Validation data sourced from multiple sites, with 100% of the data sampled from US sources.
3 (expert annotators)
Indications for Use
Viewing, post-processing, qualitative and quantitative evaluation of blood vessels and cardiovascular CT images in DICOM format.
Device Story
cvi42 Coronary Plaque Software Application is a SaMD for analyzing coronary artery CT Angiography scans. It assists physicians in visualizing and evaluating coronary plaques and luminal stenoses. Input consists of DICOM-formatted cardiac CT images. The device uses semi-automated machine learning algorithms to segment coronary vessel lumen and wall structures, converting them into editable surfaces. It provides multiplanar reconstruction (MPR), curved planar reformation (CPR), and 3D rendering. The system calculates plaque volumes (calcified, non-calcified, low-attenuation), plaque burden, remodeling index, and stenosis percentages. Results are displayed on-screen for physician review and reporting. Used in clinical settings by qualified medical professionals to aid diagnostic decision-making. The device does not replace physician judgment; users are advised to verify and correct automated outputs. It integrates with existing viewing software like cvi42.
Clinical Evidence
Bench testing only. No clinical studies were performed. Validation used a reference standard established by three expert annotators. Performance metrics for ML-based segmentation: Lumen Mean DSC 0.76, Wall Mean DSC 0.80, Lumen Mean HD 0.77 mm, Wall Mean HD 0.87 mm. Plaque volume correlation (PCC) results: Total Plaque 0.97, Calcified Plaque 0.99, Non-Calcified Plaque 0.93, Low-Attenuation Plaque 0.74. Testing confirmed consistent performance across diverse subgroups (scanner vendor, resolution, slice thickness, patient demographics).
Technological Characteristics
SaMD; DICOM-compliant; supports CT imaging modalities. Features semi-automated ML-based segmentation of coronary lumen and vessel wall. Includes MPR, CPR, and 3D rendering. Software development follows ISO 13485:2016, IEC 62304:2015, and ISO 14971:2019. Cybersecurity controls include threat modeling, SCA, static analysis, and penetration testing.
Indications for Use
Indicated for adult patients who underwent Coronary Computed Tomography Angiography (CCTA) for evaluation of CAD or suspected CAD. Assists physicians in qualitative and quantitative assessment of cardiac CT images to determine the presence and extent of coronary plaques and stenoses. Used by qualified medical professionals experienced in cardiovascular CT image examination.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Circle Cardiovascular Imaging Inc.
Omar Naimi
Regulatory Affairs & Quality Management Systems Analyst
Suite 1800, 707 8 Avenue SW
Calgary, AB T2P 1H5
Canada
October 27, 2025
Re: K251027
Trade/Device Name: cvi42 Coronary Plaque Software Application
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: September 25, 2025
Received: September 26, 2025
Dear Omar Naimi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251027 - Omar Naimi
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K251027 - Omar Naimi
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251027 | |
| Device Name cvi42 Coronary Plaque Software Application | |
| Indications for Use (Describe) cvi42 Coronary Plaque Software Application is intended to be used for viewing, post-processing, qualitative and quantitative evaluation of cardiovascular computed tomography (CT) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format. It enables a set of tools to assist physicians in qualitative and quantitative assessment of cardiac CT images to determine the presence and extent of coronary plaques and stenoses, in patients who underwent Coronary Computed Tomography Angiography (CCTA) for evaluation of CAD or suspected CAD. cvi42 Coronary Plaque’s semi-automated machine learning algorithms are intended for an adult population. cvi42 Coronary Plaque shall be used only for cardiac images acquired from a CT scanner. It shall be used by qualified medical professionals, experienced in examining cardiovascular CT images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251027
# Coronary Plaque Module 510(k) Summary

The following 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Device Act 1990 and 21 CFR 807.92(c).
## I. SUBMITTER
Submitter's Name: Circle Cardiovascular Imaging, Inc.
