Sil-Flow is indicated for Temporary inlay and onlay treatments of the cavity Sealing of openings for implant screws Relining material for temporary crowns and bridges Block-out material for retentive areas in the dental arch, e.g. before taking impressions Covering of the gingival margin Fixing of resin matrix during filling placement Temporary filling of cavities
Device Story
Sil-Flow is a flowable, light-cured resin material used by dental professionals for temporary restorative and sealing applications. It is applied to dental cavities, implant screw access channels, or gingival margins to provide temporary protection or structural support during dental procedures. The material is light-cured to achieve hardening. It functions as a temporary filling, sealing, or block-out agent. By providing a temporary barrier or space-filling function, it aids in clinical workflows such as impression taking or crown/bridge fabrication, ultimately protecting the tooth structure or implant site until a permanent restoration is placed.
Clinical Evidence
Bench testing only. Testing performed according to ISO 4049 protocols for visual appearance, depth of cure, Shore hardness D, sensitivity to ambient light, water sorption, water solubility, color stability, and shade stability. Biocompatibility testing conducted per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation and sensitization), and ISO 10993-11 (acute systemic toxicity). All results met established criteria.
Technological Characteristics
Flowable light-cured resin material. Composition includes Bis-GMA, Urethane Dimethacrylate, Hydroxyethyl methacrylate, Silicon dioxide, Camphorquinone, Ethyl-4 Dimethylamino benzoate, and Butylated hydroxytoluene. Physical properties tested per ISO 4049. Shelf life is 3 years. Form factor is a flowable paste.
Indications for Use
Indicated for temporary dental cavity treatments, sealing implant screw openings, relining temporary crowns/bridges, blocking out retentive areas, covering gingival margins, fixing resin matrices, and temporary cavity filling.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 23, 2025
DSI Dental Solutions Ltd
% Angela Blackwell
Senior Consultant
Blackwell Device Consulting
P.O. Box 718
Gresham, Oregon 97030
Re: K250778
Trade/Device Name: Sil-Flow
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth Shade Resin Material
Regulatory Class: Class II
Product Code: EBF
Dated: February 11, 2025
Received: March 14, 2025
Dear Angela Blackwell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250778 - Angela Blackwell
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250778 - Angela Blackwell
Page 3
Sincerely,

Bobak
Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250778 | |
| Device Name Sil-Flow | |
| Indications for Use (Describe) Sil-Flow is indicated for | |
| [A] Temporary inlay and onlay treatments of the cavity [B] Sealing of openings for implant screws [C] Relining material for temporary crowns and bridges [D] Block-out material for retentive areas in the dental arch, e.g. before taking impressions [E] Covering of the gingival margin [F] Fixing of resin matrix during filling placement [G] Temporary filling of cavities | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Sil-Flow
510(k) Summary
February 11, 2025
Name and Address: DSI Dental Solutions Ltd
59 haAvoda St
Light Industrial Zone
Ashdod Israel 7706300
Contact Person: Shlomi Krasner
Email: info@dsisrael.com
Telephone: US Phone 972-200-3265
Name of device: Sil-Flow
Classification Name: Tooth Shade Resin Material
CFR: 21 CFR 872.3690
Primary Product Code: EBF
Regulatory Class: II
Submission Contact:
Angela Blackwell
Blackwell Device Consulting
P.O. Box 718
Gresham, OR 97030-0172
(704)450-9934
angela@blackwelldevice.com
Device Description:
Sil-Flow is a flowable light-cured temporary filling material used for filling and sealing.
Indications for Use:
Sil-Flow is indicated for
- Temporary inlay and onlay treatments of the cavity
- Sealing of openings for implant screws
- Relining material for temporary crowns and bridges
- Block-out material for retentive areas in the dental arch, e.g. before taking impressions
- Covering of the gingival margin
- Fixing of resin matrix during filling placement
- Temporary filling of cavities
Testing Summary: Sil-Flow was tested for visual appearance, depth of cure, Shor hardness D, sensitivity to ambient light, water sorption, water solubility, color stability, shade stability according to protocols based on ISO 4049.
All test results met the criteria in the standard.
{5}
Shelf life for Sil-Flow is 3 years. All shelf life determinations use the same testing protocols as the characterization testing which are based on ISO 4049.
Cytotoxicity testing according to 10993-05, irritation testing and sensitization testing according to ISO 10993-10 and acute systemic toxicity testing according to ISO 10993-11 were conducted and passed.
**Predicate Device:** Clip Flow K153493 from Voco Gmbh
**Reference Devices:** Renew MDP K231696 from Prevest Denpro
## Substantial Equivalence:
The Sil-Flow has similar ingredients to the predicate and reference devices, the same indications for use, and similar physical parameter testing.
Sil-Flow from DSI
| | Sil-Flow | Clip Flow from Voco Gmbh K153493 | Renew MDP K231696 from Prevest Denpro |
| --- | --- | --- | --- |
| Product Code | EBF | EBF | KLE |
| Indications for Use | Sil-Flow is indicated for
• Temporary inlay and onlay treatments of the cavity
• Sealing of openings for implant screws
• Relining material for temporary crowns and bridges
• Block-out material for retentive areas in the dental arch, e.g. before taking impressions
• Covering of the gingival margin
• Fixing of resin matrix during filling placement
• Temporary filling of cavities | Clip Flow is indicated for
• Temporary inlay and onlay treatments of the cavity
• Sealing of openings for implant screws
• Relining material for temporary crowns and bridges
• Block-out material for retentive areas in the dental arch, e.g. before taking impressions
• Covering of the gingival margin
• Fixing of resin matrix during filling placement
• Temporary filling of cavities | Renew MDP is indicated for bonding of dual cure, light cure or self cure composite or compomer restorations to tooth structure, treatment of hypersensitive teeth, and intraoral repairs of fractured restorations. |
| Applicable Standards | ISO 4049 | ISO 4049 | |
{6}
| Mechanism of Action | Filling and sealing | Filling and sealing | bonding |
| --- | --- | --- | --- |
| Composition | -Bis-GMA-UrethaneDimethacrylate-Hydroxethylmethyl-acrylate-Silicon dioxide-Camphorquinone-Ethyl-4 Di methylamino benzoate-Butylated hydroxytoluene (BHT) | -UrethaneDimethacrylate-Hydroxethylmethyl-acrylate-Silicon dioxide | • Bis-GMA• UrethaneDimethacrylate• Triethylene GlycolDimethacrylate• Hydroxethylmethylacrylate• Ethanol• Camphorquinone• Ethyl-4 Di methylamino benzoate• Butylated hydroxytoluene (BHT)• 2-dimethyl aminoethylMethyacrylate• DM water• 10-MethacryloxydecylDihydogenPhosphate |
| Surface Hardness | 52 Shor D | 90 Shore A | |
| Water SorptionISO 4049 ≤ 40μg/mm3 | <40 μg/mm3 | <40 μg/mm3 | |
| Water SolubilityISO 4049 ≤7.5 μg/mm3 | <7.5 μg/mm3 | <7.5 μg/mm3 | |
| Depth of Cure(mm) | 3.1mm | 11mm | |
| Shelf Life | 3 years | 3 years | |
Conclusion: Sil-Flow is substantially equivalent to the predicate device. They have the same indications, similar testing, and very similar ingredients. Both the subject devices and the predicate device have physical parameters which meet the requirements of the relevant ISO standards. Shelf life testing is similar to the shelf life testing of predicate or reference device. Reference device is included to cover any ingredients not covered by the predicate device. Any differences in ingredients are minor and do not change the substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.