K250911 · Electrogenics Labs, Ltd. · IYE · Oct 17, 2025 · Radiology
Device Facts
Record ID
K250911
Device Name
MOSkin Radiation Measurement System
Applicant
Electrogenics Labs, Ltd.
Product Code
IYE · Radiology
Decision Date
Oct 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The MOSkin radiation measurement system intended use is a dosimeter to measure radiation dose delivered by a radiation source to the location of an Radiation Dosimeter (RD) sensor on the patient in a clinical use environment. The system is intended for the verification of the output of radiation producing devices. The output of the system is not used to directly adjust the radiation dose to the patient.
Device Story
MOSkin Radiation Measurement System provides secondary verification of radiation dose during radiotherapy and diagnostic imaging. System components: single-use MOSFET (Si-based) dosimeter (RD) placed on patient skin; reading device (HUB); MOSkin Software Application. Principle: incident radiation alters MOSFET threshold voltage; pre- and post-treatment voltage measurements calculate delivered dose. Used in clinical settings by medical physicists or radiation therapists. HUB transmits data via Bluetooth to tablet-based monitor. Software calculates, reports, and stores dose data. Output provides verification of radiation-producing device output; not used for real-time dose adjustment. Benefits: wireless connectivity, single-use dosimeter eliminates cross-contamination/disinfection needs, wider dose range (up to 400 cGy), and improved portability via tablet interface.
Clinical Evidence
No clinical studies performed. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-6), biocompatibility (ISO 10993), mechanical reliability, and simulated-use performance testing. System demonstrated accuracy of ±5 cGy (<100 cGy) and ±5% (≥100 cGy) across a 10-400 cGy range. Usability validated through three formative and one summative study with 18 participants (administrators/operators), confirming no unresolved usability issues.
Technological Characteristics
MOSFET (Si-based) dosimeter; wireless (Bluetooth) connectivity; tablet-based monitor. Dimensions: 6mm x 0.3mm x 30mm. Power: rechargeable battery. Standards: IEC 60601-1, IEC 60601-1-6, ISO 10993-1, ISTA 3A. Single-use dosimeter design. Software-based dose calculation and reporting.
Indications for Use
Indicated for patients undergoing therapeutic radiation procedures or diagnostic imaging requiring radiation dose verification. Used in clinical environments to measure radiation dose at the sensor location on the patient.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 17, 2025
Electrogenics Labs, Ltd
Grant Palmer
Head of QA/RA
Suite 1, Level 10
76 Berry Street
North Sydney, NSW 2060
Australia
Re: K250911
Trade/Device Name: MOSkin Radiation Measurement System
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical charged-particle radiation therapy system
Regulatory Class: Class II
Product Code: NZT,IYE
Dated: March 26, 2025
Received: March 26, 2025
Dear Grant Palmer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250911 - Grant Palmer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250911 - Grant Palmer
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2025.10.17 13:39:21 -04'00'
For
Lora Weidner
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250911 | |
| Device Name | |
| MOSkin Radiation Measurement System | |
| Indications for Use (Describe) | |
| The MOSkin radiation measurement system intended use is a dosimeter to measure radiation dose delivered by a radiation source to the location of an Radiation Dosimeter (RD) sensor on the patient in a clinical use environment. The system is intended for the verification of the output of radiation producing devices. The output of the system is not used to directly adjust the radiation dose to the patient. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 1 of 9
510(k) Summary
K250911
Submission Date: March 26, 2025
Summary Prepared Date: October 09, 2025
Submitter Information:
Submitted By:
Grant Palmer
Head of Quality & Regulatory
Electrogenics Laboratories Limited
Suite 1, Level 10
76 Berry Street
North Sydney, NSW
Australia, 2060
Phone: +1-818-859-2046
Email: grant.palmer@electrogenicslabs.com
Secondary Contact:
David Rundell
Chief Executive Officer
Electrogenics Laboratories Limited
Email: david.rundell@electrogenicslabs.com
Device Information:
Trade Name: MOSkin Radiation Measurement System
Common Name: Dosimetry System
Classification Name: 21 CFR 892.5050 Medical Charged-particle radiation therapy system
Device Classification: Class II
Product Code: NZT, IYE
Predicate Device(s): Portable Dosimeter, K092285
Best Medical Canada, Ltd. Class II,
Product Code IYE
21 CFR 892.5050 Medical Charged-particle radiation therapy system
03 510k Summary.docx
{5}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 2 of 9
# Device Description:
The Electrogenics Laboratories Ltd MOSkin Radiation Measurement System consists of the following components to provide secondary verification of dose from various radiotherapy and diagnostic imaging devices:
- MOSFET (Si-based metal-oxide-semiconductor field-effect transistor) Dosimeter (RD) to record radiation during radiation exposure.
