K250690 · Penumbra, Inc. · NRY · Jun 10, 2026 · Cardiovascular
Device Facts
Record ID
K250690
Device Name
Penumbra System (Thunderbolt Aspiration Tubing)
Applicant
Penumbra, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Jun 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. Penumbra Thunderbolt Aspiration Tubing As part of the Penumbra System, the Penumbra Thunderbolt Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Device Story
The Penumbra Thunderbolt Aspiration Tubing connects Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump to remove thrombus from neurovasculature. It enables both continuous and modulated aspiration; modulated aspiration alternates vacuum from the pump with positive pressure from a sterile saline IV bag at ambient pressure. Used in clinical settings by physicians during neurothrombectomy procedures. The tubing facilitates vacuum transfer to the catheter tip to aspirate clots. Modulated aspiration is intended to clear blockages and restore blood flow. The device is provided in a kit with catheters and accessories. Clinical benefit includes successful revascularization of occluded vessels in acute ischemic stroke patients.
Clinical Evidence
Prospective, multicenter, single-arm study (THUNDER) of 216 patients with acute ischemic stroke. Primary endpoint: angiographic revascularization (mTICI 2b-3) post-procedure. Success achieved in 87.5% (189/216) of ITT population, exceeding the 71.5% performance goal (p<0.001). Safety outcomes included 0.9% symptomatic intracranial hemorrhage at 24 hours and 11.7% all-cause mortality at 90 days.
Technological Characteristics
Tubing connects reperfusion catheters to aspiration pump. Features modulated (alternating vacuum/saline) and continuous aspiration modes. Materials are standard for interventional devices. Dimensions: 120 in length, 0.125 in ID. Sterilized via Ethylene Oxide (EO). Complies with IEC 60601-1, IEC 60601-1-2, ISO 10079-1/4.
Indications for Use
Indicated for patients 18-80 years old with acute ischemic stroke secondary to intracranial large vessel occlusion (internal carotid, M1/M2 middle cerebral, basilar, or vertebral arteries) within 8 hours of symptom onset. Candidates include those ineligible for or failed thrombolytic therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra System (Modified 110 Aspiration Tubing) (K182522)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 10, 2026
Penumbra, Inc.
Sindokht (Sisi) Soltanzadeh
Principal Regulatory Affairs Specialist
One Penumbra Place
Alameda, California 94502
Re: K250690
Trade/Device Name: Penumbra System (Thunderbolt Aspiration Tubing)
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: NRY
Dated: May 12, 2026
Received: May 13, 2026
Dear Sindokht (Sisi) Soltanzadeh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250690 - Sindokht (Sisi) Soltanzadeh
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250690 - Sindokht (Sisi) Soltanzadeh
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K250690
Device Name
Penumbra System (Thunderbolt Aspiration Tubing)
Indications for Use (Describe)
Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
Penumbra Thunderbolt Aspiration Tubing
As part of the Penumbra System, the Penumbra Thunderbolt Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Penumbra
K250690
# **510(k) Summary – K250690**
# **1. Submitter**
Penumbra, Inc.
One Penumbra Place
Alameda, CA 94502 USA
Contact Person:
Sindokht (Sisi) Soltanzadeh
Email: ssoltanzadeh@penumbrainc.com
Date of Preparation: June 08, 2026
# **2. Subject Device**
Device Name: Penumbra System (Thunderbolt Aspiration Tubing)
Regulatory Class: Class II
Regulation Name: Percutaneous Catheter
Regulation Number: 21 CFR 870.1250
Product Code: NRY
# **3. Predicate Device**
| 510(k) Number | Name of Device |
| --- | --- |
| K182522 | Penumbra System (Modified 110 Aspiration Tubing) |
Page 1 of 20
{5}
Penumbra
K250690
# 4. Device Description
The Penumbra System is comprised of the following devices:
- Penumbra Reperfusion Catheter
- Penumbra Aspiration Pump
- Penumbra Aspiration Pump Canister/Tubing
- Penumbra Thunderbolt Aspiration Tubing
- Penumbra Separator
The Penumbra System is designed to remove thrombus from and restore blood flow to the neurovasculature using aspiration. The Reperfusion Catheter delivers aspiration from the pump directly to the site of occlusion to remove the clot. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. The Penumbra Reperfusion Catheter is connected to the Penumbra Aspiration Pump using the Penumbra Thunderbolt Aspiration Tubing and the Penumbra Aspiration Pump Canister.
The Penumbra Thunderbolt Aspiration Tubing facilitates the transfer of vacuum between the Penumbra Reperfusion Catheter and Penumbra Aspiration Pump while providing modulated or continuous aspiration. Modulated aspiration is provided when the Penumbra Thunderbolt Aspiration Tubing alternates between connecting the Reperfusion Catheter to the Penumbra Aspiration Pump and a sterile saline intravenous (IV) bag at ambient pressure in a predetermined modulated aspiration sequence. Vacuum is applied to the Reperfusion Catheter when connected to the Penumbra Aspiration Pump. Positive pressure is applied to the Reperfusion Catheter when connected to a sterile saline IV bag at ambient pressure. In non-clinical benchtop testing simulating complete occlusion of a Penumbra Reperfusion Catheter RED 72 during modulated aspiration, connection to a sterile saline IV bag at ambient pressure applied a positive pressure of 7.6 inHg (193 mmHg) within the occluded Reperfusion Catheter tip.
The Penumbra System Thunderbolt Aspiration Tubing is provided in a kit configuration with a Penumbra Reperfusion Catheter (RED 62, RED 68, or RED 72), along with a steam shaping mandrel, rotating hemostasis valve (RHV), peelable sheath, and optionally, SENDit Technology. The Penumbra Separator is provided with an introducer and torque device. The devices are visible under fluoroscopy. Modulated aspiration may only be used with the Penumbra Reperfusion Catheters RED 62, RED 68, and RED 72, without use of any devices within the Reperfusion Catheter (e.g., no Separator or 3D Revascularization Device).
