← Product Code [MHX](/productcode/MHX) · K250854

# Multi-parameter Patient Monitor (N10);  Multi-parameter Patient Monitor (N12);  Multi-parameter Patient Monitor (N15);  Multi-parameter Patient Monitor (N10MPro);  Multi-parameter Patient Monitor (N12MPro);  Multi-parameter Patient Monitor (N15MPro) (K250854)

_Shenzhen Comen Medical Instruments Co.,Ltd · MHX · Dec 17, 2025 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250854

## Device Facts

- **Applicant:** Shenzhen Comen Medical Instruments Co.,Ltd
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Dec 17, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The N10, N12, N15, N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 12-lead selectable), Arrhythmia Analysis, ST Segment Analysis, QT Analysis, Heart Rate (HR) and Heart-Rate-Variability(HRV)), interpretations of resting 12-lead ECG, Respiration rate(Resp), Temperature(Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2). The N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are also intended for monitoring, displaying, reviewing, storing, alarming and transferring of physiological parameters including Masimo Rainbow SpO2, Anesthesia gas (AG), oxygen (O2) respiratory gas monitoring, Bispectral Index (BIS), Respiration Mechanics (RM) and Neuromuscular Transmission Monitoring (NMT). All the parameters can be monitored on single adult, pediatric, and neonatal patient except for the following: - Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population. - NIBP measurement continual mode is not applicable to neonates. - When using COMEN SpO2, the monitor is intended to be used on adult patient only. - PAWP is intended for adult and pediatric patients only. - C.O. measurement is intended for adult patients only. - BIS monitoring is intended for adult patients only. - RM is intended for adult and pediatric patients only. - NMT monitoring is intended for adult and pediatric patients only. The monitors are to be used in healthcare facilities by healthcare professionals or under their guidance. The Multi-parameter Patient monitors are not intended for emergency and transport use, aircraft environment or home use. The monitors are not intended for use as apnea monitors. The monitors are not intended for use in MRI or CT environments. The monitors are not used on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.

## Device Story

Multi-parameter patient monitors (N10, N12, N15, N10MPro, N12MPro, N15MPro) collect physiological signals including ECG, SpO2, NIBP, IBP, CO2, and respiratory gases; process data via internal algorithms for real-time display, storage, and alarm generation; used in clinical healthcare facilities by professionals. Output includes waveforms and numeric values for clinical decision support; aids in patient status assessment. Does not support emergency/transport, home, or MRI/CT environments. Provides advisory interpretations for 12-lead ECG, ST-segment, and QT intervals; clinicians must verify findings against patient history. Benefits include continuous monitoring of vital signs to detect physiological changes.

## Clinical Evidence

Bench testing only. No new clinical accuracy validation performed for SpO2, respiratory rate, or NIBP; leverages previously cleared clinical data from predicate/reference devices. ECG arrhythmia algorithm validated against standard expert-annotated databases (AHA, MIT-BIH, etc.) per AAMI/ANSI EC57:2012.

## Technological Characteristics

Multi-parameter monitor; rechargeable Li-ion battery or AC power; eMMC storage; internal thermal recorder. Sensing: trans-thoracic impedance (Resp), infrared absorption (CO2/AG), thermodilution (C.O.), oscillometric (NIBP). Connectivity: networked/standalone. Software: embedded, rule-based and algorithmic analysis. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-1, AAMI EC57.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- BeneVision N series Patient Monitors ([K213799](/device/K213799.md))

## Reference Devices

- [K211619](/device/K211619.md) - Multi-Parameter Patient Monitor
- [K193242](/device/K193242.md) - Masimo Radical-7 Pulse CO-Oximeter and Accessories
- [K123043](/device/K123043.md) - IRMA - Infrared Mainstream Gas Analyzer
- [K171121](/device/K171121.md) - Masimo Root Monitoring system and accessories
- [K042601](/device/K042601.md) - Respironics Novametrix LLC, Capnostat CO2
- [K103604](/device/K103604.md) - ISA – infrared Sidestream gas Analyzer
- [K040183](/device/K040183.md) - BISX, Aspect medical systems Inc.
- [K102084](/device/K102084.md) - STIM POD NMS45O Nerve Stimulator
- [K231160](/device/K231160.md) - Cardio Q50, Cardio Q70

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 17, 2025

Shenzhen Comen Medical Instruments Co., Ltd.
Charlotte Lin
Registration Engineer
Fl 10, Fl 11 and Sec C of Fl 12 of Bldg 1A &amp; Fl 1 to Fl 5
of Bldg 2, FIYTA Timepiece Bldg, Nanhuan Ave Matian Sub-dist
Shenzhen, 518106
China

Re: K250854

Trade/Device Name: Multi-parameter Patient Monitor (N10); Multi-parameter Patient Monitor (N12); Multi-parameter Patient Monitor (N15); Multi-parameter Patient Monitor (N10MPro); Multi-parameter Patient Monitor (N12MPro); Multi-parameter Patient Monitor (N15MPro)

Regulation Number: 21 CFR 870.1025

Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)

Regulatory Class: Class II

Product Code: MHX, DSI, MLD, DSJ, DSK, DXN, DRT, MLC, DQA, DSB, MUD, MSX, KRC, DSF, DRS, FLL, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, KOI, OLW, OLT, OMC, ORT

Dated: March 19, 2025

Received: March 21, 2025

Dear Charlotte Lin:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

U.S. Food &amp; Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993

www.fda.gov

{1}

K250854 - Charlotte Lin
Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-

{2}

K250854 - Charlotte Lin
Page 3

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JENNIFER W. SHIH -S

Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250854

Device Name
Multi-parameter Patient Monitor (N10);
Multi-parameter Patient Monitor (N12);
Multi-parameter Patient Monitor (N15);
Multi-parameter Patient Monitor (N10MPro);
Multi-parameter Patient Monitor (N12MPro);
Multi-parameter Patient Monitor (N15MPro)

