K251105 · Emed Technologies Corporation · PKP · Dec 22, 2025 · General Hospital
Device Facts
Record ID
K251105
Device Name
Tetra (TM-20); Tetra (TM-50)
Applicant
Emed Technologies Corporation
Product Code
PKP · General Hospital
Decision Date
Dec 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The EMED Tetra Micro Infusion System is intended for subcutaneous infusion for prescription liquid medicines for use in the home and hospital environment by adults and children (2 years and older), when administered according to the approved biologic or drug product labeling. The Tetra Micro Infusion System is indicated for use with: - Hizentra Immune Globulin Subcutaneous (Human) 20%, manufactured by CSL Behring.
Device Story
Tetra Micro Infuser is a battery-powered, reusable, prescription-only infusion pump; delivers liquid medication subcutaneously via compatible syringes and EMED patient administration sets. Device applies controlled pneumatic pressure to syringe stopper to drive fluid; flow rate is a resultant output determined by pressure setting, syringe size, medication viscosity, and downstream resistance. User selects one of five discrete pressure levels via membrane keypad; device provides visual/audible notifications for status, occlusion, and end-of-infusion. Used in home and hospital environments by patients or healthcare providers. Infusion disabled during USB-C charging. No wireless/network connectivity. Benefits include controlled, portable subcutaneous delivery of immunoglobulin therapies.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including: infuser operating pressure, leakage, drug-device compatibility, flow rate performance verification, cleaning/disinfection validation, useful life, drop testing, electrical safety/EMC, home use environment verification, software V&V, shipping validation, human factors/usability summative testing, and biological evaluation.
Technological Characteristics
Electro-mechanical pneumatic pump; 3.7 VDC 1000mAh rechargeable Li-ion battery. Dimensions: 64x64x41 mm; weight: 125 g. Pressure range: 2.50-7.00 psig. Connectivity: USB-C (charging/service only). Sterilization: Infuser non-sterile/reusable; administration sets sterile/single-use. Software: Embedded firmware (microcontroller-based) for pressure control and notification management. No calibration required.
Indications for Use
Indicated for subcutaneous infusion of Hizentra (Immune Globulin Subcutaneous (Human) 20%) in adults and children (≥ 2 years) in home and hospital settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
EMED Technologies Corporation
Olena Whalen
QA/RA/CA Director
1262 Hawks Flight Court
Suite 200
El Dorado Hills, California 95762
Re: K251105
Trade/Device Name: Tetra (TM-20); Tetra (TM-50)
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: PKP, FRN
Dated: November 28, 2025
Received: November 28, 2025
Dear Olena Whalen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251105 - Olena Whalen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251105 - Olena Whalen
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jake K. Lindstrom -S
Jake Lindstrom, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251105 | |
| Device Name Tetra (TM-20); Tetra (TM-50) | |
| Indications for Use (Describe) The EMED Tetra Micro Infusion System is intended for subcutaneous infusion for prescription liquid medicines for use in the home and hospital environment by adults and children (2 years and older), when administered according to the approved biologic or drug product labeling. The Tetra Micro Infusion System is indicated for use with: - Hizentra Immune Globulin Subcutaneous (Human) 20%, manufactured by CSL Behring. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
EMED Technologies
510(k) Summary
K-number: 251105
Date Prepared: December 18, 2025
I. SUBMITTER (21 CFR 807.92(a)(1))
EMED Technologies Corporation
1262 Hawks Flight Ct., Suite 200
El Dorado Hills, CA 95762
Contact Person: Olena Whalen
Contact Person Title: QA/RA/CA Director
Tel: 916.932.0071 ext. 114
Email: owhalen@emedtc.com
II. DEVICE NAME (21 CFR 807.92(a)(2))
Device Trade Name: Tetra (TM-20); Tetra (TM-50)
Common Name: Infusion Pump
Classification Name: Immunoglobulin G (IgG) Infusion System
Regulation Number: 880.5725
Product Code: PKP, FRN
III. PREDICATE DEVICES (21 CFR 807.92(a)(3))
Primary: K240148, SCIg60 Infusion System (Product Code – PKP)
Secondary: K091042, Crono S-PID 50 (Product Code – FRN)
These predicates have not been subject to a design-related recall.