# 5. Indications For Use
# Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral –
Page 2 of 20
{6}
Penumbra
K250690
M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
Penumbra Thunderbolt Aspiration Tubing
As part of the Penumbra System, the Penumbra Thunderbolt Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
6. Comparison of Technological Characteristics with the Predicate Device
| Device Attribute | Predicate Device | Subject Device |
| --- | --- | --- |
| Trade Name | Penumbra System (Modified 110 Aspiration Tubing) | Penumbra System (Thunderbolt Aspiration Tubing) |
| Classification | Class II, NRY, 21 CFR 870.1250 | SAME |
| 510(k) Number | K182522 | K250690 |
| Indications for Use | Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra 3D Revascularization Device As part of the Penumbra System, the Penumbra 3D Revascularization Device is | Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. Penumbra Thunderbolt Aspiration Tubing As part of the Penumbra System, the Penumbra Thunderbolt Aspiration Tubing is indicated to |
Page 3 of 20
{7}
Penumbra ![Penumbra logo]()
K250690
| Device Attribute | Predicate Device | Subject Device |
| --- | --- | --- |
| | indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra Aspiration Tubing As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems. | connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems. |
| Aspiration Mode | Continuous | Modulated and Continuous |
| Device Materials | Commonly used materials for interventional devices | SAME |
| Tubing Inner Diameter (ID) | 0.110 in | 0.125 in |
| Proximal Tubing Outer Diameter (OD) | 0.188 in | 0.230 in |
| Distal Tubing Outer Diameter (OD) | 0.188 in | 0.225 in |
| Overall Length | 100 in | 120 in |
| Packaging Materials | Commonly used materials for interventional devices | SAME |
Page 4 of 20
{8}
Penumbra
K250690
| Device Attribute | Predicate Device | Subject Device |
| --- | --- | --- |
| Packaging Configurations | Individual and Kit Packaged | Kit Packaged Only |
| Aspiration Pump | Penumbra Aspiration Pump | SAME |
| Condition Supplied | Sterile and Single Use | SAME |
| Sterilization Method | Ethylene Oxide (EO) | SAME |
### 7. Summary of Non-Clinical Data
The following non-clinical testing data were provided in support of the substantial equivalence determination:
- Biocompatibility
- Performance Testing – Design Verification
- Shelf Life
- Packaging Validation
- Sterilization
- Good Laboratory Practice (GLP) Animal Study
- Software
- Electrical Safety/Electromagnetic Compatibility (EMC) Testing
The subject device met all established requirements.
### 7.1 Biocompatibility
The biocompatibility evaluation for the Thunderbolt Aspiration Tubing was conducted in accordance with ISO 10993-1, USP standards, and 21 CFR Part 58, Good Laboratory Practice (GLP). The battery of testing included the following tests:
| Test | Acceptance Criteria | Conclusion |
| --- | --- | --- |
| Cytotoxicity: MEM Elution | Sample extracts must have a cytotoxic reactivity score of Grade 2 or lower. | Pass Non-cytotoxic |
| Sensitization: Magnusson-Kligman Method | Test group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control group yields Grade < 1). | Pass Non-sensitizing |
Page 5 of 20
{9}
Penumbra
K250690
| Test | Acceptance Criteria | Conclusion |
| --- | --- | --- |
| Irritation: Intracutaneous Toxicity | The difference between the average scores for the extract of the test article and the control must be ≤ 1.0. | Pass Non-irritating |
| Systemic Toxicity: Acute Systemic Injection | Sample extracts must not cause significant biological reaction greater than control. That is: • Death in 2 or more animals. • Signs of toxicity in 2 or more animals (i.e. convulsions, prostration). • Weight loss > 10% in 3 or more animals. | Pass Non-toxic |
| Systemic Toxicity: Material Mediated Pyrogen | Sample extracts must not cause a total rise in body temperature of ≥ 0.5 °C. | Pass Non-pyrogenic |
| Hemocompatibility: Hemolysis (indirect contact) | Sample extracts must be non-hemolytic (≤ 2% hemolytic index). | Pass Non-hemolytic |
Biocompatibility test results demonstrate biological safety per ISO 10993 and USP requirements.
Page 6 of 20
{10}
Penumbra
K250690
### 7.2 Performance Testing – Design Verification
The following design verification tests were performed on the Thunderbolt Aspiration Tubing.
| Test | Test Method Summary | Conclusion |
| --- | --- | --- |
| Dimensional/Visual | Confirms the units meet all dimensional and visual product specifications. | Acceptance Criteria Met |
| Valve Testing | Confirms the units meet all product specifications related to valve pressurization and functionality. | Acceptance Criteria Met |
| Leak Testing | Confirms the units meet all product specifications related to leak resistance. | Acceptance Criteria Met |
| Tensile Testing | Confirms the units meet all product specifications related to tensile strength. | Acceptance Criteria Met |
| Simulated Use | Confirms the functionality of units using clinically relevant benchtop model. | Acceptance Criteria Met |
| System Compatibility | Confirms the compatibility of units with system devices and accessory components, including reperfusion catheter integrity and Luer compatibility per ISO 80369-7. | Acceptance Criteria Met |
| System Durability | Confirms the units meet all product specifications related to durability for the procedure duration. | Acceptance Criteria Met |
| Flow Rate and Catheter Tip Pressure | The flow rate and pressures within the catheter tip were measured. | Flow rate and pressures were characterized |
| Shelf Life | Confirms expiration date based on accelerated aging test studies. | Acceptance Criteria Met |
| Packaging Validation | Confirms the units meet all packaging specifications. | Acceptance Criteria Met |
| Sterilization | Confirms the units are sterilized in accordance with ISO 11135+A1 and ISO 10993-7. | Acceptance Criteria Met |
### 7.3 GLP Animal Study
The safety of Thunderbolt Aspiration Tubing modulated aspiration with a Reperfusion Catheter when occlusive in porcine subclavian and lingual arteries was evaluated in accordance with the “Code of Federal Regulations of the Food and Drug Administration 21 CFR, Part 58 – Good Laboratory Practices for Nonclinical Laboratory Studies”. The study evaluated Thunderbolt Aspiration Tubing modulated aspiration (with saline inflow) compared to continuous aspiration with Penumbra System 110 Aspiration Tubing as a concurrent control.
