AutoDensity

K251078 · Eos Imaging · QIH · Nov 14, 2025 · Radiology

Device Facts

Record IDK251078
Device NameAutoDensity
ApplicantEos Imaging
Product CodeQIH · Radiology
Decision DateNov 14, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Spine Bone Mineral Density (BMD)Vertebral Level Identification Accuracy >= 90%; Lower boundary of 95% CI of mean Dice Coefficient >= 0.80Vertebral Level Identification Accuracy: >= 90%; Spine ROI Accuracy (Dice Coefficient): Lower boundary of 95% CI >= 0.804,679 3D reconstructions and 9,358 corresponding biplanar 2D X-ray images from 2007-2024.2 (trained technologists) + 1 (US board-certified radiologist)129 patients from EOSedge systems using low dose, single energy acquisition protocols.2 (trained technologists) + 1 (US board-certified radiologist)

Indications for Use

AutoDensity is a post-processing software intended to estimate spine Bone Mineral Density (BMD) from EOSedge dual energy images for orthopedic pre-surgical assessment applications. It is an opportunistic tool that enables immediate assessment of bone density from EOSedge images acquired for other purposes. AutoDensity is not intended to replace DXA screening. Suspected low BMD should be confirmed by a DXA exam. Clinical judgment and experience are required to properly use the software.

Device Story

AutoDensity is cloud-based post-processing software for orthopedic pre-surgical assessment. It takes dual-energy X-ray images from EOSedge systems as input. The device uses an AI algorithm to identify lumbar vertebrae (L1-L4) and compute Bone Mineral Density (BMD) by processing bone-equivalent (aluminum) and tissue-equivalent (PMMA) density images. It produces BMD measurements and image previews for clinician review. Used in clinical settings by healthcare providers to inform surgical planning; it provides opportunistic BMD assessment without replacing standard DXA screening. The output aids clinical decision-making regarding bone quality in orthopedic patients.

Clinical Evidence

Clinical performance evaluated in 253 subjects (65% US, 35% French; 68% female; 62% <50 years). Precision study (n=30) showed CV% of 2.23% [95% CI: 1.78%, 2.98%] in weight-bearing position. Bland-Altman agreement with reference device showed bias of 0.045 g/cm² and LoA [-0.088 g/cm², 0.178 g/cm²], consistent with commercial bone densitometers. Bench testing confirmed phantom precision (CV% < 1.5%) and agreement.

Technological Characteristics

Cloud-based software. Inputs: dual-energy X-ray images (aluminum/bone-equivalent and PMMA/tissue-equivalent). Processing: AI-based ROI detection and BMD computation. Output: BMD values and image previews. Connectivity: Cloud-environment. No specific hardware materials or energy sources (software-only).