Indications for Use (Describe)
The N10, N12, N15, N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 12-lead selectable), Arrhythmia Analysis, ST Segment Analysis, QT Analysis, Heart Rate (HR) and Heart-Rate-Variability(HRV)), interpretations of resting 12-lead ECG, Respiration rate(Resp), Temperature(Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2). The N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are also intended for monitoring, displaying, reviewing, storing, alarming and transferring of physiological parameters including Masimo Rainbow SpO2, Anesthesia gas (AG), oxygen (O2) respiratory gas monitoring, Bispectral Index (BIS), Respiration Mechanics (RM) and Neuromuscular Transmission Monitoring (NMT). All the parameters can be monitored on single adult, pediatric, and neonatal patient except for the following:
- Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population.
- NIBP measurement continual mode is not applicable to neonates.
- When using COMEN SpO2, the monitor is intended to be used on adult patient only.
- PAWP is intended for adult and pediatric patients only.
- C.O. measurement is intended for adult patients only.
- BIS monitoring is intended for adult patients only.
- RM is intended for adult and pediatric patients only.
- NMT monitoring is intended for adult and pediatric patients only.
The monitors are to be used in healthcare facilities by healthcare professionals or under their guidance.
The Multi-parameter Patient monitors are not intended for emergency and transport use, aircraft environment or home use.
The monitors are not intended for use as apnea monitors.
The monitors are not intended for use in MRI or CT environments.
The monitors are not used on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}

510(k) Premarket Notification
510(k) Summary
K250854

# 510(k) Summary

N10, N12, N15, N10MPro, N12MPro and N15MPro Multi-parameter Patient Monitor
This 510(k) Summary is provided in accordance with the requirements of 21 CFR 807.92.

# 1. Submitter

Date: 17 December 2025

Submitter: Shenzhen Comen Medical Instrument Co., Ltd.

Address: Floor 10, Floor 11 and Section C of Floor 12 of Building 1A &amp; Floor 1 to Floor 5 of Building 2, FIYTA Timepiece Building, Nanhuan Avenue, Matian Sub-district, Guangming District, Shenzhen, 518106, Guangdong, China.

Contact: Weiqi Qiu

Telephone: +86-13286650485

Facsimile: +86-755-3431232

E-mail: qiuweiqi@szcomen.com

# 2. Device

Device trade name: Multi-parameter Patient Monitor (N10, N12, N15, N10MPro, N12MPro and N15MPro)

Common name: Multi-parameter Patient Monitor

Classification name: 21 CFR 870.1025, Class II, Arrhythmia detector and alarm (including ST-segment measurement and alarm)

Regulatory Class: Class II

Primary Product code: MHX - Monitor, Physiological, Patient (with arrhythmia detection or alarms)

Table 1: Secondary Product Codes

|  Device Panel | Regulation number & Classification | Product code | Description  |
| --- | --- | --- | --- |
|  Cardiovascular | §870.1025, II | DSI | Arrhythmia detector and alarm (including ST-segment measurement and alarm).  |
|  Cardiovascular | §870.1025, II | MLD | Arrhythmia detector and alarm (including ST-segment measurement and alarm).  |
|  Cardiovascular | §870.1100, II | DSJ | Alarm, Blood-Pressure  |
|  Cardiovascular | §870.1110, II | DSK | Computer, Blood-Pressure  |
|  Cardiovascular | §870.1130, II | DXN | System, Measurement, Blood-Pressure, Non-Invasive  |
|  Cardiovascular | §870.2300, II | DRT | Monitor, Cardiac (Including  |

Page 1 of 22

{5}

510(k) Premarket Notification
510(k) Summary
K250854

|  Device Panel | Regulation number & Classification | Product code | Description  |
| --- | --- | --- | --- |
|   |  |  | Cardiotachometer & Rate Alarm)  |
|  Cardiovascular | §870.2340, II | MLC | Monitor, ST segment  |
|  Anesthesiology | §870.2700, II | DQA | Oximeter  |
|  Cardiovascular | §870.2770, II | DSB | Plethysmograph, Impedance  |
|  Cardiovascular | §870.2770, II | MUD | Oximeter, tissue saturation  |
|  Cardiovascular | §870.2300/II | MSX | Cardiac monitor (including cardiotachometer and rate alarm).  |
|  Cardiovascular | §870.2370, II | KRC | Tester, electrode, surface, electrocardiographic  |
|  Cardiovascular | §870.2810, I | DSF | Recorder, Paper Chart  |
|  Cardiovascular | §870.2850, II | DRS | Extravascular blood pressure transducer.  |
|  General Hospital | §880.2910, II | FLL | Thermometer, Electronic, Clinical  |
|  Anesthesiology | §868.1400, II | CCK | Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase  |
|  Anesthesiology | §868.1500, II | CBQ | Analyzer, Gas, Enflurane, Gaseous-Phase (Anesthetic Concentration)  |
|  Anesthesiology | §868.1500, II | NHO | Analyzer, Gas, Desflurane, Gaseous-Phase (Anesthetic Concentration)  |
|  Anesthesiology | §868.1500, II | NHP | Analyzer, Gas, Sevoflurane, Gaseous-Phase (Anesthetic Concentration)  |
|  Anesthesiology | §868.1500, II | NHQ | Analyzer, Gas, Isoflurane, Gaseous-Phase (Anesthetic Concentration)  |
|  Anesthesiology | §868.1620, II | CBS | Analyzer, Gas, Halothane, Gaseous-Phase (Anesthetic Conc.)  |
|  Anesthesiology | §868.1700, II | CBR | Analyzer, Gas, Nitrous-Oxide, Gaseous Phase (Anesthetic Conc.)  |
|  Anesthesiology | §868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-Phase  |
|  Anesthesiology | §868.2775, II | KOI | Electrical peripheral nerve stimulator.  |
|  Neurology | §882.1400, II | OLW | Index-Generating Electroencephalograph Software  |
|  Neurology | §882.1400, II | OLT | Non-normalizing quantitative electroencephalograph software  |
|  Neurology | §882.1400, II | OMC | Reduced- montage standard electroencephalograph  |
|  Neurology | §882.1400, II | ORT | Burst suppression detection software for electroencephalograph  |

Page 2 of 22

{6}

510(k) Premarket Notification
510(k) Summary
K250854

# 3. Predicate device

K213799 - BeneVision N series Patient Monitors (including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22), Shenzhen Mindray Bio-medical Electronics Co., LTD.

# 4. Reference devices

K211619 - Multi-Parameter Patient Monitor, Shenzhen Comen Medical Instruments Co., LTD.
K193242 - Masimo Radical-7 Pulse CO-Oximeter and Accessories, Masimo Corporation.
K123043 - IRMA - Infrared Mainstream Gas Analyzer, Phasein AB.
K171121 - Masimo Root Monitoring system and accessories, Masimo Corporation.
K042601 - Respironics Novametrix LLC, Capnostat CO₂.
K103604 - ISA – infrared Sidestream gas Analyzer, Phasein AB.
K040183 - BISX, Aspect medical systems Inc.
K102084 - STIM POD NMS45O Nerve Stimulator, XAVANT TECHNOLOGY(PTY) LTD.
K231160 - Cardio Q50, Cardio Q70, Bionet Co., Ltd.

