The DISPOSABLE BIOPSY FORCEPS FB-211D/FB-221D/FB-231D/FB-241D are intended to be used to collect tissue within the tracheobronchial tree in combination with a bronchoscope
Device Story
Disposable biopsy forceps (FB-211D, FB-221D, FB-231D, FB-241D) designed for tissue collection in tracheobronchial tree. Device consists of handle, slider, and insertion portion (sheath, metallic coil, cups). Physician inserts device through bronchoscope channel; manipulates slider to open/close distal cups to bite lesion surface; withdraws device to retrieve tissue sample. Used in clinical settings by physicians. Available in multiple cup configurations: alligator jaw-step (with/without needle) and oval type (with/without needle). Benefits include single-use design to prevent cross-contamination.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including insertion/withdrawal performance, cup operation, connection strength, and visual inspection of components. Biocompatibility testing performed per ISO 10993-1. Sterilization validated per ISO 11137. Shelf-life and packaging integrity validated per ASTM F1980-21 and ISO 11607.
Technological Characteristics
Insertion portion: stainless steel coil with polyethylene sheath. Distal end: alligator jaw-step or oval cups (with/without needle). Sterilization: gamma radiation (SAL 10^-6). Shelf-life: 5 years. Compatible with bronchoscopes having channel diameter >= 2.0 mm and working length <= 600 mm. Mechanical actuation via slider-wire mechanism.
Indications for Use
Indicated for tissue collection within the tracheobronchial tree using a bronchoscope. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Olympus Single Use Biopsy Forceps FB-433D (K172726)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30. 2025
Olympus Medical Systems Corporation
% Susan Lewandowski
Manager, Program Regulatory Affairs
Olympus Surgical Technologies of America
800 West Park Drive
Westborough, Massachusetts 01581
Re: K250957
Trade/Device Name: Disposable Biopsy Forceps (FB-211D); Disposable Biopsy Forceps (FB-221D); Disposable Biopsy Forceps (FB-231D); Disposable Biopsy Forceps (FB-241D)
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (flexible or rigid) and accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: March 30, 2025
Received: March 31, 2025
Dear Susan Lewandowski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250957 - Susan Lewandowski
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250957 - Susan Lewandowski
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Disposable Biopsy Forceps
Page 10 of 34
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250957 | ? |
| Please provide the device trade name(s). | | ? |
| Disposable Biopsy Forceps (FB-211D);
Disposable Biopsy Forceps (FB-221D);
Disposable Biopsy Forceps (FB-231D);
Disposable Biopsy Forceps (FB-241D) | | |
| Please provide your Indications for Use below. | | ? |
| The DISPOSABLE BIOPSY FORCEPS FB-211D/221D/231D/241D are intended to be used to collect tissue within the tracheobronchial tree in combination with a bronchoscope | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
OLYMPUS
Disposable Biopsy Forceps FB-series
# 510(k) Summary
## 1. General Information
Date Prepared: May 27, 2025
510(K) submitter: Olympus Medical Systems Corporation
2951 Ishikawa-cho, Hachioji-shi, Tokyo Japan 192-8507
Contact: Seiko Yunoki
Correspondent: Olympus Surgical Technologies of America
800 West Park Drive, Westborough, MA 01581
Primary Contact: Susan Lewandowski
Email: susan.lewandowski@olympus.com
## 2. Device Information
Device Name: Disposable Biopsy Forceps FB-211D
Disposable Biopsy Forceps FB-221D
Disposable Biopsy Forceps FB-231D
Disposable Biopsy Forceps FB-241D
Common Name: Biopsy Forceps
Classification: 874.4680 Bronchoscope (flexible or rigid) and accessories
Regulatory Class: II
Product Code: EOQ
Device Panel: Ear, Nose, & Throat
## 3. Predicate Device Information
Olympus Single Use Biopsy Forceps FB-433D – K172726
## 4. Device Description
The Disposable Biopsy Forceps FB-211D, FB-221D, FB-231D, and FB-241D (also known as "EndoJaw") have been designed to collect tissue within the tracheobronchial tree in combination with a flexible bronchoscope. The Disposable Biopsy Forceps are inserted into the channel of a bronchoscope to collect tissue by biting the surface of the lesion with a pair of forceps located at the distal end of the device. The Disposable Biopsy Forceps are withdrawn from the channel and the tissue sample is collected.
The Disposable Biopsy Forceps consists of a handle and an insertion portion. The handle consists of a handle body and slider. The insertion portion consists of a sheath, metallic coil
{5}
OLYMPUS
Disposable Biopsy Forceps FB-series
(operating wire), cups, and forceps cap. The slider is connected to the operating wire in the insertion portion and the cups are opened/closed by advancing/retreating the slider with the function of the connection parts. The subject device is available in multiple distal end cup shapes; FB-211D – Alligator jaw-step, FB-221D – Alligator jaw-step with needle, FB-231D – Oval type, and FB-241D – Oval type with needle.
The subject device has the same technological characteristics and similar design as the applicable predicate device.
# 5. Indications for Use
The DISPOSABLE BIOPSY FORCEPS FB-211D/FB-221D/FB-231D/FB-241D are intended to be used to collect tissue within the tracheobronchial tree in combination with a bronchoscope.
