K250839 · Changzhou Boen Zhongding Medical Technology Co., Ltd. · OAS · Jun 11, 2026 · Radiology
Device Facts
Record ID
K250839
Device Name
Dental X-ray System
Applicant
Changzhou Boen Zhongding Medical Technology Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Jun 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Dental X-ray System is a digital, extra-oral imaging system with panoramic, cephalometric, and tomographic capabilities, designed to: a. Produce orthopantomographic images of the maxillofacial region. b. Generate radiographs of the jaws, parts of the skull, and carpus for cephalometric examination, if equipped with a CEPH arm. c. Produce tomographic images of the head, including the ear, nose, and throat (ENT), the dentomaxillofacial complex, teeth, mandible, maxilla, temporomandibular joint (TMJ), and other areas of the human skull and neck, including sections of the cervical spine, for diagnostic support. If equipped with a Dental X-ray Unit, The Dental X-ray Unit is an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors (intra-oral image receptors is not included in our Dental X-ray System). The device is operated and used by physicians, dentists, radiologic technologists, and other qualified professionals.
Device Story
Digital extra-oral imaging system; panoramic, cephalometric, and CBCT capabilities. Input: X-ray radiation passing through patient anatomy; captured by flat-panel detectors (CsI/TFT/CMOS/IGZO). Operation: Motorized column and rotation arm capture multi-angle projections; reconstruction software generates 3D tomographic volumes or 2D panoramic/cephalometric images. Used in dental/clinical settings by dentists, physicians, or radiologic technologists. Output: Diagnostic images for visualization of dentition, jaws, and maxillofacial structures. Clinical decision-making: Assists in diagnosis of dental/oral diseases, treatment planning, and orthodontic assessment. Benefits: Provides high-resolution diagnostic imaging for oral/maxillofacial structures.
Clinical Evidence
Clinical evaluation of 3,015 images (CBCT, panoramic, TMJ, cephalometric). 83.6% (2,520) rated 'Pass' (diagnostically acceptable). 16.4% (495) rated 'Fail', with 16.3% attributed to patient motion, positioning, or metal artifacts, and only 0.1% (3 images) due to device-intrinsic quality failures. Results support diagnostic utility.
Indicated for adult and pediatric patients requiring diagnostic dental, maxillofacial, and cranial radiographic imaging, including panoramic, cephalometric, and CBCT tomographic examinations of teeth, jaws, skull, and cervical spine.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
June 11, 2026
Changzhou Boen Zhongding Medical Technology Co., Ltd.
% Shuo Huang
Regulatory Affairs Specialist
1st to 2nd Floor, South west Side of Building B1, NO.9
Changyang Road, West Taihu Science and Technology Industrial
Changzhou, JS 213100
CHINA
Re: K250839
Trade/Device Name: Dental X-ray System
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: OAS
Dated: March 17, 2025
Received: March 20, 2025
Dear Shuo Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250839 - Shuo Huang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250839 - Shuo Huang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K250839
Device Name
Dental X-ray System
Indications for Use (Describe)
The Dental X-ray System is a digital, extra-oral imaging system with panoramic, cephalometric, and tomographic capabilities, designed to:
a. Produce orthopantomographic images of the maxillofacial region.
b. Generate radiographs of the jaws, parts of the skull, and carpus for cephalometric examination, if equipped with a CEPH arm.
c. Produce tomographic images of the head, including the ear, nose, and throat (ENT), the dentomaxillofacial complex, teeth, mandible, maxilla, temporomandibular joint (TMJ), and other areas of the human skull and neck, including sections of the cervical spine, for diagnostic support.
If equipped with a Dental X-ray Unit, The Dental X-ray Unit is an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors (intra-oral image receptors is not included in our Dental X-ray System).