# 5. Device Description

There are six (6) models under evaluation, namely N10, N12, N15, N10MPro, N12MPro, N15MPro. All models share the same intended condition of use, the same intended patient population and operator profile, biological safety characteristic and principle of operation. All these models are the same on electric and electrical circuit and components, mechanical construction, software and alarm system. The only difference lies on the screen and configuration of with/without plug-in module slot and the number of battery packs.

# 6. Indications for Use

The N10, N12, N15, N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, and 12-lead selectable), Arrhythmia Analysis, ST Segment Analysis, QT Analysis, Heart Rate (HR) and Heart-Rate-Variability(HRV), interpretations of resting 12-lead ECG, Respiration rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO₂), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), and Carbon Dioxide (CO₂). The N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are also intended for monitoring, displaying, reviewing, storing, alarming and transferring of physiological parameters including Masimo Rainbow SpO₂, Anesthesia gas (AG), oxygen (O₂) respiratory gas monitoring, Bispectral Index (BIS), Respiration Mechanics (RM), and Neuromuscular Transmission Monitoring (NMT). All the parameters can be monitored on a single adult, pediatric, or

Page 3 of 22

{7}

510(k) Premarket Notification
510(k) Summary
K250854

neonatal patient except for the following:

- Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population.
- NIBP measurement continual mode is not applicable to neonates.
- When using COMEN SpO₂, the monitor is intended to be used on adult patient only.
- PAWP is intended for adult and pediatric patients only.
- C.O. measurement is intended for adult patients only.
- BIS monitoring is intended for adult patients only.
- RM is intended for adult and pediatric patients only.
- NMT monitoring is intended for adult and pediatric patients only.

The monitors are to be used in healthcare facilities by healthcare professionals or under their guidance.

The Multi-parameter Patient monitors are not intended for emergency and transport use, aircraft environment, or home use.

The monitors are not intended for use as apnea monitors.

The monitors are not intended for use in MRI or CT environments.

The monitors are not intended for use on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.

Please note the following use limitations:

- The ECG analysis result is not intended to be the sole means of diagnostics for conditions such as prolongation of the QT interval, syndrome associated with torsades de pointes (TDP), etc. It is offered to the clinicians on an advisory basis in conjunction with the clinicians' knowledge of the ECG pattern, patient background, clinical history, symptoms and other diagnostic information.
- The clinical use of QT interval measured by the subject device should not replace a trained clinician using diagnostic 12-lead ECG in a clinical environment.
- The clinical significance of the ST level change information should always be determined by clinicians. This feature is intended to aid the clinicians in monitoring the change in ST only and not for diagnosis of STEMI or ischemia events.
- The patient monitor provides a HRV parameter which is only a mathematical analysis and is not intended to produce any interpretation of those measurements or any kind of diagnosis including but not limited to, predicting sudden cardiac death and evaluating the pathological state related to the activity and balance of cardiac autonomic nerves.

# 7. Technological Comparison to Predicate Devices

Both the subject device and the predicate device provide a means for monitoring a patient, collecting specific physiological data, processing the data for alarm conditions and display numeric values and waveforms. In terms of indications for use, N series patient monitors are a subset of the predicate devices indications except for inclusion of the Masimo rainbow SpO₂. Although this feature is not present in the predicate devices, it is presented in the Masimo Radical-7 Pulse CO-Oximeter and Accessories (K193242, Masimo Corporation) reference device.

Both the subject devices and predicate devices can provide monitoring such as ECG(3-lead, 5-lead or 12-

{8}

510(k) Premarket Notification
510(k) Summary
K250854

lead selectable arrhythmia analysis, ST segment analysis, QT analysis, heart rate (HR) and Heart-Rate-Variability(HRV), Respiration rate(Resp), Temperature (Temp), $\mathrm{SpO}_2$, pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), PAWP, cardiac output (C.O.), carbon dioxide $(\mathrm{CO}_2)$, Oxygen $(\mathrm{O}_2)$, anesthetic gas (AG), Bispectral index (BIS), Respiration Mechanics (RM) and Neuromuscular Transmission Monitoring (NMT), and provide an interpretations of resting 12-lead ECG for a single patient.

The N10, N12, N15, N10MPro, N12MPro, NM15Pro Patient Monitors restricted the use of arrhythmia analysis, BIS monitoring and COMEN $\mathrm{SpO}_2$ on adults only, which is a subset of BeneVision N series patient monitor. These do not affect the substantial equivalence of subject device and predicate device.

Table below compares the key technological features of the subject devices to the predicate device (K213799). The features in gray are different between the predicate devices and the subject devices.

Page 5 of 22

{9}

510(k) Premarket Notification

510(k) Summary

K250854

Table 2: Device Comparison Table

|  No. | Feature | Predicate Device BeneVision N series Patient Monitors (K213799), | Subject Device N series Patient Monitor  |
| --- | --- | --- | --- |
|  1. | Model | BeneVision N22/ N19/N17/N15/N12 | N10/ N12/ N15/ N10MPro/ N12MPro/NM15Pro  |
|  2. | Power requirements | Battery or AC power | Battery or AC power  |
|  3. |   | 100 to 240 Va.c. (±10%), 50/60 Hz (±3 Hz) N22/N19: 2.8 A to 1.6 A N17/N15/N12: 2.0 A to 0.9 A | 100-240V-, 50/60Hz N10, N12, N10MPro, N12MPro: 1-0.5A N15, N15MPro: 1.5-0.75A  |
|  4. | Battery | N22/N19: Rechargeable lithium-ion battery, 10.8 V/ 5600 mAh N17/N15/N12: Rechargeable lithium-ion battery, 10.95 V/4500mAh | Standard rechargeable Lithium-ion battery: 10.8V /2500mAh for N10,N10MPro, N12, N12MPro; 10.8V/5000mAh for N15, N15MPro Optional rechargeable Lithium-ion battery: 10.8V /5000mAh  |
|  5. | Data storage | N22/ N19: Solid State Hard Drive(SDD) N17/ N15/N12: Embedded Multi Media Card (eMMC). | Embedded Multi Media Card (eMMC).  |
|  6. | Data Recorder | N22/ N19: Supports the thermal recorder module, needs to be plugged in the SMR. N17/ N15/N12: Supports the thermal recorder module and the built-in thermal recorder, but they cannot work at the same time. | Supports internal thermal recorder.  |
|  7. | Alarm system | The alarm lamp will light cyan, yellow, or red depending on alarm type. | The alarm indicator light are cyan, yellow, or red depending on alarm priority.  |
|  8. | ECG | 3-lead, 5-lead, 6-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, an interpretation of resting 12-lead ECG, J point Auto detection, Dual Channel Pace detection, adjustable QRS threshold, Multi-lead ECG synchronization analysis and heart rate (HR). interpretations of resting 12-lead ECG | Support 3-lead, 5-lead, 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, an interpretation of resting 12-lead ECG, J point Auto detection, single Channel Pace detection, Multi-lead ECG synchronization analysis, Heart Rate and Heart Rate variability (HRV), interpretations of resting 12-lead ECG  |

