K250622 · Ortho Solutions UK , Ltd. · HSB · Aug 7, 2025 · Orthopedic
Device Facts
Record ID
K250622
Device Name
Succession™ AFN System
Applicant
Ortho Solutions UK , Ltd.
Product Code
HSB · Orthopedic
Decision Date
Aug 7, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Succession™ Ankle Fusion Nail (AFN) System is a tibiotalocalcaneal (TTC) solid fusion system that has been developed for the following indications: - Failed ankle replacement - Arthritis of ankle and subtalar joint - Correcting neuromuscular imbalance of hindfoot, where bone fusion is required - Revision of failed ankle and/or subtalar fusion - Revision of failed Tibiotalocalcaneal (TTC) fusion - Talar Avascular Necrosis (AVN) - Charcot - Trauma - Neuroarthropathy - Pseudoarthrosis - Rheumatoid arthritis
Device Story
Succession™ AFN System is a cannulated intramedullary fixation rod system for tibiotalocalcaneal arthrodesis. Components include titanium (Ti-6Al-4V ELI) nails (10-12mm diameter), end caps, and cortical screws. Used by orthopedic surgeons in clinical settings to fuse tibiotalar and subtalar joints. Procedure involves preparing tibial intramedullary canal, inserting nail, and placing screws through medial-to-lateral slots for fixation in tibia, talus, and calcaneus. Dynamic compression slots allow internal/external compression across joints. Posterior-to-anterior screw provides additional calcaneal-talar fixation. End caps prevent bony ingrowth into nail cannulation. System provides hindfoot stability to facilitate bone fusion. Benefits include correction of deformities and stabilization of failed joints.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F543 (torsional yield, driving torque, pullout strength, self-tapping force). Cyclic fatigue testing performed on nail device. Comparative analyses conducted for bending/torsional strength and corrosion resistance. All results demonstrate substantial equivalence to predicate devices.
Indicated for patients requiring tibiotalocalcaneal (TTC) fusion due to failed ankle replacement, arthritis of ankle/subtalar joint, neuromuscular imbalance of hindfoot, revision of failed fusion, Talar Avascular Necrosis, Charcot, trauma, neuroarthropathy, pseudoarthrosis, or rheumatoid arthritis.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
In2Bones TriWay® Tibiotalocalcaneal (TTC) Arthrodesis System (K173811)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 7, 2025
Ortho Solutions UK Ltd.
Alex Faley
Product Development Engineer
West Station Business Park, Spital Road, Unit 5
Maldon, ESS CM9 6FF
United Kingdom
Re: K250622
Trade/Device Name: Succession™ AFN System
Regulation Number: 21 CFR 888.3020
Regulation Name: Intramedullary fixation rod
Regulatory Class: Class II
Product Code: HSB
Dated: July 7, 2025
Received: July 7, 2025
Dear Alex Faley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250622 - Alex Faley
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250622 - Alex Faley
Page 3
Sincerely,
Joseph P. Russell
-S
Digitally signed by Joseph P. Russell
Date: 2025.08.07 08:26:01 -04'00'
for: Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Succession™ AFN System
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K250622 | ? |
| Please provide the device trade name(s). | | ? |
| Succession™ AFN System | | |
| Please provide your Indications for Use below. | | ? |
| The Succession™ Ankle Fusion Nail (AFN) System is a tibiotalocalcaneal (TTC) solid fusion system that
has been developed for the following indications:
- Failed ankle replacement
- Arthritis of ankle and subtalar joint
- Correcting neuromuscular imbalance of hindfoot, where bone fusion is required
- Revision of failed ankle and/or subtalar fusion
- Revision of failed Tibiotalocalcaneal (TTC) fusion
- Talar Avascular Necrosis (AVN)
- Charcot
- Trauma
- Neuroarthropathy
- Pseudoarthrosis
- Rheumatoid arthritis | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K250622, Page 1 of 3
Prepared on: 06-August-2025
# OrthoSolutions Group
## 510(k) Summary
**Contact Details:**
Applicant Name: OrthoSolutions UK Ltd.
Applicant Address: West Station Business Park, Spital Road, Unit 5 Maldon ESS CM9 6FF United Kingdom
Applicant Contact: Mr. Alex Faley
Applicant Contact Email: alex.faley@orthosol.com
Applicant Contact Telephone: 419-577-9296
**Device Name:**
Device Trade Name: Succession™ AFN System
Common Name: Rod, Fixation, Intramedullary and Accessories
Classification Name: Intramedullary Fixation Rod
Regulation Number: 888.3020
Product Code: HSB
**Legally Marketed Predicate Devices:**
Predicate Number: K121575 (Primary Predicate)
Predicate Trade Name: Oxford™ Ankle Fusion Nail (AFN) System
Predicate Number: K173811 (Secondary Predicate)
Predicate Trade Name: In2Bones TriWay® Tibiotalocalcaneal (TTC) Arthrodesis System
**Device Description Summary:**
The Succession™ AFN System is comprised of a set of straight 10mm, 11mm, and 12 mm diameter cannulated nails, 12.5mm diameter end caps, 5mm and 6mm diameter cortical screws, and accessory instrumentation. All implants are composed of Titanium (Ti-6Al-4V ELI) and provided to the user sterile packed with a combination of both sterile and non-sterile accessory instrumentation to assist in implantation. Each of the accessory instruments provided with the subject device contain substantially equivalent materials to the Oxford™ Ankle Fusion Nail System predicate device instrumentation, including stainless steel (ASTM F899 and ASTM F138 / ISO 5832-1) and Fluorinated Ethylene Propylene direct contacting materials.
