DRX-Evolution Plus X-ray System, DRX-Compass Mobile X-ray System, Lux HD 35 detector, Lux HD 43 detector
K250954 · Carestream Health · KPR · Jan 20, 2026 · Radiology
Device Facts
Record ID
K250954
Device Name
DRX-Evolution Plus X-ray System, DRX-Compass Mobile X-ray System, Lux HD 35 detector, Lux HD 43 detector
Applicant
Carestream Health
Product Code
KPR · Radiology
Decision Date
Jan 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
DRX-Evolution Plus X-ray System The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen and other body parts. The device is not indicated for use in mammography. DRX-Compass X-ray System The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen, and other body parts. The device is not indicated for use in mammography. Lux HD Detectors Lux HD 35 Detector and Lux HD 43 Detector are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 35 Detector and Lux HD 43 Detector are not intended for mammography or dental applications.
Device Story
Stationary X-ray systems (DRX-Evolution Plus, DRX-Compass) utilize X-ray generators, tubes, and flat-panel detectors (Lux HD 35/43) to acquire radiographic images. Detectors convert X-rays to visible light via Cesium Iodide (CsI) scintillator, then to electron signals via TFT array. Systems include ImageView software with Eclipse II image processing and optional Smart Noise Cancellation (SNC) module. Digital Exposure Control (DEC) enables wireless AEC functionality on Evolution Plus. Operated by technologists in clinical settings; images reviewed by radiologists for diagnostic decision-making. Benefits include high-resolution imaging, reduced detector weight via glass-free technology, and potential dose reduction via SNC.
Clinical Evidence
Clinical study evaluated SNC image quality using live subjects, anthropomorphic phantoms, and cadaveric specimens. Four radiologists compared nominal-dose images (no SNC) to simulated 50% reduced-dose images (with SNC). Results: SNC images rated equivalent or better (p=0.000); mean RadLex score 3.4 ('Diagnostic'); >99.7% of SNC images rated same or higher than non-SNC. Non-clinical bench testing (NNPS, MTF, contrast-detail curves) confirmed significant noise reduction without loss of spatial resolution.
Technological Characteristics
Stationary X-ray system; Cesium Iodide (CsI) scintillator; glass-free epoxy laminate substrate; TFT panel design; 100-micron pixel pitch; 16-bit AD conversion. Connectivity: Wireless DEC for AEC. Software: ImageView v2.0 with Eclipse II/SNC. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for general radiographic diagnosis of human anatomy, including adult and pediatric patients. Applicable to skull, spinal column, extremities, chest, and abdomen. Contraindicated for mammography and dental applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 20, 2026
Carestream Health, Inc.
% Gina Maiolo
Sr Regulatory Affairs Manager
150 Verona St,
ROCHESTER, NY 14608
Re: K250954
Trade/Device Name: DRX-Evolution Plus X-ray System, DRX-Compass X-ray System, Lux HD 35 Detector, Lux HD 43 Detector
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary X-Ray System
Regulatory Class: Class II
Product Code: KPR, MQB
Dated: December 19, 2025
Received: December 19, 2025
Dear Gina Maiolo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250954 - Gina Maiolo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250954 - Gina Maiolo
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250954 | |
| Device Name DRX-Evolution Plus X-ray System, DRX-Compass X-ray System, Lux HD 35 Detector, Lux HD 43 Detector | |
| Indications for Use (Describe) DRX-Evolution Plus X-ray System The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen and other body parts. The device is not indicated for use in mammography. | |
| DRX-Compass X-ray System The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen, and other body parts. The device is not indicated for use in mammography. | |
| Lux HD Detectors Lux HD 35 Detector and Lux HD 43 Detector are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 35 Detector and Lux HD 43 Detector are not intended for mammography or dental applications. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
K250954
510(k) Owner Name
Carestream Health, Inc.
