← Product Code [NHA](/productcode/NHA) · K250721

# LOCATOR® Angled Abutment (K250721)

_Zest Anchors, LLC · NHA · Aug 20, 2025 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250721

## Device Facts

- **Applicant:** Zest Anchors, LLC
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Aug 20, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.

## Device Story

LOCATOR Angled Abutment is a titanium (Ti-6Al-4V ELI) dental implant abutment with a titanium nitride (TiN) coating. It features a 15-degree angle for anatomical correction. The device is used by clinicians to attach fixed or removable dental restorations to endosseous implants in the mandible or maxilla. It functions as an interface between the implant and the prosthetic restoration, utilizing retention inserts and denture attachment housings. The device is provided for end-user sterilization. It is used in a clinical dental setting to restore masticatory function for patients with full or partial edentulism. The output is a stable, angled connection point for dental prosthetics, which allows for the correction of implant angulation, improving the fit and function of the final restoration.

## Clinical Evidence

No clinical data. Bench testing only. Fatigue testing performed per ISO 14801:2016. Biocompatibility assessed per ISO 10993-1 and ISO 10993-5. MR compatibility testing conducted per ASTM F2052-21, F2213-17, F2182-19, F2119-07. TiN coating performance verified per ASTM F1044 and F1147.

## Technological Characteristics

Material: Ti-6Al-4V ELI (ASTM F136). Coating: Titanium Nitride (TiN). Design: 15-degree angled abutment. Connection types: Internal Hex, Conical, Grand Morse Taper. Sterilization: Moist heat (end-user). Connectivity: None (mechanical component).

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- LOCATOR Angled Abutment ([K243272](/device/K243272.md))

## Reference Devices

- High Retention Attachment System ([K213391](/device/K213391.md))
- LOCATOR Angled Abutment ([K233587](/device/K233587.md))
- Screw Vent Implant; Tapered Screw Vent Implant ([K013227](/device/K013227.md))
- Tapered Screw-Vent Implant ([K072589](/device/K072589.md))
- NobelActive Internal Connection Implant ([K071370](/device/K071370.md))
- Legacy Abutment System ([K060063](/device/K060063.md))
- Interactive/ SwishPlus2 Implant System ([K130572](/device/K130572.md))
- 2014 Interactive/SwishActive system ([K143011](/device/K143011.md))
- Tapered Internal Implant System ([K071638](/device/K071638.md))
- BioHorizons Simple Solutions with Laser-Lok ([K100985](/device/K100985.md))
- BioHorizons Abutments for Zimmer ([K103691](/device/K103691.md))
- BioHorizons CAD/CAM Abutments ([K151621](/device/K151621.md))
- Implant-One System ([K173701](/device/K173701.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

August 20, 2025

Zest Anchors, LLC
Maleata Hall
Director Regulatory Affairs
2875 Loker Ave E
Carlsbad, California 92010

Re: K250721
Trade/Device Name: LOCATOR® Angled Abutment
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: July 21, 2025
Received: July 22, 2025

Dear Maleata Hall:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250721 - Maleata Hall
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K250721 - Maleata Hall
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)

K250721

Device Name

LOCATOR® Angled Abutment

Indications for Use (Describe)

The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable restorations retained by endosseous implants to restore masticatory function for the patient.

