K250648 · Philips Medical Systems Nederland B.V. · JAK · Jun 27, 2025 · Radiology
Device Facts
Record ID
K250648
Device Name
Philips iCT CT system
Applicant
Philips Medical Systems Nederland B.V.
Product Code
JAK · Radiology
Decision Date
Jun 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Philips iCT CT systems is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The iCT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
Device Story
The Philips iCT CT system is a whole-body diagnostic imaging device. It utilizes a continuously rotating X-ray tube and multi-slice detector gantry to acquire X-ray transmission data from multiple angles. A Data Measurement System (DMS) captures attenuation data, which is processed by reconstruction algorithms to produce cross-sectional images. Operated by clinicians in a clinical setting, the system includes a patient support couch and an operator console for image processing and data management. The device provides high-quality diagnostic images, including specialized modes for cardiac and vascular applications, and supports low-dose lung cancer screening. Output is displayed for physician review to aid in clinical diagnosis and decision-making. The system incorporates hardware and software enhancements, including cybersecurity updates and improved image reconstruction for head scans, to maintain performance and reliability.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including phantom-based image quality evaluations (CT number accuracy, CNR, uniformity, MTF, slice width, noise), cybersecurity penetration testing, and electrical/EMC safety testing per IEC standards.
Technological Characteristics
Whole-body CT system; continuously rotating X-ray tube; Nano Panel Prism ceramic scintillator detectors; 700mm bore; 100-120kW generator. Connectivity via DICOM 3.0. Software runs on Windows 10. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, and IEC 60601-2-44.
Indications for Use
Indicated for head, whole body, cardiac, and vascular CT imaging in patients of all ages. Includes low-dose CT lung cancer screening for early detection of lung nodules in accordance with established professional or governmental protocols.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 27, 2025
Philips Medical Systems Nederland B.V.
% Yael Curtz
Regulatory Affairs Manager and Site Lead
Advanced Technology Center, Building 34, MATAM
HAIFA, 3100202
ISRAEL
Re: K250648
Trade/Device Name: Philips iCT CT system
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: February 27, 2025
Received: May 20, 2025
Dear Yael Curtz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250648 - Yael Curtz
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250648 - Yael Curtz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed
by Gabriela M. for
Rodal -S
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250648 | |
| Device Name Philips iCT CT system | |
| Indications for Use (Describe) The Philips iCT CT systems is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The iCT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250648
510(k) Summary
This 510(k) summary of safety and effectiveness is prepared in accordance with 21 CFR §807.92.
| Date Prepared | March 4th, 2025 | |
| --- | --- | --- |
| Manufacturer | Philips Medical Systems Nederland B.V.
Veenpluis 6,
5684 PC BEST
The Netherlands | |
| Establishment Registration Number | 3015777306 | |
| Primary Contact Person | Ms. Yael Curtz
Regulatory Affairs Team Lead
Phone: +972-54-7282285
E-mail: yael.curtz@philips.com | |
| Secondary Contact Person | Carmit Shmuel
Regulatory Affairs Manager and Site Lead
Phone: +972-54-2109054
E-mail: carmit.shmuel@philips.com | |
| Subject Device | Trade Name | Philips iCT CT System |
| | Common Name | iCT System |
| | Classification Name | System, X-Ray, Tomography, Computed |
| | Classification Regulation | 21 CFR 892.1750 |
| | Classification Panel | Radiology |
| | Device Class | II |
| | Product Code | JAK |
| Predicate Device | Trade Name | Philips iCT CT System |
| | Manufacturer | Philips Medical Systems (Cleveland), Inc. |
| | 510(k) Clearance | K162838 |
| | Classification Regulation | 21 CFR 892.1750 |
| | Classification Panel | Radiology |
| | Device Class | II |
| | Product Code | JAK |
Device Description
The Philips iCT CT System is a whole-body computed tomography (CT) X-ray system designed for diagnostic imaging. It features a continuously rotating X-ray tube and multi-slice detector gantry, enabling the acquisition of X-ray transmission data from multiple angles and planes. The system reconstructs these data into cross-sectional images using advanced image reconstruction algorithms, supporting a wide range of clinical applications.
