are NOT SUPPORTED)OS: Windows 10 64bit, Windows 11 64bit,Windows Server 2016/2019/2022300 GB free disk spaceDVD RW driveEthernet 100/1000 NIC |
The product is principally just the same as in the previous 510(k) #K173622.
The complete of differences of the subject device to the predicate #K173622 device is as follows
- Image format
- System requirement
# 10. Performance Testing
Software, hardware, and integration verification and validation testing was performed in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Device Software Functions".
The validation suite includes validation of implemented mitigations related to device hazards identified in the risk management procedures.
All test results have been reviewed and approved, showing the SMARTDent to be substantially equivalent to the predicate devices.
# 11. Clinical Testing
Clinical testing is not a requirement and has not been performed.
# 12. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Ray Co., Ltd. concludes that the newly SMARTDent is safe, effective and substantially equivalent to the predicate device as described herein.
Ray Co.,Ltd
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.