← Product Code [KRH](/productcode/KRH) · K250859

# TransForm McCarthy Mitral Annuloplasty Ring (TF) (K250859)

_Genesee Biomedical, Inc. · KRH · Jul 25, 2025 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250859

## Device Facts

- **Applicant:** Genesee Biomedical, Inc.
- **Product Code:** [KRH](/productcode/KRH.md)
- **Decision Date:** Jul 25, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

TransForm™ McCarthy Mitral Annuloplasty Ring (TF) is indicated for use in patients undergoing surgery for diseased or damaged mitral valves. The ring provides support for and restricts expansion of the mitral annulus.

## Device Story

Implantable, semi-rigid annuloplasty ring for mitral valve repair; used in surgical settings by cardiac surgeons. Device features MP35N alloy stiffeners within an elastic silicone core and braided polyester fabric. Stiffeners allow maintenance of 'D' saddle shape during systole and flat planar 'O' shape in diastole; provides semi-rigid elastic flexibility to accommodate annular remodeling while restricting expansion. Ring includes four green radial markers for orientation. Sizes range from 24mm to 40mm. Implanted on the mitral annulus to reduce and stabilize geometry, benefiting patients by restoring proper valve function.

## Clinical Evidence

No clinical data. Evidence consists of bench testing including braid tensile strength, radial seam tensile test, suture pull-out strength, sterilization validation (ISO 11137-1), and comprehensive biocompatibility testing (ISO 10993 series, USP standards) including cytotoxicity, pyrogenicity, sensitization, irritation, systemic toxicity, implantation, and hemocompatibility. Chemical characterization and toxicological risk assessment (ISO 10993-17/18) performed for toxicity, genotoxicity, and carcinogenicity.

## Technological Characteristics

Semi-rigid annuloplasty ring; MP35N alloy stiffeners (0.71 mm diameter), elastic silicone core, braided polyester fabric. Sizes 24-40mm (2mm increments). Four green radial markers. Sterilization per ISO 11137-1. Biocompatibility per ISO 10993 and USP standards. Hemocompatibility per ASTM F756, F2888, F2382.

## Regulatory Identification

An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

## Predicate Devices

- TransForm™ McCarthy Annuloplasty Ring (TF) ([K232599](/device/K232599.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

July 25, 2025

Genesee Biomedical Inc
James Cross
Quality Assurance Manager
700 W. Mississippi Ave Unit D-5
Denver, Colorado 80223

Re: K250859
Trade/Device Name: TransForm™ McCarthy Mitral Annuloplasty Ring (TF)
Regulation Number: 21 CFR 870.3800
Regulation Name: Annuloplasty ring
Regulatory Class: Class II
Product Code: KRH
Dated: June 26, 2025
Received: June 26, 2025

Dear James Cross:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}

K250859 - James Cross
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K250859 - James Cross
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JULIE B. MACKEL -S

For
Samuel Raben, PhD
Acting Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K250859  |   |
|  Device Name TransForm™ McCarthy Mitral Annuloplasty Ring (TF)  |   |
|  Indications for Use (Describe) TransForm™ McCarthy Mitral Annuloplasty Ring (TF) is indicated for use in patients undergoing surgery for diseased or damaged mitral valves. The ring provides support for and restricts expansion of the mitral annulus.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K250859 Page 1 of 2

