Expert clinical evaluation of CBCT images to evaluate the clinical acceptability of the default panoramic image
>1 (clinical experts)
Indications for Use
DS Core Diagnosis is a software product for use in dentistry for viewing and interacting with digital or digitized patient media as well as supporting the diagnosis of digital or digitized patient media. Intended users will be able to view and analyze different media types in one viewer. The Panoramic Curve Proposal feature is intended for patients aged 12 years and older with permanent dentition.
Device Story
Cloud-based dental imaging software; provides access to 2D/3D dental media (CBCT, bitewings, periapicals, panoramic) via internet connection. Used by clinicians to view/analyze images; includes standard tools (annotation, flip, rotate, contrast, brightness, magnification). Features AI-enabled Panoramic Curve Proposal algorithm (CNN-based) to suggest panoramic curve along dental arch in CBCT scans. Clinician reviews and manually adjusts proposed curve. Does not control life-sustaining devices or perform automated diagnosis; assists clinician interpretation. Deployed via DS Core Platform; supports clinical decision-making by providing optimized image visualization.
Clinical Evidence
Bench testing and expert clinical evaluation. Panoramic Curve Proposal Success Rate (PCPSR) measured on Dentsply Sirona and third-party CBCT scans. Dentsply Sirona scans: 98.81% (95% CI lower bound 94.57%). Combined dataset: 99% (95% CI lower bound 95.42%). Expert clinical evaluation of default panoramic images (without user modification) showed 92.5% pass rate for clinical acceptability. System validation and usability testing successfully passed.
Technological Characteristics
Cloud-based software as a medical device (SaMD). Connectivity: active internet connection required. Architecture: Convolutional Neural Network (CNN) for panoramic curve proposal. Features: 2D/3D image viewing, annotation, image manipulation tools. Platform: DS Core Platform.
Indications for Use
Indicated for dental professionals to view, interact with, and support diagnosis of digital/digitized patient media (2D/3D images, photos, documentation). Panoramic Curve Proposal feature indicated for patients aged 12+ with permanent dentition.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 12, 2025
Dentsply Sirona Inc.
% Deepthi Paknikar
Senior Manager, Regulatory Affairs
221 West Philadelphia St.
Suite 60W
YORK, PA 17401
Re: K250755
Trade/Device Name: DS Core Diagnosis
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: July 9, 2025
Received: July 9, 2025
Dear Deepthi Paknikar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250755 - Deepthi Paknikar
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250755 - Deepthi Paknikar
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250755 | |
| Device Name DS Core Diagnosis | |
| Indications for Use (Describe) DS Core Diagnosis is a software product for use in dentistry for viewing and interacting with digital or digitized patient media as well as supporting the diagnosis of digital and digitized patient media. Intended users will be able to view and analyze different media types in one viewer. The Panoramic Curve Proposal feature is intended for patients aged 12 years and older with permanent dentition. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250755_510k Summary
# 510k Summary for DS Core Diagnosis-K250755
Prepared: 8/11/2025
## Contact Details
- Applicant Name: Dentsply Sirona Inc.
- Applicant Address: 221 West Philadelphia St. Suite 60W York PA 17401 United States
- Applicant Contact Telephone: 630-201-1612
- Applicant Contact: Dr. Deepthi Paknikar
- Applicant Contact Email: deepthi.paknikar@dentsplysirona.com
## Device Name
- Device Trade Name: DS Core Diagnosis
- Common Name: System, Image Processing, Radiological
- Classification Name: Medical image management and processing system
- Regulation Number: 892.2050
- Product Code(s): QIH
## Legally Marketed Predicate Devices
- Predicate K#: K132773
- Predicate Trade Name: Sidexis 4
- Product Code: LLZ
## Device Description Summary
DS Core Diagnosis is a cloud-based dental imaging software that provides access to dental media and images via an active internet connection. It is a medical device that can only be used in combination with the DS Core Platform. Classified as a Class II dental imaging software, DS Core Diagnosis allows users to view a variety of media types, including 2D and 3D images (e.g., bitewings, periapicals, panoramic, CBCT, etc.), as well as photos and documentation.
