Standalone performance testing: 11,186 fetal ultrasound images from 296 patients.
—
Fetal Anatomical Structure Detection
Computer vision AI
—
Sensitivity (across all image qualities): 0.881 to 0.975 depending on structure group; Specificity: 0.927 to 0.991 depending on structure group
—
—
Standalone performance testing: 11,186 fetal ultrasound images from 296 patients.
—
Indications for Use
BioticsAI is intended to analyze fetal ultrasound images and frames (DICOM instances) using machine learning techniques to automatically detect views, detect anatomical structures within the views and to facilitate quality criteria verification and characteristics of the views. The device is intended for use by Healthcare Professionals as a concurrent reading aid during and after the acquisition and interpretation of fetal ultrasound images.
Device Story
BioticsAI is a cloud-based SaaS platform for prenatal ultrasound review; functions as a concurrent reading aid for sonographers and physicians. Input: DICOM instances (still images/frames) from GE ultrasound machines. Operation: AI-driven computer vision models automatically detect fetal anatomical planes, flag high-level anatomical features, and identify specific anatomical structures. Output: Visual overlays (green/red/yellow highlights) indicating detection status and quality criteria verification; automated report generation. Workflow: HCPs review AI findings, override/edit as needed, and export results to PACS via DICOM SR. Benefit: Assists HCPs in ensuring examination completeness and adherence to ISUOG/AIUM protocols; improves consistency in fetal anatomy scan documentation.
Clinical Evidence
Bench testing only. Evaluated on 11,186 fetal ultrasound images from 296 patients (18-44 years, 18-39 weeks gestation). Primary endpoints: sensitivity and specificity for high-level anatomy detection, plane classification, and anatomical structure identification. Results: High-level anatomy sensitivity 0.934 (95% CI: 0.929-0.940), specificity 0.989 (95% CI: 0.988-0.990). Plane classification sensitivity 0.960 (95% CI: 0.955-0.964), specificity 0.997 (95% CI: 0.997-0.998). Subgroup analysis performed for BMI, maternal age, gestational age, ethnicity, and twin pregnancies.
Technological Characteristics
SaaS platform; cloud-based; DICOM integration via DIMSE. Uses computer vision/AI algorithms for image classification and segmentation. Compatible with GE ultrasound systems. No hardware components. Software-based quality verification based on ISUOG/AIUM guidelines.
Indications for Use
Indicated for use by healthcare professionals (sonographers, OB/GYNs, MFMs, fetal surgeons) as a concurrent reading aid for fetal ultrasound images. Patient population: pregnant women aged 18-44 years, gestational age 18-39 weeks. Compatible with GE ultrasound systems.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
BioticsAI, Inc.
% Mary Vater
Director of Regulatory Affairs
Innolitics LLC
1101 West 34th St. #550
Austin, Texas 78705
Re: K250959
Trade/Device Name: BioticsAI
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, QIH
Dated: November 26, 2025
Received: November 26, 2025
Dear Mary Vater:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250959 - Mary Vater
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250959 - Mary Vater
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
BioticsAI
Page 8 of 96
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250959 | ? |
| Please provide the device trade name(s). | | ? |
| BioticsAI | | |
| Please provide your Indications for Use below. | | ? |
| BioticsAI is intended to analyze fetal ultrasound images and frames (DICOM instances) using machine learning techniques to automatically detect views, detect anatomical structures within the views and to facilitate quality criteria verification and characteristics of the views.
