The FreeStyle Comfort Oxygen Concentrator can deliver concentrated supplemental oxygen with either a fixed-minute pulse delivery or can be set to operate using autoSAT pulse delivery. The FreeStyle Comfort is used on a prescriptive basis by patients requiring supplemental oxygen to increase blood oxygen saturation. The device is designed for use in the home, in a vehicle, and for use during air travel. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
Device Story
Portable oxygen concentrator (POC) delivering supplemental oxygen via nasal cannula; utilizes Pressure Swing Adsorption (PSA) to extract oxygen from ambient air using molecular sieve beds; features autoSAT pulse delivery (automatic titration to respiratory rate) or fixed-minute pulse delivery; detects inhalation via pressure sensors to trigger oxygen bolus; powered by lithium-ion battery or AC/DC external sources; operated by patient via membrane keypad; includes Bluetooth connectivity for myCAIRE app to transmit usage data to home care providers; intended for home, institution, and transport use; benefits include increased portability, ambulation, and battery life compared to continuous flow systems; provides visual/audible alarms for low concentration, system faults, and flow obstructions.
Clinical Evidence
Bench testing only. Performance verified against ISO 80601-2-69 and ISO 80601-2-67 standards. Testing included pulse volume, trigger sensitivity, oxygen concentration accuracy, alarm functionality, electrical safety, EMC, and biocompatibility (ISO 18562-2/3).
Technological Characteristics
Materials: UL/CSA-approved PC/ABS plastic enclosure. Technology: Pressure Swing Adsorption (PSA) with dual molecular sieve beds. Power: 11-18 VDC or 100-240 VAC. Connectivity: Bluetooth (BLE) for myCAIRE app. Dimensions: 10.0-11.0 x 7.3 x 3.1 inches. Weight: 5-6 lbs. Software: Embedded control board for PSA cycle and pulse dosing. Sterilization: Not required.
Indications for Use
Indicated for patients requiring supplemental oxygen to increase blood oxygen saturation on a prescriptive basis. Used in home, institution, and transport modes. Not for newborn, infant, or pediatric populations.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 11, 2025
CAIRE Inc.
Ted Vlahopoulos
Sr. Regulatory Specialist
2200 Airport Industrial Drive, Ste 500
Ball Ground, Georgia 30107
Re: K250671
Trade/Device Name: FreeStyle Comfort (AS200 / FreeStyle Comfort)
Regulation Number: 21 CFR 868.5440
Regulation Name: Portable Oxygen Generator
Regulatory Class: Class II
Product Code: CAW
Dated: May 8, 2025
Received: June 30, 2025
Dear Ted Vlahopoulos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K250671 - Ted Vlahopoulos
Page
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
3K250671 - Ted Vlahopoulos
Page
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250671 | |
| Device Name FreeStyle Comfort (AS200 / FreeStyle Comfort) | |
| Indications for Use (Describe) | |
| The FreeStyle Comfort Oxygen Concentrator can deliver concentrated oxygen with either a fixed-minute pulse delivery or can be set to operate using autoSAT pulse delivery. The FreeStyle Comfort is used on a prescriptive basis by patients requiring supplemental oxygen to increase blood oxygen saturation. FreeStyle Comfort may be used in the home, institution, and transport modes. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
FreeStyle Comfort - 510(k) Summary
## Submitter:
CAIRE Inc.
2200 Airport Industrial Drive, Ste. 500
Ball Ground, GA 30107
Phone: (770) 721-7700
Fax: (770) 721-7701
Registration Number: 3004972304
## Contact person:
Ted Vlahopoulos
Senior Regulatory Specialist
Mobile: 716-339-0656
Email: ted.vlahopoulos@caireinc.com
Date Prepared: May 12, 2025
Device Trade Name: FreeStyle® Comfort® Oxygen Concentrator
Common Name: Portable Oxygen Concentrator
Classification Name: Generator, Oxygen.
Regulation Number: 21 CFR 868.5440
Product Code: CAW
Device Class: Class II
Classification Panel: Anesthesiology and Respiratory Therapy Devices
## Predicate Devices:
| Predicate | Referenced Predicate |
| --- | --- |
| Inogen Rove™ 4 Portable Oxygen Concentrator | Omni 3 Portable Oxygen System |
| Product Code: CAW | Proprietary Name: eQuinox™ Oxygen System, CAIRE |
| 510(k) number: K222086 | Product Code: CAW |
| | 510(k) number: K120785 |
FreeStyle Comfort - 510(k) Summary
{5}
Page 2 of 15
FreeStyle Comfort – 510(k) Summary
# Device Description:
The FreeStyle® Comfort® is a reusable, lightweight portable oxygen concentrator weighing 5 lbs. (2.3 kg) with the lithium-ion single battery pack, and 6 lbs. (2.7 kg) with the optional lithium-ion double battery pack.
The FreeStyle Comfort's light weight and portability allow the user to receive supplemental oxygen while traveling, shopping or while at home. The FreeStyle® Comfort® complies with FAA guidelines for use on-board commercial aircraft.
The FreeStyle® Comfort® oxygen concentrator has 5 flow settings delivering up to 1050 mL of concentrated oxygen per minute. The subject device has two delivery options, autoSAT® Pulse or Fixed Minute Pulse, providing 90% (+5.5% / -3%) of concentrated oxygen. Like the predicate device, the FreeStyle® Comfort® releases a pulse of oxygen (bolus) at the beginning of inhalation by using oxygen conserving technologies. The oxygen conserving technology detects inhalation and quickly delivers a pulse of oxygen to the user. This method of oxygen delivery increases device portability, ambulation, and battery life for extended use away from the home, as compared with continuous O2 flow oxygen concentrators.