The software includes standard image viewing tools such as annotation, flip, rotate, contrast adjustment, brightness, and magnification.
DS Core Diagnosis also features a Machine Learning enabled panoramic curve proposal algorithm for a proposed panoramic curve of a CBCT scan along the dental arch, which can be adjusted and reviewed by the user. The architecture is a Convolutional Neural Network (CNN) model.
DS Core Diagnosis does not interact directly with patients or control any life-sustaining devices. The software does not perform diagnoses; instead, it provides tools to assist qualified clinicians in interpreting the displayed images and making informed decisions.
## Intended Use/Indications for Use
DS Core Diagnosis is a software product for use in dentistry for viewing and interacting with digital or digitized patient media as well as supporting the diagnosis of digital or digitized patient media. Intended users will be able to view and analyze different media types in one viewer.
The Panoramic Curve Proposal feature is intended for patients aged 12 years and older with permanent dentition.
Page 1 of 3
{5}
K250755_510k Summary
## Indications for Use Comparison
Both the subject and predicate devices share the same essential indications for use. While there are minor distinctions—such as the subject device's absence of image acquisition and its cloud-based functionality—these differences do not affect the safety and effectiveness of the device. Both devices ultimately serve the same core purpose: to provide a platform for image review and diagnosis by qualified professionals.
## Technological Comparison
Both devices are software as a medical device and allow users to view a variety of media types, including 2D and 3D images. The subject device is a cloud-based software as a medical device, whereas the predicate device is a downloaded desktop-based software as a medical device. The subject device does not include image acquisition, whereas the predicate device does. Both devices offer the same fundamental user-defined settings for image viewing, such as brightness, contrast, zoom, gamma, moving content in the media window, and image navigation. Minor differences in the viewer exist, such as the organization of the viewing windows or the appearance of feature (brightness, contrast, etc.) icons. Both devices include a panoramic curve proposal feature that can be manually reviewed and adjusted by the user. The panoramic curve proposal feature is AI/ML enabled in the subject device and is not AI/ML enabled for the predicate device.
## Non-Clinical Tests Summary & Conclusions
Nonclinical performance testing results were reviewed as part of this submission. Software verification and validation testing was also completed.
For model-specific performance evaluation, the panoramic curve proposal success rate (PCPSR) was measured. The PCPSR represents the percentage of cases where the proposed panoramic curve projection layer includes all teeth in the scan. The evaluation results showed that the panoramic curve proposal performance met the prespecified acceptance criteria.
Dentsply Sirona Scans: PCPSR of 98.81%, with the lower bound of the 95% CI at 94.57%.
All Scans (Dentsply Sirona & Third Party): PCPSR of 99%, with 99 of 100 scans considered "successful" and the lower bound of the 95% CI at 95.42%. Various subgroups were analyzed, yielding a PCPSR range of 97.78-100%, all passing the acceptance criteria.
An expert clinical evaluation was also conducted that passed all predefined acceptance criteria. In this evaluation, clinical experts assessed a dataset made up of CBCT images, to evaluate the reconstructed CBCT panorama to determine the clinical acceptability of the default panoramic image (without user modification) with the panoramic curve proposal generated by the Panoramic Curve Proposal (PCP) algorithm. The results demonstrated that the predefined primary endpoint acceptance criteria were met, which was to determine the acceptability (pass/fail) of the CBCT reconstructed panoramic images based on the panoramic curve proposal (without user modification) to begin working per the standard of care. The results reported a 92.5% pass rate overall for the primary endpoint of this evaluation, demonstrating that the expert clinical evaluation was successful for the panoramic curve proposal algorithm.
Page 2 of 3
{6}
K250755_510k Summary
System validation and usability testing for the DS Core Diagnosis device were conducted with clinicians, including the panoramic curve proposal feature in the scope. The results demonstrated that all tests were successfully passed.
In summary, the panoramic curve proposal feature performance met the prespecified acceptance criteria, with a very high success rate and highly generalizable performance across different subgroups.
Regardless of this strong performance, the users of the device retain full autonomy and can adjust and review the proposed panoramic curve.
The results of nonclinical performance testing support the determination of substantial equivalence.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.