Medblue Skin Prick Test Applicator (AS103, AS108, AS109, AS110, AS111, AS113, AS132, AS134, AS146, AS148, AS160, AS162, AS170, AS172)

K250983 · Asistan Medikal · SCL · Jan 30, 2026 · General Hospital

Device Facts

Record IDK250983
Device NameMedblue Skin Prick Test Applicator (AS103, AS108, AS109, AS110, AS111, AS113, AS132, AS134, AS146, AS148, AS160, AS162, AS170, AS172)
ApplicantAsistan Medikal
Product CodeSCL · General Hospital
Decision DateJan 30, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2

Indications for Use

Used for percutaneous administration of diagnostic allergenic extracts.

Device Story

Medblue Skin Prick Test Applicator is a sterile, single-use, disposable device for percutaneous administration of diagnostic allergenic extracts. Device consists of multiple test heads (1, 8, 10, or 12) made of medical-grade acrylic (ps158N), with one model (AS113) featuring a 301 stainless steel tip. Each test head includes a leg with protruding tines (1, 6, or 9) that utilize capillary action to hold and deliver allergenic material to the skin surface. A stopper on each leg prevents penetration beyond the epidermis. The device is used by clinicians to perform skin hypersensitivity tests; it is applied to the skin surface to create visible indentations without piercing the skin. The applicator is designed to fit into a matching dipwell tray to load allergen extracts. Clinical benefit includes facilitating standardized, efficient allergy testing.

Clinical Evidence

Side-by-side clinical testing compared the Medblue Skin Test Applicator to the predicate (K182582). Primary endpoints included mean wheal diameters and patient pain assessment using the Wong-Baker FACES Pain Rating Scale. Results showed no statistically significant differences in allergen delivery effectiveness or pain levels between the subject and predicate devices.

Technological Characteristics

Materials: Medical grade acrylic (ps158N) and 301 stainless steel (AS113 model). Design: Multi-head applicator with capillary-action tines and depth-limiting stoppers. Sterilization: Ethylene Oxide (ISO 11135-1:2014). Biocompatibility: Conforms to ISO 10993-1, -5, -7, -10, and -11. Packaging: Polyethylene container with Tyvek paper. Connectivity: None (manual device).