The device is intended for use by Healthcare Professionals as a concurrent reading aid during and after the acquisition and interpretation of fetal ultrasound images. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
BioticsAI
510(k) Summary
Page 1 of 12
K250959
# 1. CONTACT INFORMATION
| Company Name | BioticsAI, Inc. |
| --- | --- |
| Address | 455 Market St STE 1940, United States |
| Phone Number | +1.925.320.1532 |
| Company Representative | Mr. Salman Khan |
| Email | salman@biotics.ai |
| Primary Contact | Mrs. Mary Vater |
| Primary Contact Phone Number | +1.913.523.6988 |
| Primary Contact Email | fda@innolitics.com |
| Date Summary Prepared | Dec. 2, 2025 |
# 2. DEVICE INFORMATION
| Trade Name | BioticsAI |
| --- | --- |
| Common Name | Ultrasonic pulsed doppler imaging system |
| Classification Name | System, Imaging, Pulsed Doppler, Ultrasonic |
| Regulation Number | 21 CFR 892.1550 |
| Product Code(s) | IYN, IYO, QIH |
# 3. PREDICATE DEVICE
| Predicate Device Name | Sonio Detect |
| --- | --- |
| Manufacturer | Sonio |
| 510(k) Number | K240406 |
| Product Code | IYN, IYO, QIH |
| Regulation Number | 21 CFR 892.1550 |
{5}
BioticsAI
510(k) Summary
Page 2 of 12
# 4. DEVICE DESCRIPTION
## 4.1. Description
BioticsAI is a software used by OB/GYN care centers for prenatal ultrasound review and reporting. BioticsAI uses artificial intelligence (A.I.) to automatically annotate ultrasound images with fetal anatomical planes and structures to facilitate ultrasound review and report generation for fetal ultrasound anatomical scans. It serves as concurrent reading aid for ultrasound images both during and after a fetal anatomical ultrasound examination.
BioticsAI is a Software as a Service (SaaS) solution that aims at helping sonographers, OB/GYNs, MFMs and Fetal surgeons (all three designated as healthcare professionals i.e. HCP) to perform their routine fetal ultrasound examinations in real-time.
BioticsAI can be used by Healthcare Professionals HCPs during fetal ultrasound exams for Trimester 2 of the fetus, during which a fetal anatomy exam is typically captured (typically conducted between 18-22 weeks but can be captured on gestational ages ranging from 18 up to 39 weeks). The software is intended to assist HCPs in assuring during and after their examination that the examination is complete and all images were collected according to their protocol
BioticsAI requires the following SaaS accessibility from internet browser.
BioticsAI receives DICOM instances, which consist of still fetal ultrasound images (in the form still image captures or individual frames from a multi-frame instance) from the ultrasound machine, which are submitted by the performing healthcare professional from the clinic's network, either during the screening or post-screening and performs the following:
- Automatically detect fetal anatomical planes (2D ultrasound views).
- Automatically flag high-level anatomical features (e.g., "head", "thorax", "limb detected in image", etc).
- Automatically detect specific anatomical structures within supported planes/views (i.e. "cerebellum, csp, and cisterna magna found in transcerebellar plane image").
- Facilitate quality verification of supported planes by determining whether the expected anatomical structures, as informed by the latest ISUOG and AIUM guidelines, are present in the ultrasound image. The quality assessment focuses on the presence or absence of these anatomical structures.
BioticsAI automatically identifies fetal anatomical views and anatomical structures captured during the screening. It uses green highlights to indicate successfully detected planes and structures. Red highlights are used to flag instances where the model could not detect an expected anatomical view or structure, even though it is a supported feature. Yellow highlights indicate views or structures that require manual verification (when the AI cannot determine whether anatomical features are present or absent because it is not yet supported by our product).
The end user can interact with the software to override BioticsAI's outputs. Specifically, users can unassign or assign an image to a plane or high level anatomical feature, and update the status of quality criteria for structures by changing it from "found" to "not found" or vice versa. Users have the flexibility to review and edit these assignments at any point during or after the exam.
The end user then has the ability to include the information gathered during quality and image review automatically in a final report via a button called "Confirm Screening Results", automatically filling out a
{6}
BioticsAI
510(k) Summary
report template with identified planes and structures. The report can then be further exported to the clinic's PACS over DIMSE via a populated DICOM SR.
BioticsAI also provides a standard DICOM Viewer for viewing DICOM instances, and obstetrics ultrasound report templates for manually creating ultrasound reports without the AI based functionality as described above.
