The Team3 fetal monitors are indicated for use by trained healthcare professionals in noninvasive and invasive monitoring of physiological parameters in pregnant women and fetuses, during the antepartum and intrapartum periods of pregnancy. The Team3 fetal monitors are intended for pregnant women with singleton or twin pregnancies from the 28th week of gestation, through to term and delivery. In cases of triplet, the Team3 fetal monitors are intended for use from 30th week of gestation through 35 weeks gestation. The devices are intended for use in clinical and hospital-type facilities.
Device Story
Perinatal monitoring system; monitors fetal heart rate (FHR), uterine activity, maternal heart rate, SpO2, and blood pressure. Inputs: ultrasound transducers (up to 3), FECG electrodes, Toco/intrauterine pressure sensors, pulse oximetry, and maternal ECG. Operation: directional pulsed Doppler for FHR; oscillometric for NIBP; differential optical transmission for SpO2. Output: real-time display on color TFT LCD; thermal array printer for CTG traces; Dawes-Redman analysis for singleton pregnancies. Used in clinical/hospital settings by trained healthcare professionals (e.g., midwives). Modification from predicate (K241368) adds third ultrasound port for triplet monitoring. Clinical decision-making supported by continuous physiological monitoring and automated CTG analysis; aids in assessing fetal well-being during pregnancy and labor.
Clinical Evidence
Prospective observational study; 8 patients (30–35+6 weeks GA); 15 FHR tracings. Evaluators (midwives) reported 100% correct display/print of FHR, 100% acceptable trace quality, and 93% acceptable audio performance. No major/minor concerns documented. Bench testing confirmed FHR detection accuracy within ±2 bpm across 30–240 bpm range.
Technological Characteristics
Ultrasound: 1MHz, 8-element piezo, directional pulsed Doppler. Materials: Novodur P2H-AT, Kraiburg TPE TF6 FMA. Connectivity: RS232, Ethernet, DVI, USB, wireless telemetry. Power: 85-264Vac or 5200mAh Li-ion battery. Safety: Type CF protection, complies with IEC 60601-1 and 60601-1-2. Software: Embedded firmware with Dawes-Redman CTG analysis.
Indications for Use
Indicated for pregnant women (singleton/twin from 28 weeks gestation; triplets from 30-35 weeks gestation) for noninvasive/invasive monitoring of fetal heart rate, uterine activity, maternal heart rate, oxygen saturation, and blood pressure during antepartum and intrapartum periods.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 19, 2025
Huntleigh Healthcare Ltd.
Steve Monks
QRE Director
35 Portmanmoor Road
Cardiff, CF24 5HN
UNITED KINGDOM
Re: K250777
Trade/Device Name: Sonicaid Team3
Regulation Number: 21 CFR 884.2740
Regulation Name: Perinatal Monitoring System and accessories
Regulatory Class: II
Product Code: HGM
Received: August 20, 2025
Dear Steve Monks:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250777 - Steve Monks
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K250777 - Steve Monks
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K250777
Device Name
Sonicaid Team3
## Indications for Use (Describe)
The Team3 fetal monitors are indicated for use by trained healthcare professionals in noninvasive and invasive monitoring of physiological parameters in pregnant women and fetuses, during the antepartum and intrapartum periods of pregnancy.
The Team3 fetal monitors are intended for pregnant women with singleton or twin pregnancies from the 28th week of gestation, through to term and delivery.
In cases of triplet, the Team3 fetal monitors are intended for use from 30th week of gestation through 35 weeks gestation. The devices are intended for use in clinical and hospital-type facilities.