- Reading device (HUB) for reading the radiation dose recorded by the dosimeter.
- MOSkin Software Application, a simple software tool that the user interacts with for radiation dose calculation, dose reporting, managing and storing data.
# Indications for Use:
The MOSkin radiation measurement system intended use is a dosimeter to measure radiation dose delivered by a radiation source to the location of a Radiation Dosimeter (RD) sensor on the patient in a clinical use environment. The system is intended for the verification of the output of radiation producing devices. The output of the system is not used to directly adjust the radiation dose to the patient.
# Comparison to Predicate:
The intended use of the predicate device (K092285) is: "To provide dosimetry or to detect the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of radiation producing devices."
The intended use of the MOSkin System is: "The MOSkin radiation measurement system intended use is a dosimeter to measure radiation dose delivered by a radiation source to the location of an RD sensor (RD) on the patient in a clinical use environment. The system is intended for the verification of the output of radiation producing devices. The output of the system is not used to directly adjust the radiation dose to the patient."
The two key points of both statements is that each system provides:
1) dosimetry for the delivered dose of radiation
2) the system is intended to provide verification of delivered dose.
Therefore, although the text is slightly different from the predicate, the intended use remains the same between the proposed device and predicate device (capturing and displaying radiation dose) and the proposed differences in indications are for clarity and do not have impact on safety or effectiveness.
The MOSkin System utilizes a MOSFET (Si-based metal-oxide-semiconductor field-effect transistor) as the measurement technology for determining the radiation dose delivered to a patient. The incident therapeutic radiation changes the electrical characteristics of MOSFET, specifically the threshold voltage, which can be measured. By measuring the threshold voltage before and after radiation treatment, the radiation
03 510k Summary.docx
{6}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 3 of 9
dose can be calculated. The predicate device (K092285) uses the same measurement technology and principle of operation as the MOSkin System.
The MOSkin System provides greater flexibility for treating physicians by providing twice the measure dosage range (up to 400 cGy vs 200 cGy) compared to predicate, while maintaining similar dose accuracy and angular dependence, across a similar range to therapies.
The dosimeter of the MOSkin System is single-use, removing the need for dosimeter disinfection and reducing the risk of cross-contamination between patients. This configuration also provides an over-all smaller dosimeter package, potentially facilitating a greater ease of use. The tablet interface, versus the desktop computer of the predicate, provides and smaller benchtop footprint and a degree of mobility.
The proposed device therefore employs the same fundamental scientific technology, and any differences do not raise different questions of safety or effectiveness. The device contains several advantages for use over the predicate including a wider range of measured radiation dose, wireless technology, smaller size and flexible user interface.
| Feature & Specification | Electrogenics MOSkin | Best Medical Portable Dosimeter | Comparison |
| --- | --- | --- | --- |
| Indication for Use | The MOSkin radiation measurement system intended use is a dosimeter to measure radiation dose delivered by a radiation source to the location of a Radiation Dosimeter (RD) sensor on the patient in a clinical use environment. The system is intended for the verification of the output of radiation producing devices. The output of the system is not used to directly adjust the radiation dose to the patient. | To provide dosimetry or to detect the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of radiation producing devices. | Similar |
| Measurement Technology | Metal-Oxide-Semiconductor Field-Effect Transistor (MOSFET) | Metal-Oxide-Semiconductor Field-Effect Transistor (MOSFET) | Same |
| Principle of Operation | Threshold voltage change proportional to radiation dose | Threshold voltage change proportional to radiation dose | Same |
03 510k Summary.docx
{7}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 4 of 9
| Feature & Specification | Electrogenics MOSkin | Best Medical Portable Dosimeter | Comparison |
| --- | --- | --- | --- |
| Dose Range | Up to 400 cGy | Up to 200 cGy | MOSkin has superior range |
| Accuracy | ±5 cGy (dose <100 cGy)
±5 % (dose ≥ 100 cGy) | ±3 % (at 20 cGy)
±2 % (at 200cGy) | Similar |
| Angular dependence | +/-5 % of fitted angular offset curve at angles 0-180-360°. | ±2 % for 360° | Similar |
| Width of Dosimeter | 6 mm | 2.5 mm | MOSkin provides a smaller overall dosimeter envelope, increasing ease and flexibility of use |
| Thickness of Dosimeter | 0.3 mm | 0.3 mm | |
| Length of Dosimeter | 30 mm | 200 mm | |
| Volume | ~54 mm³ | ~150 mm³ | |
| Number of Simultaneous Dosimeters Supported | Up to four | Up to five | Similar |
| Reusable vs single use | Single use dosimeter | Reusable dosimeter | Removes requirement for sterilization of dosimeters |
| Surface/Skin Dose Measurements | Yes | Yes | Same |