To further explain the AI-driven outputs provided by the device, we describe the three primary AI components below:
AI-1: High-Level Anatomy Classification
Provides a multi-label classification of the general anatomical region depicted in the image (e.g., head/brain, face, thorax/chest, abdomen, limbs). These categories correspond to standard high-level anatomy groupings used in fetal ultrasound interpretation.
AI-2: Per-Class Top-1 Fetal Plane Classification
Provides fetal anatomical plane classifications using a per-class Top-1 approach. A fetal "plane" refers to a standardized cross-sectional view defined by ISUOG and aligned with AIUM guidance for mid-trimester fetal anatomy scans. For each anatomical plane category, the model outputs the image with the single highest-confidence prediction (Top-1) associated with that class.
AI-3: Fetal Anatomical Structure Classification
Provides multi-label identification of fetal anatomical structures (e.g., cerebellum, cisterna magna, cerebral peduncles), generated from the model's segmentation head and refined through post-processing filters that enforce plane-structure consistency and remove non-intended labels.
The list of views, fetal planes, and anatomical structures automatically detected by AI-1, AI-2, and AI-3's outputs and verified by the software are detailed in tables 1, 1.1, 2, and 3 below:
Table 1: AI-1 Outputs / High Level Anatomy Classification Label Criteria
| Label | Label Criteria |
| --- | --- |
| Head (also referred to as “Brain”) | One of “Transcerebellar” plane, “Transthalamic” plane, or the “Transventricular” plane |
| Face | One of “Median Facial Profile” plane, “Coronal Plane of Upper Lip, Nose, and Nostrils” plane, or “Orbits, Lenses” plane |
| Thorax/Chest (Also referred to as “Heart Screening Planes”) | One of the “Right Ventricular Outflow Tract” plane, “Three-Vessel View” plane, “Three-Vessel and Trachea View” plane, “Left Ventricular Outflow Tract” plane, or “Thorax Four-Chamber Heart View” plane |
| Abdomen | One of “Bladder Plane”, “Kidneys Plane”, “Stomach Umbilical Plane“, or “Cord Insertion Plane” |
| Limbs | Do one or more of these elements exist: • The tibia is present • The fibula is present |
{7}
BioticsAI
510(k) Summary
Page 4 of 12
| | • The foot is present
• The ankle region (relationship between the tibia, fibula, and foot) is present
• The humerus is present
• The radius is present
• The ulna is present
• The wrist region (the relationship between the hand and the 2 forearm bones is present)
• The hand is present
• The femur is present |
| --- | --- |
Table 1.1 “Heart Screening Planes” Inclusion Categories
| Inclusion Categories |
| --- |
| Heart: Thorax 4 Chamber Heart Plane |
| Heart: 3VT Heart Plane |
| Heart: 3VV Heart Plane |
| Heart: RVOT Plane |
| Heart: LVOT Plane |
Table 2: AI-2 Outputs / Per Class Top-1 Plane Classification Labels
| Labels |
| --- |
| Abdomen: Bladder Plane |
| Abdomen: Cord Insertion Plane |
| Abdomen: Kidneys Plane |
| Abdomen: Stomach Umbilical Vein Plane |
| Face: Coronal Plane of Upper Lip, Nose, and Nostrils |
| Face: Median Facial Profile Plane |
| Face: Orbits, Lenses Plane |
| Head: Transcerebellar Plane |
| Head: Transthalamic Plane |
| Head: Transventricular Plane |
| Heart: 4 Chamber Heart Plane |
| Limbs: Femur Plane |
{8}
BioticsAI
510(k) Summary
Page 5 of 12
Spine: Sagittal Spine Plane
Table 3: AI-3 Outputs / Fetal Anatomical Structures and Regions Label Criteria
| Structure | Views the Structure is found in (hence we only predict the visibility of structures within these views) |