Address: Suite 1800 – 707 8th Ave SW, Calgary, AB, Canada, T2P 1H5
Date Prepared: October 27, 2025
Telephone Number: +1 587 686 0784
Contact Person: Omar Naimi
Email: omar.naimi@circlecvi.com
## II. DEVICE
510(k) K251027
Name of the Device: cvi42 Coronary Plaque Software Application
Common or Usual Name: Radiological Image Processing System
Classification Name: Medical Image Management and Processing System
Proposed Classification: Device Class: II
Product Code: QIH, LLZ
Regulation Number: 21 CFR 892.2050
## III. PREDICATE DEVICE
The predicate device(s) identified are cvi42 Auto Imaging Software Application manufactured by Circle Cardiovascular Imaging Inc. (K213998) as the primary predicate, and Clearly Labs manufactured by Clearly Inc. (K190868) as the secondary predicate.
The predicate device(s) have not been subject to a design-related recall.
Circle Cardiovascular Imaging Inc.
Non-Confidential
Page 1 of 9
{5}
# Coronary Plaque Module 510(k) Summary
## IV. DEVICE DESCRIPTION
Circle's cvi42 Coronary Plaque Software Application ('cvi42 Coronary Plaque' or 'Coronary Plaque Module', for short) is a Software as a Medical Device (SaMD) that enables the analysis of CT Angiography scans of the coronary arteries of the heart. It is designed to support physicians in the visualization, evaluation, and analysis of coronary vessel plaques through manual or semi-automatic segmentation of vessel lumen and wall to determine the presence and extent of coronary plaques and luminal stenoses, in patients who underwent Coronary Computed Tomography Angiography (CCTA) for the evaluation of coronary artery disease (CAD) or suspected CAD. The device is intended to be used as an aid to the existing standard of care and does not replace existing software applications that physicians use. The Coronary Plaque Module can be integrated into an image viewing software intended for visualization of cardiac images, such as Circle's FDA-cleared cvi42 Software Application. The Coronary Plaque Module does not interface directly with any data collection equipment, and its functionality is independent of the type of vendor acquisition equipment. The analysis results are available on-screen, can be sent to report or saved for future review.
The Coronary Plaque Module consists of multiplanar reconstruction (MPR) views, curved planar reformation (CPR) and straightened views, and 3D rendering of the original CT data. The Module displays three orthogonal MPR views that the user can freely adjust to any position and orientation. Lines and regions of interest (ROIs) can be manually drawn on these MPR images for quantitative measurements.
The Coronary Plaque Module implements an Artificial Intelligence/Machine Learning (AI/ML) algorithm to detect lumen and vessel wall structures. Further, the module implements post-processing methods to convert coronary artery lumen and vessel wall structures to editable surfaces and detect the presence and type of coronary plaque in the region between the lumen and vessel wall. All surfaces generated by the system are editable and users are advised to verify and correct any errors.
The device allows users to perform the measurements listed in Table 1.
Circle Cardiovascular Imaging Inc.
Non-Confidential
{6}
# Coronary Plaque Module 510(k) Summary
Table 1. Measurements in the Coronary Plaque Module
| Measurement [units] | Description | Display |
| --- | --- | --- |
| Calcified Plaque (CP) Volume [mm3] | The volume of calcified plaque (CP) in a lesion or all the lesions in a vessel. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Non-Calcified Plaque (NCP) Volume [mm3] | The volume of non-calcified plaque (NCP) in a lesion or all the lesions in a vessel. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Low-Attenuation Plaque (LAP) Volume [mm3] | The volume of low-attenuation plaque (LAP) in a lesion or all the lesions in a vessel; LAP volume is also included in NCP volume. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Total Plaque Volume [mm3] | The volume of total plaque including CP and NCP (with LAP) in a lesion or all the lesions in a vessel. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Calcified Plaque (CP) Burden [%] | The burden of CP in a lesion or the whole vessel; obtained by dividing the CP volume by the lesion vessel wall or entire vessel volume, as appropriate. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Non-Calcified Plaque (NCP) Burden [%] | The burden of NCP in a lesion or the whole vessel; obtained by dividing the NCP volume by the lesion vessel wall or entire vessel volume, as appropriate. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Low-Attenuation Plaque (LAP) Burden [%] | The burden of LAP in a lesion or the whole vessel; obtained by dividing the LAP volume by the lesion vessel wall or entire vessel volume, as appropriate. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Total Plaque Burden [%] | The burden of total plaque in a lesion or the whole vessel; obtained by dividing the total plaque volume by the lesion vessel wall or entire vessel volume, as appropriate. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Remodeling Index [no unit] | A measure of enlargement of vessel dimensions to accommodate plaque development compared to normal vessel sections; it is expressed as the ratio of the largest diameter across the lesion and reference diameter (or maximum vessel wall cross-sectional area and reference vessel wall cross-sectional area). | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Diameter-Based Stenosis [%] | Stenosis expressed as the percent amount of reduced lumen diameter compared with reference diameter. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Area-Based Stenosis [%] | Stenosis expressed as the percent amount of reduced lumen area compared with reference area. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
| Lesion Length [mm] | The length of the lesion. | • On Screen Data Panel • Coronaries (Plaque) Module Report |
Circle Cardiovascular Imaging Inc.