| Dose Report for Patient Files | Advanced patient reports in PDF format | Advanced patient reports in MSWord, Excel and PDF to R&V systems | Both provide reports in human readable format |
| Dosimeter Connection: Wireless vs Wired | Wireless | Wired | Wireless provides greater flexibility in use |
| Connection during Treatment | Wireless | Wired | Wireless provides greater flexibility in use |
| Dosimetry Display | Data transfer using Bluetooth technology to tablet-based MOSkin Monitor | Data transfer using Bluetooth technology and on-line readout from a remote PC | Similar |
| User Interface: Tablet vs PC | Tablet | PC | Tablet provides similar user experience with greater flexibility |
| Calibration | Factory calibrated Physicists can create custom calibrations to match their settings | Determined by physicists to match their settings | MOSkin provides ease of use in a "straight from the box" case. |
03 510k Summary.docx
{8}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 5 of 9
| Feature & Specification | Electrogenics MOSkin | Best Medical Portable Dosimeter | Comparison |
| --- | --- | --- | --- |
| Power | Built-in rechargeable Battery
Operable while charging | AA rechargeable Battery | Similar |
| Applications | TBI/ Central Dose
High Dose Fluoroscopy | TBI/ Central Dose
High Dose Fluoroscopy | Same |
## Performance Testing:
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
Evaluation included consideration of physical characteristics, chemical characteristics, manufacturing processes and biological testing. As per ISO 10993-1:2018 the device is in contact with intact skin, only., and the contact duration is B – Prolonged (1-30 d).
For devices in limited contact with intact skin ISO 10993-1:2018 Table A.1 (and FDA Guidance) indicates the following endpoints should be evaluated in the risk assessment. Evaluation may include review of existing data, additional endpoint specific testing, or a rationale of why the endpoint assessment does not require an additional data set. Relevant subparts of ISO 10993 which provide specific guidance are identified and annotated next to the relevant biological evaluation endpoint.
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2021)
- Irritation (ISO 10993-23:2021)
All testing passed, with no deviations.
Based on this testing, the MOSkin Radiation Dosimeter is not cytotoxic, an irritant or a sensitizer. It is biocompatible for the intended use and the indications for use.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted, consisting of the Radiation Dosimeter, MOSkin HUB and MOSkin Monitor. The system complies with the IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AMD1 (2012) + AMD2 (2020); EN 60601-1:2006 + AC1:2010 + A11:2011 + A12:2014 + AC2:2016 + A2:2021; CAN/CSA-C22.2 No. 60601-1:14; ANSI/AAMI ES60601-1:2005 + A1:2012 +A2:2021, EN 62479:2010 and JIS T 0601-1:2023 standards for safety and the IEC 60601-1-6:2010, AMD1:2013, AMD2:2020 and CSA-C22.2 No.60601-1-6:2011/A2:2021 standards for EMC.
The system was also tested for compliance to 47 Code of Federal Regulations Part 15 – Radio Frequency Devices and ETSI EN 300 328 V2.2.2: Compliance for 2.4 GHz ISM Band Devices. The system passed all testing.
## Usability Studies
The MOSkin System underwent usability four usability evaluations, three (3) formative and one (1) summative and summarized below.
03 510k Summary.docx
{9}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 6 of 9
| Test | | Result |
| --- | --- | --- |
| Formative Evaluation | 01 | This study evaluated the workflow of using the MOSkin System in the radiation treatment of simulated patients with two (2) participants, both Radiation Therapists. The evaluation resulted in several recommended improvements to the system for ease of use and clarity. There were no unresolved usability problems that could affect patient, users or environmental safety. |
| | 02 | This study evaluated the workflow of using the MOSkin System in the radiation treatment of simulated patients with two (2) participants, one Medical Physicist and one Radiation Therapist. The evaluation resulted in several recommended improvements to the system for ease of use and clarity. There were no unresolved usability problems that could affect patient, users or environmental safety. |
| | 03 | This study evaluated the workflow of using the MOSkin System in the radiation treatment of simulated patients and the administrative functions of the system. There was a total of six (6) participants, two (2) administrators, and four (4) operators. The evaluation resulted in several recommended improvements to the system for ease of use and clarity. There were no unresolved usability problems that could affect patient, users or environmental safety. |
| Summative Evaluation | | This study evaluated the workflow of using the MOSkin System in the radiation treatment of simulated patients and the administrative functions of the system. There was a total of 18 participants, eight (8) administrators, and ten (10) operators. The evaluation resulted in several recommended improvements to the system for ease of use and clarity. There were no unresolved usability problems that could affect patient, users or environmental safety |
The study reports were then evaluated and certified to comply with IEC 60601-1-6 by TÜV SÜD PSB Pte Ltd.[8]
The MOSkin System has therefore, been successfully evaluated for usability by potential users. These evaluations did not discover any unresolved usability problems that could affect patient, users or environmental safety.