The outer enclosure of the FreeStyle® Comfort® is made of UL/CSA-approved PC/ABS plastic. The innerworkings include two alloy cylinders, each filled with molecular sieve (referred to as sieve beds), an internal piston air compressor, valves, tubing assembly, electrical wiring harness, filters, and control board. FreeStyle® Comfort® does not require sterilization and is not a single use device.
# Operating Principles
The control board uses software to operate the unit. The device operating principle is a process known as Pressure Swing Adsorption, or PSA. Air is drawn into the device through the air intakes to the air compressor. Pressurized air flows from the air compressor to each of the (molecular) sieve beds in cycles. As one sieve bed is filled with pressurized air, the oxygen passes through the sieve bed. As the oxygen travels through the sieve bed, the nitrogen molecules collect on the molecular sieve. At the end of the cycle, the nitrogen is purged using enriched oxygen, then the process begins again with the alternate sieve bed (one sieve bed depressurizes while the other sieve bed is pressurized). This process continuously repeats from one sieve bed to the other, producing highly concentrated oxygen. Concentrated oxygen flows from the sieve beds to the product tank, then through sensors that measures both oxygen concentration and flow, then delivered to the patient.
# AutoSAT Feature
Traditional pulse dosing supplies the same amount of oxygen delivered to the patient per minute regardless of the number of breaths taken. If the patient inhales more often, such as during physical activity like walking, the amount of oxygen delivered decreases per breath to maintain the same volume of oxygen per minute. The autoSAT (automatic titration to patient's respiratory rate) function delivers a consistent pulse dose (bolus) volume of oxygen (up to 40 breaths per minute). As breath rate increases during physical activity, the amount of oxygen delivered will stay the same per breath, which increases the volume of oxygen delivered per minute.
The autoSAT function is a default setting in the FreeStyle Comfort. The user can change the setting to traditional dosing by using the membrane touchpad to switch off the autoSAT function. During normal operation, if after more than 15 seconds the device does not detect breath inhalation, the device will automatically begin dosing @ 20 breaths per minute until breath inhalation is detected. Upon detection, the device will resume normal operation in autoSAT mode.
# Inhalation Sensitivity
The FreeStyle® Comfort® offers an inhalation sensitivity mode. The user can adjust the sensitivity settings values from 1-5, where 1 is the least sensitive, 5 is the most sensitive. The sensitivity function has a default setting of > -0.1 cm H2O and a maximum adjustable range of -0.3 to -0.40 cm H2O, negative pressure (inhalation) to trigger the flow (bolus) of O2 to the user. In cases where shallow breathing cannot be detected at the device (default) sensitivity setting, the user can adjust the setting with the membrane keypad to increase inhalation sensitivity to trigger oxygen delivery at the desired flow setting.
# Power Options
FreeStyle Comfort is powered with either a single or double lithium-Ion battery pack or with an external AC/DC power source (A/C wall outlet or DC vehicle). FreeStyle® Comfort® includes a built-in battery charger that will automatically recharge the battery pack when the unit is plugged into an external power source, except during use on a commercial aircraft. When an AC/DC power source is not readily available for battery recharge, the battery pack can easily be removed for quick replacement by pressing the battery release button and sliding the battery pack out from the device.
{6}
Device battery is capable of up to 8 hours of operation using the 8-cell battery, and 16 hours with the 16-cell battery, depending on the selected flow setting and user's breathing rate. The battery charge status is displayed on the keypad LCD during normal operation.
## User Interface
The touch membrane keypad display allows the user to operate the device. The user can select the desired pulse dose volume and view the operating status. Once the user completes the selection process, the display screen will enter "standby" mode and then dims after 20 seconds. Pressing any button on the keypad will wake the display screen from standby mode. Oxygen delivery from the device to the user is channelled via a nasal cannula/tubing assembly, manufacturer Salter Labs (SunMed, LLC.), P/N 1600 (K911740). The end user is required to follow the cannula manufacturer's instructions for length of use, maintenance, and replacement requirements.
## Keypad/Alarms
Operating information, battery, and operating status is displayed on the keypad LCD window. The subject device comes equipped with both visual and audible alarms.
The FreeStyle Comfort features the following visual and audio safety alarms:
- Low oxygen concentration
- System malfunction due to high temperature, high system pressure or compressor fault
- System malfunction due to low temperature, low pressure or fan failure
- Flow Rate failure due to obstruction of gas pathway (cannula blockage)
- Low battery (activates at 20% battery power and intensifies alarm at 5% battery power)
- Auto-dosing until warm up is complete and breath detection is activated
- After 15 seconds of no breath detected the unit shifts from autoSAT to auto-dose.
## Wireless Connectivity
The FreeStyle Oxygen Concentrator utilizes Bluetooth technology that pairs the portable oxygen concentrator to a mobile device or tablet using the myCAIRE App.
The myCAIRE™ app connects patients with home care providers (DME), allowing DMEs to view usage hours and flow settings of a connected CAIRE oxygen concentrator. The system uses wireless technology (BLE) to transmit information from the concentrator to a smart device, which then sends it to the home care provider. Home care providers can assist patients by addressing device notifications and supplying recommendations for troubleshooting remotely.
myCAIRE™ shares data from CAIRE oxygen concentrators without altering settings, performance, or therapy. It is not intended for monitoring, diagnosis, or treatment. The connection only sends data to the app in raw format, decodable solely by Caire, with no Protected Health Information or Personally Identifiable Information transmitted.
