K250964 · Jiangsu Jumao X-Care Medical Equipment Co., Ltd. · JCX · May 11, 2026 · General, Plastic Surgery
Device Facts
Record ID
K250964
Device Name
Protable Phlegm Suction Unit (JMA D01)
Applicant
Jiangsu Jumao X-Care Medical Equipment Co., Ltd.
Product Code
JCX · General, Plastic Surgery
Decision Date
May 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is to be used to remove fluids from the airway or respiratory system. The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection canister. The fluids are trapped in the collection canister for proper disposal. It is for use on the order of a physician only.
Device Story
Portable, AC-powered medical suction device (JMA D01); utilizes oil-less piston pump to generate negative pressure (vacuum); draws airway/respiratory fluids through disposable tubing into 800 mL collection canister; canister includes overflow protection and microbial filter. Used in field, transport, and hospital/clinical settings; operated by healthcare professionals. Output is physical removal of fluids; vacuum level monitored via analog gauge; suction intensity adjusted via manual knob. Benefits patient by clearing airway obstructions.
Clinical Evidence
No clinical data. Bench testing only. Device validated against IEC 60601-1-2 (EMC), ANSI/AAMI ES 60601-1 (Electrical Safety), and ISO 10079-1 (Suction Equipment Safety).
Technological Characteristics
AC-powered (120V) oil-less piston pump; plastic casing; analog vacuum gauge; 800 mL disposable canister with microbial filter and overflow protection. Dimensions: 355x182x250mm; Weight: 3.7kg; IP22 rating. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10079-1, IEC 60529.
Indications for Use
Indicated for removal of fluids from the airway or respiratory system in patients requiring suctioning. For use on the order of a physician only.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 11, 2026
Jiangsu Jumao X-Care Medical Equipment Co., Ltd.
Yao Yu
General Manager
No.36 Danyan Road
Danyang, Jiangsu 212300
China
Re: K250964
Trade/Device Name: Protable Phlegm Suction Unit (JMA D01)
Regulation Number: 21 CFR 878.4780
Regulation Name: Powered Suction Pump
Regulatory Class: Class II
Product Code: JCX
Dated: March 30, 2026
Received: March 30, 2026
Dear Yao Yu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250964 - Yao Yu
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250964 - Yao Yu
Page 3
Sincerely,
Colin K. Chen
Digitally signed by
Colin K. Chen
Date: 2026.05.11
15:45:12 -04'00'
Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250964 | |
| Device Name Protable Phlegm Suction Unit (JMA D01) | |
| Indications for Use (Describe) The device is to be used to remove fluids from the airway or respiratory system. The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection canister. The fluids are trapped in the collection canister for proper disposal. It is for use on the order of a physician only. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250964
Prepared on:2026-04-29
# 510(k) Summary
## 1. Submitter's Identifications
Submitter's Name: Jiangsu Jumao X-Care Medical Equipment Co., Ltd.
Address: No.36 Danyan Road, Danyang, Jiangsu 212300, China.
Contact Person: Yu Yao
Contact Title: General Manager
Contact Email Address: 541640111@qq.com
Tel: Tel: +86-511-86197666
Fax: +86-511-86197033
## 2. Correspondent Identifications
Correspondent: Guangzhou Junyi Information Technology Co., Ltd.
Address: Room 304, Building A, No. 62 Nanyun 2nd Road, Science Town, Huangpu District, Guangzhou City, Guangdong, 510663, China
Official Contact Person: Shanfeng Jiang
Title: Regulation Control Manager
ZIP Code: 510663
Phone number: +86-20-82329549
Fax number: +86-20-82329549
E-mail: jiang13620586569@126.com
## 3. Name of the Device
Device Classification Name: Apparatus, Suction, Ward use, Portable, Ac-Powered
Regulation Description: Powered Suction Pump
Trade Name: Protable Phlegm Suction Unit
Model: JMA D01.
Classification Panel: General & Plastic Surgery
Product Code: JCX
Regulation Number: 21 CFR 878.4780
Device Classification: Class II
## 4. The Predicate Devices
Primary Predicate device: K182950 Acara Suction Unit
## 5. Device Description
The Protable Phlegm Suction Unit, model JMA D01 is to be used to remove fluids from the airway or respiratory system.
