The CurrentBody Skin Dual Light Hair Growth Helmet is indicated to promote hair growth in males with who have Norwood-Hamilton classifications of IIa-V or females with who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Skin Phototypes I-IV.
Device Story
Device is an over-the-counter hair growth helmet (model MZ-07A) for home use. Input: user-initiated power activation. Operation: device utilizes 100 laser diodes (655nm) and 200 red light LEDs (660nm) to deliver low-level light therapy to the scalp. Output: visible red light emission; digital display shows remaining treatment time. Device features pre-set treatment parameters and automatic shut-off. User operates device by wearing the helmet; healthcare provider role is limited to initial diagnosis of hair loss condition. Benefit: promotes hair growth in specified patient populations.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and laser safety (IEC 60825-1).
Technological Characteristics
Helmet-style device; 100 laser diodes (655±5nm) and 200 red LEDs (660±5nm). Silicone inner surface (ISO 10993 compliant). Powered by internal battery (IEC 62133-2). Standalone operation with USB charging. Pre-set treatment parameters; automatic shut-off. No external connectivity or cybersecurity interfaces.
Indications for Use
Indicated for males with Norwood-Hamilton classifications IIa-V or females with Ludwig-Savin Classifications I-II, with Fitzpatrick Skin Phototypes I-IV, to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
Predicate Devices
Neuhat Hair Growth System (or ibeauty.com Laser Cap) (K231153)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 17, 2025
Shenzhen Kaiyan Medical Equipment Co., Ltd
Alain Dijkstra
CEO
Building#3 and Building#5, 40th of Fuxin Street
Huaide Community Fuyong Town Baoan District
Shenzhen, Guangdong 518103
China
Re: K251017
Trade/Device Name: CurrentBody Skin Dual Light Hair Growth Helmet (model: MZ-07A)
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP
Dated: March 31, 2025
Received: April 2, 2025
Dear Alain Dijkstra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251017 - Alain Dijkstra
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K251017 - Alain Dijkstra
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.06.17
15:35:14 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251017 | |
| Device Name CurrentBody Skin Dual Light Hair Growth Helmet (model: MZ-07A) | |
| Indications for Use (Describe) The CurrentBody Skin Dual Light Hair Growth Helmet is indicated to promote hair growth in males with who have Norwood-Hamilton classifications of IIa-V or females with who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Skin Phototypes I-IV. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 8
K251017 - 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
1. Submitter Information
Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd
Establishment Registration Number: 3011644607
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Tel: 0755-82129361
Fax: 0755-25024651
Contact Person (including title): Alain Dijkstra (CEO)
E-mail: regulation@kaiyanmedical.com
Distributor
Company Name: CurrentBody.com Ltd
Address: Q17, Crossley Road, Stockport, Greater Manchester, SK4 5BB United Kingdom
Application Correspondent:
Contact Person: Mr. Alain Dijkstra
Company: Shenzhen Kaiyan Medical Equipment Co., Ltd
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Tel: +86 755 82129361
Fax: +86 755 25024651
Email: regulation@kaiyanmedical.com
2. Summary Prepared Date
23 May 2025
3. Subject Device Information
Trade Name: CurrentBody Skin Dual Light Hair Growth Helmet
Model: MZ-07A
Common Name: Laser, Comb, Hair
Classification Name: Infrared Lamp
Review Panel: General & Plastic Surgery
Product Code: OAP
Regulation Number: 21 CFR 890.5500
Regulation Class: II
4. Predicate Device Information
{5}
Page 2 of 8
# Primary Predicate Device Information
Sponsor: Nature Incredible Inc.