{10}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|  9. |  | HR measurement range:
15~350 bpm (neonate, pediatric), 15~300 bpm(adult); | HR measurement range:
Adult: 15-300 bpm, Neonate/pediatric:15-350 bpm  |
|  10. |   | HR accuracy:
±1 bpm or ±1%, whichever is greater. | HR accuracy:
±1 bpm or ±1%, whichever is greater.  |
|  11. |   | QT/QTc Measurement range: 200 to 800 ms | QT/QTc Measurement range: 200ms - 800ms  |
|  12. |   | QT Measurement accuracy: ±30 ms | QT Measurement accuracy: ±30ms  |
|  13. |   | ST measurement Range:
-2.0 mV to +2.0 mV RTI | ST measurement Range:
-2.5mV-+2.5mV (-25.0 mm-+25.0 mm)  |
|  14. |   | ST measurement Accuracy:
-0.8 mV- +0.8 mV: ±0.02 mV or ±10%, whichever is greater
Other range: not specified | ST measurement Accuracy:
-0.8mV-+0.8mV; measurement error: ±0.02mV or ±10%, whichever is greater;
Other range: not defined  |
|  15. | Respiration | Method: Trans-thoracic impedance | Method: Trans-thoracic impedance  |
|  16. |   | Measurement range:
Adult: 0 to 120 rpm
Pediatric, neonate: 0 to 150 rpm | Measurement range:
Adult: 7-120rpm
Pediatrics: 7-150rpm
Neonate: 7-150rpm  |
|  17. |   | Accuracy:
7 to 150 rpm: ±2 rpm or ±2%, whichever is greater
0 to 6 rpm: not specified | Accuracy: ±2 rpm  |
|  18. | SpO_{2} | Support Mindray SpO_{2}, Masimo SpO_{2}, Nellcor SpO_{2} modules | Support COMEN SpO_{2}, Masimo SpO_{2}, Masimo Rainbow SpO_{2}, Nellcor SpO_{2} modules  |
|  19. |   | Mindray SpO_{2}, range: 0-100 % | COMEN SpO_{2} range: 0% - 100%  |
|  20. |   | Mindray SpO_{2} accuracy:
70%-100%: ±2% ABS(Adult/pediatric); 70%-100%: ±3% | COMEN SpO_{2} accuracy:
70% - 100%: ±3% in Adult mode (non-motion),  |

Page 7 of 22

{11}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|   |  | ABS (neonate); 0-69%: not specified. | Within rang 0% - 69%: not defined.  |
|  21. |   | Masimo SpO_{2} range: 1% - 100% | Masimo SpO_{2} range: 1% - 100%  |
|  22. |   | Masimo SpO_{2} accuracy:
70 to 100%: ±2%ABS (adult/pediatric, without motion)
70 to 100%: ±3%ABS (adult/pediatric, with motion)
70 to 100%: ±3%ABS (neonate)
1% to 69%: Not specified. | Masimo SpO_{2} accuracy: 70% - 100%SpO_{2} :
±2% in Adu./Ped. Mode (non-motion),
±3% in Adu./Ped. mode (during motion)
±3% in Neo. mode (motion and non-motion);
within range 1% - 69%: not defined.  |
|  23. |   | Nellcor SpO_{2} range: 0 to 100% | Nellcor SpO_{2} range: 0% - 100%  |
|  24. |   | Nellcor SpO_{2} accuracy:
70 to 100%: ±2%ABS (adult/pediatric)
70 to 100%: ±3%ABS (neonate)
0 to 69%: Not specified | Nellcor SpO_{2} accuracy:
70% - 100%SpO_{2} : ±2% in Adu./Ped. mode (non-motion)
±3% in Neo. mode (non-motion);
within range 0% - 69%: not defined.  |
|  25. |   | Not supported | Masimo Rainbow SpO_{2} (only for NMPro series model)  |
|  26. | PR | PR from SpO_{2}
Range: 20-254 bpm
Accuracy: ±3 bpm | PR from SpO_{2}
Range: 20bpm-300bpm
Accuracy: ±2bpm  |
|  27. |   | PR from Masimo SpO_{2}
Range: 25 to 240 bpm
Accuracy: ±3 bpm (without motion) ±5 bpm (with motion) | PR from Masimo SpO_{2}
Range: 25bpm-240bpm
Accuracy: ±3bpm (non-motion) or ±5bpm (during motion)  |
|  28. |   | PR form Nellcor SpO_{2}
Range: 20 to 300 bpm
Accuracy: 20 to 250 bpm: ±3 bpm
251 to 300 bpm: not specified | PR form Nellcor SpO_{2}
Range: 20bpm-300bpm
Accuracy: 20bpm-250bpm: ±3bpm;
251 bpm-300 bpm: not defined  |
|  29. |   | PR form Masimo Rainbow SpO_{2} (Only for NMPro Series models) | PR form Masimo Rainbow SpO_{2} (Only for NMPro Series models)  |