IV. DEVICE DESCRIPTION SUMMARY (21 CFR 807.92(a)(4))
The Tetra Micro Infuser is a reusable multi-patient use device, battery powered, prescription-only, designed to be used as a system for subcutaneous infusion using compatible syringes through EMED patient administration sets.
The syringe connects to the Tetra Micro Infuser, and the patient administration sets connects to the syringe. Patient administration sets are sterile and single use only.
There are 2 model options for the Tetra Micro Infuser: TM-20 (designed to connect to a 20mL syringe); and TM-50 (designed to connect a 50mL syringe). The TM-50 model can be used with 5 mL, 10 mL, and 20 mL syringe sizes when used with the optional Syringe Adapter (single-use, non-sterile; sold separately). Once attached to the syringe, the Syringe Adapter is not intended to be removed.
The patient administration sets are commercially available, previously marketed devices (K140131) and support the subject device without modification.
Page 1 of 8
{5}
EMED Technologies
NOTE: Only the Tetra Micro Infuser, as the new system component, is within the scope of this 510(k). Patient administration sets and syringe models are not within the scope of this submission but are referenced when relevant. The Tetra Micro Infuser is to be used only with EMED patient administration sets.
The Tetra Micro Infuser is designed for use in home and healthcare facility environments.
The Tetra Micro Infuser functions by applying controlled pressure to the syringe stopper (gasket) to deliver medication through the connected patient administration sets.
The user interface consists of two buttons used to power the device on/off and to select discrete delivery settings. The resulting flow rate depends on the selected setting and system factors such as syringe size, medication viscosity, and downstream resistance.
The device provides visual and audible notifications for status and certain conditions. The device battery is charged through a USB-C port, and infusion is disabled during charging. There is no user-accessible wireless or networking connectivity, and USB communication is restricted to factory/service access.
The Tetra Micro Infuser does not require calibration. Periodic cleaning of external surfaces is recommended between uses, and the infuser should be disinfected by healthcare or facility personnel between each patient when transferring to a new user.
Each Tetra Micro model is packaged in a labeled box containing: one infuser (either Tetra Micro TM-20 or TM-50); one USB-C to USB-A charging cable; one Instructions for Use document; and one Quick Reference Guide document.
V. INTENDED USE/INDICATIONS FOR USE (21 CFR 807.92(a)(5))
Intended for subcutaneous infusion for prescription liquid medicines for use in the home and hospital environment by adults and children (2 years and older), when administered according to the approved biologic or drug product labeling.
The Tetra Micro Infusion System is indicated for subcutaneous infusion of:
- Hizentra, Immune Globulin Subcutaneous (Human) 20% (manufactured by CSL Behring)
| Attribute / Feature | Tetra Micro (Subject Device) | SCIg60 Infusion System (K240148) (Primary Predicate) | CRONO S-PID 50 (K091042) (Secondary Predicate) | Equivalency |
| --- | --- | --- | --- | --- |
| Intended Use | Subcutaneous infusion for prescription liquid medicines. | Subcutaneous infusion for prescription liquid medicines. | Subcutaneous infusion for prescription liquid medicines. | Equivalent |
| Indications for Use | • Hizentra, Immune Globulin Subcutaneous (Human) 20% (manufactured by CSL Behring) | • Hizentra Immune Globulin Subcutaneous (Human) 20%, manufactured by CSL Behring, | Not cleared for a specific drug. | Equivalent |
Page 2 of 8
{6}
EMED Technologies