510(k) Owner Address
150 Verona St, Rochester, New York, 14608
510(k) Owner Contact Information
Gina Maiolo
Phone (Mobile)
516 395 0597
Date Summary Prepared
January 16, 2026
Device Trade Name(s)
DRX-Evolution Plus X-ray System
DRX-Compass X-ray System
Lux HD 35 Detector
Lux HD 43 Detector
Device Common Name(s)
Stationary X-Ray System, Solid State X-ray Imager (Flat Panel/Digital Imager)
Classification Name(s)
Stationary X-ray System
Device Class
Class II
Device Code(s)
KPR, MQB
Regulation Number(s)
21 CFR 892.1680
Lux HD Detectors (K243556)
DRX-Evolution Plus X-ray System (K233381)
DRX-Compass X-ray System (K223842)
Predicate Device(s)
{5}
Page 2 | 6
# Indications for Use
## DRX-Evolution Plus X-ray System
The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen and other body parts. The device is not indicated for use in mammography.
## DRX-Compass X-ray System
The device is indicated for use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen, and other body parts. The device is not indicated for use in mammography.
## Lux HD Detectors
Lux HD 35 Detector and Lux HD 43 Detector are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 35 Detector and Lux HD 43 Detector are not intended for mammography or dental applications.
# Device Description
## Lux HD Digital Flat Panel Detectors (35x43, 43x43 sizes)
Lux HD 35x43 and 43x43 detectors are high-definition detectors (100um) designed with a Cesium Iodide (CsI) scintillator for the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by a diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics to form an image by transmitting to the device PC through the user interface. The CsI scintillator provides a high spatial resolution image which helps to increase the diagnostic confidence. The detectors use a glass free technology which makes the overall detector weight lighter and mitigates the risks related to device breakage. Lux HDs are designed with Digital Exposure Control (DEC) functionality which uses a wireless connection to communicate with a DEC control box, which in turn connects to the generator for x-ray control. This DEC feature also enables AEC functions to be used with out-of-bucky exams and mobile imaging without a physical wired connection.
## Carestream In-Room X-ray Systems
The DRX Evolution Plus (K233381), DRX Compass (K223842) are permanently installed diagnostic x-ray systems for general radiographic x-ray imaging. These systems consist of a combination of components including an x-ray generator, workstation computers, various models of patient support tables, wall-mounted image receptors/detectors for upright imaging, various models of tube support devices, x-ray tube, and collimator (beam-limiting device). Currently, the DRX Evolution Plus X-ray system is designed to use a standalone automatic exposure control (AEC) chamber placed in front of the detector to control the x-ray exposures. Lux HD will be compatible with these systems. The DRX-Compass will not support the DEC feature; however the Lux HD detectors can still be used for general radiography.
{6}
Table 1. Lux HD SE Table
| | Lux HD 35x43 Lux HD 43x43 (K243556-predicate device -iRay LLC) | Lux HD 35x43 Lux HD 43x43 (subject device-Carestream Inc.) | Evaluation |
| --- | --- | --- | --- |
| Size | 386 x 460 x 15 mm 460 x 460 x 15 mm | 384 x 460 x 14.4 mm 460 x 460 x 14.4 mm | Same |
| Scintillator | Cesium Iodide (CsI) | Cesium Iodide (CsI) | Same |
| Substrate | Glass reinforced epoxy laminate (non-glass) | Glass reinforced epoxy laminate (non-glass) | Same |
| Panel Design | TFT | TFT | Same |
| DEC Functionality | Yes | Yes | Same |
| Pixel Pitch | 100 microns | 100 microns | Same |
| AD Conversion | 16 bits | 16 bits | Same |
| Detector Resolution | 3,500 x 4,300 pixels 4,300 x 4,300 pixels | 3,500 x 4,300 pixels 4,267 x 4,267 pixels | Same |
