K250747 · Kardium, Inc. · DQK · Jun 25, 2025 · Cardiovascular
Device Facts
Record ID
K250747
Device Name
Globe® Pulsed Field System
Applicant
Kardium, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Jun 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The Globe PF System is indicated for catheter-based cardiac anatomical and electrophysiological mapping and stimulation of cardiac tissue.
Device Story
Globe PF System performs cardiac anatomical and electrophysiological mapping and stimulation. System inputs include intracardiac electrogram signals from mapping catheter electrodes and body surface ECG signals. Globe Controller acquires/processes signals and generates mapping energy/stimulation pulses. Globe Workstation (PC-based) runs proprietary software to visualize 3D geometric shells, voltage, impedance, contact, activation, and propagation maps. Clinicians use the workstation to assess catheter-to-atrial wall contact and guide diagnostic pacing. System utilizes impedance-based tracking via surface GPS electrodes and cables. Output assists physicians in identifying cardiac electrical activity and guiding diagnostic interventions. Benefits include enhanced visualization of cardiac anatomy and electrical properties to support electrophysiological procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing, biocompatibility testing, summative usability testing, electrical safety and EMC testing, software verification and validation, cybersecurity testing, and packaging validation. Testing followed FDA-recognized consensus standards and guidance documents, demonstrating performance criteria are met.
Technological Characteristics
System includes Globe Controller, Globe Workstation, and GPS electrodes/cables. Employs impedance-based tracking for navigation. Supports 3D electroanatomical mapping (voltage, impedance, contact, activation, propagation). Connectivity via workstation interface. Software-based signal processing and mapping. Single-use electrodes. Thermal sensing for contact/flow maps based on principles used in Swan-Ganz catheters.
Indications for Use
Indicated for catheter-based cardiac anatomical and electrophysiological mapping and stimulation of cardiac tissue in patients requiring such procedures.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 25, 2025
Kardium Inc.
Ricardo Romero
Vice President, RA&QA
8518 Glenlyon Parkway, Unit 155
Burnaby, BC V5J OB6
Canada
Re: K250747
Trade/Device Name: Globe® Pulsed Field System
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable diagnostic computer
Regulatory Class: Class II
Product Code: DQK
Dated: March 11, 2025
Received: March 12, 2025
Dear Ricardo Romero:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250747 - Ricardo Romero
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250747 - Ricardo Romero
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250747
Device Name
Globe® Pulsed Field System
Indications for Use (Describe)
The Globe PF System is indicated for catheter-based cardiac anatomical and electrophysiological mapping and stimulation of cardiac tissue.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Kardium
# 510(k) Summary
per 21 CFR §807.92
| Date Summary Prepared | June 24th, 2025 |
| --- | --- |
| Applicant | Kardium Inc.
155-8518 Glenlyon Parkway
Burnaby, British Columbia V5J 0B6
Canada |
| Official Correspondent | Ricardo Romero
Vice President, RA&QA
Kardium Inc.
Tel: +1 604.248.8891 |
| 510(k) Number | K250747 |
| Device Trade Name | Globe® Pulsed Field System |
| Common Name | Globe Pulsed Field System |
| Classification Name | Computer, Diagnostic, Programmable |
| Classification | Class II, 21 CFR Part 870.1425, Cardiovascular |
| Product Code | DQK |
| Predicate Device | Affera Integrated Mapping System; Impedance Localization Patch Kit (K241828) |
| Reference Device | Swan-Ganz Catheter (K160084) |
| Device Description | The Globe Pulsed Field System (Globe PF System) comprises the following components and accessories to support anatomical and electrophysiological mapping, and pacing stimulation of cardiac tissue:
• Globe Controller: Used for the acquisition and processing of signals for cardiac anatomical and electrophysiological mapping, generation of mapping energy and stimulation pulses.
• Globe Workstation: A PC workstation configured with the Globe Software, which the clinician uses to assess contact between the mapping catheter electrodes and the atrial wall, map the atrial electrical activity, and apply stimulation pulses for diagnostic purposes.
• Globe Positioning System (GPS™) Electrodes and GPS Cable: Surface electrodes and cables for localization of the mapping catheter. |
Kardium Inc. Globe® Pulsed Field System - 510(k) Summary
Page 1 of 5
{5}
Kardium
# Extended Use
Catheter-based electrophysiological mapping and stimulation.
# Indications for Use
The Globe PF System is indicated for catheter-based cardiac anatomical and electrophysiological mapping and stimulation of cardiac tissue.
