The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.
Device Story
LOCATOR Angled Abutment is a titanium (Ti-6Al-4V ELI) dental implant abutment with a titanium nitride (TiN) coating. It features a 15-degree angle for anatomical correction. The device is used by clinicians to attach fixed or removable dental restorations to endosseous implants in the mandible or maxilla. It functions as an interface between the implant and the prosthetic restoration, utilizing retention inserts and denture attachment housings. The device is provided for end-user sterilization. It is used in a clinical dental setting to restore masticatory function for patients with full or partial edentulism. The output is a stable, angled connection point for dental prosthetics, which allows for the correction of implant angulation, improving the fit and function of the final restoration.
Clinical Evidence
No clinical data. Bench testing only. Fatigue testing performed per ISO 14801:2016. Biocompatibility assessed per ISO 10993-1 and ISO 10993-5. MR compatibility testing conducted per ASTM F2052-21, F2213-17, F2182-19, F2119-07. TiN coating performance verified per ASTM F1044 and F1147.
Indicated for patients requiring attachment of full or partial, fixed or removable, dental restorations to endosseous implants to restore masticatory function.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 20, 2025
Zest Anchors, LLC
Maleata Hall
Director Regulatory Affairs
2875 Loker Ave E
Carlsbad, California 92010
Re: K250721
Trade/Device Name: LOCATOR® Angled Abutment
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: July 21, 2025
Received: July 22, 2025
Dear Maleata Hall:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250721 - Maleata Hall
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250721 - Maleata Hall
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250721
Device Name
LOCATOR® Angled Abutment
Indications for Use (Describe)
The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable restorations retained by endosseous implants to restore masticatory function for the patient.
IMPLANT COMPATIBILITY
| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15°, Straumann, BLX | Straumann | 3.5, 3.75, 4.0, 4.5 | BLX | Regular Base | 2.9 | Bone Level |
| 5.0, 5.5, 6.5 | BLX | Wide Base | 2.9 | Bone Level |
| LAA 15°, Nobel (NP) | Nobel | 3.5 | NobelActive, NobelParallel CC, NobelReplace CC | Narrow Platform | 3.0 | Conical |
| Implant Direct | 3.2, 3.7 | InterActive | N/A | 3.0 | Conical |
| 3.2, 3.7, 4.2, 4.7 | Simply Iconic | N/A | 3.0 | Conical |
| LAA 15°, Nobel (RP) | Nobel | 4.3 | NobelActive, NobelParallel CC, NobelReplace CC | Regular Platform | 3.5 | Conical |
| Implant Direct | 4.3, 5.0 | InterActive | N/A | 3.4 | Conical |
| 4.7, 5.2, 5.7 | Simply Iconic | N/A | 3.4 | Conical |
| LAA 15°, Neodent (GM) | Neodent | 3.5, 3.75, 4.0, 5.0, 6.0 | Helix GM | N/A | 3.0 | Conical / Grand Morse |
| 3.5, 4.3, 5.0 | Drive GM | N/A | Conical / Grand Morse |
| 3.5, 3.75, 4.0, 5.0 | Titamax GM | N/A | Conical / Grand Morse |
| LAA 15°, TSV | ZimVie | 3.7 & 4.1 Green | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 3.5 | Internal Hex |
| 4.7 Purple | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 4.5 | Internal Hex |
| Implant Direct | 3.7, 4.2 | Legacy 1, 2, 3, 4 | N/A | 3.5 | Internal Hex |
| 4.7, 5.2 | Legacy 2, 3, 4 | N/A | 4.5 |
| 4.7 | Legacy 1 | N/A | 4.5 |
| 4.2, 4.6 | Tapered Pro | Yellow | 3.5 | Internal Hex |
| BioHorizons | 4.6 | Tapered Plus |
| 3.8 | Tapered Internal |
| 3.0, 3.8 | Tapered Tissue Level |
| 4.6 | Tapered Short |
| 4.2 | Tapered PTG |
| 5.2 | Tapered Pro | Green | 4.5 |
| 5.8 | Tapered Plus |
| 4.6 | Tapered Internal |
| 4.6 | Tapered Tissue Level |
| 5.8 | Tapered Short |