Page 7 of 20
{11}
Penumbra
K250690
Modulated aspiration or continuous aspiration was applied continuously for five (5) minutes in each treatment vessel. Vascular response was assessed post-treatment by contrast angiography. Vascular response was also assessed at subacute and chronic timepoints by gross necropsy and histopathology of associated vasculature performed by the Study Pathologist. No significant differences were observed in vascular response angiographically or pathologically between the test and control treated vessels. The histopathological findings were considered minimal and were not considered by the Study Pathologist to be clinically relevant.
### 7.4 Software
Software verification and validation testing and documentation for the subject Thunderbolt Aspiration Tubing was provided as per FDA Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions” (issued June 14, 2023).
### 7.5 Electrical Safety/EMC Testing
Electrical Safety and EMC testing were conducted on the subject Thunderbolt Aspiration Tubing. The subject device complies with the requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, AIM 7351731, ISO 10079-1, and ISO 10079-4.
### 8. Clinical Study
#### Summary
The Acute Ischemic Stroke Study with the Penumbra System® including Thunderbolt® Aspiration Tubing (THUNDER) was a prospective, single-arm, multicenter study designed to demonstrate the safety and effectiveness of the Penumbra System including Thunderbolt Aspiration Tubing (Thunderbolt) in a population presenting with acute ischemic stroke (AIS) secondary to intracranial large vessel occlusion (LVO) who are eligible for mechanical neurothrombectomy. The study was conducted at 29 sites in the United States.
#### Patient Population
Inclusion Criteria
1. Patient age 18 - 80 years.
2. Treatment within 8 hours of symptom onset.
3. Pre-stroke modified Rankin Scale (mRS) 0-2.
4. Patient experiencing acute ischemic stroke secondary to intracranial large vessel occlusion in the internal carotid, middle cerebral (M1 or M2), basilar, or vertebral artery.
5. If patient is eligible for thrombolytic therapy, it was administered as soon as possible, and no later than 3 hours from stroke symptom onset.
6. Planned frontline treatment with Penumbra System including Thunderbolt modulated aspiration.
7. Informed consent obtained per Institution Review Board/Ethics Committee requirements.
Page 8 of 20
{12}
Penumbra
K250690
# Exclusion Criteria
1. Stenosis, excessive tortuosity, or any occlusion in a proximal vessel requiring treatment or preventing access to the thrombus.
2. Alberta Stroke Program Early Computed Tomography Score (ASPECTS) ≤6 or core infarct volume >50 mL on magnetic resonance imaging (MRI) or computed tomography (CT)-based imaging (for anterior circulation strokes).
3. Pregnant patient.
4. Life expectancy < 90 days due to comorbidities.
5. Current participation in an interventional drug or device study that may confound the results of this study. Studies requiring extended follow-up for products that were investigational but have since become commercially available are not considered investigational studies.
6. Other medical, behavioral, or psychological conditions that in the opinion of the Investigator could limit the patient's ability to participate in the study, including compliance with follow-up requirements, or that could impact the scientific integrity of the study.
# Analysis populations
A total of 216 participants were enrolled from August 2022 to September 2024. The Intent-to-Treat (ITT) population includes all participants who were enrolled. Participants were considered enrolled once they or their representative have provided informed consent and all eligibility criteria have been confirmed regardless of whether or not Thunderbolt was used. Participants who were enrolled but did not receive intervention with the Penumbra System including Thunderbolt were followed out to 72 hours of enrollment or discharge (whichever comes first). One participant in the ITT population was not treated with the Thunderbolt Aspiration Tubing because the initial Penumbra Aspiration Pump used did not power on and this participant did not have procedural data collected, including baseline modified Thrombolysis in Cerebral Infarction (mTICI) score. As pre-specified in the clinical protocol, this participant was followed out to 72 hours of enrollment or discharge (whichever came first) for safety.
The per-protocol (PP) population is defined as a subset of the ITT population. The PP population includes all participants that received intervention and did not have major protocol deviations. The eligibility criteria were not met in 13 participants, and Thunderbolt was not used in 2 participants [the Penumbra Aspiration Pump would not turn on (n=1 participant) and the Thunderbolt Aspiration Tubing could not be primed and continuous aspiration was used (n=1 participant)]. The remaining 201 participants were included in the Per-Protocol (PP) analysis. This is a secondary analysis population for the primary effectiveness outcome.
For the THUNDER study ITT population, the rate of participants with baseline mTICI 0-1 was 84.7% (183/216), baseline mTICI 2a was 11.6% (25/216), and baseline mTICI 2b-3 was 3.7% (8/216). For the Penumbra 3D study used to derive the performance goal for the THUNDER study, the rate of participants with baseline mTICI 0-1 was 93.9% (185/197), baseline mTICI 2a was 5.1% (10/197) of participants, and baseline mTICI 2b-3 was 1.0% (2/197) of participants. Additional analyses of the effectiveness endpoints
Page 9 of 20
{13}
Penumbra
K250690
were also conducted for the population of THUNDER study participants with core lab adjudicated baseline mTICI 0-1 (n=183).
### Participant Follow-up
Participants were assessed at baseline, during the procedure, 24 hours, 7-10 days or discharge, 30 days, and 90 days post-procedure. As outlined below, the 24-hour post-procedure assessment was completed in 99.1% (214/216) and the final 90-day follow-up visit was completed in 83.8% (181/216) of the ITT population.
