RhythmStar System (SL)

K250793 · RhythMedix, LLC · QYX · Apr 21, 2026 · Cardiovascular

Device Facts

Record IDK250793
Device NameRhythmStar System (SL)
ApplicantRhythMedix, LLC
Product CodeQYX · Cardiovascular
Decision DateApr 21, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K250793 · Apr 21, 2026RhythmStar System (SL)RhythMedix, LLCReal-world clinical patient datasetThe sponsor used a dataset of ECG data collected from patients in real-world clinical settings to validate that the hardware modification (lead placement) did not negatively impact the performance of the existing arrhythmia detection algorithm.Arrhythmia detection; Algorithm validation; Real-world clinical data

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective analysis of clinical dataPatients undergoing cardiac monitoringNot applicable for this studyArrhythmia detection algorithm performance

Indications for Use

The RhythmStar SL System is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers all recorded cardiac activity to the cloud server where it is presented and can be reviewed by a medical professional at a monitoring center. The RhythmStar SL is not intended for real-time monitoring. The data received from the RhythmStar SL can be used for retrospective review, arrhythmia analysis, and further evaluation, reporting and signal measurements using RhythmStar system. The RhythmStar SL is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, or provide for life support. The RhythmStar SL does not provide interpretive or diagnostic statements. Interpretation and diagnosis are the responsibility of a physician. RhythmStar SL is for prescription use only.

Device Story

Portable, battery-powered, wireless cardiac monitor; worn on chest for up to 30 days. Records ECG and activity level data; uses embedded arrhythmia detection algorithm to identify events like Bradycardia, Tachycardia, and Atrial Fibrillation. Automatically transmits data via 4G LTE modem or USB to cloud server. Used in ambulatory/outpatient settings; operated by patient. Data reviewed by medical professionals at monitoring center for retrospective analysis, arrhythmia evaluation, and signal measurement. No real-time monitoring or alarms; no therapy or life support. Benefits include continuous cardiac monitoring for patients with intermittent symptoms to aid physician diagnosis.

Clinical Evidence

No clinical trials; performance testing conducted. Evaluated algorithm performance using new lead placement on real-world clinical datasets and performed side-by-side comparison with predicate device to demonstrate identical beat detection. Biocompatibility evaluated per ISO 10993-1. Cybersecurity penetration and vulnerability testing performed per FDA guidance.

Technological Characteristics

Materials: ABS and TPU plastic. Sensing: Analog ECG front-end, accelerometer. Energy: Rechargeable 3.7V Li-ion battery. Connectivity: 4G LTE modem, USB. Form factor: 59mm x 50mm x 15mm, 38g. Sampling: 256 Hz. Bandwidth: 0.05-40 Hz. CMRR: 100 dB min. Ingress Protection: IP26. Sterilization: Non-sterile. Software: IEC 62304 compliant.