Sonicaid Team3 Antepartum is suitable for use when there is a need to monitor the following physiological applications:
- Single, twin or triplet fetal heart rates by means of ultrasound
- Uterine activity - externally sensed
- Manual fetal movement detection - maternally identified using the event marker
- Automatic fetal movement detection (defaults to OFF / Not for use in twin or triplet pregnancies)
- Maternal heart rate and oxygen saturation via pulse oximetry
- Maternal non-invasive blood pressure
- The DAWES REDMAN Cardiotocography (CTG) analysis output - advises whether a number of defined criteria indicative of a normal cardiotocograph record has been met for singleton pregnancies but not for triplet pregnancies. It provides nonspecific analysis in twin pregnancies due to nonspecific fetal movement input
Sonicaid Team3 Intrapartum is suitable for use when there is a need to monitor the following physiological applications:
- Single, twin or triplet fetal heart rates by means of ultrasound and/or FECG
- Manual fetal movement detection - maternally identified using the event marker
- Maternal heart rate via ECG electrodes
- Uterine activity - externally or internally sensed
- Maternal heart rate and oxygen saturation via pulse oximetry
- Maternal non-invasive blood pressure
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PNC Publishing Services (301) 443-6740
{4}
HUNTLEIGH
A MEMBER OF THE ARJO FAMILY
K250777
Page 1 of 7
# 510(k) Summary
Sonicaid Team 3 Fetal Monitor
Name & Address: Huntleigh Healthcare Limited
Unit 35, Portmanmoor Road
Cardiff
CF24 5HN
United Kingdom
Telephone: +44 (0)2920 485885
Fax: +44 (0)2920 492520
Prepared: September 18, 2025
Contact: Steve Monks
# 1. Device Information
Device Name: Sonicaid Team3
Common Name: Perinatal Monitoring System
Regulation Name: Perinatal monitoring system and accessories
Regulation Number: 21 CFR 884.2740
Regulatory Class: II
Product Code: HGM (system, monitoring, perinatal)
Additional Product Codes:
HEL – Monitor, Heart Rate, Fetal, Ultrasonic
HGP – Electrode, Circular (spiral), Scalp and Applicator
HFM – Monitor, Uterine Contraction, External (For Use in Clinic)
KXO – Monitor, Pressure, Intrauterine
DRT – Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)
DQA – Oximeter
DXN – System, Measurement, Blood-Pressure, Non-invasive
# 2. Predicate Device
Predicate Device: Sonicaid Team3 Fetal Monitor, manufactured by Huntleigh Healthcare Ltd., cleared under K241368
The predicate device has not been subject to a design related recall.
Huntleigh Healthcare Ltd.
Page 5.3
FDA 510(k)
{5}
HUNTLEIGH
A MEMBER OF THE ARJO FAMILY
# 3. Device Description
K250777
Page 2 of 7
The Sonicaid Team3 is a fetal monitoring device designed for perinatal monitoring. The subject device provides non-invasive and invasive monitoring of physiological parameters in pregnant women and fetuses, including singleton, twin, and triplet fetal heart rate monitoring, during antepartum and intrapartum periods. The subject device includes systems and accessories intended to perform perinatal monitoring as aligned with product code HGM.
The subject device is a modification to the Sonicaid Team3 (K241368) to allow for triplet fetal heart rate monitoring. This modification affects the hardware of the device only. An additional port has been made available on the front of the monitor to allow for a third ultrasound transducer to be connected, so three fetal heart rates can be monitored at once.
All other parameters, materials, software and technology remain identical to the predicate device.
# 4. Indications for Use
The Team3 fetal monitors are indicated for use by trained healthcare professionals in noninvasive and invasive monitoring of physiological parameters in pregnant women and fetuses, during the antepartum and intrapartum periods of pregnancy.
The Team3 fetal monitors are intended for pregnant women with singleton or twin pregnancies from the 28th week of gestation, through to term and delivery.
In cases of triplet, the Team3 fetal monitors are intended for use from 30th week of gestation through 35 weeks gestation. The devices are intended for use in clinical and hospital-type facilities.
Sonicaid Team3 Antepartum is suitable for use when there is a need to monitor the following physiological applications:
- Single, twin or triplet fetal heart rates by means of ultrasound
- Uterine activity - externally sensed
- Manual fetal movement detection - maternally identified using the event marker
- Automatic fetal movement detection (defaults to OFF / Not for use in twin or triplet pregnancies)
- Maternal heart rate and oxygen saturation via pulse oximetry
- Maternal non-invasive blood pressure
- The Dawes-Redman Cardiotocography (CTG) analysis output - advises whether a number of defined criteria indicative of a normal cardiotocograph record has been met for singleton pregnancies but not for triplet pregnancies. It provides nonspecific analysis in twin pregnancies due to nonspecific fetal movement input.
Sonicaid Team3 Intrapartum is suitable for use when there is a need to monitor the following physiological applications:
Huntleigh Healthcare Ltd.