## Associated Accessories
The device comes equipped with a rechargeable 8-cell lithium-Ion battery pack, A/C power supply/cord, DC power supply. Optional equipment consists of a 16-cell lithium-Ion battery pack for extended use and the desktop battery charger.
The following components are included with this FreeStyle® Comfort® 510(k) submission:
- 8-cell Battery Pack, lithium-Ion
- A/C power supply/power cord and plug
- Carrying case/shoulder strap
- DC Power Supply
- User manual
- 16-cell Battery Pack, lithium-Ion
- Carry All Accessory Bag (stores the device and accessories)
- Desktop Battery Charger
- Backpack
- Spare Replacement Filters
The basic technology of the FreeStyle Comfort Oxygen Concentrator is equivalent to the FDA cleared oxygen concentrator, Inogen Rove 4 Portable Oxygen Concentrator, 510(k) reference K222086.
FreeStyle Comfort – 510(k) Summary
{7}
Photos:

FreeStyle® Comfort®

with Carrying Case

Control Panel Display/Keypad

Single & Double Battery Packs
FreeStyle Comfort – 510(k) Summary
{8}
| FreeStyle Comfort Specifications | |
| --- | --- |
| Dimensions (with Single Battery) | 10.0 x 7.3 x 3.1 in (25.4 x 18.5 x 7.9 cm) |
| Dimensions (with Double Battery) | 11.0 x 7.3 x 3.1 in (27.9 x 18.5 x 7.9 cm) |
| Weight (with Single Battery Pack) | 5 lbs (2.3 kg) |
| Weight (with Double Battery Pack) | 6 lbs (2.7 kg) |
| Operating Nominal Sound Level*** | Max 43.0 dB(A) @ 2
Maximum sound power of 51 dBA
Maximum system sound pressure of 51 dBA
Typical alarm sound pressure of 56.49 dBA (sound pressures measured at 1 meter per ISO 3744) |
| Flow Settings | Pulse Setting 1, 2, 3, 4, 5 |
| O2 Output Setting (@20 bpm All Volumes +/- 15%) | Setting 1 - 210 mL/min
Setting 2 - 420 mL/min
Setting 3 - 630 mL/min
Setting 4 - 840 mL/min
Setting 5 - 1050 mL/min |
| Oxygen Concentration* | 90% (+5.5% / -3%) |
| Oxygen Monitor | Yes |
| Maximum Outlet Pressure | <30 PSI |
| Dosing Sensitivity | > -0.1 cm H^{2}O (adjustable) |
| AC Power | 100–240 VAC, 50-60Hz, 120W Max |
| DC Power | 11–18 VDC (10 max amp) |
| Battery Type | Lithium Ion |
| Battery Capacity | Single cell: 6700 mAh; Double cell: 13400 mAh |
| Battery Re-Charge Time | Single Battery: 3.5 hours
Double Battery: 6.0 hours |
| Battery Pack Durations | Single Battery Duration: Setting 2 - up to 4 hours
Double Battery Duration: Setting 2 - up to 8 hours |
| Operating Temperature** | 41°F to 104°F (5°C to 40°C) |
| Transportation and Storage Temp. (Unit) | -13°F to +158°F (-25°C to +70°C) up to 10000 ft (3048 m) |
| Transportation and Storage Temp. (Battery) | -13°F to +158°F (-25°C to +70°C) up to 10000 ft (3048 m) |
| Operating Altitude** | -1250 to 10,000 ft (-381 to 3048 m) (tested to 700 – 1060 hPa) |
| Operating Humidity | 15 - 95% relative humidity (non-condensing) |
| Storage Humidity | 0 - 90% non-condensing |
| Limited Warranty | See Warranty Statement |
* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
** The unit will function in all environments as long as the environment is with the specified temperature and humidity ranges. Operating outside of these operational specifications can limit the concentrator's ability to meet oxygen concentration specification at higher liter flow rates. Warm up screen may continue to display until the concentrator has reached target concentration or up to 15 minutes. Continue to use unit during this time.
***Sound level measured per test method Nr. 14-1 10/2018 MDS-Hi. The expected service life is a minimum of five years.
FreeStyle Comfort – 510(k) Summary
{9}
Page 6 of 15
FreeStyle Comfort – 510(k) Summary
## Intended Use:
The FreeStyle Comfort Oxygen Concentrator can deliver concentrated supplemental oxygen with either a fixed-minute pulse delivery or can be set to operate using autoSAT pulse delivery.
The FreeStyle Comfort is used on a prescriptive basis by patients requiring supplemental oxygen to increase blood oxygen saturation. The device is designed for use in the home, in a vehicle, and for use during air travel.
The device is not intended for life support, nor does it provide any patient monitoring capabilities.
## Indications for Use:
The FreeStyle Comfort Oxygen Concentrator can deliver concentrated oxygen with either a fixed-minute pulse delivery or can be set to operate using autoSAT pulse delivery. The FreeStyle Comfort is used on a prescriptive basis by patients requiring supplemental oxygen to increase blood oxygen saturation. FreeStyle Comfort may be used in the home, institution, and transport modes.
## Contraindications:
DO NOT use this oxygen concentrator until you first seek medical advice. Although oxygen is a non-addictive drug, unauthorized oxygen therapy can be dangerous. The Equipment Provider who supplies your oxygen equipment will demonstrate how to set the prescribed flow rate.