Indications for Use

Indicated for orthopedic pre-surgical assessment in patients requiring spine BMD estimation. Contraindicated for cases with supernumerary or missing vertebrae.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION November 14, 2025 EOS imaging % Mathilde Masurel Design Quality and Regulatory Affairs Specialist 32 rue blanche PARIS, 75009 FRANCE Re: K251078 Trade/Device Name: AutoDensity Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: April 8, 2025 Received: October 14, 2025 Dear Mathilde Masurel: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251078 - Mathilde Masurel Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251078 - Mathilde Masurel Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251078 | | | Device Name AutoDensity | | | Indications for Use (Describe) AutoDensity is a post-processing software intended to estimate spine Bone Mineral Density (BMD) from EOSedge dual energy images for orthopedic pre-surgical assessment applications. It is an opportunistic tool that enables immediate assessment of bone density from EOSedge images acquired for other purposes. AutoDensity is not intended to replace DXA screening. Suspected low BMD should be confirmed by a DXA exam. Clinical judgment and experience are required to properly use the software. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} EOS Imaging An OOH COMPANY Traditional 510(k) Premarket Notification AutoDensity 510(k) Summary K251078 Page 1 of 11 {5} EOS Imaging ACADES COMPANY Traditional 510(k) Premarket Notification AutoDensity This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92. 1 SUBMITTER EOS imaging 32 rue Blanche Paris, France 75009 Phone: +33 1 55 25 60 60 Fax: +33 1 55 25 60 61 Contact Person: Mathilde MASUREL Design Quality and Regulatory Affairs Specialist EOS imaging Contact Phone: +33 1 55 25 60 60 Date Summary Prepared: November 10, 2025 2 DEVICE Trade Name: AutoDensity Common or Usual Name: Medical Image processing software for Bone Mineral Density evaluation Classification Name: Medical Image Management and Processing System (21 C.F.R. § 892.2050) Regulatory Class: Class II Product Code: QIH 3 LEGALLY MARKETED PREDICATE DEVICE | 510(k) | Product Name | Clearance Date | | --- | --- | --- | | K251747 | VEA Align | August 2025 | 4 REFERENCE DEVICE | 510(k) | Product Name | Clearance Date | | --- | --- | --- | | K072847 | APEX Software for QDR X-Ray Bone Densitometers | December 2007 | 5 DEVICE DESCRIPTION Based on EOSedge™ system's images acquired with the dual energy protocols cleared in K233920, AutoDensity software provides an estimate of the Bone Mineral Density (BMD) for L1-L4 in EOSedge AP radiographs of the spine. These values are used to aid in BMD estimation in orthopedic surgical planning workflows to help inform patient assessment and surgical decisions. AutoDensity is opportunistic in nature and provides BMD information with equivalent radiation Page 2 of 11 {6} EOS Imaging ACADES COMPANY Traditional 510(k) Premarket Notification AutoDensity dose compared to the EOSedge images concurrently acquired and used for general radiographic exams. AutoDensity is not intended to replace DXA screening. ## 6 INDICATIONS FOR USE AutoDensity is a post-processing software intended to estimate spine Bone Mineral Density (BMD) from EOSedge dual energy images for orthopedic pre-surgical assessment applications. It is an opportunistic tool that enables immediate assessment of bone density from EOSedge images acquired for other purposes. AutoDensity is not intended to replace DXA screening. Suspected low BMD should be confirmed by a DXA exam. Clinical judgment and experience are required to properly use the software. ## 7 TECHNOLOGICAL COMPARISON TO PREDICATE The subject device was compared to its predicate device and reference device in intended use, indications for use, design, function and technology and it was demonstrated that it is substantially equivalent. Any technological differences within this 510(k), between the subject device and the predicate or reference device, do not impact substantial equivalence, or safety and effectiveness. Page 3 of 11 {7} EOS imaging an aBO COMPANY Traditional 510(k) Premarket Notification AutoDensity | Specification | Predicate | Reference Device | Subject Device | Substantial Equivalence Discussion | | --- | --- | --- | --- | --- | | Manufacturer | EOS imaging | Hologic | EOS imaging | | | System | VEA Align (K251747) | APEX Software for QDR X-Ray Bone Densitometers (K072847) | AutoDensity | | | Product Code | QIH | KGI | QIH | Yes, same as predicate | | Device Classification | II | II | II | Yes, same as predicate | | Device Classification Name | Medical Image Management and Processing System | Bone Densitometer | Medical Image Management and Processing System | Yes, same as predicate | | Regulation Number | §892.2050 | §892.1170 | §892.2050 | Yes, same as predicate | | Intended use / Indications | This cloud-based software is intended for orthopedic applications in both pediatric and adult populations. 