|  510(k) #: K250859 | 510(k) Summary | Prepared on: 2025-06-26  |
| --- | --- | --- |
|  Contact Details |   | 21 CFR 807.92(a)(1)  |
|  Applicant Name | Genesee Biomedical Inc |   |
|  Applicant Address | 700 W. Mississippi Ave Unit D-5 Denver CO 80223 United States |   |
|  Applicant Contact Telephone | 303-777-3000 |   |
|  Applicant Contact | Mr. James Cross |   |
|  Applicant Contact Email | jcross@geneseebiomedical.com |   |
|  Device Name |   | 21 CFR 807.92(a)(2)  |
|  Device Trade Name | TransForm™ McCarthy Mitral Annuloplasty Ring (TF) |   |
|  Common Name | Annuloplasty ring |   |
|  Classification Name | Ring, Annuloplasty |   |
|  Regulation Number | 870.3800 |   |
|  Product Code(s) | KRH, DTI |   |
|  Legally Marketed Predicate Devices |   | 21 CFR 807.92(a)(3)  |
|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
|  K232599 | TransForm™ McCarthy Annuloplasty Ring (TF) | KRH  |
|  Device Description Summary |   | 21 CFR 807.92(a)(4)  |
|  The TransForm™ McCarthy Mitral Annuloplasty Ring is an implantable, semi-rigid annuloplasty ring for mitral valve repair with the stiffened portions at the anterior and posterior segments. The ring is flexible at the commissure regions to allow physiological movement of the annulus. Stiffener placement allows the ring to maintain its “D” saddle shape during systole and a flat planar “O” shape in diastole. The stiffeners consist of MP35N 0.71 mm diameter (polished and formed) alloy. An elastic silicone core and braided polyester fabric form the ring body. The TransForm™ Annuloplasty Ring has four green radial markers: two markers at the left and right trigones and two markers at the mid-anterior and mid-posterior. The ring conforms to the natural mitral annulus throughout the cardiac cycle. The interrupted stiffeners within the silicone core provide semi-rigid elastic flexibility. Elasticity of the “D” to “O” or circular transformation, accommodates remodeling of the annulus, while creating inward forces to maintain circular annular geometry against physical expansion. The size range of TransForm™ is from 24mm to 40mm with 2mm increments. Size refers to the internal CC diameter of the ring. Rings are intended to be implanted on the patient's mitral annulus to reduce and stabilize the annulus.  |   |   |
|  Intended Use/Indications for Use |   | 21 CFR 807.92(a)(5)  |
|  TransForm™ McCarthy Mitral Annuloplasty Ring is indicated for use in patients undergoing surgery for diseased or damaged mitral valves. The ring provides support for and restricts expansion of the mitral annulus.  |   |   |
|  Indications for Use Comparison |   | 21 CFR 807.92(a)(5)  |
|  A comparison of the characteristics of the proposed device and the predicate and/or reference device show that there are no differences between the subject device and the predicate device with respect to clinical aspects, indications and intended use. Both devices have the same principle and mechanism of operation. The TransForm™ McCarthy Mitral Annuloplasty Ring is determined to be substantially equivalent to the predicate device.  |   |   |

{5}

K250859 Page 2 of 2

# Technological Comparison

21 CFR 807.92(a)(6)

Technological aspects were assessed for substantial equivalence. The subject device and the predicate devices have the same size, shape and similar material composition. The subject device, the TransForm™ McCarthy Annuloplasty Ring, will be manufactured under equivalent conditions as the predicate device.

# Non-Clinical and/or Clinical Tests Summary &amp; Conclusions

21 CFR 807.92(b)

Structural Analysis

Braid tensile strength

Radial Seam tensile test

Suture pull out/tensile strength

Sterilization validation per ISO 11137-1

Biocompatibility testing was completed as follows:

- Cytotoxicity per ISO 10993-5

- Material-Mediated Pyrogenicity per USP &lt;151&gt;

- Sensitization per USP 10993-10

- Intracutaneous Irritation per ISO 10993-23

- Acute Systemic Toxicity per USP 10993-11

- 12 week Implantation test per USP 10993-6

- Hemocompatibility per ISO 10993-4

- Hemolysis per ASTM F756

- Leukocyte and Platelet Assay per ASTM F2888

- Partial Thromboplastin Time per ASTM F2382

- Compliment Activation Assay per ISO 10993-4

Chemical characterization testing per ISO 10993-18 and toxicological risk assessment per ISO 10993-17 was completed to address the following:

- Toxicity (sub-acute/sub-chronic &amp; chronic)

- Genotoxicity

- Carcinogenicity

Genesee BioMedical, Inc. considers the TransForm™ McCarthy Mitral Annuloplasty Ring to be substantially equivalent to the predicate device. This conclusion is based upon the fact that devices have an identical intended use, and there are no clinical, technical or biological differences, in performance testing that raise new questions of safety and effectiveness. Nonclinical testing demonstrates that the construction braid tensile strength exceeds the pass/fail criteria; that the worst cases for each specific stiffener size have higher safety factors than the absolute worst case used for evaluation in this FEA and that the braid material used to construct the TransForm™ McCarthy Mitral Annuloplasty Ring provides a suture pullout strength that exceeding the minimum pass/fail criteria.

---

**Source:** [https://fda-staging.innolitics.com/device/K250859](https://fda-staging.innolitics.com/device/K250859)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).