It is intended to be used with vein puncture needle and vacuum blood collection tube for multiple collections of venous blood. It is for adult use only.
Device Story
Single-use, sterile blood collection tube holder; assists in multiple venous blood collections. Components include puncture needle hub, puncture needle, rubber boot, and holder. Used in medical institutions by trained healthcare professionals. Device facilitates connection between venipuncture needle and vacuum blood collection tube; rubber boot protects against accidental needle injury and blood splatter. Provides mechanical support for blood sampling; does not involve electronic processing or clinical decision support. Benefits include standardized, safe, and efficient venous blood collection for adult patients.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included puncture force, stiffness, resistance to breakage/corrosion, bond strength, tensile strength, and rubber leakage. Biocompatibility assessed per ISO 10993-1. Sterilization validated per ISO 11135; shelf-life validated per ASTM F1980-16.
Technological Characteristics
Materials: ABS or PP (hub), stainless steel (needle), polyisoprene (rubber boot), PP (holder). Complies with ISO 10993, ISO 7864, ISO 9626, ISO 80369-7. EO sterilization. Single-use, sterile, externally communicating (blood path, indirect) device.
Indications for Use
Indicated for adult patients requiring venous blood collection via vacuum blood collection tubes and venipuncture needles.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 4, 2025
Promisemed Hangzhou Meditech Co., Ltd.
Zearou Yang
Regulatory affairs manager
No. 1388 Cangxing Street,
Cangqian Community, Yuhang District
Hangzhou City, 311121
China
Re: K250961
Trade/Device Name: Blood collection tube holders
Regulation Number: 21 CFR 862.1675
Regulation Name: Blood Specimen Collection Device
Regulatory Class: Class II
Product Code: JKA
Dated: March 28, 2025
Received: March 31, 2025
Dear Zearou Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K250961 - Zearou Yang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250961 - Zearou Yang
Page 3
Sincerely,
David Wolloscheck -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices and Human Factors
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250961 | |
| Device Name Blood collection tube holders | |
| Indications for Use (Describe) It is intended to be used with vein puncture needle and vacuum blood collection tube for multiple collections of venous blood. It is for adult use only. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Promisemed
Promisemed Hangzhou Meditech Co., Ltd.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District,
Hangzhou City, 311121 Zhejiang, P. R. China.
# K250961-510(k) Summary
Date prepared: 2025-06-02
| 1. Manufacturer [21 CFR 807.92 (a) (1)] | |
| --- | --- |
| Name | Promisemed Hangzhou Meditech Co., Ltd. |
| Address | No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, China. |
| Contact Person | Zearou Yang |
| Phone | 86 571 88772985 |
| Fax | 86 571 88772985 |
| Email | zearou.yang@promisemed.ca |
| 2. Device [21 CFR 807.92 (a) (2)] | |
| Name | Blood collection tube holders |
| Common Name | Blood Collection Tubes, Vials, Systems, Serum Separators |
| Classification Name | Blood Specimen Collection Device |
| Regulation Number | 862.1675 |
| Class | Class II |
| Product Code | JKA |
| 3. Legally Marketed Predicate Device [21 CFR 807.92 (a) (3)] | |
| Predicate Name | Luer Adapter with Holder |
| 510(k) Number | K230080 |
| Product Code | JKA |
| Reference Devices | No reference devices were used in this submission. |
| 4. Device Description [21 CFR 807.92 (a) (4)] | |
| Blood collection tube holders is single use and sterile, it is intended to be used with vein puncture needle and vacuum blood collection tube for multiple collections of venous blood. It is sterile and to be used for assisting blood collection. The device consists of a puncture needle hub, puncture needle, rubber boot and holder. It is packaged in a sealed sterility barrier. This is a single-use device and delivered sterile. Sterilization process is validated according to EN ISO 11135. Shelf life is 5 years. | |
| 5. Indication for use [21 CFR 807.92 (a) (5)] | |
| It is intended to be used with vein puncture needle and vacuum blood collection tube for multiple collections of venous blood. It is for adult use only. | |
| 6. Indications for Use Comparison [21 CFR 807.92 (a) (5)] | |
| Other than the difference in the name of predicate product, there are no differences in the indications for use of the subject device when compared to the predicate device. | |
{5}
Promisemed
Promisemed Hangzhou Meditech Co., Ltd.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District,
Hangzhou City, 311121 Zhejiang, P. R. China.
