K250852 · Asterasys Co., Ltd. · GEI · Oct 16, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250852
Device Name
Coolfase
Applicant
Asterasys Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 16, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Coolfase is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Device Story
Coolfase is an electrosurgical unit delivering high-frequency current (6.78 MHz) to tissue for coagulation and hemostasis. System components include a main unit, handpiece, single-use tips, grounding pad, coupling gel, and footswitch. Operation involves delivering RF energy through the handpiece to the skin; electrical resistance generates heat, coagulating tissue. Used in clinical settings by healthcare professionals. Output is controlled via LCD touchscreen or footswitch. Device provides thermal coagulation; benefits include effective hemostasis in dermatologic and general surgical procedures.
Clinical Evidence
Bench testing only. Ex vivo side-by-side thermal performance testing on tissue was conducted comparing Coolfase to the predicate device. Histological analysis was performed to verify tissue effects and ensure consistent coagulation performance in accordance with FDA guidance for electrosurgical devices.
Technological Characteristics
Monopolar RF electrosurgical unit; 6.78 MHz frequency; 140W max output power. Components: main unit, handpiece, single-use tips (0.25, 4.0, 16 cm²), grounding pad, coupling gel, footswitch. Class 1/Type BF applied part. LCD touchscreen interface. Input: 100-240V, 50/60 Hz. Non-sterile delivery. Compliant with electromagnetic compatibility and medical electrical equipment safety standards.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis in patients.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 16, 2025
ASTERASYS Co., Ltd.
Chayeon Kim
Head of Regulatory Affairs Team
#1005, #508, #313, 25 Seongsuil-ro 4-gil Seongdong-gu
Seoul, 04781
Korea, South
Re: K250852
Trade/Device Name: Coolfase
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: September 15, 2025
Received: September 15, 2025
Dear Chayeon Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250852 - Chayeon Kim
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250852 - Chayeon Kim
Page 3
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2025.10.16
15:38:00 -04'00'
Colin Kejing Chen
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Coolfase
Page 9 of 42
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250852 | ? |
| Please provide the device trade name(s). | | ? |
| Coolfase | | |
| Please provide your Indications for Use below. | | ? |
| Coolfase is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K250852
PAGE 1 / 3
ASTERASYS
YOUR PARTNER for HEALTHCARE
# 510(k) Summary
## 1. General Information [21 CFR 807.92(a)(1)]
Submitter / Applicant: ASTERASYS Co., Ltd.
Rm 1005, 25 Seongsuil-ro 4-gil, Seongdong-gu, Seoul, 04781
Republic of Korea
Tel: +82-2-6953-0456
Contact Person: Dr. Chayeon Kim
Rm 1005, 25 Seongsuil-ro 4-gil, Seongdong-gu, Seoul, 04781
Republic of Korea
Phone: +82-10-4263-1721
Email: amy.kim@asterasys.com
Preparation Date: Month 07, 2025
## 2. Device Name and Code [21 CFR 807.92(a)(2) and (3)]
| Device Trade Name: | Coolfase |
| --- | --- |
| Common Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Product Code: | GEI |
| CFR References: | 21 CFR 878.4400 |
| Regulatory Class: | II |
| Predicate Device: | Thermage FLX system |
| Review Panel: | General & Plastic Surgery |
## 3. Device Description [21 CFR 807.92(a)(4)]
Coolfase is electrosurgical unit that high-frequency current for tissue coagulation during treatment. It consists of an electrosurgical unit main body, a handpiece, and a tip, and is used in conjunction with a grounding pad, coupling gel, and a foot switch. The high-frequency current generated by the main body is delivered to the skin through the handpiece connected to the tip. The electrical resistance of the skin generates heat, and this heat coagulates the tissue.
## 4. Indications for Use [21 CFR 807.92(a)(5)]
Coolfase is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
ASTERASYS Co., Ltd.