| LAA 15°, Implant Logistics, Implant-One 300 Series | Implant Logistics | 3.5, 4.1, 4.5 | Implant One | 300 Series | 2.75 | Internal Conical 6" Morse Taper |
| LAA 15°, Implant Logistics, Implant-One 400 Series | 4.0, 4.5, 5.5 | Implant One | 400 Series | 3.25 | Internal Conical 6" Morse Taper |
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PISC Publishing Services (391) 443-6740
{4}
ZD ZEST DENTAL SOLUTIONS
K250721
# 510(K) Summary – K250721
# LOCATOR Angled Abutment
## i. General Information on Submitter
Applicant: Zest Anchors, LLC
Address: 2875 Loker Avenue East
Carlsbad, CA 92010 USA
Telephone: (800) 262-2310
Contact Person: David Lin
Contact Title: Sr. Regulatory Affairs Specialist
Email: regulatoryaffairs@zestdent.com
Date Prepared: August 15, 2025
## ii. General Information on Device
Proprietary Name: LOCATOR Angled Abutment
Common Name: Dental Implant Abutment
Regulation Number: 21 CFR 872.3630
Classification Name: Endosseous dental implant abutment
Regulatory Class: Class II
Product Code: NHA (Abutment, Implant, Dental, Endosseous)
## iii. Predicate Device
| Predicate Device | 510(k) Number |
| --- | --- |
| LOCATOR Angled Abutment: Zest Anchors, LLC | K243272 |
| Reference Devices | 510(k) Number |
| High Retention Attachment System | K213391 |
| LOCATOR Angled Abutment (Various) | K233587 |
| Screw Vent Implant; Tapered Screw Vent Implant, Sulzer Dental, Inc. | K013227 |
| Tapered Screw-Vent Implant, 4.1mm, Zimmer Dental, Inc. | K072589 |
| NobelActive Internal Connection Implant, Nobel Biocare AB | K071370 |
| Legacy Abutment System, Implant Direct LLC. | K060063 |
| Interactive/ SwishPlus2 Implant System, Implant Direct LLC. | K130572 |
| 2014 Interactive/SwishActive system | K143011 |
| Tapered Internal Implant System, BioHorizons Implant Systems, Inc | K071638 |
| BioHorizons Simple Solutions with Laser-Lok® | K100985 |
| BioHorizons Abutments for Zimmer | K103691 |
| BioHorizons CAD/CAM Abutments | K151621 |
| Implant-One System, Implant Logistics, Inc. | K173701 |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{5}
ZD ZEST DENTAL SOLUTIONS
K250721
## iv. Description of Device
The purpose of this submission is to expand the Indications for Use of the LOCATOR® Angled Abutment product line (K243272 & K233587) by adding compatibility of existing abutments with various new dental implant systems from Implant Direct and Biohorizons. Additionally, the submission expands the Indications for Use of the product line with a modified version of the predicate device shown to be compatible with the Implant Logistics Implant-One Series 300 and Series 400 Implant Systems. The LOCATOR Angled Abutment is designed and intended for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants in the mandible or maxilla, as cleared to be used with LOCATOR FIXED (K213391) and LOCATOR Attachment Systems (K072878).
The expanded Indications for Use of the LOCATOR Angled Abutment demonstrates that:
1. The cleared devices per K243272 are additionally compatible with the following implant systems from Implant Direct and Biohorizons.
a. The Implant Direct Legacy and Biohorizons Tapered Pro Internal Hex implant systems are compatible with the LOCATOR Angled Abutment Zimmer (ZimVie) TSV 3.5/4.5.
b. The Implant Direct Simply Iconic and Interactive implant systems are compatible with the LOCATOR Angled Abutment Nobel NP/RP.
2. The LOCATOR® Angled Abutment, Implant Logistics 300 Series, a modified version of the predicate device, is compatible with the Implant Logistics Implant-One System Series 300 implant system.
3. The LOCATOR® Angled Abutment, Implant Logistics 400 Series, a modified version of the predicate device, is compatible with the Implant Logistics Implant-One System Series 400 implant system.