Page 8 of 22

{12}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|   |  | Range: not supported
Accuracy: not supported | Range: 25bpm-240bpm
Accuracy: ±3bpm  |
|  30. |  | PR from NIBP
Range: 30 to 300 bpm
Accuracy: ±3 bpm or ±3%, whichever is greater | PR from NIBP
Range: 30bpm-300bpm
Accuracy: ±3bpm or ±3%, whichever is greater.  |
|  31. |  | PR from IBP
Range: 25 to 350 bpm
Accuracy: ±1 bpm or ±1%, whichever is greater | PR from IBP
Range: 25bpm-350bpm
Accuracy: ±1bpm or ±1%, whichever is greater  |
|  32. | Temperature | Thermal resistance, Direct mode | Thermal resistance, Direct mode  |
|  33. |  | Range: 0 to 50 °C (32 to 122 °F) | Range: 0°C -50°C (32°F-122°F)  |
|  34. |  | Accuracy: ±0.1 °C or ±0.2 °F (without probe error) | Accuracy: ±0.1°C (±0.2°F) (main unit)  |
|  35. | CO_{2} | Method: Infrared absorption | Method: Infrared absorption  |
|  36. |  | Support using Mindray sidestream CO_{2} module, Mindray AG module or third-party CO_{2} modules, microstream module and mainstream module. | Support using Masimo mainstream/sidestream CO_{2} module, Respironics mainstream CO_{2} module  |
|  37. | CO_{2} module (Mainstream) | Mindray Mainstream CO_{2} module Range: 0-150mmHg | Masimo CO_{2} module (Mainstream)
Range: 0 - 190mmHg / 0 - 25.3 kPa / 0 - 25.0% (at 760mmHg)  |
|  38. |  | Accuracy:
Mainstream EtCO_{2} module
0-40mmHg: ±2mmHg,
41-70mmHg: ±5% of the reading,
71-100mmHg: ±8% of the reading, 101-150mmHg: ±10% of the reading; | Accuracy:
Under all conditions:
0 - 114mmHg: ± (2.25mmHg +4% of reading);
115 mmHg - 190 mmHg: not defined
Standard environment (22°C±5°C and 1013±40hPa):
0 - 114 mmHg: ± (1.52mmHg +2% of reading);
115 -190 mmHg: not defined;  |

Page 9 of 22

{13}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|  39. |  | awRR range: 0-150rpm | awRR range: 0 - 150 breaths/min. measured at I/E ratio 1:1  |
|  40. |  | awRR accuracy: ±1rpm | awRR accuracy: ±1rpm  |
|  41. | CO_{2} module (Sidestream) | Mindray Sidestream CO_{2} module:
Range: 0-150mmHg
Mindray Microstream CO_{2} module:
Range: 0 to 99 mmHg | Masimo CO_{2} module(Sidestream)
Range: 0 - 190mmHg / 0 - 25.3 kPa/ 0 - 25.0% (at 760mmHg)  |
|  42. |  | Accuracy:
Mindray Sidestream CO_{2} module:
Full accuracy mode:
0 - 40 mmHg: ±2mmHg
41 - 76 mmHg: ±5% of reading
77 - 99 mmHg: ±10% of reading
100 - 150 mmHg: ±(3mmHg + 8% of reading
ISO accuracy mode: Add ±2mmHg to the full accuracy mode
Mindray Microstream CO_{2} module
0 to 38 mmHg: ±2 mmHg
39 to 99 mmHg: ±5% of the reading (0.08% of the reading increased in error for every 1 mmHg if the reading is more than 38 mmHg) | Accuracy:
Masimo Capno sidestream:
All conditions:
0 - 114mmHg: ± (2.25mmHg +4% of reading)
115 - 190 mmHg: not defined

Standard conditions (22°C±5°C and 1013±40 hPa)
0-114 mmHg: ± (1.52mmHg +2% of reading)
114 - 190 mmHg: not defined  |
|  43. |  | awRR range: 0 to 150 rpm | awRR range: 0 - 150 breaths/min. measured at I/E ratio 1:1  |
|  44. |  | awRR accuracy:
Mindray Sidestream CO_{2} module
≤60 rpm: ±1 rpm; 61 to 150 rpm: ±2 rpm
Mindray Mircostream CO_{2} module
0 to 70 rpm: ±1 rpm; 71 to 120 rpm: ±2 rpm; 121 to 150 rpm: | awRR accuracy: ±1rpm  |

Page 10 of 22

{14}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|   |  | ±3 rpm |   |
|  45. | CO_{2} (Mainstream) | Mindray Mainstream CO_{2} module:
Range: 0-150mmHg | Respironics CO_{2} (Mainstream)
Range: 0 - 150 mmHg, 0% - 19.7%, 0 - 20.0kPa (at 760mmHg)  |
|  46. |  | Accuracy:
Mindray Mainstream CO_{2} module:
0-40mmHg: ±2mmHg,
41-70mmHg: ±5% of the reading,
71-100mmHg: ±8% of the reading, 101-150mmHg: ±10% of the reading; | Accuracy:
0 - 40 mmHg: ±2mmHg
41 - 70mmHg: ±5% of reading;
71 - 100mmHg: ±8% of reading;
101 - 150 mmHg: ±10% of reading;
Note: temperature at 35°C  |
|  47. |  | Mindray Mainstream CO_{2} module: awRR Range: 0-150rpm | awRR Range: 0rpm, 2rpm - 150rpm  |
|  48. |  | Mindray Mainstream CO_{2} module: awRR Accuracy: ±1rpm | awRR Accuracy: ±1 rpm  |
|  49. | NIBP | Adult:
Systolic: 25mmHg to 290mmHg
Diastolic: 10mmHg to 250mmHg
Mean: 15mmHg to 260mmHg | Adult mode:
SYS: 25mmHg - 290mmHg (3.3kPa-38.7kPa)
DIA: 10mmHg - 250mmHg (1.3kPa-33.3kPa)  |
|  50. |  | Pediatrics
Systolic: 25mmHg to 240mmHg
Diastolic: 10mmHg to 200mmHg
Mean: 15mmHg to 215mmHg | Pediatrics mode:
SYS: 25mmHg - 240mmHg (3.3kPa-33.3kPa)
DIA: 10mmHg - 200mmHg (1.3kPa-28.0kPa)  |
|  51. |  | Neonatal
Systolic: 25mmHg to 140mmHg
Diastolic: 10mmHg to 115mmHg
Mean: 15mmHg to 215mmHg | Neonatal mode:
SYS: 25mmHg - 140mmHg (3.3kPa-18.7kPa)
DIA: 10mmHg - 115mmHg (1.3kPa-15.3kPa)  |
|  52. | IBP | Direct invasive measurement | Direct invasive measurement  |