Participant Follow-Up Visits, ITT Population
| Participant follow-up visits | All Participants (N=216) |
| --- | --- |
| Baseline Evaluation | 100.0% (216/216) |
| Procedure | 99.5% (215/216) |
| 24 Hours Follow-Up (± 12 hours) | 99.1% (214/216) |
| 7-10 Days Follow-Up and/or Discharge (whichever comes first) | 94.9% (205/216) |
| 30 Days Follow-Up (± 7 days) | 87.0% (188/216) |
| 90 Days Follow-Up (± 14 days) | 83.8% (181/216) |
| Subject completed the study and alive^{1} | 84.3% (182/216) |
| Discontinued from the study | |
| Death | 11.1% (24/216) |
| Lost to Follow-Up | 4.2% (9/216) |
| Subject Withdrew Consent | 0.5% (1/216) |
| ^{1}One participant did not receive intervention with the Penumbra System including Thunderbolt and completed the study at 72 hours post-procedure | |
Baseline Demographics, ITT Population
| Baseline Demographics | All Participants (N=216) |
| --- | --- |
| Age | N = 216 |
| Mean (SD) | 64.8 (11.55) |
| Median [IQR] | 68.0 [56.0, 74.5] |
| Sex at birth | % (n/N) |
| Female | 45.4% (98/216) |
| Male | 54.6% (118/216) |
| Ethnicity | % (n/N) |
| Hispanic or Latino | 10.6% (23/216) |
| Not Hispanic or Latino | 78.2% (169/216) |
| Not Reported | 11.1% (24/216) |
| Race^{1} | % (n/N) |
| American Indian or Alaska Native | 0.5% (1/216) |
| Asian | 1.4% (3/216) |
| Black | 18.5% (40/216) |
| Native Hawaiian or Other Pacific Islander | 0.5% (1/216) |
| White | 74.1% (160/216) |
| Other | 2.3% (5/216) |
| Unknown | 0.9% (2/216) |
| Not Reported | 1.9% (4/216) |
| Height, Weight, Body Mass Index (BMI) and Blood Pressure (BP) | Mean (SD) |
| Height (cm) | 171.5 (10.27) |
| Weight (kg) | 87.8 (22.21) |
Page 10 of 20
{14}
Penumbra
K250690
| Baseline Demographics | All Participants (N=216) |
| --- | --- |
| BMI (kg/m^{2}) | 29.8 (6.96) |
| Systolic BP (mmHg) | 144.5 (24.88) |
| Diastolic BP (mmHg) | 82.3 (16.43) |
| **Pre-stroke mRS** | **% (n/N)** |
| 0 | 80.1% (173/216) |
| 1 | 11.6% (25/216) |
| 2 | 7.9% (17/216) |
| 4 | 0.5% (1/216) |
| 0 – 2 | 99.5% (215/216) |
| **Occlusion Location^{4}** | **% (n/N)** |
| **Anterior Circulation** | 96.7% (208/215) |
| Occlusion side | |
| Left | 43.7% (94/215) |
| Right | 53.0% (114/215) |
| Occlusion location | |
| ICA Origin | 0.5% (1/215) |
| ICA Petrous | 0.5% (1/215) |
| ICA Supraclinoid | 2.3% (5/215) |
| Carotid T | 6.0% (13/215) |
| M1 | 59.5% (128/215) |
| M2 | 27.4% (59/215) |
| Other^{2} | 0.5% (1/215) |
| **Posterior Circulation** | 3.3% (7/215) |
| Occlusion side | |
| Left | 0.0% (0/215) |
| Right | 0.5% (1/215) |
| Midline/Basilar | 2.8% (6/215) |
| Occlusion location | |
| Basilar | 2.8% (6/215) |
| Other^{3} | 0.5% (1/215) |
| **mTICI at Baseline per Core Lab^{4}** | **% (n/N)** |
| Grade 0-1 | 84.7% (183/216) |
| Grade 2a | 11.6% (25/216) |
| Grade 2b-3 | 3.7% (8/216) |
| **Baseline ASPECT score^{5}** | **N = 205** |
| Mean (SD) | 9.2 (1.00) |
| Median [IQR] | 10.0 [8.0, 10.0] |
| Minimum, Maximum | 7.0, 10.0 |
| **Baseline PC ASPECT score^{5}** | **N = 7** |
| Mean (SD) | 9.9 (0.38) |
| Median [IQR] | 10.0 [10.0, 10.0] |
| Minimum, Maximum | 9.0, 10.0 |
| **Baseline NIHSS Score** | **N = 216** |
| Mean (SD) | 14.4 (6.84) |
| Median [IQR] | 15.0 [9.0, 20.0] |
| Minimum, Maximum | 0.0, 38.0 |
| **Time from Stroke Onset to Arterial Puncture (minutes)^{4}** | **N = 215** |
| Mean (SD) | 219 (108) |
| Median [IQR] | 184 [143, 277] |
Page 11 of 20
{15}
Penumbra
K250690
| Baseline Demographics | All Participants (N=216) |
| --- | --- |
| ¹Participants who selected the option Asian Indian, Chinese, Japanese, Korean, Vietnamese, Other Asian are categorized under Asian. ²Anterior circulation Left Occlusion side in the Occlusion Location: Other, specify – Lt ICA/M1. ³Posterior circulation Right Occlusion side in the Occlusion Location: Other, specify – P1, Pre communicating segment. ⁴One participant did not receive treatment with the Penumbra System including Thunderbolt. This participant did not have procedure data reported and was assigned baseline mTICI Grade 0-1. ⁵Three participants were missing ASPECT score. Data entry is completed for the three participants on core infarct size based on perfusion imaging. Each participant is required to have ASPECT/PC ASPECT or core infarct size measurement. | |
### Devices used
Thunderbolt was used at the site of primary target vessel occlusion in 99.1% (214/216) of participants in the ITT population. Of the 216 ITT participants, 211 participants had modulated aspiration used in the first pass. Of the 211 participants with modulated aspiration used in the first pass, 61 participants (28.9%) had one or more continuous aspiration passes performed with Thunderbolt after the first pass with modulated aspiration.