The LOCATOR Angled Abutments are manufactured from titanium (Ti-6Al-4V) and are titanium nitride (TiN) coated in various abutment heights, identical to the predicate device. The LOCATOR Angled Abutment interfacing features are provided at a 15 degree angle to allow for angle correction, substantially equivalent to the predicate device of K233587 and K243272. The abutments will be used with the accessories of the LOCATOR Implant Attachment System (retention inserts, denture attachment housing, and ancillary processing parts) and LOCATOR FIXED Attachment System (fixed inserts, denture attachment housing) for the attachment of a restoration, identical to the predicate device.
## Indication for Use
The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{6}
ZD
ZEST DENTAL
SOLUTIONS
K250721
IMPLANT COMPATIBILITY
| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15°, Straumann, BLX | Straumann | 3.5, 3.75, 4.0, 4.5 | BLX | Regular Base | 2.9 | Bone Level |
| 5.0, 5.5, 6.5 | BLX | Wide Base | 2.9 | Bone Level |
| LAA 15°, Nobel (NP) | Nobel | 3.5 | NobelActive, NobelParallel CC, NobelReplace CC | Narrow Platform | 3.0 | Conical |
| Implant Direct | 3.2, 3.7 | InterActive | N/A | 3.0 | Conical |
| 3.2, 3.7, 4.2, 4.7 | Simply Iconic | N/A | 3.0 | Conical |
| LAA 15°, Nobel (RP) | Nobel | 4.3 | NobelActive, NobelParallel CC, NobelReplace CC | Regular Platform | 3.5 | Conical |
| Implant Direct | 4.3, 5.0 | InterActive | N/A | 3.4 | Conical |
| 4.7, 5.2, 5.7 | Simply Iconic | N/A | 3.4 | Conical |
| LAA 15°, Neodent (GM) | Neodent | 3.5, 3.75, 4.0, 5.0, 6.0 | Helix GM | N/A | 3.0 | Conical / Grand Morse |
| 3.5, 4.3, 5.0 | Drive GM | N/A | Conical / Grand Morse |
| 3.5, 3.75, 4.0, 5.0 | Titamax GM | N/A | Conical / Grand Morse |
| LAA 15°, TSV | ZimVie | 3.7 & 4.1 Green | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 3.5 | Internal Hex |
| 4.7 Purple | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 4.5 | Internal Hex |
| Implant Direct | 3.7, 4.2 | Legacy 1, 2, 3, 4 | N/A | 3.5 | Internal Hex |
| 4.7, 5.2 | Legacy 2, 3, 4 | N/A | 4.5 |
| 4.7 | Legacy 1 | N/A | 4.5 |
| BioHorizons | 4.2, 4.6 | Tapered Pro | Yellow | 3.5 | Internal Hex |
| 4.6 | Tapered Plus |
| 3.8 | Tapered Internal |
| 3.0, 3.8 | Tapered Tissue Level |
| 4.6 | Tapered Short |
| 4.2 | Tapered PTG |
| 5.2 | Tapered Pro | Green | 4.5 |
| 5.8 | Tapered Plus |
| 4.6 | Tapered Internal |
| 4.6 | Tapered Tissue Level |
| 5.8 | Tapered Short |
| LAA 15°, Implant Logistics, Implant-One 300 Series | Implant Logistics | 3.5, 4.1, 4.5 | Implant One | 300 Series | 2.75 | Internal Conical 6° Morse Taper |
| LAA 15°, Implant Logistics, Implant-One 400 Series | 4.0, 4.5, 5.5 | Implant One | 400 Series | 3.25 | Internal Conical 6° Morse Taper |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{7}
ZD
ZEST DENTAL
SOLUTIONS
K250721
# v. Predicate Device Comparison
The following table compares the Indications for Use and key technological characteristics of the subject and predicate device:
Device Comparison Table
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- | --- |
| | Zest Anchors, Inc. LOCATOR Angled Abutment K25XXXX | Zest Anchors, Inc. LOCATOR Angled Abutment K243272 | Implant Direct LLC. LEGACY ABUTMENT SYSTEM, K060063 | Implant Direct LLC., INTERACTIVE/ SWISHPLUS2 IMPLANT SYSTEM, , K130572 & 2014 interactive/swishactive system K143011 | BioHorizons Implant Systems, INC. Tapered Internal Implant System, K071638 & BIOHORIZONS ABUTMENTS FOR ZIMMER, K103691 & BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK K100985 | Implant Logistics, Inc., Implant-One System, , K173701 |
| Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | Implant Direct connection is compatible with TSV connection | Implant Direct connection is compatible with NP/RP connection | Biohorizons connection is compatible with TSV connection | Implant-One system, 300, 400 Series connection |