DO NOT change the flow selection unless you have been directed to do so by a licensed clinician. It is very important to select only the prescribed level of oxygen. Pulse dose mode settings must be determined for each user individually for their needs at rest, during exercise, and when traveling.
DO NOT use in the presence of flammable anesthetics.
This device should not be used without first being prescribed by a physician. In certain circumstances, the use of non-prescribed oxygen can be hazardous.
This device is not intended to be used for life support. Additional monitoring may be required for any user who is unable to communicate discomfort.
This device is not intended for use in or near magnetic resonance environments, such as MRI machines.
This device is not intended for use with a tracheotomized patient.
Some respiratory efforts of the patient might not trigger the conserving function.
## Description of Use/Patient Population:
Oxygen Concentrators are intended to supply supplemental oxygen to users suffering from ailments which affect the efficiency of one's lungs to transfer oxygen in the air to their bloodstream. Oxygen concentrator use requires a physician's prescription.
This device is not intended for and has not been tested or approved for use with newborn, infant, or pediatric patient populations.
Users typically have good cognitive abilities and must be able to communicate discomfort. If the user is unable to communicate discomfort, or unable to read and understand the concentrator labeling and instructions for use, then use is recommended only under the supervision of one who can. If any discomfort is felt while using the concentrator, users are advised to contact their healthcare provider. Users are also advised to have back-up oxygen available (i.e. cylinder oxygen) in the event of a power outage or concentrator failure. There are no other unique skills or user abilities required for concentrator use.
{10}
# Comparison of Technological Characteristics and Performance with Predicates.
The table below provides a side-by-side comparison of the FreeStyle Comfort User Interface elements with respect to the predicate device, the Inogen Rove 4. The user interface features are broken down by category and the elements of each category. All FreeStyle Comfort User Interface elements have been found to be substantially equivalent to that of the predicate device, refer to the comparison table below.
Subject Device - Predicate Device Comparison Table
| | Subject Device: FreeStyle® Comfort® | Predicate Device: Inogen Rove™ 4 | Referenced Predicate Device: eQuinox™ |
| --- | --- | --- | --- |
| 510(k) No. | K250671 | K222086 | K120785 |
| CFR | 21 CFR 868.5440 | 21 CFR 868.5440 | 21 CFR 868.5440 |
| Product Code | CAW | CAW | CAW |
| Classification | 2 | 2 | 2 |
| Rx Required | Yes | Yes | Yes |
| Patient Use | Adult | Adult | Adult |
| Intended Use | The FreeStyle Comfort Oxygen Concentrator can deliver concentrated supplemental oxygen with either a fixed-minute pulse delivery or can be set to operate using autoSAT pulse delivery.
The FreeStyle Comfort is used on a prescriptive basis by patients requiring supplemental oxygen to increase blood oxygen saturation. The device is designed for use in the home, in a vehicle, and for use during air travel.
The device is not intended for life support, nor does it provide any patient monitoring capabilities. | The Inogen Rove 4™ is used on a prescriptive basis by patients requiring supplemental oxygen to increase blood oxygen saturation. | The eQuinox Oxygen System is intended for the administration of supplemental oxygen.
The eQuinox Oxygen System is prescription legend required. |
| Indications for Use | The FreeStyle Comfort Oxygen Concentrator can deliver concentrated oxygen with either a fixed-minute pulse delivery or can be set to operate using autoSAT pulse delivery. The FreeStyle Comfort is used on a prescriptive basis by patients requiring supplemental oxygen to increase blood oxygen saturation. FreeStyle Comfort may be used in the home, institution, and transport modes | Inogen Rove™ 4 Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in home, institutions, and transport modalities. The device is to be used as an oxygen supplement and is not intended to be life-sustaining or life-supporting. | The eQuinox Oxygen System is intended for the administration of supplemental oxygen. |
| User/Patient Interface | User Interface Panel | User Interface Panel | User Interface Panel |
| | LCD Display to convey information about operating status in numbers, text, and symbols. | LCD Display to convey information about operating status in numbers and symbols. | LCD Display to convey information about operating status in numbers, text, and symbols. |
| | Alarm Indicator – yellow “Alarm/Warning” triangle symbol on LCD display is to indicate abnormal operating conditions in compliance with ISO 60601-1-8 | Alarm Indicator – yellow LED on UIP above “Alarm/Warning” triangle symbol that illuminates to indicate abnormal operating conditions in compliance with ISO 60601-1-8 | Audible ALARM (red indicator): An audible indicator and solid red light announce a reduction in oxygen concentration or power loss or interruption. |
FreeStyle Comfort - 510(k) Summary
{11}
| | Breath Detect Notification – Green LED on LCD Display illuminates when a breath is detected, and an oxygen pulse is triggered | Breath Detect Notification – Green LED on UIP illuminates when a breath is detected, and an oxygen pulse is triggered | Breath Detect Notification – Green LED illuminates when a breath is detected, and an oxygen pulse is triggered |