| Image Capture Area (usable pixel area) | 35.0 cm x 43.0 cm 43.0 cm x 43.0 cm | 35.0 cm x 43.0 cm 42.7 cm x 42.7 cm | Same |
| Sensitivity, RQA-5 Beam | 680 to 867 LSB/μGy | 680 to 867 LSB/μGy | Same |
| Minimum DQE @ 2.5 μGy, RQA-5 Beam | 68% @ 0.08 cyc/mm 54% @ 1.0 cyc/mm 39% @ 2.0 cyc/mm 27% @ 3.0 cyc/mm 16% @ 4.0 cyc/mm 8% @ 5.0 cyc/mm | 68% @ 0.08 cyc/mm 54% @ 1.0 cyc/mm 39% @ 2.0 cyc/mm 27% @ 3.0 cyc/mm 16% @ 4.0 cyc/mm 8% @ 5.0 cyc/mm | Same |
| MTF, RQA-5 Beam | >65% @ 1.0 cyc/mm Typical: 73% @ 1.0 cyc/mm | >65% @ 1.0 cyc/mm Typical: 73% @ 1.0 cyc/mm | Same |
{7}
Table 2: DRX-Evolution Plus and DRX Compass X-ray Systems
| Componen t | DRX Evolution Plus System (K233381) DRX-Compass System | DRX Evolution Plus System DRX Compass System | Evaluation |
| --- | --- | --- | --- |
| Compatible Digital Flat Panel Detectors | Evolution: • DRX Plus 2530 (K183245) • Focus HD 35 (K213646) • Focus HD | New detectors with DEC feature • Lux HD 35 (K243556) • Lux HD 43 (K243556) | Lux HD integration testing with the DR systems passed and is provided in this submission. |
| | Compass: • DRX Plus 2530 (K183245) • Focus HD 35 (K213646) • Focus HD 43 (K213529) | New detectors without DEC feature • Lux HD 35 (K243556) • Lux HD 43 (K243556) | |
| DR Application System Software | ImageView Software v2.0 with SNC | Same | Verification and Validation confirms integration of detectors does not raise any new questions of safety or effectiveness. |
| Radiograph i c | Radiographic Table | Same | No impact to safety or performance |
| X-Ray Tube(s) | Evolution: Canon XRR-4631G, Varex | Same | No impact to safety or performance |
| | Compass: Toshiba (Canon): E7254FX, E7252X | | |
| X-ray Generator | Evolution: Three Phase, CGN-80 (65kW) | Same | No impact to safety or performance |
| | Compass: Three Phase, CGF-50-2, CGF-50-3, CGF-65-3, and | | |
| Exposure Control | Evolution: Analog Automatic Exposure Control (AEC) chambers AEC is integrated inside the wall and table buckies, only support exams | Evolution: Same. Digital Exposure Control (DEC) is introduced (integrated) into the Lux HD detectors. | Verification and validation testing confirm introduction of DEC feature does not raise any new questions of safety or effectiveness. As noted above, *Compass will not support the DEC feature, however it will be compatible with Lux HD for general radiography purpose only. |
| | Compass: No AEC feature | Compass: No Digital Exposure Control (DEC) | |
{8}
Page 5 | 6
# Summary of Safety and Performance
## Technological Characteristics
The modified DRX-Evolution Plus (K233381), DRX-Compass (K223842) systems are substantially equivalent to the predicate devices currently cleared on the market. The below key points support this determination and summarizes the similarities in characteristics when comparing the modified and predicate devices. Any minor differences do not raise questions on performance or safety in regards to the radiographic systems.
- The cleared Lux HD detectors with Digital Exposure Control (DEC) functionality were cleared in a detector-level submission by Carestream’s development partner under K243556 and then subsequently transferred to Carestream for device listing. Carestream’s version of the detector is the same with the only difference being minor optimization.
- The digital radiography systems that will support the cleared Lux HD detectors consist of the same fundamental scientific technology and is designed with the same operating principles as the predicate devices. The predicate and modified systems consist of the same critical components: an x-ray generator, x-ray tube, collimator, and image acquisition software. The only difference is the minor modification to the acquisition software to support the detectors with DEC functionality.