The device is operated and used by physicians, dentists, radiologic technologists, and other qualified professionals.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Traditional 510(k) Submission
# **K250839**
# **510(K) Summary**
As required by 21 CFR Part 807.92
# **1. Date Prepared [21 CFR 807.92(a)(1)]**
June 7,2026
# **2. Submitter's Information [21 CFR 807.92(a)(1)]**
| Name: | Changzhou Boen Zhongding Medical Technology Co., Ltd. |
| --- | --- |
| Address: | 1st to 2nd Floor, South west Side of Building B1, NO.9 Changyang Road, West Taihu Science and Technology Industrial Park, Changzhou, Jiangsu, 213100 China |
| Contact person: | Shuo Huang |
| Title: | Regulatory Affairs Specialist |
| E-mail: | huang.shuo@bondent.com |
| Tel: | +86-13913615957 |
# **3. Identification of subject device [21 CFR 807.92(a)(2)]**
| 510(K) number: | K250839 |
| --- | --- |
| Trade/Device Name: | Dental X-ray System |
| Common name: | x-ray, tomography, computed, dental |
| Regulation Number: | 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Regulation Class: | Class 2 |
| Panel: | Radiology |
| Product Code: | OAS |
# **4. Identification of predicate Device [21 CFR 807.92(a)(3)]**
| 510(K) number: | K212254 |
| --- | --- |
| Trade/Device Name: | DENTRIα series |
| Common name: | x-ray, tomography, computed, dental |
| Regulation Number: | 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Regulation Class: | Class 2 |
| Panel: | Radiology |
| Product Code: | OAS |
| 510(K) number: | K223058 |
| --- | --- |
510(k) Summary
Page 1 of Page 15
{5}
Traditional 510(k) Submission
| Trade/Device Name: | EzRay Air 2 Wall (Model: VEX-S350W) |
| --- | --- |
510(k) Summary
Page 2 of Page 15
{6}
Traditional 510(k) Submission
| Common name: | unit, x-ray, extraoral with timer |
| --- | --- |
| Regulation Number: | 872.1800 |
| Regulation Name: | Extraoral source x-ray system |
| Regulation Class: | Class 2 |
| Panel: | Radiology |
| Product Code: | EHD |
### 5. Device Description [21 CFR 807.92(a)(4)]
The Dental X-ray System is a panoramic, cephalometric (Optional), cone beam computed
tomography and dental radiograph (Optional) device. It can
- producing tomographic images of the oral and maxillofacial region, for diagnostic examination of dentition (teeth), jaws, oral structures and some cranial bones.
- produce panoramic X-ray images for diagnostic examination of dentition (teeth), jaws and oral structures;
- produce radiographs of maxillofacial region and parts of the skull for cephalometric examination, if equipped with CEPH.
-produce diagnostic dental radiographs for the treatment of diseases of the teeth, jaw, and other oral structures using intra-oral image Receptors (intra-oral image receptors are not included in our Dental X-ray System), if equipped with dental X-ray Unit.
The Dental X-ray System is composed of Column, assembled mobile group, rotation arm, CEPH arm, control panel.
The Column is the main vertical support structure of the CBCT system. It provides stability and serves as the foundation to which other components. It can raise the whole assembled mobile group of the device to ensures the positioning and height adjustment of the system to accommodate patients of different sizes.
The assembled mobile group is movable parts of the CBCT system that can be raised by the column, it is the motorized components that enable the movement of the rotation arm and the patient positioning system. It provides flexibility in positioning the patient.
The rotation arm is responsible for the circular motion around the patient's head, allowing the CBCT system to capture multiple images from different angles. These images are then reconstructed into a 3D image.
The CEPH arm is a dedicated component for taking cephalometric X-rays, it captures 2D images of the patient's head from lateral (side) views.
The control panel is the user interface of the CBCT system, where the operator can set image mode, move the mobile column, turn the positioning lasers on and off, and Control Rotation arm reset.
Additionally, in model Bondream 3D-1030Pro (config 3), a wall-mounted Dental X-ray Unit (model:
510(k) Summary
Page 3 of Page 15
{7}
Traditional 510(k) Submission
Bondream 2000) is included, which is an imaging device specially used to photograph teeth and their surrounding structures.
