Flattening ratio error rate 8.05%, cross-sectional area error rate 13.11%
—
—
200 median nerve images from 20 patients
4 (anesthesiologists)
Indications for Use
The diagnostic ultrasound systems and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. (Note: V5/cV5, V4/cV4 also include Dermatology and MV-Flow Mode).
Device Story
Mobile, software-controlled diagnostic ultrasound systems; acquire ultrasound data via transducers; process signals to display 2D, Doppler (Color, Power, PW, CW, TDI, TDW), M-mode, 3D/4D, and elastography images. Operated by trained healthcare professionals in hospitals/clinics. Features AI-assisted tools (BiometryAssist, ViewAssist, HeartAssist, EzNerveMeasure) for automated view recognition, anatomical segmentation, and biometric measurements. Output displayed on-screen for clinician review to support diagnosis and anatomical measurement. Includes educational tools (EzAssist, EzTrainer) for scanning reference. Benefits include standardized measurement accuracy and workflow efficiency. System includes real-time acoustic output display (Mechanical/Thermal indices).
Clinical Evidence
Bench testing only. No clinical studies required. AI features validated using static image datasets (e.g., 320 images for BiometryAssist, 680 for ViewAssist, 280 for HeartAssist, 200 for EzNerveMeasure). Metrics include Dice-score (0.863–0.869), sensitivity (93.97–94.29%), specificity (99.62%), and measurement error rates (4–13.11%). Ground truth established by expert clinicians (obstetricians/anesthesiologists).
Technological Characteristics
Mobile ultrasound system; transducers; 2D/3D/4D/Doppler imaging; software-controlled; acoustic output display (MI/TI). Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, NEMA UD 2-2004. Deep learning-based AI algorithms for segmentation and measurement.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients across all ages and genders, including fetal, neonatal, pediatric, and adult populations, for clinical applications including obstetrics, abdominal, cardiac, vascular, musculoskeletal, and urological examinations.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
V8/cV8, V7/cV7, V6/cV6 Diagnostic Ultrasound System (K243702)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 18, 2025
Samsung Medison Co., Ltd.
Yujin Kim
Regulatory Affairs Specialist
3366, Hanseo-ro, Nam-myeon
Hongcheon-gun, Gangwon-do 25108
REPUBLIC OF KOREA
Re: K250999
Trade/Device Name: V8 Diagnostic Ultrasound System; cV8 Diagnostic Ultrasound System; V7 Diagnostic Ultrasound System; cV7 Diagnostic Ultrasound System; V6 Diagnostic Ultrasound System; cV6 Diagnostic Ultrasound System; V5 Diagnostic Ultrasound System; cV5 Diagnostic Ultrasound System; V4 Diagnostic Ultrasound System; cV4 Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, LLZ, QIH
Dated: June 17, 2025
Received: June 17, 2025
Dear Yujin Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250999 - Yujin Kim
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250999 - Yujin Kim
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARJAN NABILI -S for
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250999 | |
| Device Name V8 Diagnostic Ultrasound System; cV8 Diagnostic Ultrasound System; V7 Diagnostic Ultrasound System; cV7 Diagnostic Ultrasound System; V6 Diagnostic Ultrasound System; cV6 Diagnostic Ultrasound System; V5 Diagnostic Ultrasound System; cV5 Diagnostic Ultrasound System; V4 Diagnostic Ultrasound System;cV4 Diagnostic Ultrasound Svstem | |
| Indications for Use (Describe) V8 Diagnostic Ultrasound System; cV8 Diagnostic Ultrasound System; V7 Diagnostic Ultrasound System; cV7 Diagnostic Ultrasound System; V6 Diagnostic Ultrasound System; cV6 Diagnostic Ultrasound System: The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. V5 Diagnostic Ultrasound System; cV5 Diagnostic Ultrasound System; V4 Diagnostic Ultrasound System; cV4 Diagnostic Ultrasound System: The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac), Peripheral vessel and Dermatology. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode, MV-Flow Mode. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
K250999
510(K) Summary:
In accordance with 21 CFR 807.92, the following summary of information is provided:
1. Date Prepared – April 1st, 2025
2. Manufacturer
SAMSUNG MEDISON CO., LTD.