Page 11 of 22

{15}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|   |  | The IBP supports Pulse Pressure Variation (PPV) and Pulmonary Artery Wedge Pressure (PAWP) function | The IBP supports Pulse Pressure Variation (PPV) and Pulmonary Artery Wedge Pressure (PAWP) function  |
|  53. |   | Range:
-50 to 300 mmHg | Range:
-50mmHg-+370mmHg (-6.6kPa-+49.3kPa)  |
|  54. |   | Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor) | Accuracy: ±1mmHg or ±2%, whichever is greater (sensor error excluded);  |
|  55. | C.O. | Thermodilution method | Thermodilution method  |
|  56. |  | C.O.
Range: 0.1 to 20 L/min
Accuracy: ±5% or ±0.1 L /min, whichever is greater | C.O.
Range: 0.1-20l/min
Accuracy: ±5% or ±0.1l/min, whichever is greater  |
|  57. |   | BT:
Range: 23 to 43 °C
Accuracy: ±0.1 °C (excluding sensor) | BT:
Range: 23-43°C (73°F-109°F)
Accuracy: ±0.1°C (±0.2°F) (main unit)  |
|  58. |   | IT: Range: 0 to 27 °C
Accuracy: ±0.1 °C (excluding sensor) | IT: Range: 0-27°C
Accuracy: ±0.1°C (sensor excluded)  |
|  59. | AG (only in NMPro series models) | Infrared absorption, paramagnetic properties for O_{2} monitoring | Infrared absorption, paramagnetic properties for O_{2} monitoring  |
|  60. |  | Gas type: CO_{2}, N_{2}O, O_{2}, Des., Iso., Enf., Sev., Hal | Gas type: CO_{2}, O_{2}, N_{2}O, one of the five anesthetic gases (Enf, Iso, Sev, Hal, DES)  |
|  61. |   | CO_{2}
Range: 0 to 30%
Accuracy: 0≤CO_{2}≤1: ±0.1%ABS
1 |   |

Page 12 of 22

{16}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|   |  | CO₂>10: Not specified |   |
|  62. |  | N₂O
Range: 0 to 100%
Accuracy: 0≤N₂O≤20: ±2%ABS
20<N₂O≤100: ±3%ABS | N₂O
Range: 0%-100%
Accuracy: ±(2%+2%*reading)  |
|  63. |  | Enf, Iso, hal
Range: 0 to 30%
Accuracy: 0 to 1: ±0.15%ABS
1 to 5: ±0.2%ABS
>5: not specified | Enf, Iso, hal
Range: 0%-25%
Accuracy: 0%-8%: ±(0.15%+5%*reading);
8%-25%: not defined.  |
|  64. |  | Sev
Range: 0 to 30%
Accuracy: 0 to 1: ±0.15%ABS
1 to 5: ±0.2%ABS
5 to 8: ±0.4%ABS
>8: not specified | Sev
Range: 0%-25%
Accuracy: 0%-10%: ±(0.15%+5%*reading);
10%-25%: not defined  |
|  65. |  | DES
Range: 0 to 30%
Accuracy: 0 to 1: ±0.15%ABS
1 to 5: ±0.2%ABS
5 to 10: ±0.4%ABS
10 to 15: ±0.6%ABS
15 to 18: ±1%ABS
>18: Not specified | DES
Range: 0%-25%
Accuracy: 0%-22%: ±(0.15%+5%*reading);
22%-25%: not defined.  |
|  66. |  | O₂ | O₂  |

Page 13 of 22

{17}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|   |  | Range: 0 to 100%
Accuracy: 0 to 25: ±1%ABS; 25 to 80: ±2%ABS[80 to 100: ±3%ABS | Range: 0%-100%
Accuracy: ±(1%+2%*reading)  |
|  67. |  | awRR:
Range: 2-100 rpm
Accuracy: 2 to 60 rpm: ±1 rpm
>60 rpm: not specified | awRR:
Range: 0rpm-150rpm
Accuracy: ± 1 rpm  |
|  68. | BIS (only for NMPro series models) | Bispectral index | Bispectral index  |
|  69. |  | Measurement parameters:
BIS, SQI, EMG, SR, SEF, TP, BC, sBIS, sEMG, ASYM | Measurement parameters:
BIS, SQI, EMG,SR  |
|  70. |  | BIS, BIS L, BIS R: 0 to 100 | BIS: 0-100  |
|  71. |  | SQI, SQI L, SQI R:0 to 100% | SQI: 0-100%  |
|  72. |  | EMG, EMG L, EMG R: 0 to 100 dB | EMG: 0-100dB.  |
|  73. |  | SR, SR L, SR R: 0 to 100% | SR: 0-100%  |
|  74. |  | sBIS L, sBIS R: 0 to 10.0
SEF, SEF L, SEF R: 0.5 to 30.0 Hz
TP, TP L, TP R: 40 to 100 dB
BC, BC L, BC R: 0 to 30
sEMG L, sEMG R: 0 to 10.0
ASYM: 0 to 100% | Not supported  |
|  75. | RM (only in NMPro series modelsr) | FLOW:
Adult/Pediatric: ±(2-120)L/min;
Infant: ± (0.5 to 30) L/min
PAW: -20-120 cmH₂O;
MVe/MVi: | Flow:
Adult/pediatric: ±(2-120)L/min;
Infant: ± (0.5 to 30) L/min
Paw: -20-120 cmH₂O;
Pmean:-20-120 cmH₂O;  |

Page 14 of 22

{18}

510(k) Premarket Notification
510(k) Summary
K250854

| No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor |
| --- | --- | --- | --- |
|  |  | Adult/Pediatric: 2-60L/min;
Infant: 0.5 to 15 L/min
TVe/TVi:
Adult/Pediatric: 100-1500ml; Infant: 20 to 500ml
Calculated Parameters:
awRR:4-120rpm;
I:E:4:1-1:8;
FEV1.0: 0-100%;
Pmean:0-120 cmH_{2}O;
PEEP:0-120 cmH_{2}O;
PIP: 0-120 cmH_{2}O;
Pplat:0-120 cmH_{2}O;
PEF:2-120L/min;
PIF:2-120L/min;
Compl: 0-200ml/ cmH_{2}O;
RSBI:0-4095rpm/L; | PEEP:-20-120 cmH_{2}O;
PIP: -20-120 cmH_{2}O;
Pplat:-20-120 cmH_{2}O;
PEF:2-120L/min;
PIF:2-120L/min;
Inspiratory tidal volume (TVi) /Expiratory tidal volume (TVe): Adult/pediatric: 100-1500ml; Infant: 20-500ml,
Inspired minute volume (MVi)/Expired minute volume (MVe):
Adult/pediatric: 2-60 L/min; Infant: 0.5-15 L/min,
Respiration rate (RR): 4-120 rpm
I:E: 2:1-1:4.5
First second forced expiratory volume ratio (FEV1.0): 0-100%
Compliance (Cstat/Cdyn): 4-100ml/ cmH_{2}O;
Rapid Shallow Breathing Index (RSBI): 0-200 rpm/l |
| 76. | Flow:
Adult/Pediatric: 1.2L/min or ±10% of reading, whichever is greater;
Infant: 0.5 L/min or ±10% of the reading, whichever is greater
PAW: ±3% of reading;
MVe/MVi: ±10% of reading;
TVe/TVi: Adult/Pediatric: ±15ml or ±10% of reading, whichever is greater;
Infant: ±6 ml or ±10%×reading, whichever is greater | Flow:
Adu./Ped.: ±1.2l/min or ±10% of actual reading, which is greater
Infant. ±0.5l/min or ±10% of actual reading, which is greater
Airway Pressure (Paw):
±1 cmH_{2}O ±3% of actual reading, which is greater
PEEP, PIP, Pmean, Pplat: ±1 cmH_{2}O or ±3% of actual reading, which is greater
Inspiratory tidal volume (TVi), Expiratory tidal volume |

