Vantage Fortian/Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: - Proton density (PD) (also called hydrogen density) - Spin-lattice relaxation time (T1) - Spin-spin relaxation time (T2) - Flow dynamics - Chemical Shift Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Fortian/Orian 1.5T is a magnetic resonance imaging (MRI) system utilizing 1.5T magnet, Pianissimo/Pianissimo Zen noise reduction, and gradient coils. System acquires proton density, T1, T2, flow, and chemical shift data. New features include 4D Flow for blood flow visualization/quantification; Zoom DWI for reduced FOV diffusion imaging; and Precise IQ Engine (PIQE), a deep learning-based reconstruction algorithm for matrix expansion and artifact reduction. Operated by trained clinicians in clinical settings. PIQE processes raw data to generate higher in-plane resolution images, mitigating ringing and smoothing. Output images are reviewed by physicians to aid diagnosis. System supports clinical decision-making by providing high-quality anatomical images. Benefits include improved image sharpness, reduced scan times, and artifact mitigation.
Clinical Evidence
Multi-site, randomized, blinded clinical image review study conducted with 14 board-certified radiologists/cardiologists. 75 subjects (399 scans) across body, breast, cardiac, MSK, and neuro anatomies. 1197 reconstructions (with/without PIQE) evaluated using a 5-point Likert scale for ringing, sharpness, SNR, and overall IQ. Results showed strong agreement at 'good' and 'very good' levels, confirming PIQE generates higher spatial resolution images, reduces ringing, and maintains clinical utility.
Technological Characteristics
1.5T MRI system; 1.4m short magnet; Pianissimo/Pianissimo Zen noise reduction; XGO or standard gradient systems. Connectivity: DICOM. Software: Deep learning-based reconstruction (PIQE). Standards: IEC 60601-1, IEC 60601-2-33, ISO 10993-1, NEMA MS series.
Indications for Use
Indicated for patients requiring diagnostic cross-sectional imaging of the head or body using MRI. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Vantage Fortian/Orian 1.5T, MRT-1550, V9.0 with AiCE Reconstruction Processing Unit for MR (K240238)
Reference Devices
Vantage Galan 3T, MRT-3020, V10.0 with AiCE Reconstruction Processing Unit for MR (K243335)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 22, 2025
Canon Medical Systems Corporation
% Blake Stacy
Manager, Regulatory Affairs
Canon Medical Systems USA, Inc.
2441 Michelle Drive
Tustin, California 92780
Re: K250901
Trade/Device Name: Vantage Fortian/Orian 1.5T, MRT-1550, V10.0 with AiCE Reconstruction Processing Unit for MR
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH
Dated: July 2, 2025
Received: July 2, 2025
Dear Blake Stacy:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250901 - Blake Stacy
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250901 - Blake Stacy
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ningzhi Li
-S
for
Daniel M. Krainak, PhD
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250901
Device Name
Vantage Fortian/Orian 1.5T, MRT-1550, V10.0 with AiCE Reconstruction Processing Unit for MR
Indications for Use (Describe)
Vantage Fortian/Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical Shift
Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
Canon
CANON MEDICAL SYSTEMS USA, INC.
K250901
Made For life
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
# 1. CLASSIFICATION and DEVICE NAME
| Classification Name: | Magnetic Resonance Diagnostic Device |
| --- | --- |
| Regulation Number: | 90-LNH (Per 21 CFR § 892.1000) |
| Trade Proprietary Name: | Vantage Fortian / Orian 1.5T, MRT-1550, V10.0 with AiCE Reconstruction Processing Unit for MR |
| Model Number: | MRT-1550 |
# 2. SUBMITTER'S NAME
Canon Medical Systems Corporation
1385 Shimoishigami
Otawara-Shi, Tochigi-ken, Japan 324-8550
# 3. OFFICIAL CORRESPONDENT
Naofumi Watanabe
Senior Manager, Regulatory Affairs and Vigilance
Canon Medical Systems Corporation
# 4. CONTACT PERSON, U.S. AGENT and ADDRESS
Contact Person
Blake Stacy
Manager, Regulatory Affairs, Canon Medical Systems USA, Inc.
2441 Michelle Drive, Tustin, CA 92780
Phone: (626) 319-9444
Fax: (714) 730-1310
E-mail: bstacy@us.medical.canon
Official Correspondent/U.S. Agent
Orlando Tadeo, Jr.
Director, Regulatory Affairs, Canon Medical Systems USA, Inc.