Page 5.4
FDA 510(k)
{6}
K250777
Page 3 of 7
HUNTLEIGH
A MEMBER OF THE ARJO FAMILY
- Single, twin or triplet fetal heart rates by means of ultrasound and/or FECG
- Manual fetal movement detection - maternally identified using the event marker
- Maternal heart rate via ECG electrodes
- Uterine activity - externally or internally sensed
- Maternal heart rate and oxygen saturation via pulse oximetry
- Maternal non-invasive blood pressure
# 5. Comparison of Technological Characteristics with Predicate and Device
The following table compares the subject device to the predicate and reference devices with respect to the indications for use and technological characteristics:
| Device and Predicate Device(s): | Predicate Device (K241368) | Subject Device |
| --- | --- | --- |
| Manufacturer | Huntleigh Healthcare Ltd. | Huntleigh Healthcare Ltd. |
| Classification | II | II |
| Product Code | HGM | HGM |
| Regulation | 21 CFR 884.2740 | 21 CFR 884.2740 |
| Intended Use | System and Accessories intended to perform perinatal monitoring | System and Accessories intended to perform perinatal monitoring |
| Indications for Use | The Team3 fetal monitors are indicated for use by trained healthcare professionals in non-invasive and invasive monitoring of physiological parameters in pregnant women and fetuses, during the antepartum and intrapartum periods of pregnancy. The Team3 fetal monitors are intended for pregnant women from the 28th week of gestation, through to term and delivery.
The devices are intended for use in clinical and hospital-type facilities.
Sonicaid Team3 Antepartum is suitable for use when there is a need to monitor the following physiological applications:
• Single or twin fetal heart rates by means of ultrasound
• Uterine activity - externally sensed.
• Fetal movement – maternally sensed and externally via ultrasound.
• Maternal heart rate and oxygen saturation via pulse oximetry
• Maternal non-invasive blood pressure
• CTG analysis – advises whether a | The Team3 fetal monitors are indicated for use by trained healthcare professionals in noninvasive and invasive monitoring of physiological parameters in pregnant women and fetuses, during the antepartum and intrapartum periods of pregnancy.
The Team3 fetal monitors are intended for pregnant women with singleton or twin pregnancies from the 28th week of gestation, through to term and delivery.
In cases of triplet, the Team3 fetal monitors are intended for use from 30th week of gestation through 35 weeks gestation. The devices are intended for use in clinical and hospital-type facilities.
Sonicaid Team3 Antepartum is suitable for use when there is a need to monitor the following physiological applications:
• Single, twin or triplet fetal heart rates by means of ultrasound
• Uterine activity - externally sensed
• Manual fetal movement detection - maternally identified using the event marker
• Automatic fetal movement |
Huntleigh Healthcare Ltd.
Page 5.4
FDA 510(k)
{7}
K250777
Page 4 of 7
HUNTLEIGH
A MEMBER OF THE ARJO FAMILY
| | number of defined criteria indicative of a normal cardiotocograph has been met.
Sonicaid Team3 Intrapartum is suitable for use when there is a need to monitor the following physiological applications:
• Single or twin fetal heart rates by means of ultrasound and/or FECG.
• Maternal heart rate via ECG electrodes
• Uterine activity – externally or internally sensed
• Fetal movement – maternally sensed and externally via ultrasound.
• Maternal heart rate and oxygen saturation via pulse oximetry
• Maternal non-invasive blood pressure | detection (defaults to OFF / Not for use in twin or triplet pregnancies)
• Maternal heart rate and oxygen saturation via pulse oximetry
• Maternal non-invasive blood pressure
• The Dawes-Redman Cardiotocography (CTG) analysis output - advises whether a number of defined criteria indicative of a normal cardiotocograph record has been met for singleton pregnancies but not for triplet pregnancies. It provides nonspecific analysis in twin pregnancies due to nonspecific fetal movement input
Sonicaid Team3 Intrapartum is suitable for use when there is a need to monitor the following physiological applications:
• Single, twin or triplet fetal heart rates by means of ultrasound and/or FECG
• Manual fetal movement detection - maternally identified using the event marker
• Maternal heart rate via ECG electrodes
• Uterine activity - externally or internally sensed
• Maternal heart rate and oxygen saturation via pulse oximetry
• Maternal non-invasive blood pressure |
| --- | --- | --- |
| Ultrasound | Channels: 2
Mode: Directional Pulsed Doppler
FHR Range: 30-240 bpm
Frequency 1MHz
Safety: Type CF Protection
Type: Piezo 8 element
I_{spta}: <3mW/cm²
Material: Novodur P2H-AT | Channels: 3
Mode: Directional Pulsed Doppler
FHR Range: 30-240 bpm
Frequency 1MHz
Safety: Type CF Protection
Type: Piezo 8 element
I_{spta}: <3mW/cm²
Material: Novodur P2H-AT |
| Toco | Type: Tocodynamometer
Material: Body – Novodur P2H-AT, Faceplate – Kraiburg TPE TF6 FMA Elastomer | Type: Tocodynamometer
Material: Body – Novodur P2H-AT, Faceplate – Kraiburg TPE TF6 FMA Elastomer |
| Fetal ECG | Channels: 1 | Channels: 1 |
Huntleigh Healthcare Ltd.