The reperfusion catheter used with Thunderbolt at the site of the primary target vessel occlusion at any pass was RED 62, RED 68 and RED 72 in 40.5% (87/215), 36.7% (79/215), and 38.6% (83/215) of participants, respectively. Note that participants could have been treated with more than 1 model of the RED catheter during the procedure.
Procedural Characteristics, ITT Population
| Procedural characteristics | All Participants (N=216) |
| --- | --- |
| Use of additional therapy During Procedure | % (n/N) |
| Any additional therapy¹ | 14.4% (31/215) |
| Non-Penumbra aspiration catheter | 3.3% (7/215) |
| Stent retriever | 9.3% (20/215) |
| Penumbra 3D Revascularization Device | 1.4% (3/215) |
| Balloon angioplasty | 1.9% (4/215) |
| Stent | 3.3% (7/215) |
| Time from Arterial Puncture to mTICI 2b-3 otherwise to Final Angiogram (minutes) per Principal Investigator (PI) (mTICI assessment per Core Lab)² | N = 214 |
| Mean (SD) | 27 (26) |
| Median [IQR] | 20 [13, 29] |
| Minimum, Maximum | 4, 240 |
| Time from Initial Aspiration Start to mTICI 2b-3 otherwise to Final Angiogram (minutes)² | N = 214 |
| Mean (SD) | 11 (21) |
| Median [IQR] | 4 [2, 10] |
| Minimum, Maximum | 0, 189 |
| Time from Arterial Puncture to Final Angiogram (minutes)² | N = 214 |
| Mean (SD) | 37 (31) |
| Median [IQR] | 27 [17, 46] |
| Minimum, Maximum | 5, 240 |
| Total number of passes | N = 216 |
| Mean (SD) | 2.3 (1.82) |
| Median [IQR] | 2.0 [1.0, 3.0] |
| Minimum, Maximum | 1.0, 13.0 |
Page 12 of 20
{16}
Penumbra
K250690
| Procedural characteristics | All Participants (N=216) |
| --- | --- |
| Thrombolytic Therapy | % (n/N) |
| Received thrombolytic therapy prior to procedure | 56.9% (123/216) |
| Bridging protocol | 93.5% (115/123) |
| Waiting for improvement | 6.5% (8/123) |
| ^{1}One participant did not receive intervention with the Penumbra System including Thunderbolt and did not report devices used. Participants can have more than one type of additional therapy used during procedure. Additional therapies could have been utilized for reasons other than to achieve successful revascularization of the target vessel. ^{2}Two participants did not have an end of procedure angiographic outcome per Core Lab. One participant did not receive intervention with the Penumbra System including Thunderbolt and did not collect angiographic images; one participant had angiographic images but had “Unable to Assess” assessment for final angiogram. | |
## Results
### Effectiveness
The primary effectiveness endpoint was angiographic revascularization of the occluded target vessel at immediate post-Penumbra procedure as defined by mTICI 2b or higher. A performance goal of 71.5% was pre-defined for the THUNDER study derived from the 95% lower confidence bound of the rate of mTICI 2b or higher in the full ITT population (both study arms) of the Penumbra 3D trial. An exact binomial test, conducted at the 0.025 one-sided significance level, was used to test the null hypothesis that the population proportion of patients with post-Penumbra revascularization success is less than or equal to the performance goal of 71.5%, against the alternative hypothesis that the proportion is greater than 71.5%.
The primary effectiveness endpoint of angiographic revascularization (mTICI 2b-3) of the occluded target vessel at immediate post-Penumbra was achieved in 87.5% (189/216) of participants. The lower bound of the 95% confidence interval was 82.3%, rejecting the null hypothesis that the study rate is less than or equal to 71.5% (p-value: <0.001). Therefore, THUNDER met its pre-defined performance goal for the primary effectiveness endpoint.
The following supplementary analyses were also conducted: (1) imputing participants with first pass continuous (non-modulated) aspiration as failures; (2) imputing participants with first pass continuous (non-modulated) aspiration, the number of RED catheter passes >5, or any adjunctive device use as failures. These analyses resulted in success rates of 86.6% (187/216), and 78.7% (170/216) with corresponding lower bounds of the 95% confidence intervals of 81.3%, and 72.6%, respectively. These were all greater than the performance goal of 71.5%.
Page 13 of 20
{17}
Penumbra
K250690
# Primary Effectiveness Endpoint, ITT Population
| Primary endpoint success | ITT Participants, (95% CI)^{1} (N=216) |
| --- | --- |
| mTICI 2b-3 Post-Penumbra per Core Lab^{2} | 87.5% (189/216) (82.3%, 91.6%) |
| mTICI 2b-3 Post-Penumbra per Core Lab^{2}, imputing participants with first pass continuous (non-modulated) aspiration as failure | 86.6% (187/216) (81.3%, 90.8%) |
| mTICI 2b-3 Post-Penumbra per Core Lab^{2}, imputing participants with first pass continuous (non-modulated) aspiration, number of RED catheter passes >5, or any use of adjunctive treatment (3D Revascularization Device, non-Penumbra aspiration catheter, stent retriever, balloon angioplasty, stent) as failure | 78.7% (170/216) (72.6%, 84.0%) |
| Note: Participants not treated with Thunderbolt are analyzed as endpoint failure in all mTICI analyses. ^{1}Clopper Pearson Binomial method is used for proportion 95% CI. ^{2}Success is revascularization of the occluded target vessel defined by mTICI score of 2b to 3 following use of the Penumbra System (immediate post-Penumbra) and prior to the use of any additional therapies or adjunctive devices. Participants with intracranial stenosis/atherosclerosis (ICAD) treated with stenting or angioplasty at primary target vessel occlusion location prior to achieving mTICI 2b or greater are analyzed as effectiveness endpoint failures. | |
Additional analyses of the primary effectiveness endpoint were also conducted in the PP population. See the table below.