Trade Name: Neuhat Hair Growth System (or ibeauty.com Laser Cap)
Model: NEU300, NEU180
510(k) Number: K231153
Review Panel: General & Plastic Surgery
Regulatory Class: Class II
Regulatory Code: OAP
Regulation: 21 CFR 890.5500
# Reference device
Sponsor: Ronald E. Berglund, GRACE Consulting, LLC
Trade Name: Bosley Booster 128 Laser Cap; Bosley Booster 162 Laser Cap; Bosley Booster 288 Laser Cap
Model: 128 model, 162 model, 288 model
510(k) Number: K240456
Review Panel: General & Plastic Surgery
Regulatory Class: Class II
Regulatory Code: OAP
Regulation: 21 CFR 890.5500
# 5. Device Description
The CurrentBody Skin Dual Light Hair Growth Helmet (model:MZ-07A) is an over-the-counter device designed to promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V patterns of hair loss and to promote hair growth in females who have Ludwig-Savin Classifications of I-II; both with Fitzpatrick Skin Types I-IV.
The device consists of 100 laser diodes with wavelength at $655 \pm 5\mathrm{nm}$, and 200 red light diodes with wavelength at $660 \pm 5\mathrm{nm}$.
The CurrentBody Skin Dual Light Hair Growth Helmet includes a hair-growing cap, a device base and a USB power cord. There are LEDs and lasers diodes inside the cap to emit visible red light, the device base acts as a support when the device is not used. There is a power button on cap to turn on/off the device, a digital display to show the remaining treatment time, an power indicator to show charging status and a USB port for charging.
The treatment parameter is pre-set by manufacturer, the device will auto off after treatment.
# 6. Intended Use
{6}
The CurrentBody Skin Dual Light Hair Growth Helmet is indicated to promote hair growth in males with who have Norwood-Hamilton classifications of IIa-V or females with who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Skin Phototypes I-IV.
## 7. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of Comparison | Subject Device | Primary predicate | Reference device | Verdict |
| --- | --- | --- | --- | --- |
| Company | Shenzhen Kaiyan Medical Equipment Co., Ltd | Nature Incredible Inc. | Ronald E. Berglund, GRACE Consulting, LLC | -- |
| Trade Name | CurrentBody Skin Dual Light Hair Growth Helmet | Neuhat Hair Growth System (or ibeauty.com Laser Cap) | Bosley Booster 128 Laser Cap; Bosley Booster 162 Laser Cap; Bosley Booster 288 Laser Cap | -- |
| 510(k) Number | Applying | K231153 | K240456 | -- |
| Regulation number | 21 CFR 890.5500 | 21 CFR 890.5500 | 21 CFR 890.5500 | Same |
| Classification Name | Infrared Lamp | Infrared Lamp | Infrared Lamp | Same |
| Product Code | OAP | OAP | OAP | Same |
| Class | II | II | II | Same |
{7}
| Elements of Comparison | Subject Device | Primary predicate | Reference device | Verdict |
| --- | --- | --- | --- | --- |
| Intended Use / Indications for Use | The CurrentBody Skin Dual Light Hair Growth Helmet is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa~V or females with androgenic alopecia who have Ludwig-Savin Classifications of I~II and both with Fitzpatrick Skin Phototypes I to IV. | Neuhat Hiar Growth System is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V and females with addrogneic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I-IV.