Page 5.4
FDA 510(k)
{8}
K250777
Page 5 of 7
HUNTLEIGH
A MEMBER OF THE ARJO FAMILY
| | FHR Range: 30-240 bpm | FHR Range: 30-240 bpm |
| --- | --- | --- |
| **External Uterine Activity** | Channels: 1
Range: 0-100 Relative Units
Sensitivity: 100%FSD=120g
Offset range: ± 375g
Alerts: Toco Persistence Alert | Channels: 1
Range: 0-100 Relative Units
Sensitivity: 100%FSD=120g
Offset range: ± 375g
Alerts: Toco Persistence Alert |
| **Internal Uterine Activity** | Channels: 1
Range: 0-100mmHg or 1-13.3kPa
Sensitivity: 5μV/V/mmHg | Channels: 1
Range: 0-100mmHg or 1-13.3kPa
Sensitivity: 5μV/V/mmHg |
| **Maternal eMHR** | Reference: eMHR
Channels: 1
Electrode Type: Standard ECG
MHR Range: 30-240 bpm | Reference: eMHR
Channels: 1
Electrode Type: Standard ECG
MHR Range: 30-240 bpm |
| **Maternal Non-Invasive Blood Pressure** | Method: Oscillometric
OEM: PAR Medizintechnik
BP Ranges: Systolic 25-280mmHg,
Diastolic 10-220mmHg
Pulse Range: 30-240bpm
BP Accuracy: ±1.7mmHg
Safety Limiters: Over-inflation
300mmHg, Duration 15s | Method: Oscillometric
OEM: PAR Medizintechnik
BP Ranges: Systolic 25-280mmHg,
Diastolic 10-220mmHg
Pulse Range: 30-240bpm
BP Accuracy: ±1.7mmHg
Safety Limiters: Over-inflation 300mmHg,
Duration 15s |
| **Maternal Oximetry (MSpO2)** | Method: Differential optical transmission
Saturation Range: 1-100% MSpO₂
Accuracy: 70-100% ±1SD
Pulse Range: 30-240 bpm
Accuracy: ±3 bpm | Method: Differential optical transmission
Saturation Range: 1-100% MSpO₂
Accuracy: 70-100% ±1SD
Pulse Range: 30-240 bpm
Accuracy: ±3 bpm |
| **Dawes Redman Software Analysis** | No. of criteria: ‘Met or ‘Not Met’
Measurement Parameters: Signal Loss, Fetal Movements per hour,
Basal Heart Rate, Accelerations, Decelerations, High and Low
variation episodes, Short term variation
Firmware embedded in the Team3 (fetal monitor | No. of criteria: ‘Met or ‘Not Met’
Measurement Parameters: Signal Loss, Fetal Movements per hour,
Basal Heart Rate, Accelerations, Decelerations, High and Low
variation episodes, Short term variation
Firmware embedded in the Team3 (fetal monitor |
| **Patient Event Marker** | Channels: 1
Indication: Audio, visual & printed | Channels: 1
Indication: Audio, visual & printed |
| **Display** | Type: Colour TFT LCD
Size: 17 x 12.8cm (6.7 x 5”) | Type: Colour TFT LCD
Size: 17 x 12.8cm (6.7 x 5”) |
| **Printer** | Print Head: 128mm Thick Film Thermal Array
Speeds: 1, 2 or 3cm/min, 20cm/min fast forward | Print Head: 128mm Thick Film Thermal Array
Speeds: 1, 2 or 3cm/min, 20cm/min fast forward |
| **Controls and Indicators** | Power on/off: On/standby touch switch with indicator
Audio volume up/down: Volume up/down touch screen ‘button’ | Power on/off: On/standby touch switch with indicator
Audio volume up/down: Volume up/down touch screen ‘button’ |
| **Interfaces** | RS232-CRS: Yes
Ethernet – CRS: Yes
DVI: 1 channel, DVI-I connector
Wireless Telemetry: Yes
USB: 2 channels | RS232-CRS: Yes
Ethernet – CRS: Yes
DVI: 1 channel, DVI-I connector
Wireless Telemetry: Yes
USB: 2 channels |
Huntleigh Healthcare Ltd.