| Attribute / Feature | Tetra Micro (Subject Device) | SCIg60 Infusion System (K240148) (Primary Predicate) | CRONO S-PID 50 (K091042) (Secondary Predicate) | Equivalency |
| --- | --- | --- | --- | --- |
| | | • Cuvitru Immune Globulin Infusion (Human) 20%, manufactured by Takeda,
• Gammagard Liquid, Immune Globulin Infusion (Human) 10%, manufactured by Takeda,
• Gamunex-C Immune Globulin Subcutaneous (Human) 10%, manufactured by Grifols Therapeutics,
• Gammaked Immune Globulin Subcutaneous (Human) 10%, manufactured by Grifols Therapeutics,
• Xembify Immune Globulin Subcutaneous (Human) 20%, manufactured by Grifols Therapeutics, and
• Cutaquig Immune Globulin (Human), 16.5%, manufactured by Octapharma AG. | | |
| Prescription or Over the Counter | Prescription | Prescription | Prescription | Equivalent |
| Intended Population | Adults and pediatrics ≥ 2 years | Adults and pediatrics ≥ 2 years | Not specified | Equivalent |
| Use Environment | Home and hospital | Home and hospital | Home and hospital | Equivalent |
| Route of Administration | Subcutaneous | Subcutaneous | Subcutaneous | Equivalent |
The EMED Tetra Micro Infusion System is intended for the subcutaneous administration of immunoglobulin therapies in home or healthcare settings. The device is indicated for use with Hizentra, a drug that is also cleared for subcutaneous administration with the identified predicate devices. The Tetra Micro Infusion System is used with EMED patient administration sets that are identical to those previously cleared for use with the primary predicate device. The use of the Tetra Micro Infusion System does not alter the intended use or performance of these administration sets. The device does not modify the intended use of the drug, the route of administration, or the patient population. Based on these factors, the indications for use of the Tetra Micro Infusion System are substantially equivalent to those of the predicate devices.
Page 3 of 8
{7}
EMED Technologies
VI. TECHNOLOGICAL COMPARISON (21 CFR 807.92(a)(6))
| Attribute / Feature | Tetra Micro (Subject Device) | SCIg60 Infusion System (K240148) (Primary Predicate) | CRONO S-PID 50 (K091042) (Secondary Predicate) | Equivalency |
| --- | --- | --- | --- | --- |
| Principle of Operation | Pneumatic pump applies force to syringe gasket | Compression spring applies force to syringe plunger and gasket | Stepper motor-driven piston applies force to syringe gasket | Equivalent |
| Infusion Speed / Rate Selection | 5 discrete pressure (speed) settings (Level 1-5) | Infuset or VersaRate flow control infusion sets | Screw-drive mechanism | Equivalent |
| Nominal Infuser Generated Pressure | 2.50 -7.00 psig (level dependent) | 1 bar (14.4 psig) average | Not published. Manufacturer only published flowrate specifications. Pressure is deduced as equivalent. | Equivalent |
| Maximum Operating Pressure | 7.35 psig | 1.16 bar (16.8 psig) | Not published. Manufacturer only published flowrate specifications. Pressure is deduced as equivalent. | Equivalent |
| Flow Rate Range | ≤ 50 mL/h/site (expected rate per IFU tables; configuration-dependent) | ≤ 50 mL/h/site (expected rate per IFU tables; configuration-dependent) | 1 - 50 mL/h | Equivalent |
| Flow Rate Accuracy | EMED 26G Infusion Set: • 50mL up to ±19% • 20mL up to ±18% • 10mL up to ±27% • 5mL up to ±27% | ±14% to ±41% (based on flow controller and infusion set) | ±3% | Equivalent |
| Flow Rate Vertical Sensitivity | ± 30 cm (± 12 in) between Infuser and infusion site. When infuser is placed 30 cm (12 in.) above infusion site, up to 7% increase in flow rate. When infuser is placed 30 cm (12 in.) below infusion site, up to 7% decrease in flow rate. | Each 30.5 cm (12 in.) above infusion site: +6% from target flow rate Each 30.5 cm (12 in.) below infusion site: -4% from target flow rate | Not published | Equivalent |