The system consists of a gantry, which houses the rotating X-ray tube, detector array, and key imaging subsystems; a patient support couch, which moves the patient through the gantry bore in synchronization with the scan and is available in multiple configurations; an operator console, which serves as the primary user interface for system controls, image processing, and data management; and a Data Measurement System (DMS), which captures X-ray attenuation data to support high-quality image reconstruction.
The Philips iCT CT System maintains the same intended use and fundamental operating principles as its legally marketed predicate device (Philips iCT CT System, K162838) while incorporating hardware
{5}
and software enhancements to improve system performance, cybersecurity, and usability. These updates do not alter the device's core functionality, imaging capabilities, or clinical workflow.
## Indications for Use Statement
The Philips iCT CT systems is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The iCT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
## Indications for Use Comparison to Predicate
The proposed and predicate devices share the same intended use, target patient population, and technological principles. While the subject device includes hardware and software updates as described above, all modifications were validated to ensure they do not impact core performance or introduce new safety concerns. The enhancements support system reliability, cybersecurity, and image quality.
## Technological Comparison / Modifications
The Philips iCT CT System shares the same technological characteristics as the legally marketed predicate device, Philips iCT CT System (K162838). Both systems employ a continuously rotating X-ray tube and multi-slice detector array for image acquisition, with computer-assisted reconstruction of cross-sectional images of the body. The principle of operation, energy source, and control mechanisms remain unchanged between the two devices.
Below is a high-level overview of the Technological Characteristic Comparison between the subject and predicate device:
| | Subject Device Philips iCT CT System | Predicate Device Philips iCT CT System (K162838) | Comparison |
| --- | --- | --- | --- |
| Indications for Use | Same | Same | |
| Design and Fundamental Scientific Technology | | | |
| Application | Head/Body | Same | |
| Scan regime | Continuous Rotation | Same | |
| Scan modes | Surview
Helical
Axial | Same | |
| Spatial Resolution | Ultra-high mode 24+/-2
lp/cm
High mode 16+/-2 lp/cm | Same | |
{6}
| | Standard mode 13+/-2
Ip/cm | | |
| --- | --- | --- | --- |
| Low contrast resolution | iCT:
5 mm +/-0.3% 14mGy
CTDIvol
3 mm +/- 0.3% 25 mGy
(20 cm Catphan phantom,
10 mm slice thickness)
CTDIvol
iCT SP:
4 mm +/- 0.3% 16.4 mGy
CTDIvol (20 cm Catphan
phantom, 10 mm slice
thickness) | Same | |
| Noise | 0.27% at 120kV, 250 mAs
(10mm slice thickness) | Same | |
| Reconstruction speed | 33 images/sec | Same | |
| Image Matrix | 512², 768², 1024² | Same | |
| Display | 1024x1280 | Same | |
| Communication | DICOM – 3.0 | Same | |
| Detectors | | | |
| Type | Nano Panel Prism | Same | |
| Material | Ceramic scintillator | Same | |
| Slices | iCT configuration: 256
iCT SP configuration: 128 | Same | |
| Coverage | iCT configuration: 8 cm
iCT SP configuration: 4 cm | Same | |
| Collimation | iCT configuration:
128 x 0.625 mm
112 x 0.625 mm
96 x 0.625 mm
64 x 0.625 mm
32 x 0.625 mm
20 x 0.625 mm
16 x 0.625 mm
8 x 0.625 mm
4 x 0.625 mm
2 x 0.625 mm
64 x 1.25 mm
32 x 1.25 mm
iCT SP configuration:
64 x 0.625 mm
32 x 0.625 mm
20 x 0.625 mm
16 x 0.625 mm
8 x 0.625 mm
4 x 0.625 mm
2 x 0.625 mm
64 x 1.25 mm
32 x 1.25 mm | Same | |
| Scan field | 50 -500 mm continuous
25 - 250mm ultra-high
resolution | Same | |
| Gantry | | | |
| Bore size | 700 mm | Same | |
{7}
| Focus-isocenter distance | 570 mm | Same | |
| --- | --- | --- | --- |
| Focus-detector distance | 1040mm | Same | |