| --- | --- |
| Bladder | Abdomen Bladder |
| Cord Insertion | Abdomen Cord Insertion |
| Kidney | Abdomen Kidneys |
| Abdominal Aorta | Abdomen Stomach Umbilical |
| Inferior Vena Cava | Abdomen Stomach Umbilical |
| Stomach Shadow | Abdomen Stomach Umbilical |
| Floating Ribs | Abdomen Stomach Umbilical Vein |
| Umbilical Vein | Abdomen Stomach Umbilical vein |
| Ossification Center Of The Spine | Abdomen Stomach Umbilical, Abdomen Kidneys, Heart Screening Planes, Thorax 4 Chamber Heart, Spine Sagittal |
| Lower Lip | Face Coronal Upper lip Nose Nostrils, Face Median Facial Profile |
| Upper Lip | Face Coronal Upper lip Nose Nostrils, Face Median Facial Profile |
| Chin | Face Median Facial Profile |
| Frontal Bone | Face Median Facial Profile |
| Maxilla | Face Median Facial Profile |
| Nasal Bone | Face Median Facial Profile |
| Skin Overlying The Forehead | Face Median Facial Profile |
| Nose | Face Median Facial Profile, Face Coronal Upperlip Nose Nostrils |
| Orbits | Face Orbits Lenses |
| Cerebellum | Head Transcerebellar |
| Cerebral Peduncles | Head Transcerebellar |
| Cisterna Magna | Head Transcerebellar |
| Falx Cerebri | Head Transthalamic, Head Transventricular |
| Septum Pellucidum | Head Transthalamic, Head Transventricular |
| Atria | Head Transventricular |
| Choroid Plexus | Head Transventricular |
{9}
BioticsAI
510(k) Summary
Page 6 of 12
| Lateral Ventricle Posterior Horn Walls | Head Transventricular |
| --- | --- |
| Parito Occipital Sulcus | Head Transventricular |
| CSP | Head Transventricular, Head Transthalamic |
| Lateral Sulcus | Head Transventricular, Head Transthalamic |
| Thalami | Head Transventricular, Head Transthalamic |
| Aorta | Heart Screening Planes |
| LVOT | Heart Screening Planes |
| Pulmonary Artery | Heart Screening Planes |
| RVOT | Heart Screening Planes |
| Superior Vena Cava | Heart Screening Planes |
| Thymus Gland | Heart Screening Planes |
| Left Ventricle | Heart Screening Planes, Thorax 4 Chamber Heart |
| Lungs | Heart Screening Planes, Thorax 4 Chamber Heart |
| Right Atrium | Heart Screening Planes, Thorax 4 Chamber Heart |
| Right Ventricle | Heart Screening Planes, Thorax 4 Chamber Heart |
| Septum Primum | Heart Screening Planes, Thorax 4 Chamber Heart |
| Ventricular Septum | Heart Screening Planes, Thorax 4 Chamber Heart |
| Descending Aorta | Heart Screening Planes, Thorax 4 Chamber Heart |
| Left Atrium | Heart Screening Planes, Thorax 4 Chamber Heart |
| Ribs | Heart Screening Planes, Thorax 4 Chamber Heart |
| Femur | Limbs Femur |
| Skin Overlying The Spine | Spine Sagittal |
# 5. SUBJECT DEVICE INDICATIONS FOR USE
BioticsAI is intended to analyze fetal ultrasound images and frames (DICOM instances) using machine learning techniques to automatically detect views, detect anatomical structures within the views and to facilitate quality criteria verification and characteristics of the views.
The device is intended for use by Healthcare Professionals as a concurrent reading aid during and after the acquisition and interpretation of fetal ultrasound images.
{10}
BioticsAI
510(k) Summary
# 6. SUBSTANTIAL EQUIVALENCE COMPARISON
# 6.1. Indications for Use Equivalence Discussion
The indications for use of the devices are nearly identical. Both devices are intended to analyze fetal ultrasound images and clips using machine learning
techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria and characteristics of the views.
The devices are both intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.