Indications for Use

Indicated for percutaneous administration of diagnostic allergenic extracts in individuals suspected of having allergies.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION January 30, 2026 Asistan Medikal % Lee Strong Regulatory Consultant 510K FDA Inc. 156 E Granada Blvd Ormond Beach, Florida 32176 Re: K250983 Trade/Device Name: Medblue Skin Prick Test Applicator (AS103, AS108, AS109, AS110, AS111, AS113, AS132, AS134, AS146, AS148, AS160, AS162, AS170, AS172) Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: SCL Dated: January 30, 2026 Received: January 30, 2026 Dear Lee Strong: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250983 - Lee Strong Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250983 - Lee Strong Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # KYRAN R. GIBSON -S For Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250983 | | | Device Name Medblue Skin Prick Test Applicator (AS103, AS108, AS109, AS110, AS111, AS113, AS132, AS134, AS146, AS148, AS160, AS162, AS170, AS172) | | | Indications for Use (Describe) Used for percutaneous administration of diagnostic allergenic extracts. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} 510k FDA Consulting Medical Device Clearances 156 East Granada Blvd. Ormond Beach, FL 32176 386-506-8711 # K250983 510(k) Summary ## Applicant Asistan Medikal DBA Asistan Medikal Tibbi Analiz Lab. Um. Gida Ins. San. Tic. Ltd. Sti. Havaalani Organize Sanayi Bölgesi Körkün Mahallesi 103. Nolu Cadde No: 6 Gaziantep 27910, Turkey FDA Organization Number: 756893 SBD Number: SBR-2025-000208 Phone: +90 342 360 15 73-74 Contact: Ibrahim Muhittin Tuzcu, muhittintuzcu@asistanmedikal.com ## Consultant/Contact 510K FDA Inc. 156 E. Granada Blvd. Ormond Beach, FL 32176 Phone: (386) 846-6037 Fax: (386) 675-4621 Primary Contact: W. Lee Strong, Regulatory Dept. Mgr (lee@510kfda.com) Secondary Contact: Claude Berthoin (claude@510kfda.com) Date Prepared: January 30, 2026 ## Device Classification Trade name: Medblue Skin Prick Test Applicator (AS103, AS108, AS109, AS110, AS111, AS113, AS132, AS134, AS146, AS148, AS160, AS162, AS170, AS172) Common name: Allergen And Vaccine Delivery Needles Manufacturer: Asistan Medikal, (Gaziantep, Turkey) Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: SCL Predicate Device: Oryum and Ovem Epidermal Deri Skin Prick Test Applicator (K182582) Page 1 of 6 {5} Page 2 of 6 # Predicate Device The subject device claims equivalence to the following legally marketed predicate: 510(k) number: K182582 Clearance date: April 10, 2019 Trade name: Oryum and Ovem Epidermal Deri Skin Prick Test Applicator Common name: Allergen And Vaccine Delivery Needles Manufacturer: Allergy & Applicator Depot, LLC (Chapel Hill, NC, USA) Regulatory Class: Class II Regulation Number: 21 CFR 880.5570 Product Code: SCL # Intended Use / Indications for Use Used for percutaneous administration of diagnostic allergenic extracts. # Device Description The Medblue Skin Prick Test Applicator is a sterile, single use disposable, multiple test head applicator used to administer skin test substances to the surface of the skin. The Medblue Skin Prick Test Applicator is used for the conventional percutaneous application of substance directly onto the surface of the skin of diagnostic allergen extracts for performing skin tests for hypersensitivity reactions in individuals suspected of having allergies. The Medblue Skin Prick Test Applicator is offered in several configurations with 1, 8, 10 or 12 test heads made of medical grade acrylic (ps158N) material arranged in a symmetrical design. Model AS 113 features a metal tip made of 301 stainless steel. All other models have acrylic tips. Each of the test heads have a "leg". At the tip of each leg is an array of protruding test points (tines). Each leg has 1, 6, or 9 tines. The tines utilize capillary action to hold the allergenic material for percutaneous delivery when the applicator is applied to the skin. Each leg has a stopper to prevent going deeper than the epidermis thickness. The test heads are designed to fit into the matching asymmetrical well design of a dipwell tray. The applicator loads the allergen from each well in the tray onto each test head. When properly applied to the skin, the applicator will leave visible indentations in the patient's skin corresponding to the test heads of each applicator. The applicator is not intended to pierce the skin. {6} The Medblue Skin Prick Test Applicator has the following models: | Model | Product | | --- | --- | | AS103 | MEDBLUE ONE FINGER SKIN PRICK TEST APPLICATOR1 (LANCET) | | AS108 | MEDBLUE SKIN PRICK TEST APPLICATOR 8 PLASTIC TIP SINGLE NEEDLE | | AS109 | MEDBLUE SKIN PRICK TEST APPLICATOR 10 PLASTIC TIP SINGLE NEEDLE | | AS110 | MEDBLUE SKIN PRICK TEST APPLICATOR 12 PLASTIC TIP SINGLE NEEDLE | | AS111 | MEDBLUE ONE SKIN PRICK TEST APPLICATOR 1 PÍNLÍ (LANCET) | | AS113 | MEDBLUE SKIN PRICK TEST APPLICATOR 10 METAL TIP SINGLE NEEDLE | | AS132 | MEDBLUE SKIN PRICK