| Rotation times | 0.3, 0.33, 0.375, 0.4, 0.5, 0.75, 1.0, 1.5 seconds for full 360° scans; 0.2 for partial angle 240° scans. (Optional - 0.27 seconds for full 360° scans; 0.18 seconds for partial angle 240° scans) | Same | |
| Gantry and Couch covers color | White | Mushroom | Aesthetic Only |
| Generator and Tube Performance | | | |
| Power | 100kW (120kW optional) | Same | |
| kV Setting | 80, 100, 120, 140 | Same | |
| mA Range | 10-830 (1mA steps), optional 10-1,000) | Same | |
| Focal spot – Smart Focal Spot | small 0.6 x 0.7; large 1.1 x 1.2 | Same | |
| Couch | | | |
| Couch | Standard Bariatric Extended (aka Long) | Same | |
| Patient table scan range | Standard: 1750 mm Bariatric: 1750 mm Long: 2100 mm | Same | |
| Table Z-position accuracy | Standard: +/- 0.25 mm Bariatric: +/- 0.25 mm Long: +/- 0.25 mm | Same | |
| Table longitudinal speed | Standard: 0.5 – 185 mm/sec Bariatric: 0.5 – 185 mm/sec Long: 0.5 – 185 mm/sec | Same | |
| Table maximum load capacity | Standard: 450 lbs. (204kg) Bariatric: 650 lbs. (405kg) Long: 450 lbs. (204kg) | Same | |
| General | | | |
| CIRS Computers | Windows 10 | Windows 7 | Security/Compatibility Improvement |
| Interventional Controls | Support two types of Interventional Exams: Basic Interventional and Advanced Interventional (also called CCT) | Same | |
| Harmonized Phantom Kit | Available | Not Available | Upgraded Calibration Kit |
| Security Updates | Enhanced (RBAC, encryption, audit) | Basic | Updated |
| Software Features | | | |
| Head scan Image Quality Improvement | Included | Standard Head Scan Imaging | Algorithmic Improvement |
{8}
| (Optional Feature) Concurrent Recon in Iterative Model Reconstruction (IMR) | Included | Sequential Reconstruction of iDose/BP | Workflow Enhancement |
| --- | --- | --- | --- |
| Size Specific Dose Estimate (SSDE) | Included | Not Available | Automated calculation; Informational only |
The proposed Philips iCT CT System hardware and software modifications are detailed as follows:
- Harmonized Phantom Kit: Updated calibration phantoms with standardized dimensions and materials (PMMA) for consistency across platforms. These phantoms replace legacy versions and are used in acceptance, constancy, and quality assurance testing.
- White Covers: Aesthetic color change of the gantry and couch covers from Mushroom to White
- Operating System Upgrade: Software upgrade from Windows 7 to Windows 10 to meet current cybersecurity and compatibility standards.
- Security Enhancements: Implementation of software hardening, vulnerability mitigation, access control, and encryption.
- Head Scan Image Quality Enhancement: Optimization of the reconstruction algorithm to improve image homogeneity and sharpness.
- Concurrent Reconstruction (Optional): An optional feature enabling concurrent iDose/BP reconstruction on the IMR server for improved throughput without altering imaging output.
- Size Specific Dose Estimate (SSDE) Display: Addition of SSDE dose metric display to improve dose awareness; calculated using existing system measurements and does not affect scan output.
These modifications do not change the intended use, fundamental scientific technology, or safety and effectiveness profile of the system. The updates have been verified and validated to ensure they do not impact safety, effectiveness, or the system's intended use.
A comprehensive evaluation was conducted to confirm that the modifications introduced in the Philips iCT CT System do not raise new questions of safety or effectiveness. This included system-level verification and validation, requirements-based regression testing, usability validation (IEC 62366-1), biocompatibility assessment (ISO 10993), electromagnetic compatibility and electrical safety testing (IEC 60601-1 and IEC 60601-1-2), and a cybersecurity evaluation aligned with FDA's 2023 guidance. Risk management activities were performed in accordance with ISO 14971, and software lifecycle documentation followed IEC 62304.