| Indications for Use | The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient. A complete Implant Manufacturer's compatibility table is provided in the Instructions For Use. | The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient. The complete Indications for Use Statement with OEM implant compatibilities is provided in the 510(k) Summary for K243272. | The Legacy Abutment System is intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns, bridges, or overdentures for edentulous or partially edentulous patients. The Legacy Abutment System is compatible with implants that have mating diameters, lead-in bevels, internal hex sizes, and 1-72UNF internal threads, as shown in the Zimmer Dental Tapered Screw-Vent Surgical Manual. Implant Direct LLC will monitor the compatible implants for modifications to ensure future compatibility. In the event of any modification, Implant Direct LLC will either modify the Legacy | InterActive/SwishPlus2(K130572): InterActive/SwishPlus2 Implant System consists of two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. Compatibility: InterActive and SwishPlus2 implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) | Tapered Internal Implant System(K071638): The BioHorizons Tapered Internal Implant System is intended for use in the mandible or maxilla for use as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. The BioHorizons Tapered Internal Implant System may be restored immediately 1) with a temporary prosthesis that is not in functional occlusion or 2) when splinted together for multiple tooth replacement or when stabilized with an overdenture supported by multiple implants. - BIOHORIZONS ABUTMENTS FOR ZIMMER(K103691): BioHorizons Abutments for Zimmer® are abutments that include healing abutments for | The Implant-One™ System is indicated for surgical placement in partially or completely edentulous upper or lower jaws to provide a means for prosthetic attachment to restore a patient's chewing function. The Implant-One™ system is indicated for immediate loading only when primary stability is achieved and with the appropriate occlusal loading. |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{8}
ZD
ZEST DENTAL
SOLUTIONS
K250721
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- | --- |
| | Zest Anchors, Inc. LOCATOR Angled Abutment K25XXXX | Zest Anchors, Inc. LOCATOR Angled Abutment K243272 | Implant Direct LLC. LEGACY ABUTMENT SYSTEM, K060063 | Implant Direct LLC., INTERACTIVE/ SWISHPLUS2 IMPLANT SYSTEM, , K130572 & 2014 interactive/swishactive system K143011 | BioHorizons Implant Systems, INC. Tapered Internal Implant System, K071638 & BIOHORIZONS ABUTMENTS FOR ZIMMER, K103691 & BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK K100985 | Implant Logistics, Inc., Implant-One System, , K173701 |
| Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | Implant Direct connection is compatible with TSV connection | Implant Direct connection is compatible with NP/RP connection | Biohorizons connection is compatible with TSV connection | Implant-One system, 300, 400 Series connection |
| | | | abutment to ensure compatibility, or cease claiming compatibility to the modified Zimmer Dental Screw-Vent implants. | and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.OmmO, 8.5-18mmLength) implants. Compatibility: InterActive and SwishPlus2 implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.OmmO, 8.5-18mmLength) implants. -- interactive/swishactive(K143011): InterActive/SwishActive Implant System consists of two-piece implants for onestage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. Compatibility: InterActive and SwishActive implants are prosthetically | contouring tissue and final restorative abutments to support a prosthesis. The abutments may be used for a single or multiple unit restoration and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters. BioHorizons Titanium Base Abutments and Laser-Lok Titanium Base Abutments are intended to be used as straight abutments. -- BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK(K100985): BioHorizons Simple Solutions with Laser-Lok is an abutment system that includes healing abutments for contouring tissue and final restorative abutments for cementing a prosthesis. The abutment system may be used for a single or multiple unit restoration and is compatible for use With BioHorizons Internal and Tapered Internal implant Systems and | |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{9}
ZD
ZEST DENTAL
SOLUTIONS
K250721
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- | --- |
