Single Use Preloaded Sphincterotome V (Distal Wire Guided) (KD-VC600 Series); Single Use Sphincterotome V (Distal Wire Guided) (KD-VC400 Series); Single Use 3-Lumen Sphincterotome (KD-V Series); Disposable Triple Lumen Sphincterotome (KD Series)
K250945 · Olympus Medical Systems Corp. · KNS · Oct 17, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K250945
Device Name
Single Use Preloaded Sphincterotome V (Distal Wire Guided) (KD-VC600 Series); Single Use Sphincterotome V (Distal Wire Guided) (KD-VC400 Series); Single Use 3-Lumen Sphincterotome (KD-V Series); Disposable Triple Lumen Sphincterotome (KD Series)
Applicant
Olympus Medical Systems Corp.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Oct 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-411Q/431Q Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope. The Single Use 3-Lumen Sphincterotome V KD-V411M/V431M Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope and guidewire. The Single Use Sphincterotome V (Distal Wireguided) KD-VC411Q/VC431Q/VC433Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope. The Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC631Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope. The preloaded guidewire is used for guiding and exchanging endoscopic accessories for biliary duct, including but not limited to the common bile, cystic, pancreatic and right and left hepatic ducts.
Device Story
Single-use endoscopic sphincterotomes; used for papillotomy via high-frequency electrosurgical current. Devices feature separate lumens for guidewire, cutting wire, and contrast medium injection; pre-curved, tapered tips for papilla of Vater insertion. Some models include V-marking for insertion depth confirmation and integrated C-Hook for endoscope attachment. Distal end features PFA (perfluoroalkoxy) coating to prevent thermal injury to non-target tissue. Operated by physicians during endoscopic procedures. Input: high-frequency electrical current from compatible generators. Output: controlled tissue incision (papillotomy). Benefits: facilitates biliary duct access and therapeutic intervention while minimizing thermal damage to surrounding tissue.
Clinical Evidence
Bench testing only. No clinical or animal data collected. Performance testing included knife dimensions, insertion/withdrawal, hook/A-cord attachment, knife operation, papillotomy resistance, insulation integrity, contrast infusion, connection strength, and X-ray visibility. All tests met pre-defined acceptance criteria.
Technological Characteristics
Single-use endoscopic electrosurgical accessories. Materials include PFA (perfluoroalkoxy) coating on distal end. Energy source: high-frequency electrosurgical current. Features: multi-lumen design, pre-curved tapered tips, optional C-Hook/C-Channel. Sterilization: ethylene oxide. Connectivity: compatible with Olympus electrosurgical generators via A-cords.
Indications for Use
Indicated for patients requiring papillotomy via high-frequency electrosurgical current during endoscopic procedures. Preloaded guidewire models are indicated for guiding and exchanging endoscopic accessories within the biliary duct, including common bile, cystic, pancreatic, and hepatic ducts.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Single Use Sphincterotome V, Single Use Preloaded Sphincterotome V (Distal Wireguided) (K141991)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 17, 2025
Olympus Medical Systems Corp.
% Jillian Connery
Manager, Program Regulatory Affairs
Olympus Corporation of the Americas
800 West Park Drive
Westborough, Massachusetts 01581
Re: K250945
Trade/Device Name: Single Use Preloaded Sphincterotome V (Distal Wire Guided) (KD-VC600 Series); Single Use Sphincterotome V (Distal Wire Guided) (KD-VC400 Series); Single Use 3-Lumen Sphincterotome (KD-V Series); Disposable Triple Lumen Sphincterotome (KD Series)
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic Electrosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: KNS
Dated: September 18, 2025
Received: September 19, 2025
Dear Jillian Connery:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250945 - Jillian Connery
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250945 - Jillian Connery
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANTHONY LEE -S
Anthony C. Lee, Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Single Use Preloaded Sphincterotome V (Distal Wireguided)
Page 11 of 63
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K250945 | ? |
| Please provide the device trade name(s). | | ? |
| Single Use Preloaded Sphincterotome V (Distal Wireguided) (KD-VC600 Series);
Single Use Sphincterotome V (Distal Wireguided) (KD-VC400 Series);
Single Use 3-Lumen Sphincterotome (KD-V Series);
DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME (KD Series) | | |
| Please provide your Indications for Use below. | | ? |
| The Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC631Q Series are intended to be
used for papillotomy using high-frequency current in combination with an endoscope. The preloaded
guidewire is used for guiding and exchanging endoscopic accessories for biliary duct, including but not
limited to the common bile, cystic, pancreatic and right and left hepatic ducts.
The Single Use Sphincterotome V (Distal Wireguided) KD-VC411Q/VC431Q/VC433Q Series are intended
to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use 3-Lumen Sphincterotome V KD-V411M/V431M Series is intended to be used for
papillotomy using high-frequency current in combination with an endoscope and guidewire.
The DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-411Q/431Q Series is intended to be used for
papillotomy using high-frequency current in combination with an endoscope. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K250945 - Page 1/3
| 510(k) #: K250945 | 510(k) Summary | Prepared on: 2025-09-26 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Olympus Medical Systems Corp. | |
| Applicant Address | 2951 Ishikawa-cho Hachiochi-shi Tokyo 192-8507 Japan | |
| Applicant Contact Telephone | +81 42-642-2111 | |
| Applicant Contact | Ms. Seiko Yunoki | |
| Applicant Contact Email | seiko.yunoki@olympus.com | |
| Correspondent Name | Olympus Corporation of the Americas | |
| Correspondent Address | 800 West Park Drive Westborough MA 01581 United States | |
| Correspondent Contact Telephone | 614-817-2029 | |
| Correspondent Contact | Ms. Jillian Connery | |
| Correspondent Contact Email | Jillian.Connery@olympus.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Single Use Preloaded Sphincterotome V (Distal Wireguided) (KD-VC600 Series);
Single Use Sphincterotome V (Distal Wireguided) (KD-VC400 Series);
Single Use 3-Lumen Sphincterotome (KD-V Series);
DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME (KD Series) | |
| Common Name | Unit, Electrosurgical, Endoscopic (With Or Without Accessories) | |
| Classification Name | Endoscopic electrosurgical unit and accessories | |
| Regulation Number | 876.4300 | |
| Product Code(s) | KNS | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K141991 | Single Use Sphincterotome V, Single Use Preloaded Sphincterotome V (Distal Wireguided) | KNS |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Single Use Sphincterotome KD-VC, KD-V, and KD Series consists of the following devices:
• Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC600 Series (aka CleverCut3V)
• Single Use Sphincterotome V (Distal Wireguided) KD-VC400 Series (aka CleverCut3V)
• Single Use 3-Lumen Sphincterotome V KD-V Series (aka CleverCut3V)
• DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD Series (aka FlowCut) | | |
{5}
K250945 - Page 2/3
The subject devices have separate lumens for guidewire, cutting wire and injection of contrast medium for papillotomy. The knives are pre-curved and have a tapered tip to facilitate insertion into the papilla of Vater. The insertion portion of the KD-V, and KD-VC Series have a V-marking, which the relative insertion length into the endoscope can be confirmed by the positional relationship between the V-marking and the biopsy valve of the endoscope and feature an integrated C-Hook to attach to the endoscope. The distal end of the sphincterotome is coated with CleverCut Coating®, Olympus' PFA (perfluoroalkoxy) coating designed to prevent thermal injury to non-target tissue while cutting the papilla of Vater. The KD-VC600 Series offers models that are preloaded with VISIGLIDE guidewires.
The subject devices are provided to the user sterilized by ethylene oxide and intended for single use only.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-411Q/431Q Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope
The Single Use 3-Lumen Sphincterotome V KD-V411M/V431M Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope and guidewire.
The Single Use Sphincterotome V (Distal Wireguided) KD-VC411Q/VC431Q/VC433Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC631Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope. The preloaded guidewire is used for guiding and exchanging endoscopic accessories for biliary duct, including but not limited to the common bile, cystic, pancreatic and right and left hepatic ducts.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device has the same intended use as the predicate device. The devices are intended to be used for papillotomy using high-frequency current in combination with an endoscope and guidewire. The preloaded guidewire is used for guiding and exchanging endoscopic accessories for biliary duct, including but not limited to the common bile, cystic, pancreatic and right and left hepatic ducts.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device shares the same materials, operates on the same principles and has similar dimensional characteristics as the predicate device. Minor dimensional differences from the predicate include variants with a shorter working length and shorter knife tip length, and compatibility with additional size guidewires. The subject devices are compatible to endoscopes with the same working channel length as the predicate device, and the knife tip length is selected based on physician preference. Comparative bench testing performed demonstrates these dimensional differences do not affect safety or effectiveness of the devices.
The subject device offers models with and without a C-Channel and C-Hook as compared to the predicate, which includes both. The C-Channel and the C-Hook are usability features and not necessary for the device to achieve its intended use; therefore this difference does not raise questions on safety or effectiveness.
The subject device is compatible with additional Olympus generators and compatible A-cords that have been introduced to the market since the predicate was cleared. Compatibility with these generators is verified through electrical safety/EMC testing.
Bench testing, biological safety evaluation, sterilization validation, and shelf-life studies were undertaken on the subject device and predicate device to demonstrate substantial equivalence.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Performance testing for the subject device was determined based on the device risk assessment and industry recognized standards. Performance bench testing as listed below were conducted to demonstrate the device is safe and effective for its intended use and is substantially equivalent to the predicate device. Bench tests were performed on the Subject Single Use Sphincterotome and Predicate Olympus Single Use Sphincterotome V and Single Use Preloaded Sphincterotome V (Distal Wireguided) cleared under K141991, where applicable for comparative testing. The test results have demonstrated substantial equivalence of the subject device to the predicate device.
Performance testing included the following:
- Dimension of sphincterotome distal end (Knife length)
- Insertion/ withdrawal over the Guidewire
- Insertion/ withdrawal (Mono-rail guidewire technique)
{6}
K250945 - Page 3/3
- Attachment/detachment of hook
- Attachment/detachment of the A-cord
- Knife operation (Elevation of the knife)
- Knife operation (Knife direction)
- Papillotomy (resistance of the cutting wire)
- Papillotomy (insulation of the coated portion)
- Contrast Medium Infusion
- Connection strength (between the Tube and the handle)
- Connection strength (between the Tube and the Guidewire Port)
- Visibility test under X-ray
Animal and clinical data was not collected to support substantial equivalence.
Product performance testing of the subject device was conducted, and the device passed all the pre-defined acceptance criteria for each test item demonstrating that design verification has been met. The comparative testing demonstrates the subject device, and the predicate device are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.