# Comparison of Technological Characteristics
A comparative overview of the subject and predicate device is provided in the following table:
| Attribute | Affera Mapping System (K241828) | Subject Device: Globe® PF System (K250747) |
| --- | --- | --- |
| Mapping System | | |
| Intended Use | Catheter-based electrophysiological mapping and stimulation. | Same |
| Fundamental Scientific Technology (Navigation/ localization) | • Impedance-based tracking
• Magnetic-based tracking | • Impedance-based tracking |
| Key System Components | • Catheter/equipment interface unit (CIU)
• Workstation and software
• Magnetic field generator | • Globe Controller
• Globe Workstation and software |
| Electroanatomical Maps | 3D geometric shell, voltage, impedance, activation, and propagation maps. | 3D geometric shell, voltage, impedance, contact, activation, and propagation maps. |
| Stimulation/ Pacing | Internally (20mA max) and externally generated pacing. | Internally (10mA max). |
| ECG Signal Acquisition | Body surface ECG signals recorded directly by the Affera Integrated Mapping System. | Same |
| EGM Acquisition | Intracardiac electrogram signals measured by the catheter’s electrodes. | Same |
| Globe Positioning System (GPS) Electrodes and Cable | | |
| Configuration | 6 surface electrodes/patches. | • 6 surface electrodes/patches (GPS Electrodes)
• Adapter cable (GPS Cable) |
Kardium Inc. Globe® Pulsed Field System - 510(k) Summary
{6}
Kardium
| Attribute | Affera Mapping System (K241828) | Subject Device: Globe® PF System (K250747) |
| --- | --- | --- |
| Supported Tracking and Navigation Modalities | • Impedance-based tracking
• Magnetic-based tracking | • Impedance-based tracking |
| Electrode Sterility | Non-sterile | Same |
| Electrode Use Limitations | Single-use | Single-use |
The Globe PF System shares the following identical or similar intended use and technological characteristics with the predicate device:
- Same intended use.
- Equivalent indications for use. Both systems are indicated for catheter-based cardiac electrophysiological mapping and stimulation (pacing). The additional details on the technological characteristics provided by the predicate system are explanatory and do not affect the equivalence.
- Both devices include impedance-based tracking of compatible catheters.
- Both devices utilize a central interface unit for connection with a workstation, and compatible catheters.
- Both devices include proprietary software designed to present the user with information regarding the location of the mapping catheter and electroanatomical maps (3D geometric shell, voltage, impedance, activation, and propagation maps).
- Both devices have the capability to directly acquire ECG and electrogram signals.
- Both devices have the capability to deliver stimulation.
- The GPS electrodes and cable are equivalent to the predicate's localization kit. The GPS cable is an electrical extension of the surface electrodes that does not introduce new technological characteristics.
The differences in technological characteristics are as follows:
- The Globe PF System is intended for impedance-based tracking of catheters, whereas the predicate device includes impedance-based and magnetic-based tracking.
Kardium Inc. Globe® Pulsed Field System · 510(k) Summary
{7}
Kardium
- The Globe PF System can deliver stimulation from an internal source, whereas the predicate device can deliver stimulation for an internal and external source.
- The Globe PF system includes additional electroanatomical maps; contact maps (FLOW™ and CONTACT™). The delivery of thermal energy to tissue and sensing of the resulting thermal response used for the Globe PF System’s CONTACT and FLOW maps are based on the same principles of operation as the reference device (Swan-Ganz catheter, K160084). Since the reference device is used in the same anatomical location, and the energy delivery and sensing methods have the same principles of operation, the scientific methods used to evaluate the safety and effectiveness of the CONTACT and FLOW mapping functionality of the Globe PF System are adequate.
These differences do not raise different questions of safety and effectiveness and do not preclude a meaningful comparison with the predicate devices.
| Safety and Performance Data | The following performance testing was conducted to support the Globe PF System: • Bench testing • Biocompatibility testing • Summative usability testing • Electrical safety and EMC testing • Software verification and validation testing • Cybersecurity testing • Packaging validation Testing was performed following the applicable principles and recommendations of relevant FDA-recognized consensus standards and guidance documents. The test results demonstrate that the Globe PF System meets the performance criteria for its intended use and does not raise new questions on safety or effectiveness compared to the predicate device. |
| --- | --- |
| Statement of Equivalence | The Globe PF System and predicate device have identical or equivalent intended use, indications for use, and performance. Differences in technological characteristics do not raise any new or different questions of safety and effectiveness. Based on this and the results of performance testing, the Globe PF System is substantial equivalent to the predicate device. |
Disclaimer: All information provided herein, including comparisons of the Globe PF System with other products, is provided for the purpose of a 510(k) premarket notification and has no bearing on and has not been provided in view of any national or international patent laws, including Title 35 of the United States Code, for which different analyses apply.
Kardium Inc. Globe® Pulsed Field System · 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.