2D X-ray images acquired in EOS imaging's imaging systems is the foundation and resource to display the interactive landmarks overlayed on the frontal and lateral images. These landmarks are available for users to assess patient-specific global alignment. For additional assessment, alignment parameters compared to published normative values may be available. This product serves as a tool to aid in the analysis of spinal deformities and degenerative diseases, and lower limb alignment disorders and deformities through precise angle and length measurements. It is | The APEX for QDR X-Ray Bone Densitometers is indicated for the estimation of bone mineral density (BMD), comparison of measured variables obtained from a given QDR scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics using the Hologic QDR® X-Ray Bone Densitometers. IVA scans are intended for the visualization or quantitative assessment of vertebral body deformities. IVA also allows the visualization of abdominal aortic calcifications and, if present, clinical correlation may be advised since abdominal aortic calcification may be associated with cardiovascular disease. | AutoDensity is a post-processing software intended to estimate spine Bone Mineral Density (BMD) from EOSedge dual energy images for orthopedic pre-surgical assessment applications. It is an opportunistic tool that enables immediate assessment of bone density from EOSedge images acquired for other purposes. AutoDensity is not intended to replace DXA screening. Suspected low BMD should be confirmed by a DXA exam. Clinical judgment and experience are required to properly use the software. | Yes, the subject device intended use/indications for use are substantially equivalent to the predicate device as both devices intended/indicated to provide clinical parameters from X-ray images acquired by EOS imaging system (EOSedge) for orthopedic pre-surgical assessment applications (alignment parameters for the predicate and BMD for the subject device). | Page 4 of 11 {8} EOS imaging an aIto company Traditional 510(k) Premarket Notification AutoDensity | Specification | Predicate | Reference Device | Subject Device | Substantial Equivalence Discussion | | --- | --- | --- | --- | --- | | Manufacturer | EOS imaging | Hologic | EOS imaging | | | System | VEA Align (K251747) | APEX Software for QDR X-Ray Bone Densitometers (K072847) | AutoDensity | | | | suitable for use with adult and pediatric patients aged 7 years and older. Clinical judgment and experience are required to properly use the software | | | | | Contraindications | VEA Align is contraindicated for cases with vertebrae with severe congenital deformities (e.g., hemivertebrae, spina bifida, etc.). | Pregnancy and the use of contrast agents within the past 7 days are contraindicated. | AutoDensity is contraindicated for cases with supernumerary/missing vertebrae. | Yes, the updated contraindications do not change the Intended Use. Both devices treat the same patient population and compute clinical parameters from X-ray images acquired by EOSedge system for orthopedic applications. While contraindications are different, this is not critical to the proposed device's functionality. | | Technical Characteristics | AutoDensity software runs in a cloud-environment. | APEX software runs on QDR X-Ray Bone Densitometers. | AutoDensity software runs in a cloud-environment. | Yes, same as the predicate. | | | VEA Align generates spine landmarks from 2D X-ray images acquired with EOSedge system, used to compute alignment clinical parameters. | The Hologic APEX software generates a range of reports and images utilizing the DXA data acquired with the Hologic QDR systems. It can display the DXA measurements along with a representative color image mapping of bone and soft tissue. | AutoDensity generates bone mineral density information utilizing the dual-energy images acquired with the EOSedge system. It computes bone mineral density measurements from "aluminum" and "PMMA" images: • Aluminum image is bone-equivalent density, | Yes, the subject device is substantially equivalent to the predicate device as they both used EOSedge images to compute clinical parameters. The subject device is also equivalent to the reference device as they both enable bone mineral assessments to be identified from bone-equivalent and tissues-equivalent images. | Page 5 of 11 {9} EOS imaging an aIto company Traditional 510(k) Premarket Notification AutoDensity | Specification | Predicate | Reference Device | Subject Device | Substantial Equivalence Discussion | | --- | --- | --- | --- | --- | | Manufacturer | EOS imaging | Hologic | EOS imaging | | | System | VEA Align (K251747) | APEX Software for QDR X-Ray Bone Densitometers (K072847) | AutoDensity | | | | | | • PMMA image is soft tissue-equivalent density. | | | | The software allows: Assessment of patient spinal global alignment. | The software allows: • Assessment of bone density with lumbar spine, double hip and wrist examinations. • Diagnosis of vertebral fractures using low-dose spinal imaging. • Instant quantification of bone quality of a lumbar spine examination. Other functionalities are available on the system that are not described in this comparison table. | The software allows: Assessment of bone density with lumbar spine. | Yes, the subject device is substantially equivalent to the reference device as they both assess bone mineral density of lumbar spine. | | Principles of Operation | VEA Align workflow is divided into the following steps: • Data Acquisition • Data Computation • Report Generation | APEX software workflow is divided into the following main steps: • Data Acquisition, • Data Computation, • Report Generation. | AutoDensity workflow is divided into the following main steps: • Data Ingestion, • Data Computation, • Report Generation. | Yes, same | | Device Input | X-ray images from EOS and EOSedge systems | Dual-energy X-ray images | Dual-energy X-ray images from EOSedge systems | Yes, the subject device is substantially equivalent to the predicate device as they both uses EOSedge images as input. | | Device Output | VEA Align processes data and generates: | APEX software processes data and generates: | AutoDensity processes data and generates: | Yes, the subject device is substantially equivalent to the predicate device as they both | Page 6 of 11 {10} EOS imaging an aIso company Traditional 510(k) Premarket Notification AutoDensity | Specification | Predicate | Reference Device | Subject Device | Substantial Equivalence Discussion | | --- | --- | --- | --- | --- | | Manufacturer | EOS imaging | Hologic | EOS imaging | | | System | VEA Align (K251747) | APEX Software for QDR X-Ray Bone Densitometers (K072847) | AutoDensity | | | | • Preview of the spine used to compute landmarks and clinical parameters • Alignment clinical parameters | • Previews with vertebrae used for BMD computation, • BMD measurement. | • Previews with vertebrae used for BMD computation, • BMD measurements. | generate clinical parameters and preview of the images used to compute them. The subject device is also equivalent to the reference device as they both process data and generate previews and BMD measurements. | | Regions of Interest detection | VEA Align computes landmarks associated with vertebral centers and endplates needed to calculate coronal and sagittal clinical parameters. | Region of interest detection is done for at least the following anatomical body part: • Spine (lumbar vertebrae): ○ Frontal: L1, L2, L3 and L4 | Region of interest detection is done for the following anatomical body part: • Spine (lumbar vertebrae): ○ Frontal: L1, L2, L3 and L4. | Yes, the subject device is substantially equivalent to the reference device as they both perform region of interest detection on similar anatomical body parts. | | Provides Bone Mineral Density Value | N/A | Bone Mineral Density values are computed for at least the following anatomical body part: • Spine (lumbar vertebrae): ○ Frontal: L1, L2, L3, L4 and total BMD | Bone Mineral Density values are computed for the following anatomical body part: • Spine (lumbar vertebrae): ○ Frontal: L1, L2, L3, L4 and total BMD | Yes, the subject device is substantially equivalent to the reference device in that Bone Mineral Density values are also computed using L1 to L4. | | Information Provided in the Report | Patient information Examination information Landmark detection Alignment clinical parameters | Patient information Examination information ROI detection Bone Mineral Density Values | Patient information Examination information ROI detection Bone Mineral Density Values | Yes, same | Page 7 of 11 {11} EOS Imaging An OOH COMPANY Traditional 510(k) Premarket Notification AutoDensity # 8 PERFORMANCE DATA Non-clinical and clinical performance testing has been completed on the subject AutoDensity device to demonstrate substantial equivalence to the predicate device and reference device. The safety and effectiveness of AutoDensity's BMD output have not been established for the pediatric population. The following V&amp;V testing was performed on the subject device: A. Verification activities covering the following - Design input review - Unit testing - Software integration - System integration B. Region of Interest (ROI) Performance Evaluation To assess the standalone performance of the AI algorithm of AutoDensity, the test was performed with: - A dedicated test data set containing different data from the training data set. Specifically, the test data set is not a sampling of the training data set, has never been used for the algorithm training or for tunning the algorithm and leakage between training and test data sets did not occur. - A ground truth established for all cases following a 3 truther majority voting principle with input from a senior US board certified expert radiologist (gold standard). - A global acceptance criterion to assess the performance of the AI algorithm predictions regarding the ground truth. - Subgroup analyses to assess bias and generalizability. ## Training data information The AI algorithm was trained using 4,679 3D reconstructions and 9,358 corresponding EOS (K152788) or EOSedge (K233920) biplanar 2D X-ray images was selected to only keep relevant images with the fields of view of interest, at a variety of sites from 2007-2024. This population characteristics cover the intended use population. ## Testing data information The testing data set consisted of 129 patients and the demographic characteristics cover the intended use population. All cases were obtained from EOSedge systems (K233920) using low dose, single energy acquisition protocols. ## Subgroup definition (generalizability) Subgroup testing was performed to demonstrate generalizability of the algorithm: {12} EOS Imaging An OOH COMPANY Traditional 