3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108,
Republic of Korea
3. Contact Person
Yujin Kim
Regulatory Affairs Specialist
Phone: +82.2.2194.5486
Fax: +82.2.2194.0273
Email: yujin714.kim@samsung.com
4. Secondary Contact Person
Ninad Gujar
Vice President
Phone: +1.978.564.8632
Fax: +1.978.564.8677
Email: ngujar@neurologica.com
5. Proposed Device
- Common/Usual Name : Diagnostic Ultrasound System and Accessories
- Proprietary Name : V8 Diagnostic Ultrasound System
cV8 Diagnostic Ultrasound System
V7 Diagnostic Ultrasound System
cV7 Diagnostic Ultrasound System
V6 Diagnostic Ultrasound System
cV6 Diagnostic Ultrasound System
V5 Diagnostic Ultrasound System
cV5 Diagnostic Ultrasound System
V4 Diagnostic Ultrasound System
cV4 Diagnostic Ultrasound System
- Regulation Name : Ultrasonic pulsed doppler imaging system
- Regulatory Class : Class II
- Product Code : IYN, IYO, ITX, LLZ, QIH
- Regulation Number : 21 CFR 892.1550, 892.1560, 892.1570, 892.2050
6. Predicate / Reference Device
- V8/cV8, V7/cV7, V6/cV6 Diagnostic Ultrasound System (K243702) – Predicate
- V5/H5/XV5/XH5, V4/H4/XV4/XH4 (K242511) - Reference
{5}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
7. Device Description
- V8/cV8, V7/cV7, V6/cV6
The V8/cV8, V7/cV7, V6/cV6 Diagnostic Ultrasound Systems are general purpose, mobile, software controlled, diagnostic ultrasound systems. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. The V8/cV8, V7/cV7, V6/cV6 also give the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V8/cV8, V7/cV7, V6/cV6 have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
- V5/cV5, V4/cV4
The V5/cV5, V4/cV4 Diagnostic Ultrasound Systems are general purpose, mobile, software controlled, diagnostic ultrasound systems. Its function is to acquire ultrasound data and to display the data as B-mode, M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S-Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes. The V5/cV5, V4/cV4 diagnostic ultrasound systems also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V5/cV5, V4/cV4 diagnostic ultrasound systems have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
8. Intended Use
- V8/cV8, V7/cV7, V6/cV6
The diagnostic ultrasound systems and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
{6}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- V5/cV5, V4/cV4
The diagnostic ultrasound systems and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac), Peripheral vessel and Dermatology.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode, MV-Flow Mode.
9. Technological Comparison to Predicate Device
The V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 employ the same fundamental scientific technology as its predicate device V8/cV8, V7/cV7, V6/cV6 (K243702).
10. Determination of Substantial Equivalence
Comparison to Predicate: The V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 are substantially equivalent to the predicate device with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
- V8/cV8, V7/cV7, V6/cV6
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The proposed V8/cV8, V7/cV7, V6/cV6 and the predicate V8/cV8, V7/cV7, V6/cV6 (K243702) have the same intended use, and modes of operation.
- The proposed V8/cV8, V7/cV7, V6/cV6 have updated ‘BiometryAssist’, ‘ViewAssist’, and ‘HeartAssist’, the cleared features in the predicate V8/cV8, V7/cV7, V6/cV6 (K243702). The AI models for these features have been updated.
- The proposed V8/cV8, V7/cV7, V6/cV6 have included ‘EzAssist’. This feature provides on-screen reference information on human anatomy about ultrasound scanning. It is designed for educational purposes, and it is not a real-time guidance function and is not intended for diagnostic purpose.
- The proposed V8/cV8, V7/cV7, V6/cV6 have included ‘EzTrainer’. This feature provides users with a quick manual containing general information about the system and not intended for diagnostic purpose.
- The proposed V8/cV8, V7/cV7, V6/cV6 have included a new transducer, LM2-18D.