# 7. Comparison of Technological Characteristic [21 CFR 807.92 (a) (6)]
Blood collection tube holders is substantially equivalent to the predicate device, Luer Adapter with Holder, K230080 in that these devices have same intended use and technological characteristics. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single patient use devices. The differences between the subject device and predicate device do not affect the basic design principle, usage of the subject device.
| Items | Subject Device | Predicate Device (K230080) | Comments |
| --- | --- | --- | --- |
| Trade Name | Blood collection tube holders | Luer Adapter with Holder | |
| Manufacturer | Promisemed Hangzhou Meditech Co., Ltd | Yangzhou Medline Industry Co., Ltd. | |
| Intended use environment | Regular medical institutions | Regular medical institutions | Same |
| User | These devices are to be used by properly trained healthcare professionals only in accordance with instructions. | These devices are to be used by properly trained healthcare professionals only in accordance with instructions. | Same |
| Patient population | It is used for adult patients who need to withdraw of blood samples. | It is used for patients who need to withdraw of blood samples. | Same |
| Dimensions | needle guage:20G OD(mm):0.89-0.91; ID(mm):0.66±0.02; Effective length(mm): 14.5±1 holder length:53.3mm | Needle guage:21G,22G,23G,25G length:5/8", 3/4", 1/2", 1", 11/4", 11/2" | Analysis 1 |
| Components | The device mainly consists of puncture needle hub, puncture needle, rubber boot and holder. | The devices are composed of Needle cap, Needle cannula, Safety sheath, Needle hub, Luer adapter and Holder. | Analysis 2 |
| Operating Principle | Blood collection tube holders is intended to be used with vein puncture needle and vacuum blood collection tube for multiple collections of venous blood. The Pre-attached Holder is designed to aid in protection against accidental non-patient needle injury and blood splatter. It is provided sterile and single use. The proposed devices are composed of puncture needle hub, puncture needle, rubber boot and holder. | The Luer adapter with Holder is intended to be used with vein puncture needle and vacuum blood collection tube for multiple collections of venous blood. The Pre-attached Holder is designed to aid in protection against accidental non-patient needle injury and blood splatter. It is provided sterile and single use. The proposed devices are composed of Needle cap, Needle cannula, Safety sheath, Needle hub, Luer adapter and Holder. | Same |
| Reuse durability | Single Use | Single Use | Same |
| Materials | - puncture needle hub (AHA-02/BTH-01): Acrylonitrile Butadiene Styrene (ABS)
- puncture needle hub (AHA-01/AHA-03): Polypropylene (PP)
- puncture needle: Stainless Steel (06Cr19Ni10)
- Rubber boot: Polyisoprene rubber
- Holder: Polypropylene (PP) | - Connect base: ABS
- Non-patient end needle tube: Stainless Steel SUS304
- Rubber sleeve: Gather Isoprene Rubber
- Needle Holder (PP) | Complied with ISO 10993 series standards |
{6}
Promisemed
Promisemed Hangzhou Meditech Co., Ltd.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, P. R. China.
| Performance | Complied with ISO 7864, ISO 9626, ISO 80369-7 | Complied with ISO 7864, ISO 9626, ISO 80369-7 | Same |
| --- | --- | --- | --- |
| Sterilization method | EO Sterilization | EO Sterilization | Same |
Analysis 1- Dimensions
The subject device has different dimensions from predicate device. The subject device scope of dimension is within the scope of predicate device. The subject devices were tested according to ISO 7864:2016 and ISO 9626:2016 standards, and the results met the requirements of the standards. Therefore, this difference does not raise new or different questions of safety and effectiveness.
Analysis 2- Components
The components and configuration between subject device and predicate device is similar, and it's only a difference in the textual description of the component. Therefore, this difference does not raise new or different questions of safety and effectiveness.
8. Nonclinical test [21 CFR 807.92 (b) (1)]
Performance testing was conducted on the following items to support the marketing claims and to confirm that the safety and effectiveness of the subject device is at least equivalent to the predicate device.
- Puncture force
- Stiffness
- Resistance to breakage
- Resistance to corrosion
- Bond between hub and needle tube
- Tensile strength
- Rubber leakage test
- Characteristics of Adapter
The EO residual and ECH residual were measured after sterilization of the device to meet the criteria defined in ISO 11135 Second edition 2014 and AAMI/ANSI/ISO 10993-7:2008(R)2012.
The shelf-life for five years had been validated in accelerated testing according to ASTM F1980-16 (2016) and the requirements on packaging for terminally sterilized medical device per ISO 11607-1 and ISO 11607-2 are also met. The testing successfully demonstrated essential performance is achieved before and after the shelf-life test.
The subject device is categorized as externally communicating medical device (blood path, indirect) and limited exposure (A) per ISO 10093-1:2018. Each biocompatibility endpoint was assessed, e.g. cytotoxicity, sensitization, irritation, acute systemic & pyrogenicity and hemocompatibility.
9. Clinical test [21 CFR 807.92 (b) (2)]
No clinical test is included in this submission.
10. Conclusion [21 CFR 807.92 (b) (3)]
Based on the information provided within these 510(k) submissions, the subject device is substantially equivalent to the predicate device and is as safe, and effective as the legally marketed predicate device.
3 / 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.