| --- | --- | --- | --- |
| | Auditory Buzzer – Audible beeps are emitted to indicate alarm or status change conditions in compliance with ISO 60601-1-8. | Auditory Buzzer – Audible beeps are emitted to indicate alarm or status change conditions in compliance with ISO 60601-1-8. | Auditory alarm - An audible alarm alerts you to the operating condition of the device, either a warning or failure, and confirms a valid key press. |
| | Air Intake Filters (2) - Patient instructed to clean the air intake filters once per week. | Particle Filter – Patient instructed to clean particle filters once per week. | Air Inlet Filter - Patient instructed to clean particle filter once per week. |
| | Optional accessories - Carry bag, shoulder strap, carry-all accessory bag, external battery charger | Optional accessories - Carry bag, backpack, external battery charger | Optional accessories – Travel cart, travel case, external battery charger |
| | myCAIRE Application – Optional mobile application for viewing device settings, battery information, and current alerts, available for iOS and Android in English | Inogen Connect Mobile Application – Optional mobile application for viewing device settings, battery information, and current alerts, available for iOS and Android in English, French. | The eQuinox™ Oxygen System does not provide wireless connectivity. |
| Components | Electrical Power Source: AC/DC Power Adapter – Utilizes 100-240V, 50/60Hz AC power supply and cord for power and charging with wall adapter and barrel jack connection to concentrator. | AC/DC Power Adapter – Utilizes 100-240V, 50/60Hz AC power supply and cord for power and charging with wall adapter and barrel jack connection to concentrator. | AC/DC Power Adapter – 100-240V, 50/60Hz AC power supply and cord for power and charging with wall adapter and barrel jack connection to concentrator. |
| --- | --- | --- | --- |
| | DC Power Cable – cord and adapter to allow for connection to 12-volt DC auxiliary port connector and barrel jack connection to concentrator. | DC Power Cable – cord and adapter to allow for connection to 12-volt DC auxiliary port connector and barrel jack connection to concentrator. | DC Power Cable – cord and adapter to allow for connection to 12-volt DC auxiliary port connector and barrel jack connection to concentrator. |
| | Rechargeable single or double battery (1) 96.5 W-Hr Lithium-Ion Battery Pack (8 cell). Optional Double Battery (2) 96.5 W-Hr Lithium-Ion Battery Pack (16 cell). The battery slides into the base of the device. | Battery – utilizes a 4 or 8-cell lithium-Ion battery. To attach the battery, slide it on to the base of the concentrator. | Rechargeable single or double lithium- Ion battery pack, 89 W-Hr single battery pack 12-cell battery or 95 W-Hr double battery pack 24-cell battery |
| | Cannula - Off the shelf 7' nasal cannula | Cannula – Off-the-shelf 7' nasal cannula | Cannula - Off the shelf 7' nasal cannula |
FreeStyle Comfort - 510(k) Summary
{12}
Page 9 of 15
FreeStyle Comfort – 510(k) Summary
| Principle of Operation | The FreeStyle® Comfort® uses software to operate the unit. The device generates oxygen through a process known as "pressure swing adsorption" or PSA. Air is drawn in through the two air intake filters to the air compressor. Pressurized air flows from the air compressor to each of the sieve beds in cycles. As one sieve bed is filled with pressurized air, oxygen passes through that sieve bed. As the air travels through the sieve bed, nitrogen molecules collect on the molecular sieve. At the end of the pressure cycle, the nitrogen is purged using enriched oxygen and then the process begins again with the alternate sieve bed (one sieve bed depressurizes, the other sieve bed is pressurized). This process will continuously repeat from one sieve bed to the other producing highly concentrated oxygen. Oxygen flows from the sieve beds to the product tank, then through sensors that measure both oxygen concentration and flow. Oxygen is delivered to the patient on a demand flow basis during the inhalation breathing cycle. The oxygen conserver technology eliminates oxygen waste during the exhalation (breathing) cycle when it is not needed. The FreeStyle® Comfort® releases a pulse of oxygen (bolus) at the beginning of inhalation by using oxygen conserving technologies. The oxygen conserving technology detects inhalation and quickly delivers a pulse of oxygen from the product tank, through the O2 filter and channeled to the user with a nasal cannula. This method of oxygen delivery is controlled with pressure sensors and software. | The Inogen Rove™ 4 Portable Oxygen Concentrator uses molecular sieve / pressure swing adsorption technology. Ambient air is drawn through particle filters by a compressor and forced by pressure through molecular sieve beds, which adsorb nitrogen and allow oxygen to pass. Oxygen is collected in an accumulator reservoir. The airflow is then changed, and nitrogen is desorbed from the molecular sieve, allowing it to adsorb again during the next cycle. Waste nitrogen is exhausted back into the room. A series of sieve beds, a manifold with precision valves, sensors, and embedded software to control the cycle are used to make the system function. Oxygen is delivered to the patient on a demand flow basis during the inhalation part