IMPLANT COMPATIBILITY

| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15°, Straumann, BLX | Straumann | 3.5, 3.75, 4.0, 4.5 | BLX | Regular Base | 2.9 | Bone Level |
| 5.0, 5.5, 6.5 | BLX | Wide Base | 2.9 | Bone Level |
| LAA 15°, Nobel (NP) | Nobel | 3.5 | NobelActive, NobelParallel CC, NobelReplace CC | Narrow Platform | 3.0 | Conical |
| Implant Direct | 3.2, 3.7 | InterActive | N/A | 3.0 | Conical |
| 3.2, 3.7, 4.2, 4.7 | Simply Iconic | N/A | 3.0 | Conical |
| LAA 15°, Nobel (RP) | Nobel | 4.3 | NobelActive, NobelParallel CC, NobelReplace CC | Regular Platform | 3.5 | Conical |
| Implant Direct | 4.3, 5.0 | InterActive | N/A | 3.4 | Conical |
| 4.7, 5.2, 5.7 | Simply Iconic | N/A | 3.4 | Conical |
| LAA 15°, Neodent (GM) | Neodent | 3.5, 3.75, 4.0, 5.0, 6.0 | Helix GM | N/A | 3.0 | Conical / Grand Morse |
| 3.5, 4.3, 5.0 | Drive GM | N/A | Conical / Grand Morse |
| 3.5, 3.75, 4.0, 5.0 | Titamax GM | N/A | Conical / Grand Morse |
| LAA 15°, TSV | ZimVie | 3.7 & 4.1 Green | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 3.5 | Internal Hex |
| 4.7 Purple | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 4.5 | Internal Hex |
| Implant Direct | 3.7, 4.2 | Legacy 1, 2, 3, 4 | N/A | 3.5 | Internal Hex |
| 4.7, 5.2 | Legacy 2, 3, 4 | N/A | 4.5 |
| 4.7 | Legacy 1 | N/A | 4.5 |
| 4.2, 4.6 | Tapered Pro | Yellow | 3.5 | Internal Hex |
| BioHorizons | 4.6 | Tapered Plus |
| 3.8 | Tapered Internal |
| 3.0, 3.8 | Tapered Tissue Level |
| 4.6 | Tapered Short |
| 4.2 | Tapered PTG |
| 5.2 | Tapered Pro | Green | 4.5 |
| 5.8 | Tapered Plus |
| 4.6 | Tapered Internal |
| 4.6 | Tapered Tissue Level |
| 5.8 | Tapered Short |
| LAA 15°, Implant Logistics, Implant-One 300 Series | Implant Logistics | 3.5, 4.1, 4.5 | Implant One | 300 Series | 2.75 | Internal Conical 6" Morse Taper |
| LAA 15°, Implant Logistics, Implant-One 400 Series | 4.0, 4.5, 5.5 | Implant One | 400 Series | 3.25 | Internal Conical 6" Morse Taper |

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

PISC Publishing Services (391) 443-6740

{4}

ZD ZEST DENTAL SOLUTIONS
K250721

# 510(K) Summary – K250721
# LOCATOR Angled Abutment

## i. General Information on Submitter

Applicant: Zest Anchors, LLC
Address: 2875 Loker Avenue East
Carlsbad, CA 92010 USA
Telephone: (800) 262-2310
Contact Person: David Lin
Contact Title: Sr. Regulatory Affairs Specialist
Email: regulatoryaffairs@zestdent.com
Date Prepared: August 15, 2025

## ii. General Information on Device

Proprietary Name: LOCATOR Angled Abutment
Common Name: Dental Implant Abutment
Regulation Number: 21 CFR 872.3630
Classification Name: Endosseous dental implant abutment
Regulatory Class: Class II
Product Code: NHA (Abutment, Implant, Dental, Endosseous)

## iii. Predicate Device

|  Predicate Device | 510(k) Number  |
| --- | --- |
|  LOCATOR Angled Abutment: Zest Anchors, LLC | K243272  |
|  Reference Devices | 510(k) Number  |
|  High Retention Attachment System | K213391  |
|  LOCATOR Angled Abutment (Various) | K233587  |
|  Screw Vent Implant; Tapered Screw Vent Implant, Sulzer Dental, Inc. | K013227  |
|  Tapered Screw-Vent Implant, 4.1mm, Zimmer Dental, Inc. | K072589  |
|  NobelActive Internal Connection Implant, Nobel Biocare AB | K071370  |
|  Legacy Abutment System, Implant Direct LLC. | K060063  |
|  Interactive/ SwishPlus2 Implant System, Implant Direct LLC. | K130572  |
|  2014 Interactive/SwishActive system | K143011  |
|  Tapered Internal Implant System, BioHorizons Implant Systems, Inc | K071638  |
|  BioHorizons Simple Solutions with Laser-Lok® | K100985  |
|  BioHorizons Abutments for Zimmer | K103691  |
|  BioHorizons CAD/CAM Abutments | K151621  |
|  Implant-One System, Implant Logistics, Inc. | K173701  |

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{5}

ZD ZEST DENTAL SOLUTIONS

K250721

## iv. Description of Device

The purpose of this submission is to expand the Indications for Use of the LOCATOR® Angled Abutment product line (K243272 &amp; K233587) by adding compatibility of existing abutments with various new dental implant systems from Implant Direct and Biohorizons. Additionally, the submission expands the Indications for Use of the product line with a modified version of the predicate device shown to be compatible with the Implant Logistics Implant-One Series 300 and Series 400 Implant Systems. The LOCATOR Angled Abutment is designed and intended for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants in the mandible or maxilla, as cleared to be used with LOCATOR FIXED (K213391) and LOCATOR Attachment Systems (K072878).