Page 15 of 22

{19}

510(k) Premarket Notification
510(k) Summary
K250854

|  No. | Feature | Predicate Device
BeneVision N series Patient Monitors (K213799), | Subject Device
N series Patient Monitor  |
| --- | --- | --- | --- |
|   |  | Calculated Parameters:
awRR:4-99rpm: ±1rpm, 100-120rpm, ±2rpm;
I:E: not specified;
FEV1.0: not specified;
Pmean: ±10%;
PEEP: not specified;
PEF: ±10%;
PIF: ±10%;
PIP: ±10%;
Pplat: not specified;
Compl: not specified;
RSBI: not specified; | (TVe):
Adu./Ped.: ±15ml or ±10% of actual reading, which is greater
Infant: ±6ml or ±10% of actual reading, which is greater
Inspired minute volume (MVi), Expired minute volume
(MVe): ±10% of actual reading
Respiration rate (RR):
4 rpm - 99 rpm.: ±1 rpm; 100rpm-120rpm.: ±2 rpm
I:E, First second forced expiratory volume ratio (FEV1.0),
Compliance (Cstat/Cdyn), Rapid Shallow Breathing Index
(RSBI): not defined.  |
|  77. | NMT (only in NMPro series models) | Measurement range:
ST-Ratio: 0 to 200%
TOF-Count: 0 to 4
TOF-Ratio: 5% to 160%
TOF-T1%: 0 to 200%
DBS-Count: 0 to 2
DBS-Ratio: 5% to 160%
PTC-Count: 0 to 20 | Measurement range:
ST-Ratio: 0 to 200%
TOF-Count: 0 to 4
TOF-Ratio: 5% to 160%
TOF-T1%: 0 to 200%
DBS-Count: 0 to 2
DBS-Ratio: 5% to 160%
PTC-Count: 0 to 20  |

Page 16 of 22

{20}

510(k) Premarket Notification
510(k) Summary
K250854

# 8. Performance Data

To establish the substantial equivalence of N10, N12, N15, N10MPro, N12MPro and N15MPro Patient Monitors, functional and system level testing was conducted on the subject device. The testing showed that the devices meet specifications and the performance level is equivalent to the predicate.

## Software and Cybersecurity Verification and Validation Testing

Software testing included software design, development, verification, validation, and traceability. Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions."

Patient safety and security risks have been addressed in the design and development of the subject device including a Security Risk Assessment and Threat model, which is mapped to the General Principles and Security Capabilities as outlined in FDA's Guidance for Industry and FDA Staff, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions". The subject device has been through rigorous testing for cybersecurity vulnerability including penetration testing and code review with acceptable results. The cybersecurity controls will be monitored and updated throughout the device lifecycle.

## Biocompatibility Testing

Biocompatibility testing was performed on the subject device components in accordance with the FDA guidance on the Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (Issued: September 4, 2020).

## Usability study

Usability evaluation including critical task identification through Use-Related Risk Analysis (URRA), human factors (HF) validation testing and other activities had been completed for the proposed patient monitor. All use-related hazards have been considered and adequately mitigated and that the residual risks are acceptable.

## Bench test

The N10, N12, N15, N10MPro, N12MPro and N15MPro Patient Monitors have been tested and found to be in compliance with the following safety, performance and electromagnetic compatibility standards:

- ANSI AAMI ES60601-1:2005/(R)2012 &amp; A1:2012, C1:2009/(R)2012 &amp; A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- IEC 60601-1-2 Edition 4.1: 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC TS 60601-4-2 Edition 1.0:2024-03 Medical electrical equipment-Part 4-2: Guidance and interpretation-Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems

Page 17 of 22

{21}

510(k) Premarket Notification
510(k) Summary
K250854

- IEC 60601-1-6 Edition 3.2 2020-07 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- ANSI AMI IEC 62366-1:2015+AMD1:2020 Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1
- ANSI AAMI IEC 60601-1-8:2006 and A1:2012 [Including AMD 2:2021] Medical Electrical Equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems [Including Amendment 2 (2021)]
- ANSI AAMI IEC 60601-2-25:2011/(R)2016 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
- ANSI AAMI IEC 60601-2-27:2011(R)2016 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
- IEC 80601-2-30: Edition 2.0 2018-03 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 60601-2-34 Edition 4.0 2024-10 Medical electrical equipment - Part 2-34: Particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
- ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including: Amendment 1 (2018)].
- ISO 80601-2-61 Second edition 2017-12 (Corrected version 2018-02) Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- ISO 80601-2-55 Second edition 2018-02 [Including AMD 1:2023] Medical electrical equipment - Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors
- IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes
- IEC 81001-5-1 Edition 1.0 2021-12 Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle
- ANSI AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires
- ANSI AAMI EC 57:2012 Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms
- IEC 60601-2-40:2024 Medical electrical equipment - Part 2-40: Particular requirements for the basic safety and essential performance of electromyographs and evoked response equipment
- IEC 80601-2-49:2018+A1:2024 Medical electrical equipment - Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitors
- IEC 80601-2-26:2019+A1:2024 Medical electrical equipment - Part 2-26: Particular requirements for the basic safety and essential performance of electroencephalographs
- ISO 14971 Third Edition 2019-12 Application of Risk Management To Medical Devices

Page 18 of 22

{22}

510(k) Premarket Notification

510(k) Summary

K250854

# 9. Clinical Studies

The subject device incorporates clinical subsystem modules for SpO2 monitoring (including OEM module), respiratory rate measurement, and NIBP measurement. For these features, new clinical accuracy validation testing was not necessary because these subsystems along with their accessories have been previously validated and cleared. This 510(k) leverages the previous clinical accuracy data, as shown in Table 3, for the clinical functionalities in the subject device.