| Occlusion Detection | Visual Inspection AND Pressure threshold-based with visual and audible notification. | None – visual inspection only. | Pressure threshold-based with alarm | Equivalent |
| End of Infusion Detection | Visual Inspection AND Pressure threshold-based with visual and audible notification. | None – visual inspection only. | Time to delivery shown on LCD | Equivalent |
| Compatible Syringes | TM-50: • 50 mL Hizentra PFS (NDC# 44206-455-25) • 20 mL Hizentra PFS (NDC#: 44206-458-96) using 20mL Adapter (FP-0010214) • 10 mL Hizentra PFS (NDC#: 44206-457-95) | • 50 mL Hizentra PFS (NDC# 44206-455-25) • BD 50mL Syringe REF 309653 | Crono 50mL | Equivalent |
Page 4 of 8
{8}
EMED Technologies
| Attribute / Feature | Tetra Micro (Subject Device) | SCIg60 Infusion System (K240148) (Primary Predicate) | CRONO S-PID 50 (K091042) (Secondary Predicate) | Equivalency |
| --- | --- | --- | --- | --- |
| | using 10mL Adapter (FP-0010215)
• 5 mL Hizentra PFS (NDC#: 44206-456-94) using 5mL Adapter (FP-0010216)
TM-20:
• 20 mL Hizentra PFS (NDC#: 44206-458-96) | | | |
| Compatible Administration Sets | EMED OPTFlow Subcutaneous Administration Set | EMED OPTFlow Subcutaneous Administration Sets | Not Published | Equivalent |
| Architecture Type | Electro-mechanical | Mechanical | Electro-mechanical | Equivalent |
| Notifications | LED indicators and audible notifications | None – visual inspection only. | Acoustic + visual alarms | Equivalent |
| Software | Embedded firmware (microcontroller – controlled) | None | Embedded firmware (microcontroller – controlled) | Equivalent |
| User Interface | Membrane keypad with status LEDs and audible notifications | None | LCD, keypad, audible tones | Equivalent |
| Network Connectivity | None (USB-C for charging and restricted service/log access only) | None | None | Equivalent |
| Data Storage | Operational parameters only (e.g., usage counts and fault/notification history); no patient protected health information | None | Programmable values are automatically stored in the pump’s memory. | Equivalent |
| Sterility | Infuser: Non-sterile, reusable. Infusion set: sterile, single-use. | Infuser: Non-sterile, reusable. Infusion set: sterile, single-use. | Infuser: Non-sterile, reusable. Infusion set: sterile, single-use. | Equivalent |
| Power Source | Non-replaceable, Rechargeable, Lithium-ion, 3.7 VDC, 1000mAh | No battery - mechanical Spring | CR123A 3V lithium battery | Equivalent |
| Maintenance / Calibration | No calibration required, periodic external cleaning only | No calibration required, periodic external cleaning only | None | Equivalent |
| Operating Environment | • Room temperature of 20 – 25 °C (68 – 77 °F), in order to limit temperature effect on fluid viscosity and flow rate. A 1 °C change in room temperature can change flow rate by 4.3%. Higher temperatures increase flow rate.
• 30 – 60% relative humidity, without condensation. Relative humidity does not | • Room temperature of 20 – 25 °C (68 – 77 °F), in order to limit temperature effect on fluid viscosity and flow rate.
• Humidity range – not specified
• Altitude range - not specified | • Operating Temperature Range: +10°C to +45°C
• Humidity range: 30 to 75% RH
• Altitude range: 700 – 1060 hPa | Equivalent |
Page 5 of 8
{9}
EMED Technologies
| Attribute / Feature | Tetra Micro (Subject Device) | SCIg60 Infusion System (K240148) (Primary Predicate) | CRONO S-PID 50 (K091042) (Secondary Predicate) | Equivalency |
| --- | --- | --- | --- | --- |
| | have a significant impact to flow rate.
• 1.0 to 0.7 bar atmospheric pressure (sea level to 10,000 ft). Flow rate decreases approximately 1.2% for every 1,000 feet above sea level. | | | |
| Storage Environment | • Storage (between uses):
Store the pump in a cool, dry place.
