Easyflow (103-200; 103-300); Easyflow Duo (103-210; 103-310)
K250610 · Sorin Group Italia S.R.L. · DWF · Jun 27, 2025 · Cardiovascular
Device Facts
Record ID
K250610
Device Name
Easyflow (103-200; 103-300); Easyflow Duo (103-210; 103-310)
Applicant
Sorin Group Italia S.R.L.
Product Code
DWF · Cardiovascular
Decision Date
Jun 27, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4210
Device Class
Class 2
Indications for Use
The EasyFlow Aortic Cannula is intended for use in the perfusion of the aorta during cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass for up to 6 hours. The EasyFlow Duo Cannula with guidewire is intended for use with cardiopulmonary bypass as an arterial return femoral cannula for up to 6 hours.
Device Story
Easyflow and Easyflow Duo are single-use, sterile, non-pyrogenic cannulae for arterial return during cardiopulmonary bypass. Devices consist of PVC polymer tubes with wire-reinforced distal sections and perforated tips for fluid flow. Easyflow Duo includes a guidewire, needle, and vascular dilators for femoral access; a malleable obturator facilitates vessel placement. During surgery, the surgeon inserts the cannula into the artery, removes the obturator, and connects the proximal 3/8" barbed connector to the arterial line of the bypass circuit. The device facilitates blood flow from the bypass machine to the patient. Modifications from predicate devices include the removal of phthalate plasticizers (DEHP and DnHP) from the PVC material. Devices are intended for use by surgeons in clinical settings.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by non-clinical performance data, including verification/validation testing (visual inspection, pull strength, kink test, flow rate/pressure drop, blood pathway integrity) and comprehensive biocompatibility testing (cytotoxicity, sensitization, systemic toxicity, hemocompatibility, genotoxicity).
Technological Characteristics
Materials: PVC polymer (phthalate-free). Design: Open lumen tube with wire-reinforced distal sections and multiple perforations. Connectivity: 3/8" barbed connector. Sterilization: Ethylene oxide. Single-use. No electronic or software components.
Indications for Use
Indicated for patients undergoing cardiovascular surgery requiring extracorporeal cardiopulmonary bypass for up to 6 hours. EasyFlow Aortic Cannula is for aortic perfusion; EasyFlow Duo is for arterial return via femoral access.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Easyflow Cannula with guidewire (Easyflow Duo) (K091542)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 27, 2025
Livanova USA Inc
Martina Carlini
RA Specialist
Sorin Group Italia S.r.l.
86, Via statale 12 Nord
Mirandola, IT 41037
Italy
Re: K250610
Trade/Device Name: Easyflow (103-200; 103-300); Easyflow Duo (103-210; 103-310)
Regulation Number: 21 CFR 870.4210
Regulation Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing
Regulatory Class: Class II
Product Code: DWF
Dated: February 28, 2025
Received: May 30, 2025
Dear Martina Carlini:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250610 - Martina Carlini
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250610 - Martina Carlini
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachel E. Neubrander -S
for Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250610 | |
| Device Name Easyflow (103-200; 103-300) Easyflow Duo (103-210; 103-310) | |
| Indications for Use (Describe) The EasyFlow Aortic Cannula is intended for use in the perfusion of the aorta during cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass for up to 6 hours. The EasyFlow Duo Cannula with guidewire is intended for use with cardiopulmonary bypass as an arterial return femoral cannula for up to 6 hours. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
LivaNova
Health innovation that matters
# 510(k) Summary
(in accordance with 21 CFR 807.92)
510(k) Number: K250610
## I. Applicant Information
Applicant:
LIVANOVA USA INC
14401 West 65th Way
Arvada, CO 80004
Contact Person:
Martina Carlini
Regulatory Affairs Specialist
Tel: +39 0535 29811
e-mail: martina.carlini@livanova.com
Application Correspondent:
SORIN GROUP ITALIA S.R.L.