2441 Michelle Drive, Tustin, CA 92780
Phone: (714) 669-7459
Fax: (714) 730-1310
E-mail: otadeo@us.medical.canon
2441 Michelle Drive, Tustin, CA 92780
PHONE: 800-421-1968
https://us.medical.canon
{5}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
# 5. MANUFACTURING SITE
Canon Medical Systems Corporation
1385 Shimoishigami
Otawara-shi, Tochigi 324-8550, Japan
# 6. ESTABLISHMENT REGISTRATION
9614698
# 7. DATE PREPARED
March 25, 2025
# 8. DEVICE NAME
Vantage Fortian/Orian 1.5T, MRT-1550, V10.0 with AiCE Reconstruction Processing Unit for MR
# 9. TRADE NAME
Vantage Fortian/Orian 1.5T, MRT-1550, V10.0 with AiCE Reconstruction Processing Unit for MR
# 10. CLASSIFICATION NAME
Magnetic Resonance Diagnostic Device (MRDD)
# 11. CLASSIFICATION PANEL
Radiology
# 12. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)
# 13. PRODUCT CODE
90-LNH
# 14. PREDICATE DEVICE
Predicate Device: Vantage Fortian/Orian 1.5T, MRT-1550, V9.0 with AiCE Reconstruction Processing Unit for MR (K240238)
| System | Predicate Device |
| --- | --- |
| | Vantage Fortian/Orian 1.5T, MRT-1550, V9.0 with AiCE Reconstruction Processing Unit for MR |
| Marketed By | Canon Medical Systems USA, Inc. |
| 510(k) Number | K240238 |
| Clearance Date | April 12, 2024 |
2441 Michelle Drive, Tustin, CA 92780
PHONE: 800-421-1968
https://us.medical.canon
{6}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
# 15. REFERENCE DEVICE
Reference Device: Vantage Galan 3T, MRT-3020, V10.0 with AiCE Reconstruction Processing Unit for MR (K243335)
| System | Reference Device |
| --- | --- |
| | Vantage Galan 3T, MRT-1550, V10.0 with AiCE Reconstruction Processing Unit for MR |
| Marketed By | Canon Medical Systems USA, Inc. |
| 510(k) Number | K243335 |
| Clearance Date | January 7, 2025 |
# 16. REASON FOR SUBMISSION
Modification of a cleared device
# 17. SUBMISSION TYPE
Traditional 510(k) Premarket Notification
# 18. DEVICE DESCRIPTION
The Vantage Fortian (Model MRT-1550/WK, WM, WO, WQ)/Vantage Orian (Model MRT-1550/U3, U4, U7, U8) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. These Vantage Fortian/Orian models use 1.4 m short and 4.1 tons light weight magnet. They include the Canon Pianissimo™ $\Sigma$ and Pianissimo Zen technology (scan noise reduction technology). The design of the gradient coil and the whole-body coil of these Vantage Fortian/Orian models provide the maximum field of view of $55 \times 55 \times 50$ cm and include the standard (STD) gradient system.
The Vantage Orian (Model MRT-1550/ UC, UD, UG, UH, UK, UL, UO, UP) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Orian models use 1.4 m short and 4.1 tons light weight magnet. They include the Canon Pianissimo™ and Pianissimo Zen technology (scan noise reduction technology). The design of the gradient coil and the whole-body coil of these Vantage Orian models provide the maximum field of view of $55 \times 55 \times 50$ cm. The Model MRT-1550/ UC, UD, UG, UH, UK, UL, UO, UP includes the XGO gradient system.
This system is based upon the technology and materials of previously marketed Canon Medical Systems MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body. The Vantage Fortian/Orian MRI System is comparable to the current 1.5T Vantage Fortian/Orian MRI System (K240238), cleared April 12, 2024, with the following modifications.
2441 Michelle Drive, Tustin, CA 92780
PHONE: 800-421-1968
https://us.medical.canon
{7}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
# 19. SUMMARY OF CHANGE(S)
This submission is to report the following changes:
## Summary of Software Changes:
- 4D Flow: 4D Flow Application provides the function of visualization of blood flow conditions by combining with external analytical software. Quantitative analysis such as streamline, path line, and velocity is available. Moreover, either Cine or Retro can be used together in PS3D, and PS images with time-phase information can be obtained.
- Zoom DWI: Zoom DWI Application offers diffusion images of smaller FOV size by selective excitation and outer volume suppression (OVS). This application proposes to remove the distortion and the aliasing artifact in PE direction.
- Precise IQ Engine (PIQE): PIQE is a Deep Learning based technique that generates higher in-plane matrix images from low matrix images while mitigating the ringing artifact. PIQE has been extended to multiple scan families, weightings, and additional anatomical areas.
- System Software: Minor updates were made to the system software.
