K250841 · Baisheng Medical Co., Ltd. · GXY · Aug 13, 2025 · Neurology
Device Facts
Record ID
K250841
Device Name
Electrotherapy Electrodes
Applicant
Baisheng Medical Co., Ltd.
Product Code
GXY · Neurology
Decision Date
Aug 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Electrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient’s skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation)) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use.
Device Story
Electrotherapy Electrodes serve as disposable, conductive adhesive interface between patient skin and electrical stimulators (TENS, EMS, IF, PGF). Device consists of five-layer structure: tan fabric insulation, double-sided adhesive tape, carbon film, self-adhesive conductive hydrogel, and plastic release film. Connects to stimulators via leadwire or snap button. Used in clinical or home settings by patients or clinicians. Device transmits electrical current from stimulator to patient; adhesive hydrogel secures electrode to skin without additional materials. Benefits include reliable electrical contact for therapeutic stimulation. Single-patient, single-application use.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via electrical safety (IEC 60601-1), high-frequency surgical equipment standards (IEC 60601-2-2), shipping container testing (ASTM D4169-23), shelf-life testing, and biocompatibility (ISO 10993-5, -10, -23).
Technological Characteristics
Five-layer construction: tan fabric, double-sided adhesive, carbon film, conductive hydrogel, plastic release film. Connection via leadwire/snap button (.080"/2mm female socket). Non-sterile. A.C. impedance ≤ 160 ohms. Biocompatibility per ISO 10993. Passive conductive interface; no active electronics or software.
Indications for Use
Indicated for adult patients requiring electrical stimulation therapy via TENS, EMS, IF, or PGF devices. Used as a disposable, conductive adhesive interface between skin and stimulator for current transmission. Suitable for OTC or prescription use.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 13, 2025
Baisheng Medical Co., Ltd.
Kaimin Chen
RA Specialist
No.11, Fusheng Road, Xinhui District
Jiangmen
Guangdong, 529100
China
Re: K250841
Trade/Device Name: Electrotherapy Electrodes
Regulation Number: 21 CFR 882.1320
Regulation Name: Cutaneous Electrode
Regulatory Class: Class II
Product Code: GXY
Dated: July 14, 2025
Received: July 14, 2025
Dear Kaimin Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K250841 - Kaimin Chen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250841 - Kaimin Chen
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250841 | |
| Device Name Electrotherapy Electrodes | |
| Indications for Use (Describe) Electrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient’s skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation)) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
OBS
百生
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
510(K) SUMMARY
510(K) Number: K250841
I. GENERAL INFORMATION
Applicant:
Baisheng Medical Co., Ltd.
No.11,Fusheng Road,Xinhui District,Jiangmen,Guangdong.China. P.C:529100
Chen Xi
Tel: 86-0750-6628113
E-mail: renzheng@bs0750.com
Contact Person:
Chen Kaimin
Phone: +86-15626286758
E-mail: venus@bs0750.com
Date Prepared: Aug 13,2025
II. DEVICE INFORMATION
Trade Name: Electrotherapy Electrodes
Model: LBS-Φ2, LBS-Φ3, LBS-Φ5, LBS-Φ7, LBS-Φ9, LBS-2×2, LBS-4×4, LBS-5×5, LBS-2×3, LBS-5×3.5, LBS-6×4, LBS-6×5, LBS-8×4, LBS-9×6, LBS-10×7, LBS-12×8, LBS-12×10, LBS-8.5×5, LBS-12×7, LBS-15×5, LBS-Φ2L, LBS-Φ3L, LBS-Φ5L, LBS-Φ7L, LBS-Φ9L, LBS-2×2L, LBS-4×4L, LBS-5×5L, LBS-2×3L, LBS-5×3.5L, LBS-6×4L, LBS-6×5L, LBS-8×4L, LBS-9×6L, LBS-10×7L, LBS-12×8L, LBS-12×10L, LBS-8.5×5L, LBS-12×7L, LBS-15×5L
Common Name: Cutaneous Electrode
Classification Name: Cutaneous Electrode
Regulation Number: 21 CFR 882.1320
Regulatory Class: Class II
Product Code: GXY
Classification Panel: Neurology
Type of 510(k) submission: Traditional
510(k) Summary
Vol 002 Doc 01 Page 1 of 8
{5}
OBS
B
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
# PREDICATE DEVICE INFORMATION
## PREDICATE DEVICE INFORMATION
**Predicate device:**
510(K) Number: K180865
Company Name: ZMI Electronics, Ltd.
Trade Name: ZMI Self-Adhesive Electrodes
Common Name: Cutaneous Electrode
Product Code: GXY
## III. DEVICE DESCRIPTION
Electrotherapy Electrodes provide a means for establishing electrical contact between the lead connected to a TENS, EMS or NMES stimulation device and the skin, and are multi-layer, flexible structures composed of laminated materials commonly used in the application:
1st layer: Insulation material: Tan fabric
2nd layer: Double sides adhesive tape
3rd layer: Conductive film (Carbon Film)
4th layer: Self-adhesive conductive hydrogel
5th layer: Plastic release film
Connection: Leadwire/Snap button
The electrodes are designed for single-patient & single application use. Because of the adhesive nature of the conductive hydrogel, no securing materials are required to secure the device to the patient's skin.
510(k) Summary
Vol 002 Doc 01 Page 2 of 8
{6}
OBS
百生
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
## IV. INDICATIONS FOR USE
Electrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient’s skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use.
510(k) Summary
Vol 002 Doc 01 Page 3 of 8
{7}
OBS
百生
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
# V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
The following tables show similarities and differences of use, design, and material between our device and the predicate devices.
Comparison to predicate device:
| Description | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| 510K Number | K250841 | K180865 | - |
| Product Code | GXY | GXY | SE |
| Proprietary Name | Cutaneous Electrode | Cutaneous Electrode | - |
| Manufacturer | Baisheng Medical Co., Ltd. | ZMI Electronics, Ltd. | - |
| Intended use | Electrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient’s skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use. | ZMI Self-adhesive electrodes are intended for use as a reusable, conductive adhesive interface between the patient’s skin and the marketed electrical stimulators (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use. | SE |
510(k) Summary
Vol 002 Doc 01 Page 4 of 8
{8}
OBS
百生
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
| Target Population | Adult | Adult | SE |
| --- | --- | --- | --- |
| Design feature and material | Five layers:
1^{st} layer: Insulation material: Tan fabric
2^{nd} layer: Double sides adhesive tape
3^{rd} layer: Conductive film (Carbon Film)
4^{th} layer: Self-adhesive conductive hydrogel
5^{th} layer: Plastic release film
Connection: Leadwire/Snap button | Three layers:
1. Insulation backing material: Fabric/Foam/ Tan fabric
2. Conductive film: Aluminum foil film /Carbon film/Carbon film coated with silver/
3. Conductive hydrogel | SE |
| Electrical Connection | Leadwire
Snap button | Leadwire
Snap button
Magnetic button | SE |
| Lead Wire connector | Leadwire connector .080” (2mm) female socket connector | Leadwire connector .080” (2mm) female socket connector | SE |
| Non-sterile | Non-sterile | Non-sterile | SE |
| Disposable/ Reusable | Disposable | Reusable | - |
| Adhesive Type | Self-adhesive | Self-adhesive | SE |
| Environment of use | For OTC (Over-The-Counter) or Prescription use. | For OTC (Over-The-Counter) or Prescription use. | SE |
| Biocompatibility | Complies with ISO 10993 | Complies with ISO 10993 | SE |
| Single Patient Use | Yes | Yes | SE |
510(k) Summary
Vol 002 Doc 01 Page 5 of 8
{9}
OBS
百生
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
| Description | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Shape | Rectangular, Round | Round, Rectangular, Square, Oval, and Butterfly | SE
The shape and size are related to the treatment area. Smaller area may cause higher current density, which leads to discomfort and burns. The current densities for any shaped electrodes exceeding 2mA r.m.s /cm² may require the special attention of the user. The color is mainly for identification and marketing purpose, and does not raise any new concerns of safety or effectiveness. |
| Color | Tan | White, Red, Black, and Tan | |
| A.C. Impedance | ≤ 160 ohms | <200 ohms | SE |
## Discussion:
Our device is essentially identical to the predicate device in terms of indication for use, design feature and material, electrical connection, labeling, adhesive type, environment of use, biocompatibility and A.C. Impedance. The minor differences do not raise any safety and effectiveness concerns.
510(k) Summary
Vol 002 Doc 01 Page 6 of 8
{10}
OBS
百生
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
VI. Summary of Non-Clinical Testing
Electrotherapy Electrodes are composed of Tan fabric, Double sided adhesive tape, Conductive film (Carbon Film), self-adhesive conductive hydrogel, Plastic release film, Leadwire/Snap button. Adhesive performance was evaluated by performing a skin adhesion test. Electrode stability was evaluated by assessing electrical performance under IFU normal use conditions.
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use. The following tests were performed on the device.
- Electrical safety per IEC 60601-1
- IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, section 201.15.101.7
- ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
- Shelf life
In addition to the available device-specific guidelines, we have developed a method to test the packaging labeling, product appearance, impedance, dimension, connection between electrode and electrode wire, the cross-sectional area of the electrode connection wire, wire length and plug insertion and unplugging force surrounding the Electrotherapy Electrodes.
510(k) Summary
Vol 002 Doc 01 Page 7 of 8
{11}
OBS
B
Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841
Biocompatibility testing was performed to verify the equivalent safety of the materials that are used. According to ISO 10993-1:2018, we evaluated and conducted the compatibility test for the proposed device. The following tests were identified and done.
- ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
All tests were verified to meet acceptance criteria.
## VII. Clinical Test Conclusion
No clinical study is included in this submission
## VIII. Conclusion
Verification testing of the Electrotherapy Electrodes included electrical and mechanical tests to show that it meets its target specifications over a range of operating and storage conditions. Verification and performance testing further demonstrated that it meets user needs as reflected in the functional specification.
The submitted Electrotherapy Electrodes have the same intended use and similar technological characteristics as the predicate devices. Moreover, information contained in the submission supplied demonstrates that any differences in their characteristics do not raise new concerns of safety or effectiveness. Thus, the submitted adhesive electrodes are substantially equivalent to the predicate devices.
510(k) Summary
Vol 002 Doc 01 Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.