## Mechanical testing
The MOSkin System underwent the following mechanical testing:
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2021)
- Irritation (ISO 10993-23:2021)
- Lid reliability
- Repeatability
- RD Continuity
- Force on RD
- Port Reliability - RD Connection reliability (ref [PRD])
- On/Off push force
- Ability to open DAT Lid
- RD Adhesive Strength
03 510k Summary.docx
{10}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 7 of 9
- RD Packaging opening force
- RD Storage
- RD Protection
- Chemical Resistance
All mechanical testing passed and demonstrated that the MOSkin System is reliable and mechanically suited for use.
## Environmental Transport
The ability of the MOSkin System be shipped without compromising patient, user or environmental safety was evaluated to ISTA 3A. Representative devices were tested for function before and after environmental and transport simulations. The MOSkin System has PASSED environmental transport testing according to ISTA 3A.
## Atmospheric Pressure
The tolerance of the MOSkin System to the atmospheric pressure range that the system may be subjected to during transport, storage and use was evaluated. The evaluation concluded that the MOSkin System can operate within specification when subjected to atmospheric pressures within the range 78 kPa to 114 kPa.
## Electrical Verification
Each electrical system requirement for the MOSkin was tested. The MOSkin System was found to meet all electrical requirements.
## Storage Life
The components and materials within the MOSkin System were evaluated for any degradation due to the labelled storage conditions. The report concludes that the MOSkin System is unlikely to be degraded due to appropriate storage.
Systems have been evaluated after six-month of storage. All devices passed testing
## Software Verification and Validation Testing
Software verification and validation testing was conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” All software testing passed, and the software was found to meet all requirements.
## Cybersecurity
This cybersecurity report confirms that the Cybersecurity Management Plan has been successfully implemented for MOSkin. The applied security measures align with FDA requirements and additional cybersecurity frameworks. Security testing validated the robustness of controls, and any potential improvement identified were addressed appropriately.
03 510k Summary.docx
{11}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 8 of 9
The overall impact of wireless communication failure in this device has been assessed as negligible, meaning that any disruption in BLE or Wi-Fi functionality would result only in inconvenience or temporary discomfort, with no foreseeable risk to patient safety or device operation.
## System Performance
The MOSkin System underwent a series of tests to evaluate the requirements related to the clinical performance. All testing was completed under simulated use conditions. The tested requirements were:
- Reliably & accurately measure the radiation dose received by the sensor during patient radiotherapy treatments or dummy/phantom trials.
- Dose range 10-400 cGy
- Accuracy +/-5 % at doses => 100 cGy
- Accuracy within +/-5 cGy at doses < 100 cGy
- Accurate skin dose radiation measurement at a water equivalent depth of <0.15 mm.
- MOSkin RD provides dose accuracy measurement of +/-5 % compared with an independent calibrated radiation source in specific clinical test scenarios.
- RD and MOSkin system shall provide overall dose measurement with <1.0 % fading over 2 hours.
- RD and MOSkin System shall be characterized at a range of non-perpendicular Linac beam angles. All MOSkin RD dose measurements shall be +/-5 % to a fitted angular response curve at each angle setpoint.
The MOSkin System passed all requirements related to clinical use.
## Animal Studies
Animal studies were not necessary to support substantial equivalence of the proposed device to the predicate device.
## Clinical Studies
Clinical studies were not necessary to support substantial equivalence of the proposed device to the predicate device.
## Summary
Based on the clinical performance as documented in the pivotal clinical study, the MOSkin System was found to have a safety and effectiveness profile that is similar to the predicate device.
## Conclusion:
In conclusion, the tests conducted, as well as all verification and validation activities, demonstrate that the design specifications and technological characteristics of the MOSkin System meet applicable requirements and standards for the safety and effectiveness of the device for its intended use. The intended use and fundamental technology are the same between the subject and predicate devices. There are some
03 510k Summary.docx
{12}
Electrogenics Laboratories Ltd
Doc Title
510k Summary
Page Number
Page 9 of 9
differences in the indications for use and technological characteristics between the predicate and subject devices, but those differences allow for the MOSkin System to have additional use capabilities and improved usability. The testing and validation activities conducted demonstrate that any differences between the devices do not raise new or different questions of safety or effectiveness as compared to the predicate device. Therefore, the MOSkin System is substantially equivalent to the predicate device.
03 510k Summary.docx
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.