1) Configurations
| | X-Ray Generator | CT/Pano Detector | CEPH Detector | Dental X-ray Unit |
| --- | --- | --- | --- | --- |
| Bondream 3D-1030Pro (config 3) | Bon-1000D | Jupi0606X1 | Pluto0900X1 | Bondream 2000 |
| Bondream 3D-1030Pro (config 6) | Bon-1000D | Jupi0606X1 | Without | Without |
| Bondream 3D-1030Pro (config 8) | Bon-1000D | Jupi0606X1 | Pluto0900X1 | Without |
| Bondream 3D-1030X2 | Bon-1000DR | CareView 300RF | CareView750L | Without |
| Bondream 3D-1030X3 | Bon-1000DR | CareView 300RF | Without | Without |
2) Description of the image detectors used.
| | CT/Pano Detector | | CEPH Detector | |
| --- | --- | --- | --- | --- |
| Detector model | Jupi0606X1 | CareView 300RF | Pluto0900X1 | CareView 750L |
| Manufacturer | iRay | CareRay | iRay | CareRay |
| Detector type | Flat panel, Csl+TFT, Csl/GOS | Flat panel, Csl+IGZO+TFT | Flat panel, Csl+CMOS+TFT | Flat panel, Csl+ a-Si +TFT |
| Resolution (pixels) | 1536x1536 | 1536x2048 | 2252 × 68 | 2560×2048 |
| Pixel size (μm) | 100 | 98 | 100 | 120 |
| MTF | 56%@1lp/mm | 60%@1lp/mm | 100%@0lp/mm | 70%@1lp/mm |
| DQE | 62%@0lp/mm | 70%@0.5lp/mm, 2uGy | / | 70%@0lp/mm |
| Active area (mm) | 153.6x153.6 | 150.0x200.0 | 225.2×6.8 | 307.2×244.3 |
| A/D Conversion | 16 bits | 16 bits | 16 bits | 16 bits |
### 6. Indication for Use [21 CFR 807.92(a)(5)]
The Dental X-ray System is a digital, extra-oral imaging system with panoramic, cephalometric, and tomographic capabilities, designed to:
a. Produce orthopantomographic images of the maxillofacial region.
b. Generate radiographs of the jaws, parts of the skull, and carpus for cephalometric examination, if equipped with a CEPH arm.
510(k) Summary
Page 4 of Page 15
{8}
Traditional 510(k) Submission
c. Produce tomographic images of the head, including the ear, nose, and throat (ENT), the dentomaxillofacial complex, teeth, mandible, maxilla, temporomandibular joint (TMJ), and other areas of the human skull and neck, including sections of the cervical spine, for diagnostic support.
If equipped with a Dental X-ray Unit, The Dental X-ray Unit is an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors (intra-oral image receptors is not included in our Dental X-ray System).
The device is operated and used by physicians, dentists, radiologic technologists, and other qualified professionals.