• Transportation temperature range (in original protective packaging): –30°C to +60°C
Transportation humidity:
• At –30°C: RH uncontrolled
• At +40°C: 90% RH
• At +60°C: 15% RH | Temperature: -5°C to +40°C
Humidity – Not specified | Temperature: -10°C to +60°C
Humidity: 10 – 100% RH | Equivalent |
| Infuser Size | 64 x 64 x 41 mm
125 g | 260 x 65 x 65 mm
412 g | 84.5 x 55 x 42 mm
140 g | Equivalent |
| Useful Life | Up to 157 infusions | 4,200 cycles | Two years against manufacturing defects | Equivalent |
The Tetra Micro Infusion System achieves substantial equivalence through a combined comparison to two predicate devices.
The primary predicate, SCIg60 (K240148), is equivalent in the method of fluid delivery. Both systems apply pressure to a syringe to deliver immunoglobulin subcutaneously. In each case, the flow rate is not directly measured or adjusted based on flow feedback. Instead, flow is governed by the applied pressure/force, the viscosity of the fluid, and the resistance of the downstream subcutaneous administration set. For the Tetra Micro, pressure is regulated to the selected level via a closed-loop pressure control system; however, flow remains a resultant output and is not directly measured. The SCIg60 uses a spring to apply force to the syringe plunger rod, thereby displacing the stopper (gasket) and delivering the fluid. In contrast, the Tetra Micro uses a battery-powered pump under firmware control to generate pressure that is applied directly to the back of the stopper within the syringe barrel. In both cases, controlled pressure displaces the fluid through a passive administration set, and the actual flow rate is determined by system resistance and fluid properties.
The secondary predicate, Crono S-PID 50 (K091042), supports technological equivalence in energy source and control architecture. Like the Tetra Micro, Crono pump is a battery-operated, software-controlled infusion system. The Crono uses a stepper motor-driven piston to apply force to the syringe gasket, whereas the Tetra Micro applies pneumatic
Page 6 of 8
{10}
EMED Technologies
pressure to apply force to the syringe gasket. Both devices employ embedded firmware and use microcontroller-based control to manage system functions. For Tetra Micro, a pressure sensor provides real-time feedback to maintain the selected pressure level (infusion speed) and expected infusion rates are provided in the IFU based on configuration; the system does not directly measure flow rate. The user selects one of five pressure levels (infusion speed) to match the prescribed delivery rate. The device provides LED indicators, audible notifications, and internal diagnostics. Infusion is disabled during charging. The device is not field-programmable and does not support wireless communication. The device does not require, request, retain, display, or store patient protected health information.
While the mechanical implementation of pressure differs, the subject and predicate devices share the same general approach in which flow is not directly measured/controlled using real-time flow feedback by the infuser. Other technological characteristics are comparable as summarized in the above table, including but not limited to syringe-based delivery, reusable non-sterile infuser used with sterile single-use administration components, and overall electro-mechanical architecture. These differences do not raise new questions of safety or effectiveness. Performance, lifecycle, electrical, software, cybersecurity, and usability testing support a determination of substantial equivalence.
## VII. NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS (21 CFR 807.92(b))
No clinical study is included in this submission.
The following non-clinical tests were conducted:
- Infuser Operating Pressure
- Compatibility and Leakage
- Drug-Device compatibility
- Verification of Flow Rate Performance
- Verification of System Configurations against Drug Prescribing Information
- Cleaning and Disinfection Validation
- Useful Life Testing
- Drop Testing
- Electrical Safety and Electromagnetic Compatibility (Immunity and Emissions), including evaluation of audible levels and fluid exposure
- Home Use / Home Healthcare Environment Verification
- Software Verification and Validation (including cybersecurity evaluation as applicable)
- Shipping Validation
- Human Factors/Usability Validation (Summative Usability)
- Biological Evaluation
- Safety Assurance Case
Page 7 of 8
{11}
EMED Technologies
The results of the non-clinical testing demonstrate that the subject device performs as intended and supports substantial equivalence to the SCIg60 Infusion System (K240148) and Crono S-PID 50 (K091042) with respect to safety and performance.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.