The PADLOCK CLIP EFTR Device is an electrosurgical device that is designed for a flexible endoscope. It is intended for diagnostic tissue acquisition and full-thickness resection through the removal of lesions in the colon and rectum. The PADLOCK CLIP EFTR Device is indicated for the resection of lesions less than 3cm in the colon and rectum. Resection sizes may be less than 3cm due to targeted tissue thickness and degree of fibrosis. The disposable RAPTOR Grasping Device is intended to be used to grasp tissue and/or retrieve foreign bodies, excised tissue and stents during endoscopic procedures. The TOUCHSOFT Coagulator is used to provide monopolar therapeutic hemostasis and tissue ablation in the gastrointestinal tract through a flexible endoscope by a trained physician or clinician.
Device Story
Electrosurgical, single-use tissue resection system for flexible endoscopes; used for full-thickness resection of colorectal lesions not amenable to standard polypectomy. Kit includes: PADLOCK CLIP EFTR device (radiopaque super elastic Nitinol clip, mechanical delivery system, snare), RAPTOR grasping device, and TOUCHSOFT coagulator. Device mounts over distal endoscope tip; linking cables attach to handle. Physician uses TOUCHSOFT coagulator to mark lesion perimeter via ablation; RAPTOR grasping device retracts target lesion into tissue chamber; snare performs resection. System provides diagnostic tissue acquisition and lesion removal. Used in clinical settings by trained physicians. Benefits include enabling resection of lesions otherwise not resectable by standard techniques.
Clinical Evidence
Bench testing only. Verification testing confirmed successful clip deployment and snare closure on simulated tissue.
Indicated for adult patients requiring diagnostic tissue acquisition and full-thickness resection of lesions <3cm in the colon and rectum.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 18, 2025
STERIS Corporation
Gregory Land
Manager, Regulatory Affairs
5960 Heisley Road
Mentor, Ohio 44060-1834
Re: K250842
Trade/Device Name: PADLOCK CLIP EFTR Kit (00713229)
Regulation Number: 21 CFR 876.4400
Regulation Name: Hemorrhoidal Ligator
Regulatory Class: Class II
Product Code: PKL, FDI, KNS, OCZ
Dated: March 20, 2025
Received: March 20, 2025
Dear Gregory Land:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250842 - Gregory Land
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250842 - Gregory Land
Page 3
Sincerely,
**Shanil P. Haugen -S**
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250842
Device Name
PADLOCK CLIP EFTR Kit (00713229)
Indications for Use (Describe)
The PADLOCK CLIP™ EFTR Device is an electrosurgical device that is designed for a flexible endoscope. It is intended for diagnostic tissue acquisition and full-thickness resection through the removal of lesions in the colon and rectum.
The PADLOCK CLIP™ EFTR Device is indicated for the resection of lesions less than 3cm in the colon and rectum. Resection sizes may be less than 3cm due to targeted tissue thickness and degree of fibrosis.
The disposable RAPTOR Grasping Device is intended to be used to grasp tissue and/or retrieve foreign bodies, excised tissue and stents during endoscopic procedures.