#1005, 25 Seongsuil-ro 4-gil, Seongdong-gu, Seoul, Korea 04781
TEL: +82-2-6928-5385, FAX: +82-2-6953-0599
{5}
K250852
PAGE 2 / 3
ASTERASYS
YOUR PARTNER for HEALTHCARE
# 5. Summary of Technological Characteristics
Technical Characteristics in Comparison to Predicate Devices [21 CFR 807.92(a)(6)]
| Characteristic | Subject Device | Predicate device K170758 |
| --- | --- | --- |
| Applicant | ASTERASYS Co., Ltd. | Solta Medical, Inc. |
| Device name | Coolfase | Thermage FLX system |
| 510(k) Number | K250852 | K170758 |
| Device Classification Name | Electrosurgical Cutting and Coagulation Device and Accessories | Same |
| Classification | Class II | Same |
| Review Panel | General & Plastic Surgery | Same |
| Classification Product Code | GEI | GEI, ISA |
| Regulation Number | 21 CFR 878.4400 | Same |
| Indications for Use | Coolfase is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis | The radiofrequency-energy only delivery components of the Thermage FLX System are indicated for use in: • Dermatologic and general surgical procedures for electrocoagulation and hemostasis |
| Prescription or OTC | Prescription | Same |
| Anatomical Site | Intact skin | Same |
| Contact duration | Less than 24 hours | Same |
| Electrode type | Monopolar RF | Same |
| Frequency | 6.78 MHz ± 10 % | 6.78MHz ± 0.1% |
| Input voltage | 100-240V, 50/60 Hz, 400VA | 100-240V, 50/60 Hz, 10A |
| Different – VA The proposed device and the predicate device share the same input voltage specification (100–240 V, 50/60 Hz), although their VA ratings differ. The input power of the proposed Coolfase (1400 VA) corresponds to a lower current draw than that of the predicate device (10 A) and remains within the specified input voltage range. Accordingly, this variation does not materially affect the device’s safety or effectiveness. | | |
| Maximum output Power | 140W | 400W |
| Different – Max output power The Max output power for proposed device is different from the predicate devices. However, the Max output power of the proposed Coolfase is covered by the range of the Max output power of the predicate device. Therefore, this difference does not affect substantially equivalence on safety and effectiveness. | | |
| Mode of Operation | Manual or footswitch | Same |
| Electrode size | 0.25 cm² (CF-E tip / CF-CT05) 4.0 cm² (CF-F tip / CF-CF20) 16 cm² (CF-B tip / CF-CF40) | Same (Thermage FLX is equipped with one additional 3.0 cm² tip) |
| Sterilization | Provided as non-sterile | Same |
| Number of uses per electrode | Signle use | Same |
| Weight | 20 kg | 45 kg |
| Type & degree of protection against shock | Class 1 / type BF applied part | Same |
| Dimension | 27 cm (W) x 28.9 cm (L) x 115.4 cm (H) | 49.5 cm (W) x 43.0 cm (L) x 125.7 cm (H) |
| User interface | LCD / Touchscreen Technology for user interaction and controls | Same |
ASTERASYS Co., Ltd.
#1005, 25 Seongsuil-ro 4-gil, Seongdong-gu, Seoul, Korea 04781
TEL: +82-2-6928-5385, FAX: +82-2-6953-0599
{6}
K250852
PAGE 3 / 3
^{}[]
ASTERASYS
YOUR PARTNER for HEALTHCARE
| Demonstrated Safety and Efficacy in treated area | Performance Testing section | Provided in K170758 cleared September 22, 2017 |
| --- | --- | --- |
| Patient Contact Material | Biocompatible | Biocompatible |
| Electromagnetic compatibility standards | Compliant | Compliant |
| Medical electrical equipment Safety standards | Compliant | Compliant |
The proposed device, Coolfase, uses a monopolar electrode type, similar to the predicate device, Thermage FLX system. While the maximum output power of Coolfase differs from Thermage FLX, it still falls within the overall power range of the predicate device. This ensures that the difference does not raise new questions regarding safety or effectiveness. Both devices offer the same operational modes, supporting both manual operation and footswitch control, and they share a similar user interface, utilizing LCD touchscreen technology for user interaction and control. Taken together, these similarities demonstrate that Coolfase is substantially equivalent to Thermage FLX in terms of both technological characteristics and intended use.
# 6. Non-Clinical and/or Clinical Test Summary & Conclusions [21 CFR 807.92(b)]
Thermal performance testing was conducted to evaluate the subject device with respect to the risks of burns and incomplete tissue coagulation. Ex vivo side-by-side testing on tissue was performed comparing the subject device and the predicate device, in accordance with FDA's 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery' Guidance (March 2020). The results were further assessed by histological analysis to verify tissue effects and to demonstrate substantial equivalence.
ASTERASYS Co., Ltd.
#1005, 25 Seongsuil-ro 4-gil, Seongdong-gu, Seoul, Korea 04781
TEL: +82-2-6928-5385, FAX: +82-2-6953-0599
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.