510(k) Premarket Notification AutoDensity | Type | Subgroup | Category | | --- | --- | --- | | Demographics* | Age | Pediatric | | | | Adult | | | Sex | Male | | | | Female | | Image Acquisition* | Field of View | Full Spine | | | | Full Body | | Data Collection Site | Country | US | | | | OUS | | Disease Conditions | Degenerative – Spine | Normal-Mild | | | | Moderate | | | | Severe | | Lumbar Morphometry (Levels evaluated separately; maximum angle compared to view-plane) | Vertebral Sagittal Angle for L1, L2, L3 and L4 each, and maximum | Low | | | | Medium | | | | High | | | Vertebral Axial Rotation (VAR) for L1, L2, L3 and L4 each, and maximum | Low | | | | Medium | | | | High | ## Ground truthing method: Ground truth ROIs and level identification were established by two trained technologists under the supervision of a senior US board-certified radiologist. The radiologist reviewed the results, selected the more accurate set, and made any necessary adjustments, ensuring that the final ground truth reflected a majority vote with expert adjudication as the gold standard. ## Testing method Testing of the AutoDensity algorithm was performed against the ground truth to evaluate: - Vertebral Level Identification, assessed based on the % of vertebral levels correctly identified compared to ground truth - Spine ROI Accuracy, assessed by Dice Coefficient overlap between algorithm predicted ROIs and radiologist ground truth Testing was performed with images acquired from single energy acquisitions. A bridging study using 30 cases from dual energy acquisitions demonstrated equivalent performance of the ROI algorithm and justified utilization of single energy acquisition images for this purpose. ## Acceptance Criteria The acceptance criteria are used to characterize the performance of the AI algorithm feature. As such, the following acceptance criteria were defined: - Vertebral Level Identification Accuracy - Percent of levels correctly identified ≥ 90% Page 9 of 11 {13} EOS Imaging EUROPEAN OPTIMIZER Traditional 510(k) Premarket Notification AutoDensity - Spine ROI Accuracy: - Lower boundary of 95% CI of mean Dice Coefficient ≥ 0.80 ## Testing summary Testing confirms that the AutoDensity ROI detection algorithm meets performance thresholds. The high accuracy in both vertebral level identification and ROI detection combined with strong subgroup performance, supports its clinical utility and generalizability for the AutoDensity intended use population. ## C. Bench performance testing Precision and agreement of BMD measurements in phantoms, compared to the reference device. - BMD precision was assessed by measuring three phantoms ten times without repositioning. Results met the acceptance criterion (CV% &lt; 1.5%), indicating equivalent precision performance to the same phantoms scanned with the reference device. - BMD agreement was assessed by measuring one phantom analyzed with AutoDensity and the reference device. Bland-Altman analysis showed a maximum BMD difference of 0.057 g/cm² for the high BMD phantom vertebra, and a difference of less than 0.018 g.cm² for the vertebrae within a clinically relevant range of BMD for the specified orthopedic planning workflow for pre-surgical patient evaluation of suspected low BMD. ## D. Clinical performance testing Clinical studies were performed to demonstrate the performance of AutoDensity with respect to: - Precision of AutoDensity BMD measurements in 30 human subjects undergoing repeated measurements with repositioning. AutoDensity precision CV% was 2.23% [95% CI: 1.78%, 2.98%] with the patient in a weight-bearing standing position, in contrast to modalities where the patient is lying supine and stabilized with positioning aids which yields higher precision for DXA densitometers. Despite the difference in mode of operation, AutoDensity precision performance is within the range of acceptable clinical limits for the specified pre-surgical orthopedic patient assessment. - Bland-Altman agreement between AutoDensity and the reference device BMD measurements in a clinical population of 253 subjects representative of the intended use population, including 65% US subjects and 35% French subjects; 68% were female; the proportion of subjects aged below 50 years was 62% and above 50 was 38%; 56% of subjects had BMI below 25, 34% between 25-30 and 10% between 30-35. Bland-Altman bias was 0.045 g/cm² and limits of agreement (LoA) were [-0.088 g/cm², 0.178 g/cm²]. Bland-Altman bias and LoA for AutoDensity compared to the reference device were equivalent to published Bland- {14} EOS Imaging An OOH COMPANY Traditional 510(k) Premarket Notification AutoDensity Altman agreement between other commercial bone densitometers in similar clinical study populations. The results demonstrate that the subject AutoDensity device is substantially equivalent to the predicate and reference device for nonclinical and clinical testing. # 9 CONCLUSION Based on the information provided in this 510(k) submission, it was determined that the subject device, AutoDensity, is substantially equivalent to the legally marketed predicate device and reference device with regards to indications for use, intended use, design, technology, and performance. Page 11 of 11
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...