{5}
K250622, Page 2 of 3
Prepared on: 06-August-2025
During the procedure, following preparation of the tibial intramedullary canal (using the device accessory instrumentation) and placement of the intramedullary fixation nail, the provided 5mm screws are inserted into the medial to lateral (M/L) slots of the implant, allowing for fixation in the tibia, talus, and calcaneus. In both the talus and calcaneus, dynamic compression slots are included in the nail and are to be used with the internal and/or external methods of compression across the tibiotalar and subtalar joints. In addition, a 6mm screw is placed through the only angled posterior to anterior (P/A) slot located at the distal end of the nail implant, allowing for additional fixation between the calcaneus and talus. Following the placement of all screws, the end caps are implanted to add stability to the distal end of the nail and prevent excess bony ingrowth from occurring within the nail cannulation. Overall, the Succession™ Ankle Fusion Nail is intended to effectively fuse the tibiotalar and subtalar joints together, providing stability in the hindfoot region to facilitate tibiotalocalcaneal arthrodesis.
**Intended Use/Indications for Use:** The Succession™ Ankle Fusion Nail (AFN) System is a tibiotalocalcaneal (TTC) solid fusion system that has been developed for the following indications:
- Failed ankle replacement
- Arthritis of ankle and subtalar joint
- Correcting neuromuscular imbalance of hindfoot, where bone fusion is required
- Revision of failed ankle and/or subtalar fusion
- Revision of failed Tibiotalocalcaneal (TTC) fusion
- Talar Avascular Necrosis (AVN)
- Charcot
- Trauma
- Neuroarthropathy
- Pseudoarthrosis
- Rheumatoid arthritis
**Indications for Use Comparison:** Other than the system name, there are no changes to the device Indications for Use Statement in comparison to the predicate.
**Technological Comparison:** The Subject device is substantially equivalent to the OrthoSolutions Oxford™ (now called Oxbridge™) Ankle Fusion Nail System (Primary Predicate - K121575) and In2Bones TriWay® Tibiotalocalcaneal (TTC) Arthrodesis System predicates (Secondary Predicate - K173811) and the new components do not raise any different questions regarding the device safety or effectiveness. The subject device has the same technological characteristics as the original Oxbridge™ device including the same materials, chemical composition, manufacturing methods, and a similar size range. Other than the change in trade name, there are no substantial changes to the device labeling.
The subject device utilizes the same sterilization method (gamma irradiation) and equivalent implant packaging as the original primary predicate device. Likewise, the new sterile accessories will utilize the same sterilization method (gamma irradiation) and equivalent implant packaging as the Oxbridge™ sterile accessories. The new non-sterile Succession™ class 1 accessory
{6}
K250622, Page 3 of 3
Prepared on: 06-August-2025
instrumentation will utilize the same method of sterilization as the non-sterile predicate device accessory instrumentation - Steam Sterilization.
The shelf-life of the subject Succession™ device is equivalent to the predicate OrthoSolutions Oxbridge™ device.
The subject device will utilize a similar set of class 1 accessory instrumentation with the addition of an extraction adapter, new Hudson connection handles, an impactor, and an external compression instrument (for optional use).
There are no new materials that directly contact the body and/or tissue, new points of contact between dissimilar metals, or novel manufacturing techniques/processes being introduced into the subject system. Therefore, there are no significant changes to device biocompatibility.
## Non-Clinical and/or Clinical Tests
## Summary:
Mechanical Testing was completed on the screw implant devices per ASTM F543 Annex A1 through Annex A4. The results of these tests indicated that the Succession™ device cortical screws are substantially equivalent with regards torsional yield strength, driving torque, pullout strength, and self-tapping force.
Cyclic Fatigue Testing was completed on the nail device (Oldenburg G, Kaschner H, DiCiccoJ, et al.). The results of this analysis indicated that the Succession™ nail device is substantially equivalent with regards to fatigue strength.
Comparative analyses were completed to address the bending and torsional strength of the nail, the bending strength of the screws, and the corrosion resistance of the worst-case components, with all results indicating that the subject Succession™ device is substantially equivalent to the predicate device.
## Conclusions:
In conclusion, the components that are the subject of this 510(k) are substantially equivalent to the K121575 Ortho Solutions Oxford™ Ankle Fusion Nail System (now called the Oxbridge™ Ankle Fusion Nail System) and the new Succession™ components do not raise any different questions regarding the safety or effectiveness of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.