of the breathing cycle. This conserver technology eliminates waste of unused oxygen at other times in the breathing cycle when it is not needed. The Inogen Rove™ 4 Portable Oxygen Concentrator senses the beginning of the inhalation cycle and releases a specified dose of oxygen enriched gas from the accumulator reservoir, through a final filter, into the connected nasal cannula and onto the patient. Pressure sensors and embedded software are used to control the conserving function. | The eQuinox™ uses a passive system to separate oxygen from air. Clean air flows into the eQuinox™ where it enters a compressor. Pressurized air flows from the compressor into the ATF® Concentrator Module (molecular sieve bed) where it is separated into oxygen and nitrogen components. The air separation process uses a rotary valve system in the ATF module to force air through a series of pressurized sieve beds. Through a process known as "pressure swing adsorption," oxygen molecules are collected on an absorbent material within the ATF and forced through a sieve into the product tank. The remaining nitrogen molecules are purged using a vacuum pressure cycle. Oxygen flows from the product tank through a HEPA filter and past a sensor that measures flow and concentration. A flow control valve regulates the flow of purified oxygen presented to the patient. The process is continuously repeated during operation. |
| --- | --- | --- | --- |
| Operating | Portable. For use in home, institutions, and transport modalities. | Portable. For use in home, institutions, and transport modalities. | Portable. For use in home, institutions, and transport modalities. |
| --- | --- | --- | --- |
{13}
FreeStyle Comfort – 510(k) Summary
| Environment | | | |
| --- | --- | --- | --- |
| Power Source | The AC/DC Power Adapter – Utilizes 100-240V, 50/60Hz AC power supply and cord for power and charging with wall adapter and barrel jack connection to concentrator. | AC/DC Power Adapter – Utilizes 100-240V, 50/60Hz AC power supply and cord for power and charging with wall adapter and barrel jack connection to concentrator. | AC/DC Power Adapter – 100-240V, 50/60Hz AC power supply and cord for power and charging with wall adapter and barrel jack connection to concentrator. |
| | The DC Power Cable – cord and adapter to allow for connection to 12-volt DC auxiliary port connector and barrel jack connection to concentrator. | DC Power Cable – cord and adapter to allow for connection to 12-volt DC auxiliary port connector and barrel jack connection to concentrator. | DC Power Cable – cord and adapter to allow for connection to 12-volt DC auxiliary port connector and barrel jack connection to concentrator. |
| | Battery - Utilizes a Lithium-Ion Battery Pack (8 cell) or Optional Double Battery (16 cell). Battery attaches by sliding it on to the base of the concentrator. | Battery – utilizes a 4 or 8-cell lithium-Ion battery. To attach the battery, slide it on to the base of the concentrator. | Battery – Utilizes a 12-cell battery pack 12-cell battery or 24-cell battery pack. Battery slides into the front of the unit |
| Flow Control Settings | Pulse dose settings 1, 2, 3, 4, 5 | Pulse dose settings 1, 2, 3, 4 | Pulse dose settings 1, 2, 3, 4, 5, 6, 7, 8, 9
Continuous lpm settings 0.5, 1, 1.5, 2.0, 2.5, 3.0 |
| Oxygen Concentration | 90% +5.5% /-3% at all flow settings | 87% to 96% at all flow settings | 90% +5.5%/-3% at all flow settings |
Page 10 of 15
{14}
| Oxygen Flow Rate | Fixed-Minute Pulse: | Fixed Minute Pulse: | Fixed Minute Pulse is N/A in this device |
| --- | --- | --- | --- |
| Setting 1 - 210 mL per minute, 10.5 mL @20 bpm | Setting 1 - 210 mL per minute, 10.5 mL @20 bpm |
| Setting 2 - 420 mL per minute, 21.0 mL @20 bpm | Setting 2 - 420 mL per minute, 21.0 mL @20 bpm |
| Setting 3 - 630 mL per minute, 31.5 mL @20 bpm | Setting 3 - 630 mL per minute, 31.5 mL @20 bpm |
| Setting 4 - 840 mL per minute, 42.0 mL @20 bpm | Setting 4 - 840 mL per minute, 42.0 mL @20 bpm |
| Setting 5 - 1050 mL per minute, 52.5 mL @20 bp Dose volume adjusts based on breath rate. | Dose volume does adjusts based on breath rate |
| Fixed Minute Pulse Volume (mL) Setting (Tolerance +/-15%) | Breath rate | 1 | 2 | 3 | 4 |
| Breath Rate | 1 | 2 | 3 | 4 | 5 | 10 | 21.0 | 42.0 | 63.0 |
| 15 | 14.0 | 28.0 | 42.0 | 56.0 | 70.0 | 15 | 14.0 | 28.0 | 42.0 |
| 20 | 10.5 | 21.0 | 31.5 | 42.0 | 52.5 | 20 | 10.5 | 21.0 | 31.5 |
| 25 | 8.4 | 16.8 | 25.2 | 33.6 | 42.0 | 25 | 8.4 | 16.8 | 25.2 |
| 30 | 7.0 | 14.0 | 21.0 | 28.0 | 35.0 | 30 | 7.0 | 14.0 | 21.0 |
| 35 | 6.0 | 12.0 | 18.0 | 24.0 | 30.0 | 35 | 6.0 | 12.0 | 18.0 |
| 40 | 5.3 | 10.5 | 15.8 | 21.0 | 26.3 | 40 | 5.25 | 10.5 | 15.75 |
| Total Volume Per Min (mL/min) | 210 | 420 | 630 | 840 | 1050 | Total Volume Per Min (ml/min) | | | |
FreeStyle Comfort - 510(k) Summary
{15}
Page 12 of 15
FreeStyle Comfort – 510(k) Summary
autoSAT® Pulse:
Setting 1 - 10.5 mL nominal dose
Setting 2 - 21.0 mL nominal dose
Setting 3 - 31.5 mL nominal dose
Setting 4 - 42.0 mL nominal dose
Setting 5 - 52.5 mL nominal dose
autoSAT® Pulse Volume (mL) Setting
(Tolerance +/- 15%)
| Breath Rate | 1 | 2 | 3 | 4 | 5 |
| --- | --- | --- | --- | --- | --- |
| 15 | 10.5 | 21.0 | 31.5 | 42.0 | 52.5 |
| 20 | 10.5 | 21.0 | 31.5 | 42.0 | 52.5 |
| 25 | 10.5 | 21.0 | 31.5 | 42.0 | 42.0 |
| 30 | 10.5 | 21.0 | 31.5 | 35.0 | 35.0 |
| 35 | 10.5 | 21.0 | 30.0 | 30.0 | 30.0 |
| 40 | 10.5 | 21.0 | 26.3 | 26.3 | 26.3 |
Total system output adjusts based on breath rate.