Table 3: Clinical Accuracy Data

|  Monitoring subsystem | Physiological monitoring unit |   | Subsystem Cleared in 510(k)  |   |
| --- | --- | --- | --- | --- |
|   |  Technology | Component | 510K number | Device name/manufacturer  |
|  SpO2 monitoring | Masimo | Module, Algorithm | K053269 | Masimo SET Rad-8 Pulse Oximeter, Model Rad-8, Kestrel/Masimo Corporation  |
|   |   |  Sensor cable extender (interconnection cable) | K053269 | Masimo SET Rad-8 Pulse Oximeter, Model Rad-8, Kestrel/Masimo Corporation  |
|   |   |  Probes | K051212 K111888 K180046 K101896 | Masimo Rad-97 and Accessories/Masimo Corporation  |
|  Masimo Rainbow SpO2 monitoring | Masimo | Module, Algorithm | K193242 | Masimo Radical-7 Pulse CO-Oximeter and Accessories/Masimo Corporation  |
|   |   |  Sensor cable extender (interconnection cable) | K080238 | MASIMO RAINBOW SET RADICAL 7 PULSE CO-OXIMETER, MASIMO RAINBOW SET RAD 87 PULSE CO-OXIMETER  |
|   |   |  Probes | K081659 K080238 | RAINBOW ADHESIVE CO-OXIMETRY SENSOR, MODELS RI-25-L, RI-25, RI-20 AND RI-20-L  |
|  SpO2 monitoring | Nellcor | Module, Algorithm | K060576 | OxiMax Pulse Oximetry System with N-600x Pulse Oximeter and  |
|   |  |  |  | Accessories  |
|  NIBP measurement | NIBP | Module, Algorithm | K060576 | NIBP Oximeter, Model Rad-8, Kestrel, OxiMax  |
|   |   |  Sensor cable extender (interconnection cable) | K060576 | NIBP Oximeter, Model Rad-8, Kestrel, OxiMax  |

{23}

510(k) Premarket Notification
510(k) Summary
K250854

| Monitoring subsystem | Physiological monitoring unit | Subsystem Cleared in 510(k) |
| --- | --- | --- |
| Technology | Component | 510K number | Device name/manufacturer |
|  |  |  |  | OxiMax Sensors and Cables/Nellcor Puritan Bennett Incorporated |
| Sensor cable extender (interconnection cable) | K060576 | OxiMax Pulse Oximetry System with N-600x Pulse Oximeter and OxiMax Sensors and Cables/Nellcor Puritan Bennett Incorporated |
| Probes | K012891 | OxiMax Pulse Oximetry System with N-595x Pulse Oximeter and OxiMax Sensors and Cables/Nellcor Puritan Bennett Incorporated |
| SpO_{2} monitoring | COMEN | Module, Algorithm | K211619 | Shenzhen COMEN NC10 NC12 multi-parameter patient monitor / Shenzhen Comen Medical Instruments Co., Ltd. |
|  |  | Sensor cable extender (interconnection cable) | K211619 | Shenzhen COMEN NC10 NC12 multi-parameter patient monitor / Shenzhen Comen Medical Instruments Co., Ltd. |
| Probes | K211619 | Shenzhen COMEN NC10 NC12 multi-parameter patient monitor / Shenzhen Comen Medical Instruments Co., Ltd. |
| NIBP | COMEN | Module, | K211619 | Shenzhen COMEN |

Page 20 of 22

{24}

510(k) Premarket Notification
510(k) Summary
K250854

|  Monitoring subsystem | Physiological monitoring unit |   | Subsystem Cleared in 510(k)  |   |
| --- | --- | --- | --- | --- |
|   |  Technology | Component | 510K number | Device name/manufacturer  |
|  measurement |  | Algorithm |  | NC10 NC12 multi-parameter patient monitor / Shenzhen Comen Medical Instruments Co., Ltd.  |
|   |   |  NIBP hose and Cuff | K211619 | Shenzhen COMEN NC10 NC12 multi-parameter patient monitor / Shenzhen Comen Medical Instruments Co., Ltd.  |
|  Respiratory rate measurement | COMEN | Module, Algorithm | K211619 | Shenzhen COMEN NC10 NC12 multi-parameter patient monitor / Shenzhen Comen Medical Instruments Co., Ltd.  |
|   |  Masimo Mainstream and Sidestream CO_{2} module | Module, Algorithm and accessories | K123043 | IRMA - Infrared Mainstream Gas Analyzer/Masimo Corporation  |
|   |  Respironics Mainstream CO_{2} module | Module, Algorithm and accessories | K042601 | Capnostat CO_{2}./Respironics Novametrix LLC,  |

# ECG

The performance of the ECG arrhythmia analysis algorithm has been comprehensively evaluated using standard, expert-annotated ECG databases (i.e., AHA, MIT-BIH, NST, CU, ESC, QT database, MIT-BIH Malignant Ventricular Fibrillation Database, MIT-BIH Supraventricular Arrhythmia Database(MIT-SV), MIT-BIH ST Change Database, MIT-BIH Atrial Fibrillation Database (AFDB), INCART Database, Long Term AF Database (LTAFDB), Apnea-ECG Database, Sudden Cardiac Death Holter Database, MGH/MF Waveform Database, Long-Term ST Database, Comen Pacemaker Database and Comen supraventricular Arrhythmia database) in accordance with the requirements of AAMI/ANSI EC57:2012 “Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms”.

Page 21 of 22

{25}

510(k) Premarket Notification
510(k) Summary
K250854

# 10. Conclusion

The N10, N12, N15, N10MPro, N12MPro and N15MPro Multi-parameter Patient Monitor are equivalent to the predicate device, the BeneVision N series Patient Monitors, (K213799), in terms of indications for use and technological characteristics. Performance testing and compliance with FDA-recognized consensus standards demonstrate that the N10, N12, N15, N10MPro, N12MPro and N15MPro Patient Monitors are as safe and effective as the predicate device.

---End of Document---

---

**Source:** [https://fda-staging.innolitics.com/device/K250854](https://fda-staging.innolitics.com/device/K250854)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