# Primary Effectiveness Endpoint, PP Population
| Primary endpoint success | PP Participants, (95% CI)^{1} (N=201) |
| --- | --- |
| mTICI 2b-3 Post-Penumbra per Core Lab^{2} | 89.1% (179/201) (83.9%, 93.0%) |
| mTICI 2b-3 Post-Penumbra per Core Lab^{2}, imputing participants with first pass continuous (non-modulated) aspiration as failure | 88.6% (178/201) (83.3%, 92.6%) |
| mTICI 2b-3 Post-Penumbra per Core Lab^{2}, imputing participants with first pass continuous (non-modulated) aspiration, number of RED catheter passes >5, or any use of adjunctive treatment (3D Revascularization Device, non-Penumbra aspiration catheter, stent retriever, balloon angioplasty, stent) as failure | 80.6% (162/201) (74.4%, 85.8%) |
| Note: Participants not treated with Thunderbolt are analyzed as endpoint failure in all mTICI analyses. ^{1}Clopper Pearson Binomial method is used for proportion 95% CI. ^{2}Success is revascularization of the occluded target vessel defined by mTICI score of 2b to 3 following use of the Penumbra System (immediate post-Penumbra) and prior to the use of any additional therapies or adjunctive devices. Participants with intracranial stenosis/atherosclerosis (ICAD) treated with stenting or angioplasty at primary target vessel occlusion location prior to achieving mTICI 2b or greater are analyzed as effectiveness endpoint failures. | |
Page 14 of 20
{18}
Penumbra
K250690
Primary Effectiveness Endpoint, mTICI 0-1 at Baseline Population
| Primary endpoint success | mTICI 0-1 at Baseline Participants, (95% CI)1(N=183) |
| --- | --- |
| mTICI 2b-3 Post-Penumbra per Core Lab2 | 86.9% (159/183) (81.1%, 91.4%) |
| mTICI 2b-3 Post-Penumbra per Core Lab2, imputing participants with first pass continuous (non-modulated) aspiration as failure | 85.8% (157/183) (79.9%, 90.5%) |
| mTICI 2b-3 Post-Penumbra per Core Lab2, imputing participants with first pass continuous (non-modulated) aspiration, number of RED catheter passes >5, or any use of adjunctive treatment (3D Revascularization Device, non-Penumbra aspiration catheter, stent retriever, balloon angioplasty, stent) as failure | 77.0% (141/183) (70.3%, 82.9%) |
Note: Participants not treated with Thunderbolt are analyzed as endpoint failure in all mTICI analyses. One participant without procedure data reported was assigned baseline mTICI Grade 0-1.
\( ^{1} \) Clopper Pearson Binomial method is used for proportion 95% CI.
\( ^{2} \) Success is revascularization of the occluded target vessel defined by mTICI score of 2b to 3 following use of the Penumbra System (immediate post-Penumbra) and prior to the use of any additional therapies or adjunctive devices. Participants with intracranial stenosis/atherosclerosis (ICAD) treated with stenting or angioplasty at primary target vessel occlusion location prior to achieving mTICI 2b or greater are analyzed as efficacy endpoint failures.
Other effectiveness outcomes for the ITT population are provided in the table below.
After the first pass, mTICI 2b or higher per Core Lab was achieved in 64.8% (140/216).
Post-Penumbra, mTICI 2c or higher per Core Lab was achieved in 70.4% (152/216) of participants. At the end of the procedure, mTICI 2b or higher per Core Lab was achieved in 96.3% (206/214). Two participants did not have an end of procedure angiographic outcome per Core Lab. The median time to revascularization was 20 minutes. At 90 days, 55.4% (112/202) of participants had mRS 0-2.
Other Effectiveness Outcomes, ITT Population
| | ITT Participants, (95% CI)1(N=216) |
| --- | --- |
| Angiographic outcomes | |
| Angiographic revascularization of the occluded target vessel after first pass as defined by mTICI 2b or higher per Core Lab2,3 | 64.8% (140/216) (55.5%, 73.4%) |
| Angiographic revascularization of the occluded target vessel Post-Penumbra as defined by mTICI 2c or higher per Core Lab2,3 | 70.4% (152/216) (61.3%, 78.4%) |
| Angiographic revascularization of the occluded target vessel at end of procedure as defined by mTICI 2b or higher per Core Lab4 | 96.3% (206/214) |
| Time to revascularization (time from arterial puncture to mTICI 2b-3 otherwise to final angiogram, minutes) | |
| N4 | 214 |
| Mean (SD) (95% CI)3,5 | 27 (26) (59, 122) |
| Median [IQR] | 20 [13, 29] |
| Range (Min, Max) | 4, 240 |
| Clinical outcome | |
| Modified Rankin Scale (mRS) 0-2 at 90 days3,6 | 55.4% (112/202) (45.7%, 64.9%) |
Page 15 of 20
{19}
Penumbra
K250690
| | ITT Participants, (95% CI)^{1} (N=216) |
| --- | --- |
| ^{1}Clopper Pearson Binomial method is used for proportion 95% CI. ^{2}Participants not treated with Thunderbolt are analyzed as endpoint failure. ^{3}Confidence intervals are adjusted for multiplicity using the Bonferroni method (k=7). ^{4}Two participants did not have an end of procedure angiographic outcome per Core Lab. One participant did not receive intervention with the Penumbra System including Thunderbolt and did not collect angiographic images; one participant had angiographic images but had “Unable to Assess” assessment for final angiogram. ^{5}Wald normal approximation to the binomial distribution confidence intervals are presented. ^{6}Denominator is out of all completed mRS 90-day follow-up assessments and includes reported deaths. Missing data for 90-day mRS includes 9 participants lost-to-follow-up, 1 participant withdrew consent, 3 participants with missed mRS assessment at 90 days. One participant did not receive intervention with the Penumbra System including Thunderbolt and completed the study at 72 hours. | |
Other effectiveness outcomes for the PP population are provided in the table below. After the first pass, mTICI 2b or higher per Core Lab was achieved in 65.2% (131/201). Post-Penumbra, mTICI 2c or higher per Core Lab was achieved in 71.6% (144/201) of participants. At the end of the procedure, mTICI 2b or higher per Core Lab was achieved in 96.0% (192/200). One participant did not have an end of procedure angiographic outcome per Core Lab. The median time to revascularization was 19 minutes. At 90 days, 55.9% (105/188) of participants had mRS 0-2.