510(k) Summary
Page 5 of Page 15
{9}
Traditional 510(k) Submission
## 7. Determination of Substantial Equivalence
Summary of technological characteristics of the device compared to the predicate device. [21 CFR 807.92(a)(6)]
Compared to the predicate devices, the subject device has the similar indication for use, same technical principle, similar performance specification, same safety as the predicate device, the summarized comparison information is listed in the following table
| SE Comparisons | Subject Devices | | Predicate Device | Similarities /Differences |
| --- | --- | --- | --- | --- |
| | Bondream 3D-1030Pro (config 3) Bondream 3D-1030Pro (config 6) Bondream 3D-1030Pro (config 8) | Bondream 3D-1030X2 Bondream 3D-1030X3 | DENTRIα series | |
| 510(k) Number | K250839 | K250839 | K212254 | / |
| Common/Usual Name | x-ray, tomography, computed, dental | x-ray, tomography, computed, dental | x-ray, tomography, computed, dental | / |
| Regulation Number | 892.1750 | 892.1750 | 892.1750 | / |
| Product Code | OAS | OAS | OAS | / |
| Class | Class II | Class II | Class II | / |
| Indications for Use | The Dental X-ray System is a digital, extra-oral imaging system with panoramic, cephalometric, and tomographic capabilities, designed to: a. Produce orthopantomographic images of the maxillofacial region. b. Generate radiographs of the jaws, parts of the skull, and carpus for cephalometric examination, if equipped with a CEPH arm. | The Dental X-ray System is a digital, extra-oral imaging system with panoramic, cephalometric, and tomographic capabilities, designed to: a. Produce orthopantomographic images of the maxillofacial region. b. Generate radiographs of the jaws, parts of the skull, and carpus for cephalometric examination, if equipped with a CEPH arm. | The DENTRIα series is a Computed Tomography X-Ray imaging device specialized in diagnosing general dental treatments and orthodontic purpose using Panoramic and Cephalometric images respectively. In addition DENTRIα series is used in the field of Otolaryngology by capturing 360 | similar |
510(k) Summary
Page 6 of Page 15
{10}
Traditional 510(k) Submission
| | c. Produce tomographic images of the head, including the ear, nose, and throat (ENT), the dentomaxillofacial complex, teeth, mandible, maxilla, temporomandibular joint (TMJ), and other areas of the human skull and neck, including sections of the cervical spine, for diagnostic support. If equipped with a Dental X-ray Unit, The Dental X-ray Unit is an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors (intra-oral image receptors is not included in our Dental X-ray System). The device is operated and used by physicians, dentists, radiologic technologists, and other qualified professionals. | c. Produce tomographic images of the head, including the ear, nose, and throat (ENT), the dentomaxillofacial complex, teeth, mandible, maxilla, temporomandibular joint (TMJ), and other areas of the human skull and neck, including sections of the cervical spine, for diagnostic support. If equipped with a Dental X-ray Unit, The Dental X-ray Unit is an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors (intra-oral image receptors is not included in our Dental X-ray System). The device is operated and used by physicians, dentists, radiologic technologists, and other qualified professionals. | degree rotation sequence of the head and neck areas, including the ENT and dentomaxillofacial areas for a dental treatment in adult and pediatric dentistry and obtains x-ray images from different angles and calculate through computer-processed to produce 3D xray tomographic images. The DENTRIα series used by physicians, dentists, and x-ray technologists. | |
| --- | --- | --- | --- | --- |
| Operation Mode | CBCT Mode PANO Mode CEPH Mode(optional) Dental Mode(optional) | 1. CBCT Mode 2. PANO Mode 3. CEPH Mode(optional) | 1. CT 2. Panorama 3. Ceph 1) One-Shot type 2) Scan type 4. Model Scan | similar |