The expanded Indications for Use of the LOCATOR Angled Abutment demonstrates that:

1. The cleared devices per K243272 are additionally compatible with the following implant systems from Implant Direct and Biohorizons.

a. The Implant Direct Legacy and Biohorizons Tapered Pro Internal Hex implant systems are compatible with the LOCATOR Angled Abutment Zimmer (ZimVie) TSV 3.5/4.5.
b. The Implant Direct Simply Iconic and Interactive implant systems are compatible with the LOCATOR Angled Abutment Nobel NP/RP.

2. The LOCATOR® Angled Abutment, Implant Logistics 300 Series, a modified version of the predicate device, is compatible with the Implant Logistics Implant-One System Series 300 implant system.
3. The LOCATOR® Angled Abutment, Implant Logistics 400 Series, a modified version of the predicate device, is compatible with the Implant Logistics Implant-One System Series 400 implant system.

The LOCATOR Angled Abutments are manufactured from titanium (Ti-6Al-4V) and are titanium nitride (TiN) coated in various abutment heights, identical to the predicate device. The LOCATOR Angled Abutment interfacing features are provided at a 15 degree angle to allow for angle correction, substantially equivalent to the predicate device of K233587 and K243272. The abutments will be used with the accessories of the LOCATOR Implant Attachment System (retention inserts, denture attachment housing, and ancillary processing parts) and LOCATOR FIXED Attachment System (fixed inserts, denture attachment housing) for the attachment of a restoration, identical to the predicate device.

## Indication for Use

The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{6}

ZD

ZEST DENTAL

SOLUTIONS

K250721

IMPLANT COMPATIBILITY

| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15°, Straumann, BLX | Straumann | 3.5, 3.75, 4.0, 4.5 | BLX | Regular Base | 2.9 | Bone Level |
| 5.0, 5.5, 6.5 | BLX | Wide Base | 2.9 | Bone Level |
| LAA 15°, Nobel (NP) | Nobel | 3.5 | NobelActive, NobelParallel CC, NobelReplace CC | Narrow Platform | 3.0 | Conical |
| Implant Direct | 3.2, 3.7 | InterActive | N/A | 3.0 | Conical |
| 3.2, 3.7, 4.2, 4.7 | Simply Iconic | N/A | 3.0 | Conical |
| LAA 15°, Nobel (RP) | Nobel | 4.3 | NobelActive, NobelParallel CC, NobelReplace CC | Regular Platform | 3.5 | Conical |
| Implant Direct | 4.3, 5.0 | InterActive | N/A | 3.4 | Conical |
| 4.7, 5.2, 5.7 | Simply Iconic | N/A | 3.4 | Conical |
| LAA 15°, Neodent (GM) | Neodent | 3.5, 3.75, 4.0, 5.0, 6.0 | Helix GM | N/A | 3.0 | Conical / Grand Morse |
| 3.5, 4.3, 5.0 | Drive GM | N/A | Conical / Grand Morse |
| 3.5, 3.75, 4.0, 5.0 | Titamax GM | N/A | Conical / Grand Morse |
| LAA 15°, TSV | ZimVie | 3.7 & 4.1 Green | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 3.5 | Internal Hex |
| 4.7 Purple | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 4.5 | Internal Hex |
| Implant Direct | 3.7, 4.2 | Legacy 1, 2, 3, 4 | N/A | 3.5 | Internal Hex |
| 4.7, 5.2 | Legacy 2, 3, 4 | N/A | 4.5 |
| 4.7 | Legacy 1 | N/A | 4.5 |
| BioHorizons | 4.2, 4.6 | Tapered Pro | Yellow | 3.5 | Internal Hex |
| 4.6 | Tapered Plus |
| 3.8 | Tapered Internal |
| 3.0, 3.8 | Tapered Tissue Level |
| 4.6 | Tapered Short |
| 4.2 | Tapered PTG |
| 5.2 | Tapered Pro | Green | 4.5 |
| 5.8 | Tapered Plus |
| 4.6 | Tapered Internal |
| 4.6 | Tapered Tissue Level |
| 5.8 | Tapered Short |
| LAA 15°, Implant Logistics, Implant-One 300 Series | Implant Logistics | 3.5, 4.1, 4.5 | Implant One | 300 Series | 2.75 | Internal Conical 6° Morse Taper |
| LAA 15°, Implant Logistics, Implant-One 400 Series | 4.0, 4.5, 5.5 | Implant One | 400 Series | 3.25 | Internal Conical 6° Morse Taper |