## Summary of Accessory Changes:
- Breast SPEEDER pad kit: Newly added, optional patient support to be used with Breast SPEEDER RF coil.
# 20. SAFETY PARAMETERS
| Item | Subject Device: Vantage Fortian/Orian 1.5T, MRT-1550, V10.0 with AiCE Reconstruction Processing Unit for MR | Predicate Device: Vantage Fortian/Orian 1.5T, MRT-1550, V9.0 with AiCE Reconstruction Processing Unit for MR | Notes |
| --- | --- | --- | --- |
| Static field strength | 1.5T | 1.5T | Same |
| Operational Modes | Normal and 1st Operating Mode | Normal and 1st Operating Mode | Same |
| i. Safety parameter display | SAR, dB/dt | SAR, dB/dt | Same |
| ii. Operating mode access requirements | Allows screen access to 1st level operating mode | Allows screen access to 1st level operating mode | Same |
| Maximum SAR | 4W/kg for whole body (1st operating mode specified in IEC 60601-2-33: 2010+A1:2013+A2:2015) | 4W/kg for whole body (1st operating mode specified in IEC 60601-2-33: 2010+A1:2013+A2:2015) | Same |
| Maximum dB/dt | 1st operating mode specified in IEC 60601-2-33: 2010+A1:2013+A2:2015 | 1st operating mode specified in IEC 60601-2-33: 2010+A1:2013+A2:2015 | Same |
| Potential emergency condition and means provided for shutdown | Shutdown by Emergency Ramp Down Unit for collision hazard for ferromagnetic objects | Shutdown by Emergency Ramp Down Unit for collision hazard for ferromagnetic objects | Same |
2441 Michelle Drive, Tustin, CA 92780 PHONE: 800-421-1968
https://us.medical.canon
{8}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
# 21. IMAGING PERFORMANCE PARAMETERS
No change from the previous predicate submission, K240238.
# 22. INDICATIONS FOR USE
Vantage Fortian/Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical Shift
Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
*Note: No change from the predicate submission, K240238
# 23. SUMMARY OF DESIGN CONTROL ACTIVITIES
Risk Management activities for this modification are included in this submission. The test methods used are the same as those submitted in the previously cleared submission of the predicate device, Vantage Fortian/Orian 1.5T, MRT-1550, V9.0 with AiCE Reconstruction Processing Unit for MR (K240238). A declaration of conformity with design controls is included in this submission.
2441 Michelle Drive, Tustin, CA 92780
PHONE: 800-421-1968
https://us.medical.canon
{9}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
## 24. SAFETY
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.
This device is based upon the same technologies, materials and software as the predicate device. Risk activities were conducted in concurrence with established medical device development standards and guidance. Additionally, testing was done in accordance with applicable recognized consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA):
### LIST OF APPLICABLE STANDARDS
- ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]
- IEC 60601-1:2005, A1:2012, A2:2020
- IEC 60601-1-2:2014+A1:2020
- IEC 60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020)
- IEC 60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015)
- IEC 60825-1 (2014)
- IEC 62304 (2006), Amd.1 (2015)
- IEC 62366-1 (2020)
- ISO 10993-1 (2018)
- NEMA MS 1:2008 (R2020)
- NEMA MS 2:2008 (R2020)
- NEMA MS 3:2008 (R2020)
- NEMA MS 4 (2010)
- NEMA MS 5 (2018)
## 25. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrate that the system requirements have been met. Additionally, image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria.
**4D Flow** was evaluated utilizing phantom images and representative volunteer images. Bench testing included velocity measurement with and without PIQE of a phantom with known flow values. Additionally, images in volunteers demonstrated velocity stream lines. The testing confirmed that the flow velocity of the 4DFlow sequence met the acceptance criteria.
**Zoom DWI** was evaluated utilizing phantom images and representative volunteer images. Testing confirmed that Zoom DWI is effective for suppressing wraparound artifacts in the PE direction; setting a smaller PE-FOV in Zoom DWI scan can reduce the image distortion level; and the ADC values can be measured accurately.
**PIQE (Precise IQ Engine)** for MR was not altered from the previously cleared version, but was extended to include additional scanning techniques, image weightings, and anatomical locations. PIQE underwent the same performance testing established in the predicate submission (K240238) as applied to each of the new anatomical regions. The same methodology for extension was applied in the reference submission, K243335. PIQE testing on the subject device included bench testing, performed on typical clinical images from Body, Breast, Cardiac, musculoskeletal (MSK), and Neuro, wherein the metrics of Edge Slope Width (to evaluate image sharpness), Ringing Variable Mean (to evaluate ringing artifacts), Signal-to-Noise ratio, and Contrast Ratio were evaluated. Comparing images with PIQE at various
2441 Michelle Drive, Tustin, CA 92780 PHONE: 800-421-1968
https://us.medical.canon
{10}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
scaling factors to standard clinical techniques, such as zero-padding interpolation and typical clinical filters, confirmed that PIQE generates images with sharper edges while mitigating the smoothing and ringing effects and maintaining similar or better contrast and SNR.