| X-ray generator | Bon-1000D | Bon-1000DR | PXD-140CT | different |
| X-ray tube | D-054SB | D-054SB | OPX/105 (C.E.I.) | different |
| Target angle | 5° | 5° | 5° | same |
510(k) Summary
Page 7 of Page 15
{11}
Traditional 510(k) Submission
| Focal spot size | 0.5mm | 0.5mm | 0.5 mm | same |
| --- | --- | --- | --- | --- |
| Inherent filtration | 0.8mm Al | 0.8mm Al | 0.5mm Al | same |
| Anode material | Tungsten | Tungsten | Tungsten | same |
| Total filtration | 2.8mm Al | 2.8mm Al | > 2.5 mmAl > 2.5 mmAl + 0.5 mmCu (Optional) | similar |
| Range of X-ray Tube Voltage | 60kV-100kV | 60kV-100kV | PXD-140CT (Optional) 1. CT 60-110 kV ±8% 2. Panorama 60-90 kV ±8% 3. Cephalo 1) One-Shot type 60-110 kV ±8% 2) Scan type 60-90 kV ±8% | similar |
| Range of X-ray Tube Current | 4mA-10mA | 4mA-10mA | 4-10 mA ±10% | same |
| Range of Irradiation Time | 8s,24s,36s (CBCT mode) 14s,9.6s (PANO mode) 14s,8.4s (CEPH mode)-- - applicable to Bondream 3D- 1030Pro(config 3&8) 0.01s-2.0s, using R'10 numerical system (Dental X-ray Unit)-- - applicable to Bondream 3D- 1030Pro(config 3) | 8s,16s,20s (CBCT mode) 5s,9.6s ,12s,14s(PANO mode) 0.5s, 0.8s, 1.6s (CEPH mode)- -applicable to Bondream 3D- 1030X2 | 1. CT 8.0-36.0 s ± (5 % + 50 ms) 2. Panorama 1.2-14.0 s ± (5 % + 50 ms) 3. Cephalo 1) One-Shot type 0.5 s to 2.0 s ± (5 % + 50 ms) (in 0.5 s increments) 2) Scan type 2.5-8.0 s ± (5 % + 50 ms) 4. Model Scan 24 s | similar |
510(k) Summary
Page 8 of Page 15
{12}
Traditional 510(k) Submission
| Detector type | CBCT/PANO:Flat panel, CsI+TFT, CsI/GOS ; CEPH:Flat panel, CsI+CMOS+TFT | CBCT/PANO:Flat panel, CsI+TFT+IGZO; CEPH : Flat panel, Flat panel, CsI+ a-Si +TFT | 1. CT CMOS or TFT:a-Si 2. Panorama CMOS or TFT:a-Si 3. Cephalo 1) One-shot type TFT:a-Se 2) Scan type CMOS | similar |
| --- | --- | --- | --- | --- |
| Detector Pixel size (um) | 100 um (CBCT/PANO/CEPH) | CBCT/PANO:100 um CEPH:120 um | 1. CT 100.1 um or 119 um 2. Panorama 100.1 um or 119 um 3. Cephalo 1) One-shot type 129.0 um 2) Scan type 99.0 um | different |
| Active area(mm) | CBCT/PANO:153.6mm x153.6mm CEPH:225.2mm × 6.8mm | CBCT/PANO:204.8mmx204.8mm CEPH:307.2mm × 244.3mm | 1. CT 131mm x 131mm or 149.464 mmx 149.464mm 2. Panorama 6mm x 130mm or 6mm x 149.5mm 3. Cephalo 1) One-shot type 198mm x 264mm 2) Scan type 6.7mm x 228mm | different |
| MTF | CBCT/PANO:56%@1lp/mm CEPH:100%@0lp/mm | CBCT/PANO:60%@1lp/mm CEPH:70%@1lp/mm | 1. CT: - 57% at 1 lp/mm or - 60% at 1 lp/mm 2. Panorama: - 57% at 1 lp/mm or - 60% at 1 lp/mm 3. Cephalo 1) One shot type | different |
510(k) Summary
Page 9 of Page 15
{13}
Traditional 510(k) Submission
| | | | - 83.3% at 2 lp/mm 2) Scan type: - 65% at 1 lp/mm | |
| --- | --- | --- | --- | --- |
| DQE | CBCT/PANO:62%@0lp/mm | CBCT/PANO:70%@0.5lp/mm,2u Gy CEPH:70%@0lp/mm | 1. CT: - 70% at 0 lp/mm or - 60% at 1 lp/mm 2. Panorama: - 70% at 0 lp/mm or - 60% at 1 lp/mm 3. Cephalo 1) One shot type - 38.5% at 0 lp/mm 2) Scan type: - 57% at 1 lp/mm | different |
| Source Image Distance | 600mm (CBCT mode) 560mm (PANO mode) 1765mm (CEPH mode) | 600mm (CBCT mode) 560mm (PANO mode) 1629mm (CEPH mode) | 1. CT 600 mm 2. Pano 560 mm 3. Cephalo 1) One-Shot Ceph 1792 mm 2) Scan Ceph 1,782 mm | similar |
| Format Compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible | same |
### Substantial Equivalence Discussion
The Bondream 3D-1030Pro series and Bondream 3D-1030series are substantially equivalent to the predicate device identified above with respect to intended use, principles of operation, and technological characteristics. From the information provided in table above, it is understood that the subject device does not introduce any new technology and/or indications of use. Therefore, the Bondream 3D-1030Pro series and Bondream 3D-1030series are considered substantially equivalent to the predicate device.