TEST APPLICATOR 8 SIX NEEDLE | | AS134 | MEDBLUE SKIN PRICK TEST APPLICATOR 8 NINE NEEDLE | | AS146 | MEDBLUE SKIN PRICK TEST APPLICATOR 10 SIX NEEDLE | | AS148 | MEDBLUE SKIN PRICK TEST APPLICATOR 10 NINE NEEDLE | | AS160 | MEDBLUE SKIN PRICK TEST APPLICATOR 12 SIX NEEDLE | | AS162 | MEDBLUE SKIN PRICK TEST APPLICATOR 12 NINE NEEDLE | | AS170 | MEDBLUE SKIN PRICK TEST APPLICATOR 8 SINGLE NEEDLE PEDIATRIC | | AS172 | MEDBLUE SKIN PRICK TEST APPLICATOR 10 SINGLE NEEDLE PEDIATRIC | # Packaging Medblue Skin Prick Test Applicator is packaged in a polyethylene container with Tyvek paper heat-welded to top edge. ![img-0.jpeg](img-0.jpeg) Figure 1 - Diagrams of device in polyethylene container {7} ![img-1.jpeg](img-1.jpeg) Figure 2 - Photos of Tyvek label on container ![img-2.jpeg](img-2.jpeg) # Comparison of Technological Characteristics with Predicate The following table compares technological and other characteristics of the subject and predicate device. | | Oryum and Ovem Epidermal Skin Prick Test Applicator (K182582) | Medblue Skin Prick Test Applicator (Subject Device) | Comparison | | --- | --- | --- | --- | | Product code | SCL | SCL | Same | | Indications for use | Percutaneous administration of diagnostic allergenic extracts. | Percutaneous administration of diagnostic allergenic extracts. | Same | | Applicator for Allergenic Extracts | YES | YES | Same | | Acrylic Polymer Construction | YES | YES | Same | | SAL | 10-6 | 10-6 | Same | | Sterilization Method | Ethylene Oxide (applicator); AAMI Guidelines | Ethylene Oxide (applicator); AAMI Guidelines | Same | | Use Area | Skin Surface | Skin Surface | Same | | Prick Size | 1.2-1.6mm | 1.10-1.5mm | Different: Subject device has a smaller prick size | | Lancet Intervals | 2.0-2.5cm | 20.0-30.5mm | Different: Subject device has a larger lancet interval | | Packaging Method | Sterilization Pouch | Sterilization Pouch | Same | {8} | | Oryum and Ovem Epidermal Skin Prick Test Applicator (K182582) | Medblue Skin Prick Test Applicator (Subject Device) | Comparison | | --- | --- | --- | --- | | Packaging Materials | PET Plastic | PET Plastic + Tyvek Paper | Different: Subject device uses PET with the additional use of Tyvek paper | | Shelf Life | 3 years | 3 years | Same | Differences between the subject and predicate devices are seen in the shape, size, design of the device and those are in cm and mm measurements of slight difference. The differences of the device do not raise new or different question of safety and effectiveness when compared to the predicate device. A combination of bench and clinical performance testing were performed to support that the differences in technological characteristics were adequately verified and validated. ## Clinical Performance Evaluation Clinical testing of the Medblue Skin Test Applicator and Oryum and Ovem Epidermal Deri Skin Prick Test Applicator (predicate device K182582) was conducted to establish equivalence to the predicate device through side-by-side testing. Mean wheal diameters were compared showing no statistically significant differences in effectiveness of allergen delivery between subject and predicate devices. Pain assessment, using the Wong-Baker FACES Pain Rating Scale, showed no significant differences between Medblue and Oryum & Ovem applicators. It was observed that both devices were able to administer the allergenic extract percutaneously to elicit comparable reactions. ## Biocompatibility The device was tested for conformance to the following standards: ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity ISO 10993-7:2008/Amd 1:2019 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals Amendment 1: Applicability of allowable limits for neonates and infants ISO 10993-10:2010 Tests for irritation and skin sensitization ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity ## Sterilization The device conforms to the following standards for Ethylene Oxide Sterilization: ISO 11135-1:2014 Sterilization of health-care products. Ethylene oxide. Requirements for the development, validation and routine control of a sterilization process for medical devices Page 5 of 6 {9} ISO 11737-1:2018/AMD 1:2021 Sterilization of health care products—Microbiological methods Part 1: Determination of a population of microorganisms on products ISO 11737-2:2019 Sterilization of health care products – Microbiological Methods Part 2: Tests of Sterility Performed in the definition, validation and maintenance of a sterilization process Package integrity validation was confirmed after shelf-life and simulated shipping. ## Conclusion In conclusion, the comparison of intended use and technological characteristics, combined with the subject device testing, supports that the subject Medblue Skin Prick Test Applicator raises no new or different questions of safety or effectiveness, and is substantially equivalent to the predicate Oryum and Ovem Epidermal Deri Skin Prick Test Applicator. Page 6 of 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...