# PIQE Performance Testing - Clinical Image Review Study
Additionally, a multi-site, randomized, blinded clinical image review study was conducted with 14 USA board-certified radiologists/cardiologists (3 per anatomy). Using the conventional method (i.e., matrix expansion with Fine Reconstruction and processing with typical clinical filter) as a reference, the images reconstructed with either the conventional method or the new PIQE method were randomized, blinded to the reviewers, and scored by 3 reviewers per anatomy in various clinically-relevant categories (including ringing, sharpness, SNR, overall image quality (IQ), and feature conspicuity) using a modified 5-point Likert scale. A total of 75 unique subjects, from two sites in USA and Japan, were scanned in body, breast, cardiac, musculoskeletal, or neuro to provide the test data sets (separate from the training data sets) using Vantage Orian or Fortian 1.5T and anatomy-appropriate coils, comprising a total of 399 scans, in multiple orientations, and multiple contrast weightings within the multiple families of pulse sequences. Each scan was reconstructed multiple ways with or without PIQE, totaling 1197 reconstructions to be scored. Reviewer scoring data was analyzed for reviewer agreement and differences between reconstruction techniques using Gwet's Agreement Coefficient and Generalized Estimating Equations (GEE), respectively. The resulting reconstructions were scored on average at, or above, clinically acceptable. Exhibiting a strong agreement at the "good" and "very good" level in the IQ metrics, the Reviewers' scoring confirmed that: (a) PIQE generates higher spatial in-plane resolution images from lower resolution images (with the ability to triple the matrix dimensions in both in-plane directions, i.e. a factor of 9x); (b) PIQE contributes to ringing artifact reduction, denoising and increased sharpness; (c) PIQE is able to accelerate scanning by reducing the acquisition matrix only, while maintaining clinical matrix size and image quality; and (d) PIQE benefits can be obtained on regular clinical protocols without requiring acquisition parameter adjustment.
Although the dataset of this multinational study includes subjects from outside the USA, the population is expected to be representative of the intended US population as it pertains to the intended use of the PIQE product. PIQE is an image post-processing algorithm that is not disease-specific and is not dependent on factors that would be relevant for scan acquisition or diagnosis-specific algorithms. The PIQE algorithm is not directly dependent on acquisition parameters that might be affected by population variation (such as body habitus). Indeed, the current study design is based on reconstructing each k-space acquisition multiple times (both with and without PIQE) such that all comparisons are internal, with each subject as its own control. So, patient-variable influences should be removed as a factor. For PIQE, where matrix expansion (pixel size) and edge sharpness are key, the relevant characteristics are primarily tissue structure, feature definition, and edge definition; and to some extent matrix size and FOV. Thus, related to the anatomies of interest, relevant features do not vary significantly from country to country. Further, even if the internal structures themselves are different, or differently distributed, across countries, the edge definition of such structures are not expected to vary differently patient-to-patient due to nationality.
2441 Michelle Drive, Tustin, CA 92780 PHONE: 800-421-1968
https://us.medical.canon
{11}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
Software Documentation for a Basic Documentation Level, per the FDA guidance document, "Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, is included in this submission. This documentation includes justification for the Basic Documentation Level determination as well as testing which demonstrates that the verification and validation requirements have been met.
Cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" issued on September 27, 2023, is also included as part of this submission.
# 26. SUBSTANTIAL EQUIVALENCE
Canon Medical Systems Corporation believes that the changes incorporated into the Vantage Fortian/Orian 1.5T, MRT-1550, V10.0, Magnetic Resonance Imaging (MRI) System with AiCE Reconstruction Processing Unit for MR, are substantially equivalent to the previously cleared predicate device.
# 27. CONCLUSION
The modifications incorporated into the Vantage Fortian/Orian 1.5T, MRT-1550, V10.0, Magnetic Resonance Imaging (MRI) System with AiCE Reconstruction Processing Unit for MR, do not change the indications for use or the intended use of the device. Based upon the evidence presented, Canon Medical Systems Corporation considers the subject Vantage Fortian / Orian 1.5T device to be as safe, as effective, and with performance that is substantially equivalent to the predicate device.
2441 Michelle Drive, Tustin, CA 92780 PHONE: 800-421-1968
https://us.medical.canon
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.