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{7}

ZD

ZEST DENTAL

SOLUTIONS

K250721

# v. Predicate Device Comparison

The following table compares the Indications for Use and key technological characteristics of the subject and predicate device:

Device Comparison Table

|   | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device  |
| --- | --- | --- | --- | --- | --- | --- |
|   | Zest Anchors, Inc. LOCATOR Angled Abutment K25XXXX | Zest Anchors, Inc. LOCATOR Angled Abutment K243272 | Implant Direct LLC. LEGACY ABUTMENT SYSTEM, K060063 | Implant Direct LLC., INTERACTIVE/ SWISHPLUS2 IMPLANT SYSTEM, , K130572 & 2014 interactive/swishactive system K143011 | BioHorizons Implant Systems, INC. Tapered Internal Implant System, K071638 & BIOHORIZONS ABUTMENTS FOR ZIMMER, K103691 & BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK K100985 | Implant Logistics, Inc., Implant-One System, , K173701  |
|  Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | Implant Direct connection is compatible with TSV connection | Implant Direct connection is compatible with NP/RP connection | Biohorizons connection is compatible with TSV connection | Implant-One system, 300, 400 Series connection  |
|  Indications for Use | The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient. A complete Implant Manufacturer's compatibility table is provided in the Instructions For Use. | The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient. The complete Indications for Use Statement with OEM implant compatibilities is provided in the 510(k) Summary for K243272. | The Legacy Abutment System is intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns, bridges, or overdentures for edentulous or partially edentulous patients. The Legacy Abutment System is compatible with implants that have mating diameters, lead-in bevels, internal hex sizes, and 1-72UNF internal threads, as shown in the Zimmer Dental Tapered Screw-Vent Surgical Manual. Implant Direct LLC will monitor the compatible implants for modifications to ensure future compatibility. In the event of any modification, Implant Direct LLC will either modify the Legacy | InterActive/SwishPlus2(K130572): InterActive/SwishPlus2 Implant System consists of two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. Compatibility: InterActive and SwishPlus2 implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) | Tapered Internal Implant System(K071638): The BioHorizons Tapered Internal Implant System is intended for use in the mandible or maxilla for use as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. The BioHorizons Tapered Internal Implant System may be restored immediately 1) with a temporary prosthesis that is not in functional occlusion or 2) when splinted together for multiple tooth replacement or when stabilized with an overdenture supported by multiple implants. - BIOHORIZONS ABUTMENTS FOR ZIMMER(K103691): BioHorizons Abutments for Zimmer® are abutments that include healing abutments for | The Implant-One™ System is indicated for surgical placement in partially or completely edentulous upper or lower jaws to provide a means for prosthetic attachment to restore a patient's chewing function. The Implant-One™ system is indicated for immediate loading only when primary stability is achieved and with the appropriate occlusal loading.  |

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{8}

ZD

ZEST DENTAL

SOLUTIONS

K250721

|   | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device  |
| --- | --- | --- | --- | --- | --- | --- |
|   | Zest Anchors, Inc. LOCATOR Angled Abutment K25XXXX | Zest Anchors, Inc. LOCATOR Angled Abutment K243272 | Implant Direct LLC. LEGACY ABUTMENT SYSTEM, K060063 | Implant Direct LLC., INTERACTIVE/ SWISHPLUS2 IMPLANT SYSTEM, , K130572 & 2014 interactive/swishactive system K143011 | BioHorizons Implant Systems, INC. Tapered Internal Implant System, K071638 & BIOHORIZONS ABUTMENTS FOR ZIMMER, K103691 & BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK K100985 | Implant Logistics, Inc., Implant-One System, , K173701  |
|  Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | Implant Direct connection is compatible with TSV connection | Implant Direct connection is compatible with NP/RP connection | Biohorizons connection is compatible with TSV connection | Implant-One system, 300, 400 Series connection  |
|   |  |  | abutment to ensure compatibility, or cease claiming compatibility to the modified Zimmer Dental Screw-Vent implants. | and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.OmmO, 8.5-18mmLength) implants. Compatibility: InterActive and SwishPlus2 implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.OmmO, 8.5-18mmLength) implants. -- interactive/swishactive(K143011): InterActive/SwishActive Implant System consists of two-piece implants for onestage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. Compatibility: InterActive and SwishActive implants are prosthetically | contouring tissue and final restorative abutments to support a prosthesis. The abutments may be used for a single or multiple unit restoration and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters. BioHorizons Titanium Base Abutments and Laser-Lok Titanium Base Abutments are intended to be used as straight abutments. -- BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK(K100985): BioHorizons Simple Solutions with Laser-Lok is an abutment system that includes healing abutments for contouring tissue and final restorative abutments for cementing a prosthesis. The abutment system may be used for a single or multiple unit restoration and is compatible for use With BioHorizons Internal and Tapered Internal implant Systems and |   |