Non-Confidential
{7}
Coronary Plaque Module 510(k) Summary
## V. INTENDED USE / INDICATIONS FOR USE
### Intended Use
Viewing, post-processing, qualitative and quantitative evaluation of blood vessels and cardiovascular CT images in DICOM format.
### Indications for Use
cvi42 Coronary Plaque Software Application is intended to be used for viewing, post-processing, qualitative and quantitative evaluation of cardiovascular computed tomography (CT) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format.
It enables a set of tools to assist physicians in qualitative and quantitative assessment of cardiac CT images to determine the presence and extent of coronary plaques and stenoses, in patients who underwent Coronary Computed Tomography Angiography (CCTA) for evaluation of CAD or suspected CAD.
cvi42 Coronary Plaque’s semi-automated machine learning algorithms are intended for an adult population.
cvi42 Coronary Plaque shall be used only for cardiac images acquired from a CT scanner. It shall be used by qualified medical professionals, experienced in examining cardiovascular CT images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process.
## VI. COMPARISON WITH PREDICATE DEVICE
The detailed analysis of the subject device and the predicate devices (shown in Table 2 and Table 3) demonstrates that the subject device is substantially equivalent in indications for use / intended use, technological characteristics, functionality, and operating principles with the primary (K213998) and secondary (K190868) predicates. Of the three characteristics (technical, biological, and clinical) required for the demonstration of equivalence, biological characteristics are not applicable since both the subject device and predicate device(s) are software as a medical device applications with no tangible component interfacing with the body.
Circle Cardiovascular Imaging Inc.
Non-Confidential
{8}
# Coronary Plaque Module 510(k) Summary
Table 2. Intended use and indications comparison.
| | Subject Device cvi42 Coronary Plaque Software Application Manufactured by Circle Cardiovascular Imaging Inc. | Primary Predicate cvi42 Auto Imaging Software Application (K213998) Manufactured by Circle Cardiovascular Imaging Inc. | Secondary Predicate Cleerly Labs (K190868) Manufactured by Cleerly Inc. |
| --- | --- | --- | --- |
| Intended Use | Viewing, post-processing, qualitative and quantitative evaluation of blood vessels and cardiovascular CT images in DICOM format. | Viewing, post-processing, qualitative and quantitative evaluation of blood vessels and cardiovascular MR and CT images in DICOM format. | N/A |
| Indications for Use | cvi42 Coronary Plaque Software Application is intended to be used for viewing, post-processing, qualitative and quantitative evaluation of cardiovascular computed tomography (CT) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format. It enables a set of tools to assist physicians in qualitative and quantitative assessment of cardiac CT images to determine the presence and extent of coronary plaques and stenoses, in patients who underwent Coronary Computed Tomography Angiography (CCTA) for evaluation of CAD or suspected CAD. cvi42 Coronary Plaque's semi-automated machine learning algorithms are intended for an adult population. cvi42 Coronary Plaque shall be used only for cardiac images acquired from a CT scanner. It shall be used by qualified medical professionals, experienced in examining cardiovascular CT images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. | cvi42 Auto is intended to be used for viewing, post-processing, qualitative and quantitative evaluation of cardiovascular magnetic resonance (MR) and computed tomography (CT) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format. It enables a set of tools to assist physicians in qualitative assessment of cardiac images and quantitative measurements of the heart and adjacent vessels; perform calcium scoring; and to confirm the presence or absence of physician-identified lesion in blood vessels. The target population for cvi42 Auto's manual workflows is not restricted; however, cvi42 Auto's semi-automated machine learning algorithms are intended for an adult population. cvi42 Auto shall be used only for cardiac images acquired from an MR or CT scanner. It shall be used by qualified medical professionals, experienced in examining cardiovascular MR or CT images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. | Cleerly Labs is a web-based software application that is intended to be used by trained medical professionals as an interactive tool for viewing and analyzing cardiac computed tomography (CT) data for determining the presence and extent of coronary plaques and stenosis in patients who underwent Coronary Computed Tomography Angiography (CCTA) for evaluation of CAD or suspected CAD. This software post processes CT images obtained using any Computed Tomography (CT) scanner. The software provides tools for the measurement and visualization of coronary arteries. The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present. |
Circle Cardiovascular Imaging Inc.