Other Effectiveness Outcomes, PP Population
| | PP Participants, (95% CI)^{1} (N=201) |
| --- | --- |
| Angiographic outcomes | |
| Angiographic revascularization of the occluded target vessel after first pass as defined by mTICI 2b or higher per Core Lab^{2} | 65.2% (131/201) (55.6%, 74.0%) |
| Angiographic revascularization of the occluded target vessel Post-Penumbra as defined by mTICI 2c or higher per Core Lab^{2} | 71.6% (144/201) (62.3%, 79.8%) |
| Angiographic revascularization of the occluded target vessel at end of procedure as defined by mTICI 2b or higher per Core Lab^{3} | 96.0% (192/200) |
| Time to revascularization (time from arterial puncture to mTICI 2b-3 otherwise to final angiogram, minutes) | |
| N^{1} | 200 |
| Mean (SD) (95% CI)^{2,4} | 27 (26) (59, 151) |
| Median [IQR] | 19 [13, 29] |
| Range (Min, Max) | 4, 240 |
| Clinical outcome | |
| Modified Rankin Scale (mRS) 0-2 at 90 days^{2,5} | 55.9% (105/188) (45.8%, 65.6%) |
| ^{1}Clopper Pearson Binomial method is used for proportion 95% CI. ^{2}Confidence intervals are adjusted for multiplicity using the Bonferroni method (k=7). ^{3}One participant had “Unable to Assess” assessment for final angiogram. ^{4}Wald normal approximation to the binomial distribution confidence intervals are presented. ^{5}Denominator is out of all completed mRS 90-day follow-up assessments and includes reported deaths. Missing data for 90-day mRS includes 8 participants lost-to-follow-up, 1 participant withdrew consent, 3 participants with missed mRS assessment at 90 days. One participant did not receive treatment with the Penumbra System including Thunderbolt and completed the study at 72 hours. | |
Other Effectiveness Outcomes, mTICI 0-1 at Baseline Population
| | mTICI 0-1 at Baseline Participants, (95% CI)^{1} (N=183) |
| --- | --- |
| Angiographic outcomes | |
| Angiographic revascularization of the occluded target vessel after first pass as defined by mTICI 2b or higher per Core Lab^{2} | 60.1% (110/183) (49.9%, 69.7%) |
Page 16 of 20
{20}
Penumbra
K250690
| | mTICI 0-1 at Baseline Participants, (95% CI)^{1} (N=183) |
| --- | --- |
| Angiographic revascularization of the occluded target vessel Post-Penumbra as defined by mTICI 2c or higher per Core Lab^{2} | 68.3% (125/183) (58.3%, 77.2%) |
| Angiographic revascularization of the occluded target vessel at end of procedure as defined by mTICI 2b or higher per Core Lab^{3} | 95.6% (173/183) |
| Time to revascularization (time from arterial puncture to mTICI 2b-3 otherwise to final angiogram, minutes) | |
| N^{1} | 181 |
| Mean (SD) (95% CI)^{2,4} | 28 (28) (66, 122) |
| Median [IQR] | 21 [13, 30] |
| Range (Min, Max) | 6, 240 |
| Clinical outcome | |
| Modified Rankin Scale (mRS) 0-2 at 90 days^{2,5} | 54.1% (92/170) (46.3%, 61.8%) |
| Note: One participant without procedure data reported was assigned baseline mTICI Grade 0-1. ^{1}Clopper Pearson Binomial method is used for proportion 95% CI. ^{2}Confidence intervals are adjusted for multiplicity using the Bonferroni method (k=7). ^{3}One participant had “Unable to Assess” assessment for final angiogram. One participant did not receive treatment with the Penumbra System including Thunderbolt. ^{4}Wald normal approximation to the binomial distribution confidence intervals are presented. ^{5}Denominator is out of all completed mRS 90-day follow-up assessments and includes reported deaths. Missing data for 90-day mRS includes 8 participants lost-to-follow-up, 1 participant withdrew consent, 3 participants with missed mRS assessment at 90 days. One participant did not receive treatment with the Penumbra System including Thunderbolt and completed follow-up at 72 hours. | |
## Safety
The safety of the Penumbra System including the Thunderbolt Aspiration Tubing was evaluated through a review of adverse events (AEs) recorded during the study. A summary of key safety metrics is provided below.
In comparing THUNDER with the Penumbra 3D study, the rate of participants that had additional passes due to additional clot in distal vasculature was 20.8% (45/216) in THUNDER and 16.7% (33/198) in the Penumbra 3D study. The rate of serious adverse events (SAEs) classified as embolization to a new territory (ENT) per Clinical Events Committee (CEC) was 0.5% (1/216) in THUNDER. ENT was not assessed by the CEC in the Penumbra 3D study.
