Single-use Ureteral Access Sheath

K250695 · Shenzhen HugeMed Medical Technical Development Co., Ltd. · FED · Jul 8, 2025 · Gastroenterology, Urology

Device Facts

Record IDK250695
Device NameSingle-use Ureteral Access Sheath
ApplicantShenzhen HugeMed Medical Technical Development Co., Ltd.
Product CodeFED · Gastroenterology, Urology
Decision DateJul 8, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Single-use Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract.

Device Story

Single-use Ureteral Access Sheath establishes conduit in urinary tract during endoscopic urological procedures; facilitates passage of endoscopes and instruments. Device comprises access sheath (sheath tube, connector) and dilator (dilator tube, dilator base). Used in clinical settings by physicians. Provides physical pathway for instrumentation; aids in maintaining access during procedures. Benefits patient by enabling minimally invasive urological interventions.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench performance testing, including appearance, dimensions, patency, fitness, seal leakage, peak tensile force, connection, bending resistance, deformation resistance, suction, friction, flow rate, ray detectability, and corrosion. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic, pyrogenicity).

Technological Characteristics

Materials: Pebax, SUS304, PTFE, PC, Silicone, ABS, PP, PE, POM, PVP coating. Dimensions: 9-12Fr ID, 40-50cm length. Sterilization: Ethylene Oxide (ISO 11135). Shelf life: 3 years (ASTM F1980). Packaging validated per ASTM F1140, F1886, F88, F1929. Mechanical properties: bending resistance, tensile force, friction force.

