Nanox.ARC X is a stationary X-ray system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus indications, adjunctive to conventional radiography, on adult patients. This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists. Digital Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions. This device is not intended for mammographic, angiographic, cardiac, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.
Device Story
Nanox.ARC X is a stationary, floor-mounted digital tomosynthesis system. It utilizes a multisource, tiltable arc gantry containing five cold-cathode X-ray tubes and a flat panel scintillator-photodetector. During a single tomographic sweep, the system sequentially activates the five tubes to acquire multiple low-dose projection images. A motorized table supports the patient in prone, supine, or lateral positions. Acquired data is anonymized and reconstructed into tomographic slices parallel to the table plane, reducing overlying structure effects and providing depth information. Reconstruction and DICOMization occur locally or via the secured Nanox.CLOUD. The system is operated by trained radiographers, radiologists, or physicists in clinical environments using a handheld console. Output images are reviewed by healthcare providers to assist in diagnostic decision-making, providing anatomical depth information to supplement conventional radiography.
Clinical Evidence
No clinical tests were performed. Substantial equivalence is supported by extensive bench testing, including system electrical qualification, performance, longevity/reliability, functional verification, motion control, mechanical properties, image quality, human factors, phantom validation, and software verification and validation.
Technological Characteristics
Stationary, floor-mounted tomosynthesis system. Features a multisource arc gantry with five sequentially switched cold-cathode X-ray tubes (MEMS electron source). Employs a flat panel scintillator-photodetector. System electronics are integrated internally. X-ray tube housing is glass. Connectivity includes local or cloud-based (Nanox.CLOUD) reconstruction and DICOM protocol support. Motorized patient table supports up to 150kg.
Indications for Use
Indicated for adult patients requiring tomographic imaging of the musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus regions as an adjunct to conventional radiography. Contraindicated for pediatric/neonatal patients and for mammographic, angiographic, cardiac, intra-cranial, interventional, or fluoroscopic applications.
Regulatory Classification
Identification
A tomographic x-ray system is an x-ray device intended to be used to produce radiologic images of a specific cross-sectional plane of the body by blurring or eliminating detail from other planes. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 16, 2025
Nano-x Imaging Ltd.
% Noa First
Regulatory Affairs Manager
94 Em Hamoshavot Road, Brosh building, Ofer Tech Park,
PO Box 3486
PETAH TIKVA, 4970602
ISRAEL
Re: K250850
Trade/Device Name: Nanox.ARC X
Regulation Number: 21 CFR 892.1740
Regulation Name: Tomographic x-ray system
Regulatory Class: Class II
Product Code: IZF, MQB
Dated: March 20, 2025
Received: March 21, 2025
Dear Noa First:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250850 - Noa First
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250850 - Noa First
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

for
Lu Jiang
Assistant Director
Diagnostic X-ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250850
Device Name
Nanox.ARC X
Indications for Use (Describe)
Nanox.ARC X is a stationary X-ray system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus indications, adjunctive to conventional radiography, on adult patients.
This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists.
Digital Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions.
This device is not intended for mammographic, angiographic, cardiac, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
NANOX
04/10/2024
K250850
# 510(k) SUMMARY
## I. SUBMITTER
Applicant Name: NANO-X Imaging Ltd.
Applicant Address: 94 Em Hamoshavot Road, Brosh building, Ofer Tech Park,
PO Box 3486, Petah Tikva, 4970602 Israel
Applicant Contact Telephone: +972-54-5225792
Applicant Contact: Ofir Koren, General Manager ARC Division
Applicant Contact Email: ofir.k@nanox.vision
Correspondent Contact Telephone: +972-50-4040698
Correspondent Contact: Noa First, Regulatory Affairs Manager
Correspondent Contact Email: noa.f@nanox.vision
## II. DEVICE
Device Trade Name: Nanox.ARC X
Common Name: System, X-Ray, Tomographic
Classification Name: Tomographic x-ray system
Regulation Number: 892.1740
Product Code(s): IZF, MQB
## III. PREDICATE DEVICE
Predicate #: K242395
Predicate Trade Name: Nanox.ARC
Product Code: IZF, MQB
## IV. DEVICE DESCRIPTION
Nanox.ARC X is a stationary, floor-mounted, stand-alone digital tomosynthesis system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus indications, from a single tomographic sweep. It serves as an adjunct to conventional radiography, for adult patients in recumbent positions. The system is intended for use in professional healthcare settings such as hospitals, clinics, and imaging centers by trained radiographers, radiologists, and physicists
The Nanox.ARC X includes a secured, dedicated off-the-shelf handheld operator console, a multisource, tiltable arc gantry with five identical tubes, a motorized patient table, and a flat panel detector of a scintillator-photodetector type. The image reconstruction service and DICOMization services can be hosted either locally or as part of the secured Nanox.CLOUD, according to customer preference. Nanox.CLOUD also hosts a protocol database service package.