ibeauty.com Laser Cap is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V and females with addrogneic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I-IV. | Indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of IIa-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I – II and Fitzpatrick Classification of Skin Phototypes I to IV. | Same |
| Location for use | OTC | OTC | OTC | Same |
| Intended User | Both sex | Both sex | Both sex | Same |
| Type of light | Low-level laser diodes and light emitting diodes | Low-level laser diodes and light emitting diodes | Low-level laser diodes and light-emitting diodes | Same |
| Wavelength | Laser: 655 ±5nm
LED: 660 ± 5nm | Laser: 655nm
Red light LED: 650nm | 650nm | Similar Note1 |
| Number of diodes | Laser diodes: 100
LED diodes: 200 | NEU300
Laser diodes: 100
LED diodes: 200
NEU180
Laser diodes: 50
LED diodes: 130 | 288 model: 115 LDs, 173 LEDs | Same |
Page 4 of 8
{8}
| Elements of Comparison | Subject Device | Primary predicate | Reference device | Verdict |
| --- | --- | --- | --- | --- |
| Energy of per laser diode | 5 mw | 5mw | 5mW per light | Same |
| Distribution | Uniform | Not public available | Uniform | Same |
| Treatment Time | Every day 20 mins | Each treatment: 25 mins
Total Treatment: 16 weeks, every other day (indefinite) | 25 minutes every day | Same
Note 2 |
| Irradiance | 2.85 mw/cm² | Not public available | 2.47 mw/cm² | Same
Note 2 |
| Applicable people | Norwood-Hamilton IIa~V (males);
Ludwig-Savin I~II (females) | Norwood-Hamilton IIa~V (males);
Ludwig-Savin I~II (females) | Norwood-Hamilton IIa~V (males);
Ludwig-Savin I~II (females) | Same |
| Applicable skin | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Same |
| Helmet/Cap design | Yes | Yes | Yes | Same |
| Biocompatibility feature | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10 | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10 | Not public available | Same |
Comparison in Detail(s):
Note 1:
Though the wavelength of subject device is a little different from predicate devices, it is very close to predicate devices. Therefore, this slight difference will not raise any effectiveness or safety issue.
Note 2:
The treatment time and irradiance of the subject device are a little difference from the predicate devices but very close. Therefore, this slight difference will not raise any effectiveness or safety issue.
8. Test Summary
7.1 Summary of Non-Clinical Performance Testing
1) Performance Testing Summary
The CurrentBody Skin Dual Light Hair Growth Helmet (Model: MZ-07A) has been evaluated the safety and performance by lab bench testing as following:
{9}
| Standards No. | Standard Title | Version | Recognition Number |
| --- | --- | --- | --- |
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Edition 3.2 2020-08 | 19-49 |
| IEC 60601-1-11 | Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment | Edition 2.1 2020-07 | 19-38 |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | Edition 4.1 2020-09 | 19-36 |
| IEC TS 60601-4-2 | Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems | IEC TS 60601-4-2: Edition 1.0 2024-03 | 19-50 |
| IEC 60601-2-57 | Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic diagnostic monitoring and cosmetic/aesthetic use | Edition 2.0 2023-07 | 12-355 |
| IEC 60825-1 | Safety of laser products - Part 1: Equipment classification and requirements [Including: Technical Corrigendum 1 (2008) Interpretation Sheet 1 (2007) Interpretation Sheet 2 (2007)] | Edition 2.0 2007-03 | 12-273 |
Page 6 of 8
{10}
Page 7 of 8
| Standards No. | Standard Title | Version | Recognition Number |
| --- | --- | --- | --- |
| IEC 62471 | Photobiological safety of lamps and lamp systems | First edition 2006-07 | 12-249 |
| IEC 62133-2 | Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them for use in portable applications - Part 2: Lithium systems | Edition 1.0 2017-02 | 19-33 |
## 2) Biocompatibility
The patient directly contracting materials in the subject device are showed in the following list.
| Components of Subject Device | Material of Components | Body Contact Category (ISO 10993-1) | Contact Duration (ISO 10993-1) |
| --- | --- | --- | --- |
| Inner surface of cap | Silicone | Surface-contacting device: skin | A-limited(<24h) |
The Nature of body contact is Surface medical device-Intact skin contact. And the contact duration is less than 24 hours. According to Table 1 - Initial evaluation tests for consideration in ISO 10993-1, the applicable biological effect is:
- Cytotoxicity
- Irritation or intracutaneous reactivity
- Sensitization
The inner surface of cap (silicone) is identical to cleared device K230336 in formulation, processing, and cleaning, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents. Etc.).
## 3) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
## 4) Cybersecurity
The subject device no any external interfaces, according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", no need cybersecurity evaluation.
{11}
Page 8 of 8
## 7.2 Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
## 9. Conclusion
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device K231153 and K240456.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.