Indications for Use

Indicated for ambulatory patients with or at risk of cardiac disease, or those with intermittent symptoms requiring continuous cardiac monitoring.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 21, 2026 RhythMedix LLC % Rita King Chief Executive Officer MethodSense, Inc. 1 Copley Parkway Suite 130 Morrisville, North Carolina 27560 Re: K250793 Trade/Device Name: RhythmStar System (SL) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: QYX Dated: March 27, 2026 Received: March 27, 2026 Dear Rita King: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250793 - Rita King Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250793 - Rita King Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JENNIFER W. SHIH -S Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250793 | | | Device Name RhythmStar System | | | Indications for Use (Describe) The RhythmStar SL System is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers all recorded cardiac activity to the cloud server where it is presented and can be reviewed by a medical professional at a monitoring center. The RhythmStar SL is not intended for real-time monitoring. The data received from the RhythmStar SL can be used for retrospective review, arrhythmia analysis, and further evaluation, reporting and signal measurements using RhythmStar system. The RhythmStar SL is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, or provide for life support. The RhythmStar SL does not provide interpretive or diagnostic statements. Interpretation and diagnosis are the responsibility of a physician. RhythmStar SL is for prescription use only. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} RhythmStar System (SL) Traditional 510(k) K250793 # Summary of 510(k) RhythMedix, LLC This 510(k) Summary is in conformance with 21 CFR 807.92. Submitter: RhythMedix, LLC 5000 Atrium Way Ste. 1 Mount Laurel, NJ 08054 United States Primary Contact: Rita King, CEO MethodSense, Inc. Email: rltaking@method sense.com Phone: 919-313-3961 Fax: 919-313-3979 Company Contact: Stan Biletsky, CTO RhythMedix, LLC Email: sbiletsky@rhythmedix.com Phone: 856-282-1080 ext. 223 # Device Name and Classification Trade Name: RhythmStar System Common Name: Outpatient cardiac telemetry Classification: Class II Regulation Number: 21 CFR 870.1025 Classification Panel: Cardiovascular Product Code: QYX # Predicate Device: | Primary Predicate | | | --- | --- | | Trade Name | RhythmStar System | | Common Name | Outpatient Cardiac Telemetry | | 510(k) Submitter / Holder | RhythMedix, LLC | | 510(k) Number | K233584 | | Regulation Number | 21 CFR Part 870.1025 | | Classification Panel | Cardiovascular | | Product Code | QYX | The predicate device has not been subject to a design-related recall. RhythMedix, LLC {5} RhythmStar System (SL) Traditional 510(k) K250793 ## Device Description The RhythmStar system (model: SL) consists of the RhythmStar device and the server. The RhythmStar device is a portable, battery-powered, wireless cardiac monitor which may be worn by a patient to record ECG and activity level data for up to 30 consecutive days. The device can capture patient activated and auto-triggered events such as Bradycardia, Tachycardia, and Atrial Fibrillation as identified by an embedded arrhythmia detection algorithm. The device is capable of automatically delivering the collected ECG data to the server using a built-in 4G LTE wireless data modem, or the data can be transferred from the device using a USB connection. The data transmitted by the RhythmStar device can be stored, evaluated, and presented for review, analysis and reporting by a medical professional using a server, such as the RhythmStar System server. ## Indications for Use The RhythmStar SL system is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers all recorded cardiac activity to the cloud server where it is presented and can be retrospectively reviewed by a medical professional at a monitoring center. The RhythmStar SL is not intended for real-time monitoring. The data received from the RhythmStar SL can be used for retrospective review, arrhythmia analysis, and further evaluation, reporting and signal measurements using RhythmStar system. The RhythmStar SL is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, or provide for life support. The RhythmStar SL does not provide interpretive or diagnostic statements. Interpretation and diagnosis are the responsibility of a physician. RhythmStar SL is for prescription use only. ## Risk Analysis Method The updated RhythmStar monitoring device was assessed to determine the risks of the device as it relates to a hardware modification. A risk analysis was conducted in accordance with ISO 14971:2019, Medical devices – Application of risk management to medical devices. No new risks or changes to existing risks were identified. ## Substantial Equivalence The table below provides a side-by-side comparison between the RhythmStar SL and the predicate device with respect to the intended use, technological characteristics, and principles of operation. RhythmMedix, LLC {6} RhythmStar System (SL) Traditional 510(k) K250793 Table 1. Substantial Equivalence Table | Characteristic | Subject Device RhythmStar SL | Primary Predicate Device RhythmStar RS-10003 (K233584) | Comparison | | --- | --- | --- | --- | | Intended Use | The RhythmStar System is intended for continuous ECG monitoring and arrhythmia detection. | The RhythmStar System is intended for continuous ECG monitoring and arrhythmia detection. | Identical | | Indications for Use | The RhythmStar SL System is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers all recorded cardiac activity to the cloud server where it is presented and can be reviewed by a medical professional at a monitoring center. The RhythmStar SL is not intended for real-time monitoring. The data received from the RhythmStar SL can be used for retrospective review, arrhythmia analysis, and further evaluation, reporting and signal measurements using RhythmStar system. The RhythmStar SL is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, or provide for life support. The RhythmStar SL does not provide interpretive or diagnostic statements. Interpretation and diagnosis are the responsibility of a physician. RhythmStar SL is for prescription use only. | The RhythmStar System is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers all recorded cardiac activity to the cloud server where it is presented and can be reviewed by a medical professional at a monitoring center. The RhythmStar system is not intended for real-time monitoring. The data received from the RhythmStar device can be used for retrospective review, arrhythmia analysis, and further evaluation, reporting and signal