# 6.2. Device Comparison
Table 4 provides a comparison of the Technological Characteristics of BioticsAI to the predicate Sonio Detect cleared in K240406.
Table 4: Comparison of Technological Characteristics
| Feature/Function | Proposed Device: BioticsAI | Predicate Device: Sonio Detect (K240406) |
| --- | --- | --- |
| Manufacture Name | BioticsAI, Inc. | Sonio |
| Device Name | BioticsAI | Sonio Detect |
| Regulation Number | 21 CFR 892.1550 - accessory to Ultrasonic Pulsed Doppler Imaging System21 CFR 892.1560 - accessory to Ultrasonic Pulsed Echo Imaging System21 CFR 892.2050 - Medical Image Management and Processing System | 21 CFR 892.1550 - accessory to Ultrasonic Pulsed Doppler Imaging System21 CFR 892.1560 - accessory to Ultrasonic Pulsed Echo Imaging System21 CFR 892.2050 - Medical Image Management and Processing System |
| Product Code | IYN (primary)IYO, QIH (Secondary) | IYN (primary)IYO, QIH (Secondary) |
| Intended Users | Qualified and trained healthcare professional personnel in a professional prenatal ultrasound (US) imaging environment (this includes sonographers, MFMs, OB/GYN, and Fetal surgeons) | Qualified and trained healthcare professional personnel in a professional prenatal ultrasound (US) imaging environment (this includes sonographers, MFMs, OB/GYN, and Fetal surgeons) |
| Features | - BioticsAI automatically detects views-BioticsAI automatically detects anatomical structures within the supported views-BioticsAI provides an interface to the operator that shows BioticsAI's automatic anatomical findings, allowing the operator to verify the quality criteria and characteristics of the supported views. | - Sonio Detect automatically detects views-Sonio Detect automatically detects anatomical structures within the supported views-Sonio Detect automatically verifies the quality criteria and characteristics of the supported views. |
| Algorithm Methodology | Artificial Intelligence (Computer Vision) | Artificial Intelligence (Computer Vision) |
{11}
BioticsAI
510(k) Summary
Page 8 of 12
| Platform | Secure cloud-based software compatible with standard DIMSE integration with ultrasound systems from GE Medical | Secure cloud-based and stand-alone software compatible with ultrasound systems from GE Medical, Samsung, Canon and Philips |
| --- | --- | --- |
| Real Time? | No | No |
| System Compatibility | BioticsAI requires the following:
- Clinical support for exporting DICOM over DIMSE to BioticsAI’s DICOM Adapter.
- SaaS accessibility from an internet browser (recommended browser: Google Chrome). | Sonio Detect requires the following:
- Edge Software (described below) to install on a server on the same network as the Ultrasound Machine;
- SaaS accessibility from any internet browser (recommended browser: Google Chrome). |
BioticsAI and its predicate differs in the following:
- the platform: BioticsAI and its predicate differ in their compatibility with ultrasound machine manufacturers, however, both devices support ultrasound systems from GE Medical.
- BioticsAI can integrate with a standard DIMSE connection that the client provides for integration within the same network as the ultrasound machine. The predicate ships a separate standard alone edge software to install on a server on the same network as an ultrasound machine for connectivity.
However, these differences do not raise new questions regarding safety and effectiveness of the device when used as labeled.
# 7. DISCUSSION OF PERFORMANCE TESTING
The following performance data were provided in support of the substantial equivalence determination.
- Software Verification and Validation Testing
- Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions.”
The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Design Reviews
- Software Development Lifecycle
- Algorithm Verification (Algorithm internal validation)
- Software Verification (including unit, integration, interoperability, UI/UX testing, load testing)
- Simulated use testing (Validation)
- Performance testing
- Cybersecurity testing
# Bench Testing
BioticsAI conducted a standalone performance testing on a dataset of 11,186 fetal ultrasound images in the United States across 296 patients across varying ethnicities, patient BMIs, patient ages, and gestational ages, twin pregnancies, and presence of abnormalities representative of the intended use population from a single site (across multiple ultrasound screening units and machine instances) in the United States. This dataset was independent of the data used during model development (training/fine tuning/internal validation) and establishment of device operating points.