510(k) Summary
Page 10 of Page 15
{14}
Traditional 510(k) Submission
| SE Comparisons | Subject Devices | Predicate Device | Similarities/Differences |
| --- | --- | --- | --- |
| | Dental X-ray Unit of Bondream 3D-1030Pro (config 3)(Model:Bondream 2000) | EzRay Air 2 Wall (Model: VEX-S350W) | |
| 510(k) Number | K250839 | K223058 | / |
| Common/Usual Name | Extraoral source x-ray system | Extraoral source x-ray system | / |
| Regulation Number | 21 CFR 872.1800 | 21 CFR 872.1800 | / |
| Product Code | EHD | EHD | / |
| Class | II | II | / |
| Intended Use | Dental X-ray Unit (Model:Bondream 2000) is intended to produce diagnostic dental radiographs for the treatment of diseases of the teeth, jaw, and other oral structures using intra-oral image receptors. | The EzRay Air 2 Wall (Model: VEXS350W) is a dental X-ray system intended for use by a trained and qualified dentist or dental technician for both adult and pediatric subjects for producing diagnostic dental radiographs for the treatment of diseases of the teeth, jaw, and other oral structures using intra-oral image receptors. | similar |
| Patient | adult and pediatric | adult and pediatric | same |
| **Mechanical** | | | |
| Minimum source to skin distance | 200mm | 200 mm | same |
| X-ray field Size | 60mm round | 60 mm round | same |
| Focal spot | 0.4 mm | 0.4 mm | same |
| Exposure time | 0.01-2.0s | 0.05 - 2.0 seconds | similar |
| Tube current | 5mA | 5mA | same |
| Tube | 70kV | 70 kV | same |
510(k) Summary
Page 11 of Page 15
{15}
Traditional 510(k) Submission
| voltage | | | |
| --- | --- | --- | --- |
| Electric Power Voltage | 230VAC | AC 100-240 V | similar |
| Rated Current | 8A | 10 A | different |
### **Substantial Equivalence Discussion**
The Bondream 2000 is substantially equivalent to the predicate device identified above with respect to intended use, principles of operation, and technological characteristics. From the information provided in table above, it is understood that the subject device does not introduce any new technology and/or indications of use. Therefore, the Bondream 2000 is considered substantially equivalent to the predicate device.
510(k) Summary
Page 12 of Page 15
{16}
Traditional 510(k) Submission
### 8. Performance Data
#### Clinical test:
Subject assessed
5 clinical images (3000 diagnostic radiographic images) of the CBCT image (covering all listed anatomical regions),
5 clinical images Panoramic image (full dentition and jaw structure),
5 clinical images Panoramic image of the TMJ,and
5 clinical images Lateral cephalometric image
Acceptance criteria
PASS - Diagnostically Acceptable:
The image demonstrates sufficient spatial resolution, contrast resolution, and structural alignment to successfully fulfill all assigned diagnostic tasks. Minor non-systemic anomalies - including localized beam hardening or streak artifacts caused by the patient's own existing metallic restorations – are permissible only where they do not obliterate the primary regions of diagnostic interest and do not prevent formulation of a complete clinical diagnosis and treatment plan without supplemental imaging.