- The proposed V8/cV8, V7/cV7, V6/cV6 have included two caster brake options.
{7}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- The proposed V8/cV8, V7/cV7, V6/cV6 and the predicate V8/cV8, V7/cV7, V6/cV6 (K243702) have the same capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The proposed V8/cV8, V7/cV7, V6/cV6 and the predicate V8/cV8, V7/cV7, V6/cV6 (K243702) have been designed in compliance with approved electrical and physical safety standards.
- The system is manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The system has acoustic power levels which are below the applicable FDA level.
■ V5/cV5, V4/cV4
- The proposed V5/cV5, V4/cV4 and the predicate V8/cV8, V7/cV7, V6/cV6 (K243702) have the same intended use, and modes of operation.
- The proposed V5/cV5, V4/cV4 have included a clinical application for Transesophageal (Cardiac) already cleared in V8/cV8, V7/cV7, V6/cV6 (K243702) by adding the TEE transducer.
- The proposed V5/cV5, V4/cV4 have updated ‘BiometryAssist’, ‘ViewAssist’, and ‘HeartAssist’, the cleared features in the reference device V5/H5/XV5/XH5, V4/H4/XV4/XH4 (K242511). The AI models for these features have been updated.
- The proposed V5/cV5, V4/cV4 have included ‘EzAssist’. This feature provides on-screen reference information on human anatomy about ultrasound scanning. It is designed for educational purposes, and it is not a real-time guidance function and is not intended for diagnostic purpose.
- The proposed V5/cV5, V4/cV4 have included ‘EzTrainer’. This feature provides users with a quick manual containing general information about the system and not intended for diagnostic purpose.
- The proposed V5/cV5, V4/cV4 have included ‘EzNerveMeasure’ previously cleared in the predicate V8/cV8, V7/cV7, V6/cV6 (K243702), as a sub-function of NerveTrack based on AI technology.
- The proposed V5/cV5, V4/cV4 have included ‘RFA Viewer’, the cleared function in the predicate V8/cV8, V7/cV7, V6/cV6 (K243702). This function displays information in real-time, such as total ablation time and total energy, generated by the RFA Generator.
- The proposed V5/cV5, V4/cV4 systems have updated the 'Strain+' function, which
{8}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
was previously cleared in the predicate, by expanding its measurement items to include the left atrium (LA) and right ventricle (RV). This change has already been cleared in the predicate device V8/cV8, V7/cV7, V6/cV6 (K243702) and it has been migrated to V5/cV5, V4/cV4.
- The proposed V5/cV5, V4/cV4 have updated ‘AutoEF’ function, which was previously cleared in the predicate, by expanding its measurement item to include ‘Global Longitudinal Strain(GLS)’. This change has already been cleared in the predicate device V8/cV8, V7/cV7, V6/cV6 (K243702), and it has been migrated to V5/cV5, V4/cV4.
- The proposed V5/cV5, V4/cV4 have included the five transducers: PA3-9B, miniER7, MMPT3-7, TA2-9, LA3-16AD.
- For marketing purpose, we’ve included the cV series product names(cV5 and cV4), which have the same specifications with V series.
- The proposed V5/cV5, V4/cV4 and the reference device V5/H5/XV5/XH5, V4/H4/XV4/XH4 (K242511) have the same capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The proposed V5/cV5, V4/cV4 and the reference device V5/H5/XV5/XH5, V4/H4/XV4/XH4 (K242511) have been designed in compliance with approved electrical and physical safety standards.
- The system is manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The system has acoustic power levels which are below the applicable FDA level.
## 11. Summary of Non-Clinical Testing
The device has been evaluated for acoustic output, biocompatibility, software function, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable FDA guidances and medical device safety standards. The V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 and its applications comply with the following FDA-recognized standards.
| Reference No. | Title |
| --- | --- |
| IEC 60601-1 | IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 |
| IEC 60601-1-2 | IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020 |
| IEC 60601-2-37 | IEC 60601-2-37 (ed.2), am1 for use in conjunction with IEC60601-1 (ed.3), am1 with Corr1 and Corr2 |
| IEC 60601-4-2 | IEC TS 60601-4-2:2024 |
| ISO10993-1 | ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process |
{9}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
| Reference No. | Title |
| --- | --- |
| ISO14971 | ISO 14971:2019, Medical devices - Application of risk management to medical devices |
| NEMA UD 2-2004 | NEMA UD 2-2004 (R2009) Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3 |
All software tests met the acceptance criteria for verification and validation testing.