Key Safety Outcomes, ITT Population
| | ITT Participants, (95% CI)^{1} (N=216) |
| --- | --- |
| Occurrence of symptomatic intracranial hemorrhages (sICH) at 24 Hours per CEC | 0.9% (2/216) (0.0%, 4.4%) |
| All-cause mortality at 90 days^{2} | 11.7% (24/205) (6.4%, 19.0%) |
| Incidence of device related, and/or procedure related Serious Adverse Events (SAEs) within 24 hours post-procedure per CEC^{3} | 2.8% (6/216) (0.7%, 7.3%) |
Page 17 of 20
{21}
Penumbra
K250690
| | ITT Participants, (95% CI)^{1} (N=216) |
| --- | --- |
| ^{1}Clopper Pearson Binomial method is used for proportion 95% CI. Confidence intervals are adjusted for multiplicity using the Bonferroni method (k=7). ^{2}Denominator is out of all completed mRS 90-day follow-up assessments and includes reported deaths. Missing data for 90-day mRS includes 9 participants lost-to-follow-up, 1 participant withdrew consent, 3 participants with missed mRS assessment at 90 days. One participant did not receive treatment with the Penumbra System including Thunderbolt and completed follow-up at 72 hours. ^{3}Includes symptomatic intracerebral hemorrhage, embolization to a new territory, and intracranial vascular injury (e.g. dissection or perforation). | |
Images were reviewed and imaging findings of ICH within 24 hours – irrespective of clinical significance – were identified by the core lab and classified using ECASS classification.
ICH by ECASS Classification (Core Lab), ITT Population
| | ITT Participants (N=214) |
| --- | --- |
| Total Intracranial Hemorrhage within 24 hours per Core Lab | 26.6% (57/214) |
| IPH: HI-2 | 8.4% (18/214) |
| IPH: HI-1 | 6.1% (13/214) |
| SAH | 4.7% (10/214) |
| IPH: HI-2 + SAH | 2.8% (6/214) |
| IPH: PH-1 | 0.9% (2/214) |
| IPH: PH-1 + IVH + SAH | 0.9% (2/214) |
| IPH: HI-1 + SAH | 0.5% (1/214) |
| IPH: HI-2 + IVH + SAH | 0.5% (1/214) |
| RIH | 0.5% (1/214) |
| IPH: PH-2 + SAH | 0.5% (1/214) |
| IPH: PH-2 + IVH | 0.5% (1/214) |
| IPH: PH-2 | 0.5% (1/214) |
| Note: Two participants with missing 24-hour imaging are excluded from intracranial hemorrhage summary. | |
Adverse events categorized under the nervous system disorders MedDRA System Organ Class are summarized below for the ITT population.
MedDRA Classification of Adverse Events, Nervous System Disorders, ITT Population
| MedDRA classification (Participants, N = 216)^{1} | | |
| --- | --- | --- |
| | Events (participants) | Rate of participants with event |
| Nervous system disorders | 140 (105) | 48.6% |
| Basal ganglia haemorrhage | 2 (2) | 0.9% |
| Brain compression | 2 (2) | 0.9% |
| Brain oedema | 7 (7) | 3.2% |
| Carotid artery dissection | 2 (2) | 0.9% |
| Carotid artery occlusion | 2 (2) | 0.9% |
| Carotid artery stenosis | 1 (1) | 0.5% |
| Cerebellar infarction | 1 (1) | 0.5% |
| Cerebellar stroke | 1 (1) | 0.5% |
Page 18 of 20
{22}
Penumbra
K250690
| Cerebral artery embolism | 3 (3) | 1.4% |
| --- | --- | --- |
| Cerebral artery occlusion | 5 (5) | 2.3% |
| Cerebral artery perforation | 1 (1) | 0.5% |
| Cerebral artery stenosis | 1 (1) | 0.5% |
| Cerebral artery thrombosis | 1 (1) | 0.5% |
| Cerebral haematoma | 1 (1) | 0.5% |
| Cerebral haemorrhage | 13 (13) | 6.0% |
| Cerebral infarction | 7 (7) | 3.2% |
| Cerebral microangiopathy | 1 (1) | 0.5% |
| Cerebral vasoconstriction | 9 (9) | 4.2% |
| Cerebrovascular accident | 2 (2) | 0.9% |
| Dysarthria | 1 (1) | 0.5% |
| Encephalopathy | 2 (2) | 0.9% |
| Haemorrhage intracranial | 3 (3) | 1.4% |
| Haemorrhagic transformation stroke | 13 (13) | 6.0% |
| Headache | 16 (15) | 6.9% |
| Hemiparesis | 1 (1) | 0.5% |
| Hyperaesthesia | 1 (1) | 0.5% |
| Intracranial artery dissection | 1 (1) | 0.5% |
| Ischaemic cerebral infarction | 1 (1) | 0.5% |
| Ischaemic stroke | 2 (2) | 0.9% |
| Metabolic encephalopathy | 3 (3) | 1.4% |
| Neuropathy peripheral | 1 (1) | 0.5% |
| Partial seizures | 1 (1) | 0.5% |
| Restless legs syndrome | 1 (1) | 0.5% |
| Seizure | 7 (7) | 3.2% |
| Stroke in evolution | 15 (15) | 6.9% |
| Subarachnoid haemorrhage | 5 (5) | 2.3% |
| Syncope | 2 (2) | 0.9% |
| Toxic encephalopathy | 2 (2) | 0.9% |
| ¹MedDRA reflects site-reported adverse events. Therefore, some core-lab identified ICH findings (which are imaging-based, and irrespective of clinical significance) may not be captured as MedDRA events. Site-reported hemorrhagic events, as coded using MedDRA terminology, are independent of core lab imaging assessments and do not necessarily correspond to core-lab identified ICH. | | |
### THUNDER Study Conclusions
The THUNDER study demonstrated that the Penumbra System including Thunderbolt Aspiration Tubing has substantially equivalent safety and effectiveness for the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusion compared to the predicate. The THUNDER study met its prespecified primary effectiveness endpoint.
### 9. Conclusions
The subject Penumbra System (Thunderbolt Aspiration Tubing) is substantially equivalent to the predicate Penumbra System (Modified 110 Aspiration Tubing). The subject device has the same intended use and similar operating principle, design concept, and fundamental technology as the predicate. The differences between the subject and predicate devices do not raise new or different questions of safety or effectiveness. The
Page 19 of 20
{23}
Penumbra
K250690
device testing described in this 510(k) Summary demonstrates the subject device performs as intended.
Page 20 of 20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.