{5}
NANOX
K250850 Nanox.ARC X
The Nanox.ARC X X-ray tubes are operated sequentially, one at a time, generating multiple low-dose images acquired from different angles, during a single sweep, dividing the overall power requirements among the tubes. The sweep is performed over a motorized patient table. Patients can be placed in prone, supine, and lateral positions.
The acquired projection imaging data is anonymized and automatically reconstructed to form tomographic slices of the imaged object, with each slice parallel to the table plane. The Tomosynthesis image result reduces the effect of overlying structures and provides depth information on structures of interest. The resultant images are re-identified and sent using the DICOM protocol.
## V. INDICATIONS FOR USE
Nanox.ARC X is a stationary X-ray system intended to produce tomographic images for general use including human musculoskeletal system, pulmonary, intra-abdominal, and paranasal sinus indications, adjunctive to conventional radiography, on adult patients.
This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists.
Digital Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions.
This device is not intended for mammographic, angiographic, cardiac, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.
## VI. INDICATIONS FOR USE COMPARISON
The subject and the predicate devices have the same indications for use.
## VII. TECHNOLOGICAL CHARACTERISTICS COMPARISON
Both the subject and predicate devices share the same fundamental operating and technological principles.
Both devices are tomosynthesis-based imaging systems designed to produce tomographic images of human anatomy, from a single tomographic sweep, adjunctive to conventional radiography, of adult patients in recumbent positions.
Both devices utilize sequentially activated cold-cathode X-ray sources, with identical internal tube geometry and spatial alignment with the same detector, and the Nanox.CLOUD reconstruction software.
The subject device maintains functional and technological equivalence to the predicate device while incorporating system-level refinements.
{6}
NANOX
K250850 Nanox.ARC X
Technological Characteristics Comparison Table
| Item | Subject Device K250850 | Predicate Device K242395 | SE Justification |
| --- | --- | --- | --- |
| System Architecture | System electronics internally arranged | System electronics externally arranged | Similar - Both system architectures retain the same functionality and components. Internalization of system electronics previously housed externally to improve system integration and fit within clinical environments; enhanced internal layout with no impact on system functionality |
| ARC Imaging Ring | Gantry-based arc structure with square-rounded design | Gantry-based arc structure with rounded design | Similar -Spatial configuration and internal geometry preserved; enclosure shape adjusted to optimize integration |
| Number of X-ray sources | Five (5), alternately-switched X-ray sources | Same | - |
| X-ray Tube Housing Material | Glass | Ceramic | Similar -Both systems utilize identical cathode technology, MEMS electron source, anode design, focal spot size, exposure parameters, X-ray field size, and target angle. Modification in housing material; no impact on emission characteristics, cathode technology, exposure range, or performance |
| Patient Weight Allowance | 150kg / 330 lbs. | Same | - |
| Patient Table Footprint | Optimized footprint | Larger footprint | Similar -Both tables serve the same function and are motor-driven |
All modifications have been verified and validated and do not impact safety or effectiveness.
{7}
NANOX
K250850 Nanox.ARC X
# VIII. NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS
In order to support substantial equivalence, the following bench testing were performed on the subject device:
- System Electrical Qualification
- System Performance
- Longevity and Consistency
- Tube Longevity and Reliability
- Functional Verification
- Motion Control
- Dimensional and Mechanical Properties
- Image Quality
- Tube Comparison CEI and Nanox Korea
- Human Factors Summary
- Phantom Validation
- Weight Considerations
- Transportation
- Software Verification and Validation
In all instances, Nanox.ARC X System functioned as intended.
No clinical tests were performed for the subject device.
Based on performance data, the subject device was found to have a safety and effectiveness profile that is similar to the predicate device, demonstrating that the subject device is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.