Collectively, these activities confirm that the system modifications are safe, effective, and compliant with FDA-recognized consensus standards. Therefore, based on the technological comparison, risk analysis, and non-clinical testing results, the modified Philips iCT CT System is substantially equivalent to the predicate device.
## Non-Clinical / Clinical Test Summary
The non-clinical testing strategy followed an established framework to verify that all system changes, both hardware and software, did not introduce new risks and remain in compliance with regulatory requirements. No clinical testing was required, as the intended use, core imaging algorithms, and diagnostic claims remain unchanged from the predicate (K162838).
## Verification Testing
Requirements-based testing was performed to verify that all modified system components met their respective requirements through structured test execution. Regression testing confirmed that
{9}
modifications did not impact existing system functionality. The following verification activities were conducted to ensure compliance with regulatory and safety expectations:
- Cybersecurity testing, including penetration testing, vulnerability assessment, and access control validation, was performed per FDA's Cybersecurity in Medical Devices Guidance (2023).
- Static code analysis and software review were conducted to confirm adherence to secure coding practices.
- Electromagnetic Compatibility (EMC) testing was performed in accordance with IEC 60601-1-2.
- Electrical safety testing was conducted per IEC 60601-1 and IEC 60601-2-44.
- Biocompatibility evaluation followed ISO 10993 to confirm safety of unchanged patient-contacting materials.
## Validation Testing
Validation activities confirmed that system updates aligned with intended use and user needs. Testing included:
- Usability testing in accordance with IEC 62366-1, evaluating user interaction and workflow.
- System-level validation under simulated clinical conditions, confirming functionality of modified hardware and software.
- Software validation ensuring reliable system performance after updates, including:
- Operating system transition
- Cybersecurity enhancements
## Performance Testing
To support the modifications in the Philips iCT CT System, a comprehensive non-clinical performance testing strategy was executed. Testing included both bench testing and phantom-based evaluations, addressing software, imaging, usability, and cybersecurity enhancements.
Testing activities included:
- Automated Image Quality (IQ) Testing using standard phantoms to assess CT number accuracy, contrast-to-noise ratio (CNR), uniformity, modulation transfer function (MTF), slice width accuracy, and noise characteristics.
- Phantom Testing using anthropomorphic head phantoms to evaluate head image quality enhancements, specifically improvements in homogeneity and sharpness near bone-brain interfaces.
- Traceability to system requirements was maintained across all testing, with test results documented in the system verification report and traceability matrices.
These tests confirmed the modified device performs equivalently to the predicate and supports the substantial equivalence determination.
## Compliance with FDA-Recognized Consensus Standards
Testing protocols ensured compliance with the following FDA-recognized standards:
- IEC 60601-1: Edition 3.2 (2020-08) – FDA Recognition Number: 19-49, General Requirements for Basic Safety and Essential Performance
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] – FDA Recognition Number: 19-36, Electromagnetic Compatibility (EMC) – Requirements and Tests
- IEC 60601-1-3: Edition 2.2 (2021-01) – FDA Recognition Number: 12-336, Radiation Protection in Diagnostic X-ray Equipment
- IEC 60601-2-44: Edition 3.0 (2017-06) – FDA Recognition Number: 12-309, Particular Requirements for the Safety of X-ray Equipment for Computed Tomography
- IEC 62366-1:2015+AMD1:2020 (Consolidated Text) – FDA Recognition Number: 5-129. Application of Usability Engineering to Medical Devices
{10}
Conclusion
All verification and validation activities were successfully completed in accordance with Philips' quality system. Testing confirmed that the Philips iCT CT System continues to meet its intended use and satisfies all applicable user, performance, and safety requirements. The results demonstrated that the modifications do not introduce new risks or adversely affect the safety or effectiveness of the system. This comprehensive non-clinical evaluation supports the determination of substantial equivalence to the predicate device under 21 CFR 807.92(b).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.