Indications for Use

Indicated for patients undergoing endoscopic urological procedures requiring a conduit for the passage of endoscopes and instruments into the urinary tract.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 8, 2025 Shenzhen HugeMed Medical Technical Development Co., Ltd. Cathy Shi Regulatory Engineer 401, 501, Building 4, Haizhi Technology Park, Fortis No. 17 Bulan Road, Xialilang Community, Nanwan Street Shenzhen, Guangdong 518112 China Re: K250695 Trade/Device Name: Single-use Ureteral Access Sheath Regulation Number: 21 CFR § 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: May 5, 2025 Received: June 17, 2025 Dear Cathy Shi: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1} K250695 - Cathy Shi Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical- {2} K250695 - Cathy Shi Page 3 devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark J. Antonino -S Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250695 Device Name Single-use Ureteral Access Sheath Indications for Use (Describe) The Single-use Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K250695 Page 1 of 6 # 510(K) Summary Prepared in accordance with the requirements of 21 CFR Part 807.92 ## I. Submitter Shenzhen HugeMed Medical Technical Development Co., Ltd 401, 501, Building 4, Haizhi Technology Park, Fortis, No. 17, Bulan Road, Xialilang Community, Nanwan Street, Longgang District, Shenzhen, Guangdong, 518112, China Date of preparation: Mar. 5, 2025 Contact person: Cathy.shi E-mail: Cathy.shi@hugemed.cn Tel: +86-13534156592 ## II. Subject Device Information Trade/Device Name: Single-use Ureteral Access Sheath Common Name: Endoscope and accessories Classification Name: Endoscopic Access Overtube Regulation Name: Endoscope and accessories Regulation Number: 21 CFR 876.1500 Regulatory Class: Class II Product Code: FED Panel: Gastroenterology/Urology Submission type: Traditional 510(K) ## III. Predicate Device 510(k) number: K241181 Product name: Disposable Ureteral Guide Sheath Regulatory Class: Class II Product Code: FED Manufacturer: Dongguan ZSR Biomedical technology Company Limited ## IV. Device Description The Single-use Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. The device models are Y-09-40, Y-09-50, Y-10-40, Y-10-50, Y-11-40, Y-11-50, Y-12-40, and Y-12-50. The device is composed of access sheath and dilator. The access sheath consists of sheath tube and connector. The dilator consists of dilator tube and dilator base. Single-use Ureteral Access Sheath is sterilized by Ethylene Oxide Gas to achieve a SAL of $10^{-6}$ and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 3 years. {5} K250695 Page 2 of 6 # V. Intended use & Indication for use The Single-use Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. # VI. Comparison to the Predicate Device | Characteristics | Subject Device | Predicate Device | Remark | | --- | --- | --- | --- | | K number | / | K241181 | / | | Product Code | FED | FED | Same | | Regulation Description | Endoscope and accessories | Endoscope and accessories | Same | | Regulation No. | 21 CFR 876.1500 | 21 CFR 876.1500 | Same | | Regulatory Class | II | II | Same | | Device trade name | Single-use Ureteral Access Sheath | Disposable Ureteral Guide Sheath | / | | Indication for use/Intended use | The Single-use Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. | The Disposable Ureteral Guide Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. | Same | | Single use/ Reuse | Single-use | Single-use | Same | | Sterile | Yes | Yes | Same | | Sterilization Method | EO sterilization | EO sterilization | Same | | Sheath ID | 9Fr,10Fr, 11Fr, 12Fr | 10Fr, 11Fr, 12Fr, 13Fr, 14Fr, 15Fr, 16Fr | Different 1 | | Sheath length | 40cm, 50cm | 10Fr, 11Fr, 12Fr, 13Fr, 14Fr:35cm, 40cm, 45cm, 50cm, 55cm 15Fr, 16Fr:40cm, 45cm | Different 2 | | Primary structure | The device is composed of access sheath and dilator. The access sheath consists of sheath tube and connector. The dilator consists of dilator tube and dilator base. | The device is composed of guide sheath and dilator. The guide sheath consists of guide sheath tube and connector. | Same | {6} K250695 Page 3 of 6 | Materials | Sheath tube: Pebax, SUS304, PTFE Sheath connector: Polycarbonate (PC), Silicone, ABS, Polypropylene (PP) Dilator tube: polyethylene (PE) Dilator base: Polyoxymethylene (POM) Hydrophilic coating: Polyvinylpyrrolidone (PVP) | Sheath tube: Pebax, US304, PTFE Connector: polyamide (PA), silica gel Dilator connector: polypropylene (PP) Dilator tube: polyethylene (PE) Hydrophilic coating: Polyvinylpyrrolidone (PVP) | Different 3 | | --- | --- | --- | --- | | Package | Single-use EO sterilized pouch with one device per pouch | Single-use EO sterilized pouch with one device per pouch | Same | | Biocompatible | Yes | Yes | Same | | Shelf Life | 3 years | 3 years | Same | | Bending Resistance | The access sheath tube and the dilator tube were bent 90 ° in both directions by grasping the head and the tail. After 20 times of repetition, the tube body was free of cracks, creases, cracks and fractures, the coating was free of falling off, and the steel wire was not separated from the inner and outer rubber layers. | The guiding sheath tube and the dilating tube were bent 90 ° in both directions by grasping the head and the tail. After 20 times of repetition, the tube body was free of cracks, creases, cracks and fractures, the coating was free of falling off, and the steel wire was not separated from the inner and outer rubber layers. | Same | | Friction Force | After being dipped in water, the surface is smooth, and the friction force is not more than 0.5 N; The average dynamic | After being dipped in water, the surface is smooth, and the friction force is not more than 0.5 N; The average dynamic | Same | {7} K250695 Page 4 of 6 | | friction force of 10 repeated tests shall not be greater than 0.5 N. | friction force of 10 repeated tests shall not be greater than 0.5 N | | | --- | --- | --- | --- | | Peak tensile force | The peak tensile force between the sheath tube and sheath connector, and between the dilator tube and the dilator base shall be ≥ 20 N. The fracture force test shall be carried out when the sheath connector and the dilator base are locked. The minimum fracture force shall be ≥ 15N and shall not break for 15 s. | The joint between the guiding sheath tube and the tube body of the dilating tube and the connector shall be able to bear a force of ≥ 20 N and shall not break for 15 s; The fracture force test shall be carried out when the guiding sheath tube connector and the expansion tube connector are locked. The minimum fracture force shall be ≥ 15N and shall not break for 15 s. | Same | Different 1 – Sheath ID This difference is in sheath ID. Different sheath inner diameter device will be selected by physician per patient's condition, this difference will not raise any issues in safety and effectiveness. Different 2 – Sheath length Different sheath length device will be selected by physician per patient's condition, this difference will not raise any issues in safety and effectiveness. Different 3 – Materials Biocompatibility of the Disposable Ureteral Guide Sheath was evaluated in accordance with the FDA guidance "Use of International Standard ISO 10993-1", this difference will not raise any issues in safety and effectiveness. VII. Performance Data The following performance data were provided in support of the substantial equivalence determination. Biocompatibility testing Biocompatibility of the Single-use Ureteral Access Sheath was evaluated in accordance {8} K250695 Page 5 of 6 with the FDA guidance “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. 1) Cytotoxicity 2) Sensitization 3) Irritation 4) Acute Systemic 5) Pyrogenicity ## Sterility and Shelf -life The Single-use Ureteral Access Sheath is provided sterile and its shelf-life is 3 years. Sterilization Process has been validated accordance with ISO 11135:2014, ISO 11737-1:2018, ISO11737-2:2019. EO/ECH residual test was performed according to ISO 10993-7:2008. The shelf life is determined based on product performance testing after accelerated aging test according to ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Package validation was conducted according to ASTM F1140/F1140M -13, ASTM F1886/F1886M-16, ASTM F88/F88M-23, ASTM F 1929-23. ## Bench performance testing The following bench Comparison tests were performed: - Appearance - Dimensions - Patency - Fitness - Seal leakage - Peak tensile force - Connection - Bending resistance - Deformation resistance - Suction - Friction - Flow rate - Ray detectability - Corrosion ## VIII. Clinical study Not applicable. {9} K250695 Page 6 of 6 ## IX. Conclusion The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device. 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...