The AI summaries for ‘BiometryAssist’, ‘ViewAssist’ and ‘HeartAssist’ are applicable to V8/cV8, V7/cV7, V6/cV6 and V5/cV5, V4/cV4 diagnostic ultrasound systems.
[The AI Summary of Testing for BiometryAssist]
☐ Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance. We tested on two areas: Segmentation and Size measurement.
■ Segmentation test
- A deep learning based view recognition algorithm was validated using 320 fetal biometry images collected at hospitals (South Korea and United States).
- The average dice-score is 0.869 (threshold 0.8)
■ Size measurement test
- We use same datasets of segmentation test.
- The error rate of circumference measured value is 8% or less.
- The error rate of distance measured value is 4% or less.
- The error rate of NT measured value is 1mm or less.
☐ The number of individual patients, images were collected from:
- A total of 52 individual patients contributed to the validation dataset (17 from South Korea, 35 from the US).
- From each of these 52 patients, multiple static images were randomly selected and used for the validation, resulting in a total of 320 images.
☐ The number of samples, if different from above, and the relationship between the two:
- Each individual contributed at least one static image across all views.
- Validation dataset includes a total of 320 static images.
☐ Demographic distribution:
- Gender: Female
- Age: Reproductive age, specific age not collected
- BMI: Normal (~24.9), Overweight (25~29), Obese (30~) categories.
- Ethnicity/Country: United States and South Korea
☐ Information about clinical subgroups and confounders present in the dataset:
- We divided the fetal ultrasound images, depending on the ISUOG and AIUM guidelines, into 8 views.
{10}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- BMI: Distributed across standard range, overweight or obesity categories.
- Gestational age: Weeks from 11 - 38, distributed across 1st trimester, 2nd trimester or 3rd trimester categories
☐ Information about equipment and protocols used to collect images
- We acquired the data set with Samsung ultrasound systems (HERA W10, HERA W9, WS80A, V series (V8/V7/V6/V5/V4)) in order to secure diversity of the data set: Mix of data from retrospective data collection and prospective data collection in clinical practice.
☐ Information about how the reference standard was derived from the dataset (i.e. the “Truthing” process):
- All images used for training and tuning were first classified into the correct ultrasound views by three participating experts. Following view classification, the corresponding anatomical structures were manually annotated for each image.
- For validation, the ground truth was established using the target ultrasound device. Expert reviewers manually performed view classification and measurements following the same protocol as used during actual device operation. This ensured that the reference standard closely reflects the clinical use environment and output behavior of the device.
- The participating experts included one obstetrician with over 20 years of experience in fetal cardiology and two sonographers, each with more than 10 years of experience. The entire process was supervised by a senior obstetrician with over 25 years of clinical experience, who reviewed and corrected all annotations to ensure consistency and accuracy.
☐ Description of how the independence of test data from training data was ensured:
- Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three.
[The AI Summary of Testing for ViewAssist]
☐ Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
We tested on two areas: View recognition and Segmentation.
- View recognition test
- A deep learning based view recognition algorithm was validated using 680 fetal biometry images collected at hospitals (South Korea and United States).
- The achieved sensitivity is 93.97% and specificity is 99.62% (thresholds: 75.9%, 88.2%, respectively)
- Segmentation test
- We use same datasets of view recognition test.
- The average dice-score is 0.863 (threshold 0.8)
☐ The number of individual patients, images were collected from:
{11}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- A total of 102 individual patients contributed to the validation dataset (42 from South Korea, 60 from the US).
- From each of these 102 patients, multiple static images were randomly selected and used for the validation, resulting in a total of 680 images.