Page 5.4
FDA 510(k)
{9}
K250777
Page 6 of 7
HUNTLEIGH
A MEMBER OF THE ARJO FAMILY
| Physical | Size(WxDxH): 32x23x23.4cm
Weight (Net): 6kg | Size(WxDxH): 32x23x23.4cm
Weight (Net): 6kg |
| --- | --- | --- |
| Mains Power Supply | Voltage: 85-264Vac
Frequency: 50 or 60 Hz
Consumption (max): <140VA | Voltage: 85-264Vac
Frequency: 50 or 60 Hz
Consumption (max): <140VA |
| Internal Battery | Rechargeable 5200mAh Li-ion Battery | Rechargeable 5200mAh Li-ion Battery |
| Environmental | Operating temp: +10-40°C
Storage temp: -20-50°C
Operating Press: 70-106kPa
Storage Press: 70-106kPa
Operating Humidity: 15-90% RH
Storage Humidity: 10-90% RH | Operating temp: +10-40°C
Storage temp: -20-50°C
Operating Press: 70-106kPa
Storage Press: 70-106kPa
Operating Humidity: 15-90% RH
Storage Humidity: 10-90% RH |
| EMC and Electrical Safety | 60601-1-2: Comply
60601-1: Comply | 60601-1-2: Comply
60601-1: Comply |
The Sonicaid Team3 subject device has the same intended use as the predicate device - a system and accessories intended to perform perinatal monitoring. As noted in the table above, the subject device has identical technological characteristics as the predicate device, except for the inclusion of a third ultrasound transducer port, to allow for triplet fetal heart rate monitoring. This different technological characteristic does not raise different questions of safety and effectiveness and can be addressed through performance testing.
## 6. Performance Data
| Testing Conducted | Discussion |
| --- | --- |
| Functional Performance Bench Testing | Performance bench testing was performed on the Sonicaid Team3 device with the incorporation of the new technological characteristic – triplet fetal heart rate monitoring. Ultrasound performance bench testing for triplet heart rate detection was included. Testing demonstrated accurate and simultaneous detection of three independent fetal heart rates across the full operating range (30–240 bpm), with accuracy within ±2 bpm, meeting the predefined acceptance criteria. No concerns were noted. The performance verification, along with the risk analysis confirmed there are no new safety, effectiveness or performance concerns. |
| Clinical Validation | Clinical validation was assessed via a prospective observational study and carried out by trained healthcare professionals, such as Midwives, in a variety of UK hospital settings and feedback was given. The validation testing was to confirm the device meets the user requirements and functions as intended, with the focus |
Huntleigh Healthcare Ltd.
Page 5.4
FDA 510(k)
{10}
K250777
Page 7 of 7
HUNTLEIGH
A MEMBER OF THE ARJO FAMILY
| | of the validation to confirm triplet functionality. In this study of 15 FHR tracings from 8 patients (30–35+6 weeks GA), 100% of evaluators reported correct display and print of fetal heart rates, 100% rated trace quality acceptable, and 93% rated audio performance acceptable, with no major or minor concerns documented. It was concluded that the Sonicaid Team3 meets the intended use and is fit for purpose in regards to the additional feature of monitoring triplets. |
| --- | --- |
| EMC | Further EMC testing was conduction on the product with triplet fetal heart rate monitoring functionality option, by a third party laboratory. Testing confirms device complies with EN 60601-1-2:2020 |
# 7. Conclusion
The results of the performance testing described above demonstrate that the Sonicaid Team3 is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Huntleigh Healthcare Ltd.
Page 5.4
FDA 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.