The TOUCHSOFT Coagulator is used to provide monopolar therapeutic hemostasis and tissue ablation in the gastrointestinal tract through a flexible endoscope by a trained physician or clinician.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
STERIS Special 510(k) PREMARKET NOTIFICATION
Modification to the PADLOCK CLIP EFTR Kit
K250842
Page 1 of 4
# 1. Device Name
| Trade Name: | PADLOCK CLIP EFTR Kit |
| --- | --- |
| Device Class: | Class II |
| Common/usual Name: | Endoscopic Full-Thickness Resection Device |
| Classification Name: | Hemorrhoidal Ligator |
| Classification Number: | 21 CFR 876.4400 |
| Product Code: | PKL |
| Associated Product Codes: | FDI, KNS, and OCZ |
# 2. Predicate Device
K240274
PADLOCK CLIP EFTR Kit
Table 1. Device Comparison Table for proposed and Predicate devices
| Features | PADLOCK CLIP EFTR Device K240274 (Predicate Device) | PADLOCK CLIP EFTR Device (Proposed Device) | Comparison |
| --- | --- | --- | --- |
| Indications for Use | The PADLOCK CLIP EFTR Device is an electrosurgical device that is designed for flexible endoscopy. It is intended for diagnostic tissue acquisition and full-thickness resection through the removal of lesions in the colon and rectum. | The PADLOCK CLIP EFTR Device is an electrosurgical device that is designed for flexible endoscopy. It is intended for diagnostic tissue acquisition and full-thickness resection through the removal of lesions in the colon and rectum. | Same |
| | The PADLOCK CLIP EFTR Device is indicated for the resection of lesions less than 3 cm in the colon and rectum. Resection sizes may be less than 3 cm due to targeted tissue thickness and degree of fibrosis. | The PADLOCK CLIP EFTR Device is indicated for the resection of lesions less than 3 cm in the colon and rectum. Resection sizes may be less than 3 cm due to targeted tissue thickness and degree of fibrosis. | |
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}
STERIS Special 510(k) PREMARKET NOTIFICATION
Modification to the PADLOCK CLIP EFTR Kit
K250842
Page 2 of 4
| Features | PADLOCK CLIP EFTR Device K240274 (Predicate Device) | PADLOCK CLIP EFTR Device (Proposed Device) | Comparison |
| --- | --- | --- | --- |
| Construction | A -Molded handle (components consisting of:)
1. Thumb ring
2. Handle shaft
3. Spool
B -Connector
C-Handle stay
D-Linking wires
E -Clip actuator
F -Clip housing assembly
G -Clip
H -Snare
I -Drive Cable
N/A -Safety cap | A -Molded handle (components consisting of:)
1. Thumb ring
2. Handle shaft
3. Spool
B -Connector
C-Handle stay
D-Linking wires
E -Clip actuator
F -Clip housing assembly
G -Clip
H -Snare
I -Drive Cable
N/A -Safety cap | Same |
| Sterile/ Non-sterile | Sterile | Sterile | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Sterilization Assurance Level | 10-6 | 10-6 | Same |
| Usage | Single-Use | Single-Use | Same |
| Materials (by component) | -Molded handle (components consisting of:)
Thumb ring – ABS & Colorant
Handle shaft- ABS & Colorant
Spool – ABS & Colorant
-Connector – ABS & Colorant, Stainless steel
-Handle stay – ABS & Colorant
-Linking wires – HDPE, PTFE & Stainless steel
-Clip actuator – Stainless steel & PTFE Sandvik Glidefilm coating
-Clip housing assembly – Polycarbonate & TPE -Clip – Super Elastic Nitinol
-Snare – Stainless Steel
-Safety cap – ABS & Colorant | -Molded handle (components consisting of:)
Thumb ring – ABS & Colorant
Handle shaft- ABS & Colorant
Spool – ABS & Colorant
-Connector – ABS & Colorant, Stainless steel
-Handle stay – ABS & Colorant
-Linking wires – HDPE, PTFE & Stainless steel
-Clip actuator – Stainless steel & PTFE Sandvik Glidefilm coating
-Clip housing assembly – Polycarbonate & TPE -Clip – Super Elastic Nitinol
-Snare – Stainless Steel
-Safety cap – ABS & Colorant | Same |
| Target Population | Adult patients requiring diagnostic tissue acquisition | Adult patients requiring diagnostic tissue acquisition | Same |
STERIS Corporation ▪ 5960 Heisley Road ▪ Mentor, OH 44060-1834 USA ▪ 440-354-2600
{6}
STERIS Special 510(k) PREMARKET NOTIFICATION
Modification to the PADLOCK CLIP EFTR Kit
K250842
Page 3 of 4
| | and full-thickness resection through the removal of lesions in the colon and rectum. | and full-thickness resection through the removal of lesions in the colon and rectum. | |
| --- | --- | --- | --- |
| Energy Used/Delivered | Monopolar High Frequency | Monopolar High Frequency | Same |
| Device Specs: | Overall Length: 230cm
Cap outer diameter: 20mm
Cap inner diameter: 13.3mm
Cap depth: 23mm | Overall Length: 230cm
Cap outer diameter: 20mm
Cap inner diameter: 13.3mm
Cap depth: 23mm | Same |
| Compatible Endoscopes | Suitable for endoscopes with a distal tip O.D. of 10.9mm - 12.5mm | Suitable for endoscopes with a distal tip O.D. of 10.9mm - 12.5mm | Same |
| MRI safety status | MR Conditional | MR Conditional | Same |
| Number of Devices/Box | 1 | 1 | Same |
The proposed device has the same intended use as the predicate with the same technological characteristics. Based on the testing, which can be viewed in summary format in this submission, the change to the labeling does not impact safety and effectiveness of the proposed device.
## 3. Description of Device
The PADLOCK CLIP Endoscopic Full Thickness Resection (EFTR) Device is an electrosurgical, single use tissue resection system. The system is part of a kit that consists of a preloaded, radiopaque, Super Elastic NiTi (Nitinol) single use clip within a mechanical delivery system along with a preloaded snare for tissue resection. The PADLOCK CLIP EFTR Device is mounted over the distal tip of a flexible endoscope. The linking cables run along the outside of the endoscope and are attached to the handle. The PADLOCK CLIP EFTR Device includes a tissue chamber that resides on the distal tip of the endoscope.
The kit's two additional devices, the TOUCHSOFT Coagulator Monopolar Single Use Device, and RAPTOR Grasping Device are used in conjunction with the PADLOCK CLIP EFTR Device through the flexible endoscope's accessory channel. The TOUCHSOFT Coagulator Monopolar Single Use Device (ablation device) will be used to mark a lesion by ablating the lesion's perimeter, thereby indicating the area for resection. Instructions for Use for the TOUCHSOFT Coagulator Monopolar Single Use Device are included within the kit. The RAPTOR Grasping Device will be used to grasp and retract the target lesion into the tissue chamber. Instructions for Use for the RAPTOR are included within the kit.
These devices are used for colorectal lesions that would otherwise not be resectable by standard polypectomy techniques.
The PADLOCK CLIP EFTR Device includes a resection clip made of super elastic nitinol and a snare made of stainless steel. The PADLOCK CLIP EFTR Device also includes handle components molded from ABS material, Stainless Steel drive wires, HDPE catheter, and a scope attachment assembly made of TPE and Polycarbonate.
STERIS Corporation ▪ 5960 Heisley Road ▪ Mentor, OH 44060-1834 USA ▪ 440-354-2600
{7}
STERIS Special 510(k) PREMARKET NOTIFICATION Modification to the PADLOCK CLIP EFTR Kit
K250842 Page 4 of 4
## 4. Intended Use/Indications for Use
The PADLOCK CLIP EFTR Device is an electrosurgical device that is designed for a flexible endoscope. It is intended for diagnostic tissue acquisition and full-thickness resection through the removal of lesions in the colon and rectum.
The PADLOCK CLIP EFTR Device is indicated for the resection of lesions less than 3cm in the colon and rectum. Resection sizes may be less than 3cm due to targeted tissue thickness and degree of fibrosis.
The disposable RAPTOR Grasping Device is intended to be used to grasp tissue and/or retrieve foreign bodies, excised tissue and stents during endoscopic procedures.
The TOUCHSOFT Coagulator is used to provide monopolar therapeutic hemostasis and tissue ablation in the gastrointestinal tract through a flexible endoscope by a trained physician or clinician.
## 5. Summary of Verification Testing
The proposed and predicate devices are the same device with an updated Instructions for Use. Verification testing was performed to show that following the updated Instructions for Use does not raise any new questions or concerns of safety and/or effectiveness as compared to the predicate device.
| Testing | Acceptance Criteria | Result |
| --- | --- | --- |
| Verification Testing | Clip deploys, and the snare closes on the simulated tissue | Pass. |
## 6. Conclusion
Based on the non-clinical verification testing results, and updated risk analysis, the subject device is as safe and effective and performs at least as well as the legally cleared predicate device K240274, Class II (21 CFR 876.4400), product code PKL and associated codes FDI, KNS, and OCZ.
STERIS Corporation ▪ 5960 Heisley Road ▪ Mentor, OH 44060-1834 USA ▪ 440-354-2600
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.