K250665 · Konica Minolta, Inc. · MQB · Jun 17, 2025 · Radiology

Device Facts

Record IDK250665
Device NameSKR 3000
ApplicantKonica Minolta, Inc.
Product CodeMQB · Radiology
Decision DateJun 17, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2
AttributesPediatric

Indications for Use

This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, and angiography applications.

Device Story

SKR 3000 (AeroDR 3) is a digital radiography system using an X-ray planar detector (FPD) to capture human anatomical images. The system transforms X-ray inputs into digital signals, processed by the Console CS-7 software, and transmitted to filing, printing, or display systems. Used in clinical settings by healthcare professionals, the device replaces traditional film/screen systems. The CS-7 software performs image processing, including gradation, frequency, dynamic range compression, smoothing, and Intelligent-Grid (scattered radiation correction). Data transmission occurs via wired Ethernet or wireless LAN (IEEE802.11a/n) with WPA2-PSK (AES) encryption. The system supports pediatric imaging. Clinicians use the output images for diagnostic decision-making, benefiting patients through digital imaging efficiency and reduced radiation exposure compared to film.

Clinical Evidence

Bench testing only. Image quality evaluation was performed using small-size phantoms to simulate pediatric imaging conditions. Comparative evaluation demonstrated that the SKR 3000 with P-53 detector provides image performance substantially equivalent to the predicate AeroDR System 2 with P-52 detector.

Technological Characteristics

Stationary X-ray system; planar detector (FPD) technology; wired Ethernet and wireless LAN (IEEE802.11a/n) connectivity; WPA2-PSK (AES) encryption. Software-based image processing includes gradation, frequency, dynamic range compression, smoothing, rotation, reversing, zooming, and Intelligent-Grid (scattered radiation correction).

Indications for Use

Indicated for generating radiographic images of human anatomy, including pediatric patients, for general-purpose diagnostic procedures. Not indicated for mammography, fluoroscopy, or angiography.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 17, 2025 Konica Minolta, Inc. % Jan Maniscalco Executive Vice President QA/RA Konica Minolta Inc. 411 Newark Pompton Turnpike WAYNE, NJ 07470 Re: K250665 Trade/Device Name: SKR 3000 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB, LLZ Dated: May 13, 2025 Received: May 13, 2025 Dear Jan Maniscalco: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250665 - Jan Maniscalco Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2} K250665 - Jan Maniscalco Page 3 See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250665 | | | Device Name SKR 3000 | | | Indications for Use (Describe) | | | This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, and angiography applications. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} KONICA MINOLTA # 510(k) Summary K250665 Complying with 21 CFR R §807.92 ## I. SUBMITTER KONICA MINOLTA, INC. 1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Contact Person: Jan Maniscalco, Executive Vice President of QA/RA Konica Minolta Healthcare Americas, Inc. Contact Telephone: 973.633.1500 Contact Email: jan.maniscalco@konicaminolta.com Mailing Address: 411 Newark Pompton Turnpike, Wayne, New Jersey 07470 Date Prepared: June 13, 2025 ## II. DEVICE Name of Device: SKR 3000 Common Name: Digital Radiography Classification Name: Stationary x-ray system (21 CFR 892.1680) Regulatory Class: Class II Product Code: MQB, LLZ ## III. PREDICATE DEVICES Primary Predicate Device: K241319 - SKR 3000 (KONICA MINOLTA, INC.) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Codes: MQB, LLZ Giving Shape to Ideas {5} KONICA MINOLTA # Predicate Device: K151465 – AeroDR System2 (KONICA MINOLTA, INC.) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Codes: MQB, LLZ # IV. DEVICE DESCRIPTION The digital radiography SKR 3000 performs X-ray imaging of the human body using an X-ray planar detector that outputs a digital signal, which is then input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data. - This device is not intended for use in mammography - This device is also used for carrying out exposures on children. The Console CS-7, which controls the receiving, processing, and output of image data, is required for operation. The CS-7 is a software with basic documentation level. CS-7 implements the following image processing; gradation processing, frequency processing, dynamic range compression, smoothing, rotation, reversing, zooming, and grid removal process/scattered radiation correction (Intelligent-Grid). The Intelligent-Grid is cleared in K151465. The FPDs used in SKR 3000 can communicate with the image processing device through the wired Ethernet and/or the Wireless LAN (IEEE802.11a/n and FCC compliant). The WPA2-PSK (AES) encryption is adopted for a security of wireless connection. The SKR 3000 is distributed under a commercial name AeroDR 3. The purpose of the current premarket submission is to add pediatric use indications for the SKR 3000 imaging system. Giving Shape to Ideas {6} KONICA MINOLTA # V. INDICATIONS FOR USE This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, and angiography applications. The subject device has the same Indications for Use statement as the primary predicate device except for the explicit inclusion of pediatric applications for the detector P-53 of the SKR 3000. This difference does not alter the intended use of the device. The subject device has the identical technological characteristics as the primary predicate device. # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The technological characteristics are identical in the proposed subject device and the primary predicate device. The sole purpose of this submission is to support the conditions of use for a pediatric population for the detector P-53 of the SKR 3000. # VII. PERFORMANCE DATA Since this submission is to support the conditions of use for a pediatric population for the detector P-53 of the SKR 3000, image quality evaluation was conducted in accordance with the "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices". Pediatric image evaluation using small-size phantoms was performed on the P-53. The comparative image evaluation demonstrated that the SKR 3000 with P-53 provides substantially equivalent image performance to the comparative device, AeroDR System 2 with P-52, for pediatric use. Giving Shape to Ideas {7} KONICA MINOLTA # VIII. CONCLUSIONS It is concluded that the subject device is as safe and effective as the predicates and is substantially equivalent to the predicate devices to support the indication for pediatric imaging for the detector P-53 of the SKR 3000. Giving Shape to Ideas
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...