The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection canister.
{5}
K250964
Prepared on:2026-04-29
The fluids are trapped in the collection canister for proper disposal. The device is a rugged, compact and portable electrically operated medical suction device, designed for multi-source powering from AC (External 120 V) mains supply and is ideally suited for field, transport and hospital/clinical use applications.
The Acare canister is an accessory to the Acare Pump and the removed body fluids are collected in it.
The Acare Pump is an AC powered oil-less positive displacement pump capable of delivering up to 600 mmHg of vacuum to draw fluid and small particles.
The Acare Pump has a plastic casing fitted with a power (ON/OFF) button, power indicator, one suction adjustment knob, a suction gauge and a canister holder.
The Acare canister is a preassembled 800 mL canister including canister tubing fitted with a microbial filter at the distal end.
The canister is fitted with a removable lid for fluid containment and an overflow protection device.
The Acare canister should be placed in the in-built canister holder and connect the Acare canister with the Acare Pump via the connectors of the canister tubing prior to use.
## 6. Indications for Use
The device is to be used to remove fluids from the airway or respiratory system. The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection canister. The fluids are trapped in the collection canister for proper disposal. It is for use on the order of a physician only.
## 7. Technology Comparison and Substantial Equivalence Discussion:
| | Proposed device | Primary predicate device | Comparison |
| --- | --- | --- | --- |
| 510k Number | K259064 | K182950 | |
| Product Code | JCX | JCX | Same |
| Proprietary Name | Protable Phlegm Suction Unit | Acare Suction Unit | |
| Model | JMA D01 | ASU-200 | |
| Manufacturer | Jiangsu Jumao X-Care Medical Equipment Co., Ltd. | ACARE TECHNOLOGY CO., LTD. | |
| Regulation No. | 21 CFR 878.4780 | 21 CFR 878.4780 | Same |
| Classification Name | Apparatus, Suction, Ward Use, Portable, AC-Powered | Apparatus, Suction, Ward Use, Portable, AC-Powered | Same |
| Class | Class II | Class II | Same |
{6}
K250964
Prepared on:2026-04-29
| Indications for use | The device is to be used to remove fluids from the airway or respiratory system. The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection canister. The fluids are trapped in the collection canister for proper disposal. It is for use on the order of a physician only. | The device is to be used to remove fluids from the airway or respiratory system. The device creates a negative pressure (vacuum) that draws fluids through disposable tubing that is connected to a collection canister. The fluids are trapped in the collection canister for proper disposal. It is for use on the order of a physician only. | Same |
| --- | --- | --- | --- |
| Product Type | AC-Powered suction pump | AC-Powered suction pump | Same |
| Electrical requirement | AC 120V, 60Hz | AC 100-240V, 50/60Hz DC 18V | Similar |
| Power Consumption | 135 VA | 35 VA | Different |
| Indication of Vacuum Level | Analogue Vacuum Gauge/760mmHg/100kPa | Analogue Vacuum Gauge/760mmHg/100kPa | Same |
| Maximum Vacuum (Pressure) | <600mmHg | 600 mmHg | Similar |
| Maximum Flow Rate | 24 LPM | 24 LPM | Same |
| Sound Level (Noise level) | Pass ISO 10079-1 | Pass ISO 10079-1 | Same |
| Type of Pump | Oil-less Piston pump | Oil-less Piston pump | Same |
| Pressure Control | adjusted by tuning the knob of vacuum adjusting set | adjusted by tuning the knob of vacuum adjusting set | Same |
| Disposable Canister | Yes | Yes | Same |
| Connection tube | Yes | Yes | Same |
| Patient Tube | Yes | Yes | Same |
| Overflow protection | Yes | Yes | Same |
| Bacterial filter | Yes | Yes | Same |
| Operating Environment | Temp:40°F~100°F (5°C-40°C) Humidity: ≤ 80%RH | Temp: 50°F~104°F (10°C ~40°C) Humidity:10%-90% RH | Similar |
| Operating Atmospheric | 86KPa~106KPa | 700~1060hPa | Similar |
| Storage Environment | Temp: -20°F~128°F(-29°C-55°C) Humidity:10-90%RH | Temp: 59°F~122°F (15°C ~50°C), Humidity: 10% ~ 90% | Similar |
{7}
K250964
Prepared on:2026-04-29
| | | RH | |
| --- | --- | --- | --- |
| Dimensions | (L)355x(W)182x(H)250mm | 19.0cm x 11.0cmx 17.5cm | Different |