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{9}

ZD

ZEST DENTAL

SOLUTIONS

K250721

|   | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device  |
| --- | --- | --- | --- | --- | --- | --- |
|   | Zest Anchors, Inc. LOCATOR Angled Abutment K25XXXX | Zest Anchors, Inc. LOCATOR Angled Abutment K243272 | Implant Direct LLC. LEGACY ABUTMENT SYSTEM, K060063 | Implant Direct LLC., INTERACTIVE/ SWISHPLUS2 IMPLANT SYSTEM, , K130572 & 2014 interactive/swishactive system K143011 | BioHorizons Implant Systems, INC. Tapered Internal Implant System, K071638 & BIOHORIZONS ABUTMENTS FOR ZIMMER, K103691 & BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK K100985 | Implant Logistics, Inc., Implant-One System, , K173701  |
|  Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | Implant Direct connection is compatible with TSV connection | Implant Direct connection is compatible with NP/RP connection | Biohorizons connection is compatible with TSV connection | Implant-One system, 300, 400 Series connection  |
|   |  |  |  | compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) titanium abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5- 5.0mmD, 8.5-18mmLength) implants. | Zimmer Dental ScrewVent and Tapered ScrewVent implants with 3.5mm, 4.5mm and 5.7mm internal hex-,connection mating platform diameters. |   |
|  Design |  |  |  |  |  |   |
|  Abutment Cuff Height | 2.5 - 7.5 mm | 2.5 - 7.5 mm | 4.5, 5.7, and 6.5 mm | - | - | 0.5-6 mm  |
|  Abutment Type | Angled | Angled | Straight, Angled | Straight, Angled | Straight | Straight, Angled  |
|  Abutment connection | Internal Hex Conical Conical Taper | TorcFit (Hexalobe) Internal Hex Internal Tri-Channel Conical Grand Morse Taper | Internal hex | Conical | Internal hex | Internal Hex Conical Taper  |
|  Abutment Angle | 15° | 15° | - | 15° | - | 15°  |
|  Sterilization Method | Moist heat end user sterilization | Moist heat end user sterilization | Sterilized by end user | Non-sterile inside a vial sealed with a cap | Abutments provided sterile | -  |

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{10}

ZD

ZEST DENTAL

SOLUTIONS

K250721

|   | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device  |
| --- | --- | --- | --- | --- | --- | --- |
|   | Zest Anchors, Inc.
LOCATOR Angled
Abutment
K25XXXX | Zest Anchors, Inc.
LOCATOR Angled Abutment
K243272 | Implant Direct LLC.
LEGACY ABUTMENT
SYSTEM,
K060063 | Implant Direct LLC.,
INTERACTIVE/ SWISHPLUS2
IMPLANT SYSTEM, ,
K130572 &
2014 interactive/swishactive system
K143011 | BioHorizons Implant Systems,
INC.
Tapered Internal Implant System,
K071638 &
BIOHORIZONS ABUTMENTS
FOR ZIMMER,
K103691 &
BIOHORIZONS SIMPLE
SOLUTIONS WITH LASER-LOK
K100985 | Implant Logistics, Inc.,
Implant-One System, ,
K173701  |
|  Reason for
Predicate/Reference | n/a | Design of Locator Angled
Abutment | Implant Direct connection is
compatible with TSV
connection | Implant Direct connection is compatible
with NP/RP connection | Biohorizons connection is
compatible with TSV connection | Implant-One system,
300, 400 Series
connection  |
|  Compatibility | Zimmer (ZimVie) TSV
3.5ZimVie TSV 3.5
Zimmer (ZimVie)TSV 4.5
Nobel RP
Nobel NP
Neodent Grand Morse
Implant Direct Legacy
Implant Direct InterActive
Implant Direct Simply
Biohorizons
(Tapered Internal Implant
System &
LASER-LOK)
Iconic Implant-One system,
300&400 series | Straumann BLX
Zimmer (ZimVie) TSV
3.5ZimVie TSV 3.5
Zimmer (ZimVie)TSV 4.5
Nobel RP
Nobel NP
Neodent Grand Morse | Implant Direct Legacy | Implant Direct InterActive
Implant Direct Simply Iconic | Biohorizons
(Tapered Internal Implant System
&
LASER-LOK) | Implant One System,
(300&400series)  |
|  Materials |  |  |  |  |  |   |
|  Abutment | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Titanium 6AL4V ELI | Titanium alloy | Titanium alloy | Titanium alloy (Ti-
6Al-4V ELI)  |
|  Abutment
Coating | TiN | TiN | - | - | - | -  |