FAIL - Diagnostically Not Acceptable:
The image fails to meet the minimum threshold for clinical diagnostic utility due to severe artifact degradation, extreme motion blur, structural distortion or improper patient positioning. A FAIL is assigned when key anatomical landmarks or pathologies are completely obscured, when unacceptable geometric distortion is present, or when image quality would necessitate a mandatory retake in standard clinical practice.
Results
| Metric | Count | Percentage |
| --- | --- | --- |
| Total | 3015 | 100% |
| Pass | 2520 | 83.6% |
| Fail | 495 | 16.4% |
| Of which - Device-Intrinsic Quality Failures(Resolution / Contrast only) | 3 | 0.1% |
| Of which - Metal Artifact / Patient Motion / PositioningErrors | 492 | 16.3% |
#### Conclusion
The conclusion show that the clinical images provided support the safety and effectiveness of the device for its intended use.
510(k) Summary
Page 13 of Page 15
{17}
Traditional 510(k) Submission
## 9. Non-clinical data
The Bondream 3D-1030Pro series, Bondream 3D-1030series and Bondream 2000 are verified and validated according to the FDA design control requirements, 21 CFR 820. The subject device had been subjected to the applicable safety and performance testing before release to ensure the device meets all its specifications. The quality assurance measures applied to the design and development of the subject device include, but not limited to risk analysis, verification and validation, product specifications and design reviews.
### 1) Thermal, electrical, mechanical safety & Electromagnetic Compatibility
The Bondream 3D-1030Pro series, Bondream 3D-1030series and Bondream 2000 comply with the electrical safety and electromagnetic compatibility requirements established by the standards below:
- IEC 60601-1:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-3:2021 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- IEC 60601-2-63:2021 Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment
- IEC 60601-2-65:2021 Medical electrical equipment - Part 2-65: Particular requirements for the basic safety and essential performance of dental intra-oral-X-ray equipment
- IEC 60825-1:2007 Safety of laser products - Part 1: Equipment classification and requirements
- IEC 60601-1-2:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
### 2) Software Verification and Validation
Software information is provided in accordance with FDA guidance: “Content of Premarket Submissions for Device Software Functions, issued on June 14, 2023.”
Cybersecurity information is provided in accordance with FDA guidance: “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued on September 27, 2023”
### 3) Biocompatibility
Biocompatibility evaluation is provided in accordance with FDA guidance: “Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', issued on: September 8, 2023”
Biocompatibility tests were conducted in accordance with
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
510(k) Summary
Page 14 of Page 15
{18}
Traditional 510(k) Submission
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation
#### **4) Performance Test**
Bench Testing were conducted for CBCT Imaging in accordance with IEC 61223-3-7: 2021 Evaluation and routine testing in medical imaging departments - Part 3-7: Acceptance and constancy tests - Imaging performance of X-ray equipment for dental cone beam computed tomography. In bench testing, spatial resolution, low-contrast resolution, image uniformity, voxel accuracy, and reconstruction efficiency were verified, the verification results show that the subject devices meet claims.
Bench Testing were conducted for 2D X-ray Imaging in accordance with FDA guidance: 'Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices'. In bench testing, the physical characteristics, operational functions, functional characteristics, exposure characteristics and safety feature were verified, the verification results shows that the subject devices meet claims.
#### **9. Conclusion**
The conclusions drawn from the nonclinical tests demonstrate that the subject device, Dental X-ray System (Bondream 3D-1030Pro (config 3), Bondream 3D-1030Pro (config 6), Bondream 3D-1030Pro (config 8), Bondream 3D-1030X2, Bondream 3D-1030X3), are as safe, as effective, and performs as well as the legally marketed predicated device DENTRIα series (K212254).
The conclusions drawn from the nonclinical tests demonstrate that the subject device, Dental X-ray Unit of Bondream 3D-1030Pro (config 3) (Model: Bondream 2000), is as safe, as effective, and performs as well as the legally marketed predicated device EzRay Air 2 Wall (Model: VEX-S350W) (K223058).
510(k) Summary
Page 15 of Page 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.