Non-Confidential
{9}
# Coronary Plaque Module 510(k) Summary
Table 3. Regulatory and technological features comparison.
| Feature | Subject Device Coronary Plaque Module Manufactured by Circle Cardiovascular Imaging Inc. | Primary Predicate cvi42 Auto (K213998) Manufactured by Circle Cardiovascular Imaging Inc. | Secondary Predicate Cleerly Labs (K190868) Manufactured by Cleerly Inc | Comparison |
| --- | --- | --- | --- | --- |
| Device Class | II | II | II | Same as predicate device(s) |
| Product Code | QIH, LLZ | QIH, LLZ | LLZ | Same as primary predicate device, secondary predicate uses AI/ML functionality. |
| Regulation Name | Medical image management and processing system | Medical image management and processing system | Picture archiving and communications system | Same as primary predicate device, secondary predicate uses AI/ML functionality. |
| Regulation Number | 21 CFR § 892.2050 | 21 CFR § 892.2050 | 21 CFR § 892.2050 | Same as predicate device(s) |
| Computer operating system | macOS, Microsoft Windows | macOS, Microsoft Windows | Client-Server Google Chrome Application | Same as primary predicate device. Core functionality of subject device unaffected by OS. No safety or effectiveness concerns |
| Imaging Modalities | CT | MR and CT | CT | Same as the predicate device(s) with the exception that the primary predicate also allows MR inputs. |
| DICOM Compliant | Yes | Yes | Yes | Same as predicate device(s) |
| Import and display CT images | Yes | Yes | Yes | Same as predicate device(s) |
| Post process CCT images | Yes | Yes | Yes | Same as predicate device(s) |
| Images can be displayed by study and series | Yes | Yes | Yes | Same as predicate device(s) |
| 2D Imaging | Yes | Yes | Yes | Same as predicate device(s) |
| 2D Measurement | Yes | Yes | Yes | Same as predicate device(s) |
| 3D Imaging | Yes | Yes | Yes | Same as predicate device(s) |
Circle Cardiovascular Imaging Inc.
Non-Confidential
{10}
# Coronary Plaque Module 510(k) Summary
| Multiplanar reformatting (MPR) | Yes | Yes | Yes | Same as predicate device(s) |
| --- | --- | --- | --- | --- |
| Segmentation of Region of Interest | Manual and Semi-Automatic (using Machine Learning technique)
Segmentation of Coronary Vessels, Including Lumen and Vessel Wall Structures. | Manual and Semi-Automatic (using Machine Learning technique)
Centerline Generation of the Coronary Vessels. | Manual and Semi-Automatic (using Machine Learning Technique)
Segmentation of Coronary Vessels, Including Lumen and Vessel Wall Structures. | Same as the predicate device(s), with the exception that the primary predicate does not offer segmentation of the lumen and vessel wall structures.