☑ The number of samples, if different from above, and the relationship between the two:
- Each individual contributed at least one static image across all views.
- Validation dataset includes a total of 680 static images.
☑ Demographic distribution:
- Gender: Female
- Age: Reproductive age, specific age not collected
- BMI: Normal (~24.9), Overweight (25~29), Obese (30~) categories.
- Ethnicity/Country: United States and South Korea
☑ Information about clinical subgroups and confounders present in the dataset:
- We divided the fetal ultrasound images, depending on the ISUOG and AIUM guidelines, into 17 views.
- BMI: Distributed across standard range, overweight or obesity categories.
- Gestational age: Weeks from 11 - 38, distributed across 1st trimester, 2nd trimester or 3rd trimester categories
☑ Information about equipment and protocols used to collect images
- We acquired the data set with Samsung ultrasound systems (HERA W10, HERA W9, WS80A, V series (V8/V7/V6/V5/V4)) in order to secure diversity of the data set: Mix of data from retrospective data collection and prospective data collection in clinical practice.
☑ Information about how the reference standard was derived from the dataset (i.e. the “Truthing” process):
- All images used for training and tuning were first classified into the correct ultrasound views by three participating experts. Following view classification, the corresponding anatomical structures were manually annotated for each image.
- For validation, the ground truth was established using the target ultrasound device. Expert reviewers manually performed view classification and measurements following the same protocol as used during actual device operation. This ensured that the reference standard closely reflects the clinical use environment and output behavior of the device.
- The participating experts included one obstetrician with over 20 years of experience in fetal cardiology and two sonographers, each with more than 10 years of experience. The entire process was supervised by a senior obstetrician with over 25 years of clinical experience, who reviewed and corrected all annotations to ensure consistency and accuracy.
☑ Description of how the independence of test data from training data was ensured:
{12}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three.
[The AI Summary of Testing for HeartAssist(fetus)]
☑ Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
We tested on three areas: View recognition, Segmentation and Size measurement.
☐ View recognition test
- A deep learning based view recognition algorithm was validated using 280 fetal biometry images collected at hospitals (South Korea and United States).
- The achieved sensitivity is 94.29% and specificity is 99.62% (thresholds: 75.9%, 88.2%, respectively)
☐ Segmentation test
- We use same datasets of view recognition test.
- The average dice-score is 0.865 (threshold 0.8)
☐ Size measurement test
- We use same datasets of segmentation test.
- The error rate of area measured value is 8% or less.
- The error rate of angle measured value is 4% or less.
- The error rate of circumference measured value is 11% or less.
- The error rate of diameter measured value is 11% or less.
☑ The number of individual patients, images were collected from:
- A total of 70 individual patients contributed to the validation dataset (26 from South Korea, 44 from the US).
- From each of these 70 patients, multiple static images were randomly selected and used for the validation, resulting in a total of 280 images.
☑ The number of samples, if different from above, and the relationship between the two:
- Each individual contributed at least one static image across all views.
- Validation dataset includes a total of 280 static images.
☑ Demographic distribution:
- Gender: Female
- Age: Reproductive age, specific age not collected
- BMI: Normal (~24.9), Overweight (25~29), Obese (30~) categories.
- Ethnicity/Country: United States and South Korea
☑ Information about clinical subgroups and confounders present in the dataset:
- We divided the fetal ultrasound images, depending on the ISUOG and AIUM guidelines, into 7 views.
- BMI: Distributed across standard range, overweight or obesity categories.
{13}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
- Gestational age: Weeks from 11 - 38, distributed across 1st trimester, 2nd trimester or 3rd trimester categories
☐ Information about equipment and protocols used to collect images
- We acquired the data set with Samsung ultrasound systems (HERA W10, HERA W9, V series (V8/V7/V6/V5/V4)) in order to secure diversity of the data set: Mix of data from retrospective data collection and prospective data collection in clinical practice.
☐ Information about how the reference standard was derived from the dataset (i.e. the “Truthing” process):
- All images used for training and tuning were first classified into the correct ultrasound views by three participating experts. Following view classification, the corresponding anatomical structures were manually annotated for each image.
- For validation, the ground truth was established using the target ultrasound device. Expert reviewers manually performed view classification and measurements following the same protocol as used during actual device operation. This ensured that the reference standard closely reflects the clinical use environment and output behavior of the device.
- The participating experts included one obstetrician with over 20 years of experience in fetal cardiology and two sonographers, each with more than 10 years of experience. The entire process was supervised by a senior obstetrician with over 25 years of clinical experience, who reviewed and corrected all annotations to ensure consistency and accuracy.
☐ Description of how the independence of test data from training data was ensured:
- Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three.
The AI summary of ‘EzNerveMeasure of NerveTrack’ is applicable to V5/cV5, V4/cV4 diagnostic ultrasound systems.
[ The AI Summary of testing for EzNerveMeasure of NerveTrack ]
☐ Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
We tested on the flattening ratio (FR) and cross-sectional area (CSA) of NerveTrack EzNerveMeasure.
☐ A deep learning-based segmentation algorithm was validated using 200 median nerve images collected at a hospital.
☐ FR and CSA measure test
- The average error rate of FR was 8.05% (95% Confidence Interval: [7.64, 8.46]), and the average error rate of CSA was 13.11% (95% Confidence Interval: [11.83, 14.38]). The standard deviation error rate of FR was 0.87, and the standard deviation error
10
{14}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
rate of CSA was 2.73. Error Rate (%) is calculated as follows: (GT Value - Measured Value) / GT Value x 100.
☐ The number of individual patients, images were collected from:
☐ A total of 20 individual patients contributed to the validation dataset (10 from South Korea, 10 from the US).
☐ From each of these 20 patients, 10 static images were randomly selected and used for the validation, resulting in a total of 200 images.
☐ The number of samples, if different from above, and the relationship between the two:
☐ Each individual contributed at 10 static images. In other words, the 10 static images from a 2D sequence data acquired from an individual were used for validation.
☐ Validation dataset includes a total of 200 static images.
☐ Demographic distribution:
☐ Gender: Female and male
☐ Age: 22–68 (mean: 36.2)
☐ BMI: 16.0–29.0 (mean: 22.75)
☐ Ethnicity/Country: Korean and American
☐ Information about clinical subgroups and confounders present in the dataset:
☐ BMI: Distributed across underweight, standard range or overweight categories.
☐ Gender: Female and male.
☐ Information about equipment and protocols used to collect images
☐ We acquired the dataset using the Samsung ultrasound systems (RS80A and V8) in order to secure diversity of ultrasound images and prospective data in clinical practice was collected.
☐ Information about how the reference standard was derived from the dataset (i.e. the “Truthing” process):
☐ The ground truth (GT) for median nerve locations was established by three anesthesiologists, each with over 10 years of experience. One doctor performed the initial ultrasound scans and drew the GT, which was then verified by the other two doctors.
☐ Disagreements during review were resolved by a senior anesthesiologist with over 10 years of extensive clinical experience in regional anesthesia and ultrasound-guided procedures, considering both the initial annotations and the two review annotations. The clinical evaluation has been conducted to get the clinical
11
{15}
SAMSUNG MEDISON CO., LTD.
510(k) Premarket Notification – Traditional
significance of the developed EzNerveMeasure by doctors having experience in respective fields more than 10 years.
☐ Description of how the independence of test data from training data was ensured:
☐ Data used for training, tuning and validation purpose are completely separated from the ones during training process, and there is no overlap among the three
12. Summary of Clinical Tests
The proposed device V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 diagnostic ultrasound systems did not require clinical studies to demonstrate substantial equivalence.
13. Conclusion
Since the predicate device and the proposed device have a similar intended use and key technological features, the non-clinical data support the safety of the device and demonstrate that the V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4 diagnostic ultrasound systems should perform as intended in the specified use conditions. Therefore, SAMSUNG MEDISON CO., LTD. considers the proposed device to be as safe, as effective, and substantially equivalent in performance to the predicate device that is currently marketed for the same intended use.
- END of 510(k) Summary -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.