| Weight | 3.7 kg (2.52 lb) | 1.7 kg | Different |
| Electromagnetic compatibility (EMC) IEC 60601-1-2 | Passed | Passed | Same |
| Basic safety and Essential performance IEC 60601-1 (ANSI/AAMI ES 60601-1) | Passed | Passed | Same |
| Electrically powered suction equipment - Safety requirements ISO 10079-1 | Passed | Passed | Same |
| Degrees of Protection provided by enclosures (IP Code) IEC 60529 | IP 22 | IP 21 | Different |
| Substantially Equivalent (SE) Discussions | The proposed device Protable Phlegm Suction Unit products has the same purpose as the predicate device: product code, Regulation No., Classification Name , indications for use, Product Type, Electrical requirement, Indication of Vacuum Level, Maximum Vacuum(Pressure), Maximum Flow Rate , Sound Level(Noise level), Type of Pump, Pressure Control, Disposable Canister, Connection tube, Patient Tube, Overflow protection, Bacterial filter, Operating Environment , Operating Atmospheric, Storage Environment, Compliance with EMC standard IEC 60601-1-2 / Electrical Safety standard IEC 60601-1 and ISO10079-1 for Electrically powered suction equipment -Safety requirements. The difference only exists in such contents: Power Consumption, Dimensions, Weight and Degrees of Protection provided by enclosures (IP Code) IEC 60529. The differences between the proposed and predicate devices do not cause new safety and effectiveness problems. According to the non-clinical test results, the proposed device is as safe, effective and has good performance as the predicate device. The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. | | |
{8}
K250964
Prepared on:2026-04-29
## 8. Non-Clinical Tests
The Protable Phlegm Suction Unit, model JMA D01 has the same intended use and indication for use as the predicate devices. The subject device was compared to the predicate device by testing the electrical specifications, vacuum range, flow rate, power options and operation environment. The results of the Protable Phlegm Suction Unit, model JMA D01 validation studies demonstrate that the Suction Unit perform as intended. The results are summarized as follows:
- The electromagnetic compatibility (EMC) testing performed as described in IEC 60601-1-2 Edition 4.0 2014-02, “Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests”. The testing results support the device EMC with the intended use environment.
- The protection against electrical hazards from equipment, Protection against excessive temperature and other hazards, hazards situations and fault conditions, construction of equipment and electromagnetic compatibility of equipment and systems, etc. testing performed as described in ANSI/AAMI ES 60601-1:2005+AMD1:2012 + AMD2:2020, “Medical electrical equipment – Part 1: General requirements for basic safety and essential performance”. The testing results demonstrated the Protable Phlegm Suction Unit, model JMA D01 met safety and essential performance testing acceptance criteria for each test mentioned above.
- The Measurement of the patient leakage current, Measurement of the vibration and noise of equipment other than the low vacuum equipment, Liquid overflow/ spillage test, Air Leakage test, Interruption of the power supply, Abnormal operation and fault conditions, Environmental tests for equipment for field and/or transport use, Degree of collapse test, Airflow and Vacuum tests, and Resistance to implosion, etc. testing performed as described in ISO 10079-1:2015, “Medical suction equipment Part 1: Electrically powered suction equipment — Safety requirements”. The testing results demonstrated the Protable Phlegm Suction Unit, model JMA D01 met Safety requirements testing acceptance criteria for each test mentioned above.
- The classification of degrees of protection provided by enclosures for electrical equipment testing performed as described in IEC60601-1 and IEC 60529-2013, “Degrees of protection provided by enclosures (IP Code)”. The testing results demonstrate of the Protable Phlegm Suction Unit, model JMA D01 met IP 22 testing acceptance criteria for each test mentioned above.
## 8. Conclusions drawn from clinical and non-clinical tests submitted:
The nonclinical tests demonstrate that the proposed device is as safe and effective as the predicate device K182950.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.