- The ImageView software is the image acquisition software that is installed on the digital x-ray systems. The base version of the software (2.0) has been cleared with the following Carestream DR systems
- K233381 (DRX Evolution Plus)
- K223842 (DRX Compass System)
ImagineView 2.0 (B2.2) supports the digital radiography rooms systems (DRX Compass, DRX Evolution Plus) with the Lux HD detectors with DEC, SNC and Smart Virtual LLI features. The Smart Virtual LLI feature is built on top of the existing cleared LLI functionality. It is designed to enhance (not replace) the operator’s workflow by providing an optimized starting point for LLI exams for the technologist. Instead of relying on the existing (cleared) LLI default settings, Smart Virtual LLI intelligently suggests the initial parameter for the regions (top, bottom, collimation width, and OID), allowing for more efficiency in adjustments.
The image processing software (Eclipse II) includes the optional Smart Noise Cancellation (SNC) module and is the same as the predicate devices. *SNC is not installed on the detectors. It is important to note that the image processing software and the imaging chain remains the same, we are only adding cleared detectors, Lux HD 35, Lux HD 43 (K243556), Focus HD 35, Focus HD 43 (K213646, K213529) for use with the previously cleared Eclipse II w/SNC.
### Smart Noise Cancellation clearances:
- Eclipse II with Enhanced Noise Reduction (K202441)
- Eclipse II SNC Claims (K213307)
- DRX Evolution Plus (K233381)
- DRX Compass (K223842)
## Clinical and Non Clinical Testing
A clinical study was designed to evaluate the image quality performance of SNC on images acquired from the Carestream Focus HD 35/43 and Lux HD 35/43 detectors. Performance testing included imaging performance of the Eclipse II (rendering software) with and without Smart Noise Cancellation, capturing images with the Focus HD 35/43
{9}
and Lux HD 35/43 detectors for use on stationary systems. Images were acquired using live subjects, anthropomorphic phantoms, and adult cadaveric specimens. All processed images were delivered to a diagnostic workstation with high-resolution monitors for comparative evaluation by board-certified radiologists. The study demonstrated that images processed with SNC are equivalent to or better than images processed without SNC (the predicate device). The following key points summarize the outcomes of the testing for the in-rooms systems:
- Overall preference is in favor of SNC at a meaningful level (p = 0.000).
- Image quality for SNC is as good or better than images without SNC (p = 0.000) regardless of image type, with a mean RadLex score of 3.4 ("Diagnostic") which supports the claim of SNC being safe and effective. Greater than 99.7% of the SNC RadLex images were rated the same of higher than the non-SNC images.
- SNC produces images with hardware appearance as good as or better than the predicate device.
- A CDRAD phantom was used to compare the images acquired at nominal dose without SNC to the images generated at 50% reduced dose images with SNC which produced similar scores of inverse of Image Quality Figure (IQFinv).
- A reader study conducted by four independent, board-certified diagnostic radiologists compared pairwise live patient images acquired at nominal dose without SNC to simulation-generated images representing approximately 50% dose reduction processed with SNC. Actual live patient images acquired at reduced dose were not obtained or evaluated; therefore, the comparison was limited to simulated reduced-dose images and did not include a direct comparison using real patient images acquired at reduced exposure. The study demonstrated that simulated reduced-dose images processed with SNC were of diagnostic quality comparable to nominal-dose images processed without SNC based on RadLex ratings.
System verification and validation was performed. This testing demonstrates the safety and performance of the new Lux HD detectors integrated with the DRX-Evolution and DRX-Compass systems.
Additional IQ testing was performed, which specifically characterizes the SNC performance using quantitative methods of analysis of non-clinical data. The "predicate" device processed images without SNC and the "subject" device processed images with SNC. The predicate and the subject devices were evaluated using captures of flat fields and physics phantoms. The images from both devices were quantitatively compared using Normalized Noise Power Spectra (NNPS), standard deviation (noise) of flat sections of the images, Modulation Transfer Function (MTF), and the contrast-detail curve. This testing demonstrated that there is significant noise reduction without impacting sharpness and resulting in retention of image spatial resolution between the predicate and the subject devices.
In conclusion both clinical and non-clinical (bench) testing demonstrates the DRX Evolution and DRX Compass systems (modified) with integration of the Lux detectors with DEC and SNC is the same as or equivalent to the cleared room systems (predicate) currently on the market.
Page 6 | 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.