{12}
BioticsAI
510(k) Summary
The results of the standalone performance testing demonstrate that BioticsAI:
Table 5: Results of Standalone Performance Testing for AI-1: High-Level Anatomy Classification
| Items (fetal ultrasound views detected) | Sensitivity | | Specificity | |
| --- | --- | --- | --- | --- |
| | Point Estima te | Bootstrappi ng CI (95%) | Point Estima te | Bootstrappi ng CI (95%) |
| AI-1 High Level Anatomy: Fetal “Abdomen” View | 0.953 | (0.942, 0.962) | 0.986 | (0.984, 0.989) |
| Detection across any of: (“Bladder Plane”, “Kidneys Plane”, “Stomach Umbilical Plane“, “Cord Insertion Plane”) | | | | |
| AI-1 High Level Anatomy: Fetal “Face” View, | 0.944 | (0.932, 0.956) | 0.993 | (0.991, 0.994) |
| Detection across any of: (“Median Facial Profile Plane”, “Coronal Plane of Upper Lip, Nose, and Nostrils Plane”, “Orbits, Lenses Plane”) | | | | |
| AI-1 High Level Anatomy: Fetal “Head” Planes, | 0.955 | (0.946, 0.964) | 0.996 | (0.995, 0.997) |
| Detection across any of: (“Transthalamic Plane”, “Transventricular Plane”, “Transcerebellar Plane”) | | | | |
| AI-1 High Level Anatomy: Automatic detection of the Fetal “Limbs” characterized across, 10 potential visual elements | 0.919 | (0.895, 0.943) | 0.983 | (0.981, 0.985) |
| Detection across any of: -tibia -fibula -foot -ankle region, “relationship between the tibia, fibula, and foot” -humerus -radius -ulna -wrist region, “relationship between the hand, radius, and ulna” -hand -femur | | | | |
{13}
BioticsAI
510(k) Summary
Page 10 of 12
# AI-1 High Level Anatomy:
Automatic detection "Heart Screening" Planes, defined by any of the 5 Fetal Heart Planes
("Right Ventricular Outflow Tract Plane", "Three-Vessel View", "Three-Vessel and Trachea View", "Left Ventricular Outflow Tract Plane", "Four-Chamber Heart Plane")
0.912 (0.895, 0.928) 0.990 (0.988, 0.992)
Table 6: Results of Standalone Performance Testing for AI-2: Per-Class Top-1 Fetal Plane Classification
| Class | Sensitivity Bootstrapping CI (95%) | Specificity Bootstrapping CI (95%) |
| --- | --- | --- |
| abdomen_bladder | 0.960 (0.940, 0.977) | 0.998 (0.997, 0.998) |
| abdomen_cord_insertion | 0.965 (0.947, 0.983) | 0.998 (0.997, 0.999) |
| abdomen_kidneys | 0.953 (0.927, 0.973) | 0.998 (0.997, 0.999) |
| abdomen_stomach_umbilical_vein | 0.990 (0.982, 0.997) | 1.000 (1.000, 1.000) |
| face-coronal_of_upperlip_nose_nostrils | 0.981 (0.968, 0.993) | 0.999 (0.999, 1.000) |
| face_median_facial_profile | 1.000 (1.000, 1.000) | 0.999 (0.998, 1.000) |
| face_orbits_lenses | 0.897 (0.863, 0.927) | 0.999 (0.999, 1.000) |
| head_transcerebellar | 0.998 (0.994, 1.000) | 1.000 (0.999, 1.000) |
| head_transthalamic | 0.923 (0.899, 0.945) | 0.992 (0.991, 0.994) |
| head_transventricular | 0.975 (0.964, 0.984) | 1.000 (1.000, 1.000) |
| limbs_femur | 0.955 (0.944, 0.966) | 0.992 (0.990, 0.994) |
| spine_sagittal | 0.909 (0.891, 0.927) | 0.995 (0.993, 0.996) |
| thorax_lungs_four-heart_chambers | 0.969 (0.954, 0.983) | 0.997 (0.996, 0.998) |
{14}
BioticsAI
510(k) Summary
Table 7: Summary of Results for "AI-1: High-Level Anatomy Classification", "AI-2: Per-Class Top-1 Fetal Plane Classification", and "AI-3: Fetal Anatomical Structure Classification"