Note: Bolus volume testing is the average of 30 consecutive measurements and required to be within +/- 15% of the target bolus volume.
Continuous flow is setting is N/A
autoSAT Pulse N/A in this device
AutoSAT® Pulse:
Setting 1 - 16 mL nominal dose @ 15 bpm
Setting 2 - 32 mL nominal dose @ 15 bpm
Setting 3 - 48 mL nominal dose @ 15 bpm
Setting 4 - 64 mL nominal dose @ 15 bpm
Setting 5 - 80 mL nominal dose @ 15 bpm
Setting 6 - 96 mL nominal dose @ 15 bpm
Setting 7 - 128 mL nominal dose @ 15 bpm
Setting 8 - 160 mL nominal dose @ 15 bpm
Setting 9 - 192 mL nominal dose @ 15 bpm
autoSAT® Pulse Volume (mL) Setting
(Tolerance +/-15%)
| Breath Rate | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 15 | 16 | 32 | 48 | 64 | 80 | 96 | 128 | 160 | 192 |
| 18 | 16 | 32 | 48 | 64 | 80 | 96 | 128 | 160 | 160 |
| 23 | 16 | 32 | 48 | 64 | 80 | 96 | 128 | 128 | 128 |
| 31 | 16 | 32 | 48 | 64 | 80 | 96 | 96 | 96 | 96 |
| 37 | 16 | 32 | 48 | 64 | 80 | 80 | 80 | 80 | 80 |
| 40 | 16 | 32 | 48 | 64 | 74 | 74 | 74 | 74 | 74 |
| mL/Min | 240 | 576 | 864 | 1152 | 1440 | 1728 | 2304 | 2880 | |
Note: Bolus volume testing is the average of 30 Consecutive measurements and required to be within +/- 15% of the target bolus volume.
Continuous Flow Settings:
- 0.5 liters per minute
- 1.0 liters per minute
- 1.5 liters per minute
- 2.0 liters per minute
- 2.5 liters per minute
- 3.0 liters per minute
Continuous flow setting is N/A
{16}
FreeStyle Comfort – 510(k) Summary
| Oxygen Outlet Pressure | Maximum Outlet Pressure <30 PSI | <22 PSI Max
18.2 PSI (129 kPa) +/-10% | 5.0 psig (34.5 kPa) Nominal (continuous flow) |
| --- | --- | --- | --- |
| Operating Conditions | Operating Temperature: 41 F° - 104 F°
Humidity: 15-95% RH noncondensing
Altitude: -1250 to 10,000 Ft. | Operating Temperature: 41 F° - 104 F°
Humidity: 0 – 95% RH noncondensing
Altitude: Up to 10,000 Ft. | Operating Temperature: 32 F° - 104 F°
Humidity: 10%-95% @ an 82.4° F dew point
Altitude: 0-13,133 Ft |
| Dose Sensitivity | > -0.1 cm H2o (adjustable) | -0.12 cm H2o | 0.135 - 0.37 cm H2o (adjustable) |
| Bluetooth Technology | myCAIRE App – BLE Connection to Android or iPhone.
The FreeStyle Comfort Oxygen Concentrator is capable of Bluetooth functionality with the myCAIRE App. | Inogen Connect App – BLE Connection to Android or iPhone.
The Inogen Rove 4 Oxygen Concentrator is capable of Bluetooth functionality with the Inogen Connect App. | Not equipped with Bluetooth technology |
| Weight | • 5 lbs. (2.3 kg) w/ single battery pack
• 6 lbs. (2.7 kg) w/ double battery pack | • 3.0 lbs. (1.4 kg) w/ 4-cell battery
• 3.3 lbs. (1.5 kg) w/ 8-cell battery | • 14 lbs. (6.53 kg) w/ 12 cell lithium-Ion battery pack
• 16 lbs. (7.26 kg) w/ 24 cell lithium-Ion battery pack |
| Dimensions | With Single Battery:
10.0" H. 7.3" W.3.1" D (25.4 cm. 18.5 cm. 7.9 cm)
With Double Battery:
11.0" H. 7.3" W.3.1" D (27.9 cm. 18.5 cm. 7.9 cm) | With 4-Cell Battery:
7.5" H. 6" W. 2.7" D (19.1 cm. 15.2 cm. 6.9 cm)
With 8-Cell Battery:
7.8" H. 6" W. 2.7" D (19.7 cm. 15.2 cm. 6.9 cm) | 13.6" H X 10.6" W X 7.4" D (34.5 cm.26.9 cm.18.8 cm) |
| Operating Nominal Sound Level | 43 dB(A) @ setting 2 (MDS-Hi)
Maximum sound power of 51 dBA
Maximum system sound pressure of 51 dBA
Typical alarm sound pressure of 56.49 dBA (Sound pressures measured at 1 meter per ISO 3744) | 39 dB(A) @ setting 2 (MDS—Hi)
Maximum sound power of 59 dBA
Maximum system sound pressure of 51 dBA
Typical alarm sound pressure of 53 dBA (Sound pressures measured at 1 meter per ISO 3744 | 46 dB(A) @ 3LPM (continuous flow mode) 40 dB(A) @ setting 3 (pulse mode) |
| Performance and Electrical Safety and EMC | • IEC 60601-1:2012+AMD1:2020
• IEC 60601-1-2:2015+A1:2021 Ed.4.1
• IEC 60601-1-6:2020
• IEC 60601-1-8:2020
• IEC 60601-1-11:2020
• ISO 80601-2-69:2020 Ed. 2
• ISO 80601-2-67:2020 Ed. 2
• IEC 62366-1:2015 | • IEC 60601-1:2012
• IEC 60601-1-2:2020
• IEC 60601-1-6:2020
• IEC 60601-1-8:2012
• IEC 60601-1-11:2015
• ISO 80601-2-69:2020
• ISO 80601-2-67:2020
• IEC 62366-1 | • IEC 60601-1:1988+A1:1991+A2:1995
• IEC-60601-1-2:2001+A1:2004
• ISO 8359:1996 |
| Biocompatibility | Externally Communicating, Tissue Permanent Duration (>30 days)
ISO 18562-2:2017 - Particulate matter
ISO 18562-3:2017 - Volatile organic compounds | Externally Communicating, Tissue Permanent Duration (>30 days)
ISO 18562-2:2017 - Particulate matter
ISO 18562-3:2017 - Volatile organic compounds | No testing to ISO 18562 1 and 2 – FDA clearance predates standard requirements. |
Page 13 of 15
{17}
SUBSTANTIAL EQUIVALENCE DISCUSSION