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{11}

ZD ZEST DENTAL SOLUTIONS

K250721

## vi. Summary of Non-Clinical Performance Testing

The critical features were identified on the OEM component specifications (implant connections, abutment connections, and abutment fixation screws specifications), which are required for proper function. Using engineering analysis, critical areas for implant engagement were analyzed using the implant drawing, abutment drawing and/or the abutment screw drawings provided by the manufacturer. Using the data collected, including the variations, the specifications for the design were created. These specifications were reviewed for manufacturability. All critical tolerances were verified functionally in OEM implants.

Fatigue testing according to ISO 14801: 2016 was performed for the tallest abutment cuff height LOCATOR Angled Abutment along with the smallest diameter OEM implant for the predicate devices cleared in K243272. Compatibility of the predicate LOCATOR Angled Abutment with the various implant systems from Implant Direct and Biorizons, was conducted through engineering analysis, cooperatively with the implant manufacturer (OEM). The LOCATOR Abutments for Implant Logistics ONE 300/400 Series implants have been designed according to the manufacturer's specifications. This abutment design including the implant connection has been cleared per Implant Logistics 510(k) (K173701 and K102822), which required ISO 14801 testing in order to approve the worst conditions, for cuff height and connection. As the LOCATOR connection is identical to the specifications provided by the OEM, the design of the LOCATOR Abutments for Implant Logistics ONE 300/400, a modification of the predicate device, does not create a new worst case.

The LOCATOR Angled Abutments are made of titanium alloy Ti-6Al-4V ELI, conforming to ASTM F136, and have a TiN (Titanium Nitride) coating, identical to the predicate device K233587 and K243272. TiN coating performance was tested per ASTM F1044 and ASTM F1147 in K233587 and being leveraged in the current submission.

The packaging of the LOCATOR Angled Abutments is similar to the packaging of the predicate device, consisting of the LOCATOR angled abutment placed in a vial and sealed in a polybag, along with a parallel post (class I device) used to visually confirm the desired abutment orientation. Packaging and shipping validation testing was completed where the LOCATOR Angled Abutment worst case device and packaging were undamaged after the test, as desired.

The cleaning and sterilization are identical to the predicate device cleared under K233587 and K243272. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.

MR compatibility testing was conducted previously per ASTM F2052-21, ASTM F2213-17, ASTM F2182-19, ASTM F2119-07, and FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" on abutment and implant components made of Ti-6Al-4V and designed with similar features as the LOCATOR Angled Abutments of this 510(K) submission. The tests that were conducted are Force: static magnetic field induced displacement force, Torque: static magnetic field induced torque, Heating: Radiofrequency field (RF) induced heating, Image Quality: susceptibility induced image artifacts, Heating: Gradient field induced heating, and Vibration: Gradient field induced vibration. The results have been

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

{12}

ZD ZEST DENTAL SOLUTIONS
K250721

leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.

An assessment for biocompatibility per ISO 10993-1 was conducted using testing from K072878 and additional cytotoxicity testing per ISO 10993-5 cleared under K233587 was provided in this submission. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.

No other new testing was performed as a part of this submission for the determination of substantial equivalence.

## vii. Substantial Equivalence

As this is a modification to the manufacturer's own cleared and marketed device, the risk based analysis and results of the design control activities performed provide reasonable assurance that the subject devices have demonstrated substantial equivalence to the predicate devices in that they share the same intended use and principles of operation, use the same materials and manufacturing processes, and utilize the same fundamental design including identical prosthetic attachment features; thus the indications for use has been expanded to include compatibility with additional implant systems from Implant Direct, Biohorizons, and Implant One.

Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA

---

**Source:** [https://fda-staging.innolitics.com/device/K250721](https://fda-staging.innolitics.com/device/K250721)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