Via Statale 12 Nord, 86
Mirandola, MO 41047 Italy
Contact Person:
Martina Carlini
Regulatory Affairs Specialist
Tel: +39 0535 29811
e-mail: martina.carlini@livanova.com
Date Prepared: June 27th 2025
## II. Subject Device Identification
Device Trade Name: Easyflow (103-200; 103-300)
Easyflow Duo (103-210; 103-310)
Classification Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing
Regulation Number: 21 CFR 870.4210
Product Code: DWF
Classification: Class II
Classification Panel: Cardiovascular
Special 510(k) – Easyflow & Easyflow Duo cannulae
{5}
LivaNova
Health innovation that matters
## III. Predicate Device
The Easyflow & Easyflow Duo are substantially equivalent to the following cleared predicate device. Both modified and unmodified models have the same fundamental scientific technology and intended use:
| Easyflow | 510(k) Number: | K060101 |
| --- | --- | --- |
| | Device Trade Name: | Easyflow Aortic Cannula |
| | Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, or tubing |
| | Regulation Number: | 21 CFR 870.4210 |
| | Product Code: | DWF |
| | Classification: | Class II |
| | Classification Panel: | Cardiovascular |
| Easyflow Duo | 510(k) Number: | K091542 |
| | Device Trade Name: | Easyflow Cannula with guidewire (Easyflow Duo) |
| | Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, or tubing |
| | Regulation Number: | 21 CFR 870.4210 |
| | Product Code: | DWF |
| | Classification: | Class II |
| | Classification Panel: | Cardiovascular |
## IV. Device Description
Easyflow & Easyflow Duo are single-use, non-toxic, non-pyrogenic fluid path devices and supplied sterile and individually packaged.
The Easyflow & Easyflow Duo are composed of the cannula and a malleable obturator inserted into the cannula to allow the placement of the cannula along the Artery.
The cannula is an open lumen PVC polymer tube incorporating wire reinforcement in distal sections. The distal sections of the cannula are perforated with multiple holes at multiple stages to allow fluid flow. The clear proximal section is not reinforced to allow clamping; the proximal end terminates in a 3/8" barbed connector for standard cardiopulmonary bypass tubing
In the Easyflow Duo models the device has a 0.035" hole punched in the very distal tip of the cannula and the obturator has an approx. 0.035" internal diameter central lumen to accommodate a 0.035" guide wire used to drive the cannula along the vessel till its final placement. Once the cannula is positioned onto the vessel the surgeon removes the obturator from the cannula tubing and connects it to the arterial line.
The Easyflow is packaged as sterile stand-alone device while the Easyflow Duo is packaged into a tray containing also a Guidewire, a Needle and Vascular Dilators in order to have a kit for vascular access packaged together with the cannula.
The Easyflow & Easyflow Duo are the modified version of the disposables currently marketed as the Easyflow cannula (K060101) and the Easyflow Duo cannula (K091542).
Both modified and unmodified cannulae models are recommended for use as an arterial return cannula during cardiopulmonary bypass up to six hours
Special 510(k) – Easyflow & Easyflow Duo cannulae
{6}
LivaNova
Health innovation that matters
V. Indications for Use
The EasyFlow Aortic Cannula is intended for use in the perfusion of the aorta during cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass for up to 6 hours
The EasyFlow Duo Cannula with guidewire is intended for use with cardiopulmonary bypass as an arterial return femoral cannula for up to 6 hours
VI. Summary of Technical Characteristics
The Easyflow & Easyflow Duo have the same fundamental technological characteristics, principles of operation and control mechanisms as the unmodified devices.
The PVC materials of the body, tip and suture ring/ ring stop of the Easyflow & Easyflow Duo were changed within the modified Easyflow & Easyflow Duo in order to remove the phthalates (DEHP and DnHP) used as plasticizers currently present in the device.
No other design changes have been made to the devices.
The devices are ethylene oxide sterilized and have a non-pyrogenic fluid path. They are for single use only.
VII. Non-Clinical Performance Data
Livanova USA Inc. has conducted extensive verification and validation testing of the Easyflow & Easyflow Duo cannula.
The devices passed all the testing in accordance with national and international standards.
In particular, the performed tests are the following:
| TEST CLASSIFICATION | TEST DESCRIPTION |
| --- | --- |
| Verification/Validation | Visual inspection |
| Verification/Validation | Connector testing 180° pull test |
| Verification/Validation | Cannula clamp test |
| Verification/Validation | Blood pathway integrity |
| Verification/Validation | