K250997 · Anika Therapeutics, Inc. · OWX · Jun 17, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250997
Device Name
Integrity™ Implant
Applicant
Anika Therapeutics, Inc.
Product Code
OWX · General, Plastic Surgery
Decision Date
Jun 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Integrity Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Device Story
Integrity™ Implant is a knitted, porous, resorbable/non-resorbable composite mesh designed for tendon injury management. Composed of Hyaff-11 multifilament fibers and polyethylene terephthalate (PET) fibers. Used in surgical environments (open or arthroscopic) by trained surgeons to provide an augmentation layer over injured tendons. Secured to tendon or tendon/bone via fixation. Acts as a protective scaffold; supports healing in injuries without substantial tissue loss. Provided sterile for single use.
Clinical Evidence
No clinical study conducted. Substantial equivalence supported by bench testing of new dimensions and reliance on biocompatibility and performance data from predicate device K223538.
Technological Characteristics
Knitted porous mesh; composite of resorbable Hyaff-11 multifilament and non-resorbable PET multifilament fibers. Pliable, nonfriable. Dimensions: 25x60mm and 40x60mm. Sterilization: Gamma irradiation. Single-use. Standalone implant.
Indications for Use
Indicated for management and protection of tendon injuries without substantial loss of tendon tissue in patients requiring surgical tendon repair.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 17, 2025
Anika Therapeutics, Inc.
Shajunath Nirupama
Sr. Regulatory Affairs Specialist
32 Wiggins Ave
Bedford, Massachusetts 01730
Re: K250997
Trade/Device Name: Integrity™ Implant
Regulation Number: 21 CFR 878.3300
Regulation Name: Surgical Mesh
Regulatory Class: Class II
Product Code: OWX
Dated: April 21, 2025
Received: April 22, 2025
Dear Shajunath Nirupama:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250997 - Shajunath Nirupama
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250997 - Shajunath Nirupama
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, MS
Assistant Director
DHT6C: Division of Restorative, Repair and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Integrity™ Implant
Page 11 of 37
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250997 | ? |
| Please provide the device trade name(s). | | ? |
| Integrity™ Implant | | |
| Please provide your Indications for Use below. | | ? |
| The Integrity Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
ANIKA
K250997
Traditional 510(k)- Integrity™ Implant
# 510(k) Summary
[As required 21 CFR 807.92]
Date Prepared: June 16, 2025
510(k) Number: K250997
## Submitter Name
Anika Therapeutics, Inc.
32 Wiggins Avenue
Bedford, MA 01730
Establishment No: 3007093114
## Contact Person
Shajunath Nirupama
Sr. Regulatory Affairs Specialist
Phone: 781-457-9230
Email: snirupama@anika.com
## General Information
| General Information of Subject Device | |
| --- | --- |
| Trade Name | Integrity™ Implant |
| Common Name | Tendon Protector |
| 510(k) Submitter | Anika Therapeutics, Inc |
| Regulatory Class | II |
| Classification Name | Mesh, Surgical |
| Regulation | 878.3300 |
| Product Code | OWX |
| Review Panel | Orthopedic |
| Predicate 510(k) Number | Integrity™ Implant- K223538 |
## Reason for Submission
The purpose of this 510(k) is to obtain clearance for new sizes for Integrity Implant with same indication for use as predicate device as stated herein.
## Device Description
The Integrity™ Implant is designed to provide an augmentation layer over an injured tendon. The implant is a knitted porous mesh comprised of a single composite layer of resorbable Hyaff-11 multifilament fibers and non-resorbable polyethylene terephthalate (PET) multifilament fibers. The implant is easy to handle, pliable, nonfriable, and porous in both the dry and hydrated state. The implant is provided sterile, for single use only, in a thermoformed tray with peelable lid and outer polymer packaging.
The subject of the 510(k) is the addition of two new sizes:
510(k) Summary
Page 1 of 3
{5}
ANIKA
K250997
# Traditional 510(k)- Integrity™ Implant
- 25mm x 60mm
- 40mm x 60mm
The device will be used in a surgical environment by a trained surgeon. It will be implanted using a standard open or arthroscopic access surgical procedure. It will be secured to the tendon or tendon/bone by applying fixation.
## Indications for Use
The Integrity Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
## Substantial Equivalence Summary
The key features of the subject and predicate device, Integrity Implant is compared and provided in the Table below:
| Device Name | Integrity Implant (Subject Device) | Integrity Implant (Predicate Device) | Discussion |
| --- | --- | --- | --- |
| 510(K) No | K250997 | K223538 | N/A |
| Manufacturer | Class II | Class II | Same |
| Device Class | Mesh, Surgical | Mesh, Surgical | Same |
| Product Code | OWX | OWX | Same |
| Regulation | 878.3300 | 878.3300 | Same |
| Indications for Use | Management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. | Management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. | Same |
| Material Type | Hyaff + PET | Hyaff +PET | Same |
| Resorbable | Partial | Partial | Same |
| Dimensions/ Sizes | 25 X 60 mm
(2.5 X 6.0 cm)
40 X 60 mm
(4.0 X 6.0 cm) | 20 X 25 mm
(2.0 X 2.5 cm)
25 X 30 mm
(2.5 X 3.0 cm) | Larger |
| Reusable | No | No | Same |
| Packaging | Tray within a peel pouch | Tray within a peel pouch | Same |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Same |
## Performance Data (Bench)
A risk analysis was conducted for the new dimensions/sizes of Integrity Implant and following performance testing completed to demonstrate the substantial equivalence:
- Dimensions/Sizes
- Mesh/Device Thickness
- Mesh Basis Weight/Density
- Device Stiffness
- Mesh Compliance
- Sterility
510(k) Summary
Page 2 of 3
{6}
ANIKA
K250997
# Traditional 510(k)- Integrity™ Implant
All other past verification testing: suture retention, tear resistance, hydration fluid compatibility, bacterial endotoxin, residual solvents, and Hyaff Degree of esterification, conducted for predicate device(K223538) remains applicable for the subject device.
## Performance Data (Animal Testing)
Results of original animal study submitted in 510(k) K223538 is applicable to the subject device since the subject device and predicate device have same indication for use, material composition and sterilization.
## Performance Data (Clinical Testing)
No clinical study was conducted for the Predicate device and clinical study not deemed necessary for subject device.
## Biocompatibility
The subject device and the predicate device have same material, manufacturing process, and sterilization parameters. A biological risk assessment on the subject device confirmed that the new sizes meet all biological endpoints for a tissue/bone contacting implant device for permanent contact duration for its intended use. Therefore, past biocompatibility tests included in 510(k) K223538 are still applicable.
## Conclusions
The new dimensions/sizes of Integrity Implant are considered substantially equivalent to the predicate device, Integrity Implant. The devices have the same indications for use, materials, principle of operation, and characteristics. Testing data also indicated performance characteristics of the proposed dimensions of the Integrity Implant is substantially equivalent to the predicate device. The differences in dimension of the devices do not raise any additional questions of safety or effectiveness.
510(k) Summary
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.