# 6. Predicate Comparison
| Description | Subject Device (SD)
Single Use Biopsy Forceps
FB-211D, FB-221D, FB-231D, and FB-241D | Predicate Device (PD)
Single Use Biopsy Forceps
FB-433D
(K172726) | Comparison |
| --- | --- | --- | --- |
| Design of Insertion portion of Forceps | The insertion portion is a stainless steel coil covered by a polyethylene sheath. | The insertion portion is a stainless steel coil. | DIFFERENT
Non-clinical bench testing was performed to demonstrate that this difference in technology does not impact the effectiveness of the device. |
| Cup Shape | FB-211D
Alligator Jaw-step
Manipulation wire
FB-221D
Alligator jaw-step with needle
Manipulation wire
FB-231D
Oval type
Manipulation wire
FB-241D | Standard (Oval type),
Fenestrated
Fenestrated | SIMILAR
Non-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device. |
{6}
OLYMPUS
Disposable Biopsy Forceps FB-series
| Description | Subject Device (SD)
Single Use Biopsy Forceps
FB-211D, FB-221D, FB-231D, and FB-241D | Predicate Device (PD)
Single Use Biopsy Forceps
FB-433D
(K172726) | Comparison |
| --- | --- | --- | --- |
| | Oval type with Needle Manipulation wire | | |
| Compatible Endoscopes | Bronchoscopes with channel diameter 2.0 mm and working length 600 mm or less (exclude ultrasound endoscopes). | Bronchoscopes with channel diameter 1.7 mm and working length 600 mm or less (exclude ultrasound endoscopes). | SIMILAR
There is a difference in the compatible endoscope channel diameter between the SD and PD. Non-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device. |
| Product Specifications | Maximum insertion portion diameter: 1.9 mm
Working length: 1150 mm | Maximum insertion portion diameter: 1.5 mm
Working length: 1150 mm | SIMILAR
There is a difference in the maximum insertion portion channel diameter between the SD and PD. Non-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device. |
| Single Use/Reusable | Single Use | Single Use | SAME |
| Sterilization | Gamma | Ethylene Oxide (EtO) | DIFFERENT
There is a difference in sterilization methods between the SD and PD. Sterilization validation testing was performed to demonstrate the efficacy of the sterilization dose. |
| Shelf-Life | Five years | Three years | DIFFERENT
There is a difference in shelf-life expiry dating between the SD and PD. |
{7}
OLYMPUS
Disposable Biopsy Forceps FB-series
| Description | Subject Device (SD)
Single Use Biopsy Forceps
FB-211D, FB-221D, FB-231D, and FB-241D | Predicate Device (PD)
Single Use Biopsy Forceps
FB-433D
(K172726) | Comparison |
| --- | --- | --- | --- |
| | | | Stability / Shelf-life
testing was performed to demonstrate the difference
does not impact the effectiveness of the subject device. |
# 7. Non-Clinical/Clinical Tests Summary and Conclusion
Olympus performed bench testing to demonstrate substantial equivalence to the predicate device. Test samples were final, finished devices subjected to the full manufacturing process including sterilization.
The following performance side-by-side bench tests were conducted to demonstrate substantial equivalence between the subject and predicate devices. All test samples passed pre-defined acceptance criteria.
- Insertion Performance into the Bronchoscope
- Withdrawal Performance from the Bronchoscope
- Opening and Closing of the Forceps Cups
- Visual Inspection of the Insertion Portion
- Performance After Repeated Insertion and Withdrawal Operations
- Performance After Repeated Open/Close Operations
- Connection Strength Between the S-Cover and the Coil Sheath
- Visual Inspection for Defects of the Needle and Forceps Teeth
# 8. Biocompatibility Testing
Biocompatibility testing for the Disposable Biopsy Forceps was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." The biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity.
# 9. Sterilization and Shelf Life
The devices are sterilized via gamma radiation. Sterilization validation (VDmax25) was conducted in accordance with ISO 11137-1:2006/Amd1:2013 and Amd2:2018 Sterilization of Health Care Product – Radiation Part 1 – Requirements for development, validation and routine control of a sterilization process for medical devices and ISO11137-2:2013 Sterilization of Health Care Product – Radiation Part 2 – Establishing the sterilization dosage. The sterility assurance level is $10^{-6}$.
{8}
OLYMPUS
Disposable Biopsy Forceps FB-series
The shelf-life of the subject device is five years. Olympus conducted stability testing to demonstrate packaging integrity and product performance for the stated expiration date. The accelerated aging test was conducted in accordance with ASTM F1980-21. Package integrity testing was conducted in accordance with ASTM F1980-21, ISO 11607-1:2019 and ISO 11607-2:2019 after accelerated aging and simulated distribution. Package Integrity testing included Visual Inspection of Package, Peel Strength of Package, and Integrity of the Package. All samples successfully met the acceptance criteria and passed testing.
## 10. Human Factors Testing
Olympus conducted a Human Factors Evaluation and determined that the Disposable Biopsy Forceps FB-series are safe for intended users, uses, and use environments.
## 11. Conclusion
Based on the comparison to the predicate device, intended use, technological characteristics, and performance testing, the Disposable Biopsy Forceps FB-211D, FB-221D, FB-231D, and FB-241D are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.