## Intended Use/Indications for Use
The FreeStyle Comfort and Predicate device have similar Intended Use/Indications for use. The Subject and Predicate device provide a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. The FreeStyle Comfort and Predicate device are to be used as an oxygen supplement, may be used in the home, institution, transport modalities and are not intended for life support. FreeStyle Comfort has included this statement as part of the Intended Use/Indications for Use statement.
## Technological Characteristics and Operating Principles
The Subject and Predicate device operate on Pressure Swing Adsorption (PSA) technology to produce oxygen and deliver it to patient via a standard oxygen nasal cannula. The devices deliver a bolus of oxygen upon sensing a pressure change at the start of inhalation. The oxygen concentration produced is the same with the Subject device and Predicate devices.
There are no differences which raise questions regarding the safety or effectiveness.
## Non-clinical Testing
Testing included:
- ISO 80601-2-69 - Medical Electrical Equipment - Part 2-69: Particular Requirements for Basic Safety and Essential Performance of Oxygen Concentrator Equipment
- ISO 80601-2-67 - Medical Electrical Equipment - Part 2-67: Particular Requirements for Basic Safety and Essential Performance of Oxygen-Conserving Equipment.
- Pulse, volume; Pulse Time; Trigger Sensitivity; Oxygen Concentration Performance; Oxygen Sensor Accuracy; Visible/Audible Alarms
- Software verification and validation
- Electrical Safety / Electro-Magnetic Compatibility (EMC) / RFID
- Biocompatibility
- ISO 18562-2:2017 - Particulate matter
- ISO 18562-3:2017 - Volatile organic compounds
- Sound testing.
The FreeStyle Comfort and Predicate device meet safety, collateral, and performance standards required for portable oxygen concentrators verified through both internal and third-party testing to FDA recognized consensus standards.
The FreeStyle Comfort Oxygen Concentrator met all requirements.
## Other Similarities
- The oxygen concentration produced by the Subject and Predicate device is the same.
- Operating environmental conditions are similar for the Subject and Predicate device.
The user interface for the Subject device is similar to the Predicate device.
The Subject and Predicate device use the same sources of power, wall outlet, battery, and DC adapter.
The Subject device can operate in both the autoSAT and fixed minute pulse dose modes. The Referenced Predicate uses the autoSAT feature to deliver a bolus of oxygen to the patient. The Predicate device uses a "fixed" minute feature to deliver a bolus of oxygen to the patient. Flow settings 1-4 are identical with both the Subject and Predicate devices.
The Subject and Predicate device are intended for adult use in homes, institutions, and transport modalities.
Page 14 of 15
Summary
FreeStyle Comfort – 510(k)
{18}
The remarks column in the Substantially Equivalence Device Comparison Table notes the similarities, substantial equivalence, and differences to the line items listed in the table
## Discussion of Differences
The major differences between the Subject device and the Predicate devices are:
- The device dimensions and weight - The Predicate device is smaller and lighter than the Subject device.
- Battery pack cells - The Predicate uses either a 4-cell or 8-cell battery whereas the Subject device uses either an 8-cell or 16-cell battery pack.
- Flow settings – The Predicate device offers four (4) fixed oxygen flow settings. The Subject device offers five (5) fixed flow settings plus the autoSAT flow functions similar to the Referenced Predicate, eQuinox. The cleared Referenced Predicate device, eQuinox, has nine settings and for breath rates shown, at least 4 of the setting are higher than setting 5 (max) of the Subject Device.
The above differences noted between the Subject and Predicate devices do not raise any additional concerns regarding safety or effectiveness
## Substantial Equivalence Conclusion
The Subject and Predicate device have been found to be substantially equivalent. The FreeStyle Comfort and Predicate device have similar fundamental technology, operating systems, components/materials, operating principles and the same delivered concentrated oxygen, 90%+5.5% / -3%. The indications for use are the same between the FreeStyle® Comfort® and the Predicate device.
The above comparison demonstrates that the FreeStyle® Comfort® has no significant technological differences from the Predicate device that would raise any additional concerns regarding product safety and effectiveness.
Page 15 of 15
Summary
FreeStyle Comfort – 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.