| Items (fetal ultrasound views, anatomical structures and characteristics automatically detected) | Sensitivity Across Diagnostically Acceptable Images | | Sensitivity Across all Image Qualities | | Specificity Across all Image Qualities | |
| --- | --- | --- | --- | --- | --- | --- |
| | Point Estimate | Bootstrapping CI (95%) | Point Estimate | Bootstrapping CI (95%) | Point Estimate | Bootstrapping CI (95%) |
| Automatic detection of 5 High-Level Fetal Anatomy Sections (Abdomen, Face, Head, Limbs, Thorax) | - | - | 0.934 | (0.929, 0.94) | 0.989 | (0.988, 0.99) |
| Automatic detection of 13 fetal ultrasound planes (Per-class Top-1 classification) | - | - | 0.960 | (0.955, 0.964) | 0.997 | (0.997, 0.998) |
| Automatic detection of 12 fetal head anatomical structures on the views "Transthalamic", "Transventricular", "Transcerebellar" | 0.948 | (0.935, 0.959) | 0.881 | (0.871, 0.891) | 0.991 | (0.99, 0.992) |
| Automatic detection of 9 fetal abdomen anatomical structures on the views "Bladder", "Kidneys", "Stomach Umbilical Vein", "Cord Insertion" | 0.953 | (0.941, 0.964) | 0.919 | (0.909, 0.93) | 0.983 | (0.982, 0.984) |
| Automatic detection of 9 fetal face anatomical structures on the views "Coronal Plane of Upper Lip, Nose, and Nostriles", "Median Facial Profile", "Orbits, Lenses" | 0.983 | (0.976, 0.989) | 0.958 | (0.951, 0.965) | 0.991 | (0.99, 0.992) |
| Automatic detection of 2 fetal spine | 0.992 | (0.989, 0.996) | 0.975 | (0.97, 0.98) | 0.927 | (0.921, 0.931) |
{15}
BioticsAI
510(k) Summary
Page 12 of 12
| anatomical structures on the views
“Sagittal Spine” | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Automatic detection of 16 fetal thorax and heart anatomical structures on the views
“Four Chamber”,
“LVOT”,
“RVOT”,
“3VV”,
“3VT” | 0.978 | (0.969, 0.985) | 0.925 | (0.911, 0.939) | 0.989 | (0.988, 0.99) |
Additionally, the performance for the detection of high level anatomy, planes/views, and structures was also validated for subgroups including: BMI, maternal age, gestational age, race/ethnicity, twin vs singleton pregnancies, and presence of abnormalities when appropriate.
BioticsAI was validated only with GE Ultrasound devices and is intended only to be used with these Ultrasound vendors. The device was validated on patients aged 18 to 44 years and is not intended for use on patients younger than 18 or older than 44. The device was validated on fetal ultrasound images from 18 to 39 weeks' gestation and is not intended for use on gestational ages below 18 weeks.
The results of verification and performance testing demonstrate the safe and effective use of BioticsAI.
# 8. CONCLUSION
BioticsAI's intended users, clinical outcome and clinical applications are similar to those of the predicate device (K240406).
The technological characteristics differences identified and discussed in Section “Comparison of Technological Characteristics with the Predicate Device” do not raise any different questions of safety and effectiveness of the device.
Furthermore, results of successful verification and validation activities and additional bench performance testing do not raise any new issue regarding the safety and effectiveness of the Device.
Thus, BioticsAI is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.