| | Zest Anchors, Inc. LOCATOR Angled Abutment K25XXXX | Zest Anchors, Inc. LOCATOR Angled Abutment K243272 | Implant Direct LLC. LEGACY ABUTMENT SYSTEM, K060063 | Implant Direct LLC., INTERACTIVE/ SWISHPLUS2 IMPLANT SYSTEM, , K130572 & 2014 interactive/swishactive system K143011 | BioHorizons Implant Systems, INC. Tapered Internal Implant System, K071638 & BIOHORIZONS ABUTMENTS FOR ZIMMER, K103691 & BIOHORIZONS SIMPLE SOLUTIONS WITH LASER-LOK K100985 | Implant Logistics, Inc., Implant-One System, , K173701 |
| Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | Implant Direct connection is compatible with TSV connection | Implant Direct connection is compatible with NP/RP connection | Biohorizons connection is compatible with TSV connection | Implant-One system, 300, 400 Series connection |
| | | | | compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) titanium abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5- 5.0mmD, 8.5-18mmLength) implants. | Zimmer Dental ScrewVent and Tapered ScrewVent implants with 3.5mm, 4.5mm and 5.7mm internal hex-,connection mating platform diameters. | |
| Design | | | | | | |
| Abutment Cuff Height | 2.5 - 7.5 mm | 2.5 - 7.5 mm | 4.5, 5.7, and 6.5 mm | - | - | 0.5-6 mm |
| Abutment Type | Angled | Angled | Straight, Angled | Straight, Angled | Straight | Straight, Angled |
| Abutment connection | Internal Hex Conical Conical Taper | TorcFit (Hexalobe) Internal Hex Internal Tri-Channel Conical Grand Morse Taper | Internal hex | Conical | Internal hex | Internal Hex Conical Taper |
| Abutment Angle | 15° | 15° | - | 15° | - | 15° |
| Sterilization Method | Moist heat end user sterilization | Moist heat end user sterilization | Sterilized by end user | Non-sterile inside a vial sealed with a cap | Abutments provided sterile | - |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{10}
ZD
ZEST DENTAL
SOLUTIONS
K250721
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- | --- |
| | Zest Anchors, Inc.
LOCATOR Angled
Abutment
K25XXXX | Zest Anchors, Inc.
LOCATOR Angled Abutment
K243272 | Implant Direct LLC.
LEGACY ABUTMENT
SYSTEM,
K060063 | Implant Direct LLC.,
INTERACTIVE/ SWISHPLUS2
IMPLANT SYSTEM, ,
K130572 &
2014 interactive/swishactive system
K143011 | BioHorizons Implant Systems,
INC.
Tapered Internal Implant System,
K071638 &
BIOHORIZONS ABUTMENTS
FOR ZIMMER,
K103691 &
BIOHORIZONS SIMPLE
SOLUTIONS WITH LASER-LOK
K100985 | Implant Logistics, Inc.,
Implant-One System, ,
K173701 |
| Reason for
Predicate/Reference | n/a | Design of Locator Angled
Abutment | Implant Direct connection is
compatible with TSV
connection | Implant Direct connection is compatible
with NP/RP connection | Biohorizons connection is
compatible with TSV connection | Implant-One system,
300, 400 Series
connection |
| Compatibility | Zimmer (ZimVie) TSV
3.5ZimVie TSV 3.5
Zimmer (ZimVie)TSV 4.5
Nobel RP
Nobel NP
Neodent Grand Morse
Implant Direct Legacy
Implant Direct InterActive
Implant Direct Simply
Biohorizons
(Tapered Internal Implant
System &
LASER-LOK)
Iconic Implant-One system,
300&400 series | Straumann BLX
Zimmer (ZimVie) TSV
3.5ZimVie TSV 3.5
Zimmer (ZimVie)TSV 4.5
Nobel RP
Nobel NP
Neodent Grand Morse | Implant Direct Legacy | Implant Direct InterActive
Implant Direct Simply Iconic | Biohorizons
(Tapered Internal Implant System
&
LASER-LOK) | Implant One System,
(300&400series) |
| Materials | | | | | | |
| Abutment | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Titanium 6AL4V ELI | Titanium alloy | Titanium alloy | Titanium alloy (Ti-
6Al-4V ELI) |
| Abutment
Coating | TiN | TiN | - | - | - | - |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{11}
ZD ZEST DENTAL SOLUTIONS
K250721
## vi. Summary of Non-Clinical Performance Testing
The critical features were identified on the OEM component specifications (implant connections, abutment connections, and abutment fixation screws specifications), which are required for proper function. Using engineering analysis, critical areas for implant engagement were analyzed using the implant drawing, abutment drawing and/or the abutment screw drawings provided by the manufacturer. Using the data collected, including the variations, the specifications for the design were created. These specifications were reviewed for manufacturability. All critical tolerances were verified functionally in OEM implants.