The successful passing of the V&V performance testing has demonstrated that the addition of semi-automatic segmentation of the lumen and vessel wall structures does not raise new questions of safety. |
| Plaque Thresholds | Yes | No | Yes | Both the subject device and secondary predicate device provide plaque thresholds. |
| Measurements | | | | |
| Signal Density [HU] | Yes | Yes | Yes | Same as predicate device(s) |
| Distance Measurements (Vessel, Lesion, Length) [mm] | Yes | Yes | Yes | Same as predicate device(s) |
| Volumetric Plaque Measurements [mm³] | - Non-Calcified Plaque (NCP)
- Low-Attenuation Plaque (LAP)
- Calcified Plaque (CP)
- Total Plaque | - Calcified Plaque (CP) | - Total Vessel
- Total Lumen
- Non-Calcified Plaque (NCP)
- Low-Density Non-Calcified Plaque (LD-NCP)
- Calcified Plaque (CP)
- Total Plaque | The subject device and the secondary predicate display the same measurements, with the exception of Total Vessel and Total Lumen – which are calculated but not reported as a device output. |
| Remodeling Index | Yes | No | Yes | Same as the secondary predicate device. |
| Stenosis [%] | - Diameter-Based Stenosis
- Area-Based Stenosis | - Diameter-Based Stenosis
- Area-Based Stenosis | - Diameter-Based Stenosis
- Area-Based Stenosis | Same as predicate device(s) |
Circle Cardiovascular Imaging Inc.
Non-Confidential
{11}
Circle Cardiovascular Imaging Inc.
Non-Confidential
Page 8 of 9
# Coronary Plaque Module 510(k) Summary
## VII. PERFORMANCE DATA AND TESTING
Performance testing was conducted to verify compliance with specified design requirements in accordance with ISO 13485:2016, IEC 62304:2015, ISO 14971:2019, and DICOM standards.
Verification and validation testing were conducted to ensure specifications and performance of the device and were performed per the FDA Guidance "Technical Performance Assessment of Quantitative Imaging in Radiological Device Premarket Submission". No clinical studies were necessary to support substantial equivalence.
The Coronary Plaque Module has been tested according to the specifications that are documented in a Master Software Test Plan. Testing is an integral part of Circle Cardiovascular Imaging Inc.'s software development as described in the company's product development process.
## Cybersecurity Testing
The Coronary Plaque Module has been designed to meet cybersecurity requirements in accordance with FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", through the use of threat modeling, software composition analysis (SCA), static analysis, vulnerability testing, and a risk assessment based on exploitability and severity of harm.
Penetration testing was conducted to ensure that there were no unidentified vulnerabilities and that the appropriate risk control measures were implemented to protect from known vulnerabilities when the device subject to a source of threat.
Circle maintains a cybersecurity risk management plan for post-market monitoring and response.
## Validation of non-Machine Learning Outputs
The validation of non-ML and mathematical outputs in the Coronary Plaque Module have been performed through general in-house software QA verification and validation processes including unit tests, content/phantom tests, quantitative and qualitative analysis, and visual/manual evaluation.
## Validation of Machine Learning Derived Outputs
The Machine Learning (ML) algorithms of the Coronary Plaque Module have been trained and tested on images acquired from major vendors of CT imaging devices. All data used for validation were not used during the development of the ML algorithms. Image information for all samples was anonymized and limited to ePHI-free DICOM headers. The validation data was sourced from multiple sites, with 100% of the data sampled from US sources.
{12}
# Coronary Plaque Module 510(k) Summary
The performance of the ML-based coronary vessel and lumen wall segmentation algorithm was evaluated against pre-defined acceptance criteria and compared to a reference standard established from three expert annotators. Segmentation accuracy was quantified using the Dice Similarity Coefficient (DSC) and Hausdorff Distance (HD), while the correlation of derived plaque volumes across plaque compositions was assessed with the Pearson Correlation Coefficient (PCC). All performance testing results met Circle's pre-defined acceptance criteria.
| Endpoint | Results | Pass / Fail |
| --- | --- | --- |
| Lumen Mean DSC | 0.76 | Pass |
| Wall Mean DSC | 0.80 | Pass |
| Lumen Mean HD | 0.77 mm | Pass |
| Wall Mean HD | 0.87 mm | Pass |
| TP PCC | 0.97 | Pass |
| CP PCC | 0.99 | Pass |
| NCP PCC | 0.93 | Pass |
| LAP PCC | 0.74 | Pass |
The ML algorithms of the Coronary Plaque Module demonstrated consistent and comparable performance across diverse subgroups, including variations in scanner vendor, tube current, pixel resolution, slice thickness, patient age, sex, vessel location, and plaque lesion volume.
# VIII. CONCLUSION
The information submitted in this premarket notification, including the performance testing and predicate device comparison, support the safety and effectiveness of the Coronary Plaque Module as compared to the predicate device(s) when used for the defined intended use.
Circle Cardiovascular Imaging Inc.
Non-Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.