measurements using RhythmStar system or a compatible arrhythmia analysis software that has been FDA cleared for Lead II analysis using non-traditional wet electrode placement. The RhythmStar system is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, or provide for life support. The RhythmStar system does not provide interpretive or diagnostic statements. Interpretation and diagnosis are the responsibility of a physician. RhythmStar is for prescription use only. | Equivalent | RhythmMedix, LLC {7} RhythmStar System (SL) Traditional 510(k) K250793 | Characteristic | Subject Device RhythmStar SL | Primary Predicate Device RhythmStar RS-10003 (K233584) | Comparison | | --- | --- | --- | --- | | Prescription / OTC | Prescription | Prescription | Identical | | Intended Population | Patients requiring cardiac monitoring who are ambulatory and without life-threatening conditions | Patients requiring cardiac monitoring who are ambulatory and without life-threatening conditions | Identical | | Environment of Use | Ambulatory, Outpatient | Ambulatory, Outpatient | Identical | | Dimensions | 59mm x 50mm x 15mm | 59mm x 50mm x 15mm | Identical | | Weight | 38 g | 38 g | Identical | | Material | ABS and TPU plastic | ABS and TPU plastic | Identical | | Duration of Use | Up to 30 days | Up to 30 days | Identical | | Type of Wear | Device worn on chest | Device worn on chest | Identical | | Basic Technology | Analog ECG front-end, accelerometer, MCU, flash data storage, built-in cellular modem for data transmission | Analog ECG front-end, accelerometer, MCU, flash data storage, built-in cellular modem for data transmission | Identical | | Battery Powered | Yes | Yes | Identical | | Energy Source | Rechargeable 3.7V Li-ion battery | Rechargeable 3.7V Li-ion battery | Identical | | Battery Life | 72 hours | 72 hours | Identical | | Removable Battery | No | No | Identical | | Bandwidth | 0.05 ... 40 Hz | 0.05 ... 40 Hz | Identical | | Common-mode rejection ratio (CMRR) | 100 dB minimum, 115 dB typical | 100 dB minimum, 115 dB typical | Identical | | Sampling Rate | 256 Hz | 256 Hz | Identical | | ECG leads | 2 | 2 | Identical | | ECG Acquisition | Body worn monitoring device | Body worn monitoring device | Identical | | Distance between the electrodes | 5.5" | 7.5" | Different – This change does not affect the intended use of the device. Substantial Equivalence of the device have been demonstrated through performance testing. | | Lead off Detection | Yes | Yes | Identical | | Accelerometer | Yes | Yes | Identical | | Display | LED lights | LED lights | Identical | | User Event Trigger | Double tap on monitoring device | Double tap on monitoring device | Identical | RhythMedix, LLC Page 4 of 6 {8} RhythmStar System (SL) Traditional 510(k) K250793 | Characteristic | Subject Device RhythmStar SL | Primary Predicate Device RhythmStar RS-10003 (K233584) | Comparison | | --- | --- | --- | --- | | Wireless Communication Technology | Yes - 4G mobile network using embedded cellular data modem | Yes - 4G mobile network using embedded cellular data modem | Identical | | System Design | PEMS and software | PEMS and software | Identical | | Sterile | Non-sterile | Non-sterile | Identical | | Use Type | Multi-Patient Use | Multi-Patient Use | Identical | | Arrhythmia Detection Algorithm | Yes | Yes | Identical | | Alarms | None | None | Identical | | Adjustable Device Programming Parameters | Yes | Yes | Identical | | Review of ECG Data | ECG Technicians at 24/7 Monitoring Center | ECG Technicians at 24/7 Monitoring Center | Identical | | Ingress Protection (IP) rating | IP26 (product is protected from touch by fingers and objects larger than 12.5 millimeters, and it's also protected against powerful water jets from any direction) | IP26 (product is protected from touch by fingers and objects larger than 12.5 millimeters, and it's also protected against powerful water jets from any direction) | Identical | RhythMedix, LLC Page 5 of 6 {9} RhythmStar System (SL) Traditional 510(k) K250793 # Testing The RhythmStar SL performed the following evaluations in K233584. The change to the subject device from the predicate device included shortening the leadwire and revising the placement of the ECG electrodes. Information was provided to support the applicability of prior evaluations for the following: - Electrical Safety and EMC Testing Electrical safety of the RhythmStar device was performed in accordance with the applicable standards and guidance documents in K233584. - Software Verification and Validation The RhythmStar software was developed and is maintained in accordance with IEC 62304 Medical device software – Software life cycle processes. All relevant processes including the software development process, risk management activities, verification & validation, configuration management, maintenance, and problem resolution procedures have been implemented and maintained in accordance with IEC 62304 requirements (see K233584). - Performance Testing The RhythmStar System was tested in accordance with ANSI/AAMI EC57 (see K233584). Performance testing confirmed the RhythmStar device met the requirements of ANSI/AAMI EC53 and ANSI/AAMI EC12 in K233584. The following evaluations were performed for the RhythmStar SL in support of this submission: - Biocompatibility Biocompatibility of the RhythmStar device was evaluated in accordance with ISO 10993-1. - Performance Testing Additional performance testing was conducted as part of this submission to demonstrate that the hardware modification and adjusted device placement do not affect performance of the unchanged arrhythmia detection algorithm. This performance testing included evaluating algorithm performance using the new lead placement on a dataset derived from patients in real-world clinical settings as well as demonstrating identical beat detection compared to the predicate using a side-by-side comparison with patients wearing the subject device and predicate simultaneously. - Cybersecurity Testing Cybersecurity penetration and vulnerability testing was performed to identify vulnerabilities and exploits in accordance with FDA Guidance, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions. # Conclusion In conclusion, the intended use for RhythmStar SL is identical to the intended use of the primary predicate (RhythmStar System (K233584)) with regard to continuous ECG monitoring and arrhythmia detection. Additionally, the technological characteristics of the subject device are equivalent to those of the predicate (RhythmStar System (K233584)). Testing demonstrates that the RhythmStar SL is substantially equivalent to the predicate device (RhythmStar System (K233584)). RhythmMedix, LLC
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...