Fatigue testing according to ISO 14801: 2016 was performed for the tallest abutment cuff height LOCATOR Angled Abutment along with the smallest diameter OEM implant for the predicate devices cleared in K243272. Compatibility of the predicate LOCATOR Angled Abutment with the various implant systems from Implant Direct and Biorizons, was conducted through engineering analysis, cooperatively with the implant manufacturer (OEM). The LOCATOR Abutments for Implant Logistics ONE 300/400 Series implants have been designed according to the manufacturer's specifications. This abutment design including the implant connection has been cleared per Implant Logistics 510(k) (K173701 and K102822), which required ISO 14801 testing in order to approve the worst conditions, for cuff height and connection. As the LOCATOR connection is identical to the specifications provided by the OEM, the design of the LOCATOR Abutments for Implant Logistics ONE 300/400, a modification of the predicate device, does not create a new worst case.
The LOCATOR Angled Abutments are made of titanium alloy Ti-6Al-4V ELI, conforming to ASTM F136, and have a TiN (Titanium Nitride) coating, identical to the predicate device K233587 and K243272. TiN coating performance was tested per ASTM F1044 and ASTM F1147 in K233587 and being leveraged in the current submission.
The packaging of the LOCATOR Angled Abutments is similar to the packaging of the predicate device, consisting of the LOCATOR angled abutment placed in a vial and sealed in a polybag, along with a parallel post (class I device) used to visually confirm the desired abutment orientation. Packaging and shipping validation testing was completed where the LOCATOR Angled Abutment worst case device and packaging were undamaged after the test, as desired.
The cleaning and sterilization are identical to the predicate device cleared under K233587 and K243272. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.
MR compatibility testing was conducted previously per ASTM F2052-21, ASTM F2213-17, ASTM F2182-19, ASTM F2119-07, and FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" on abutment and implant components made of Ti-6Al-4V and designed with similar features as the LOCATOR Angled Abutments of this 510(K) submission. The tests that were conducted are Force: static magnetic field induced displacement force, Torque: static magnetic field induced torque, Heating: Radiofrequency field (RF) induced heating, Image Quality: susceptibility induced image artifacts, Heating: Gradient field induced heating, and Vibration: Gradient field induced vibration. The results have been
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{12}
ZD ZEST DENTAL SOLUTIONS
K250721
leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.
An assessment for biocompatibility per ISO 10993-1 was conducted using testing from K072878 and additional cytotoxicity testing per ISO 10993-5 cleared under K233587 was provided in this submission. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.
No other new testing was performed as a part of this submission for the determination of substantial equivalence.
## vii. Substantial Equivalence
As this is a modification to the manufacturer's own cleared and marketed device, the risk based analysis and results of the design control activities performed provide reasonable assurance that the subject devices have demonstrated substantial equivalence to the predicate devices in that they share the same intended use and principles of operation, use the same materials and manufacturing processes, and utilize the same fundamental design including identical prosthetic attachment features; thus the indications for use has been expanded to include compatibility with additional implant systems from Implant Direct, Biohorizons, and Implant One.
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.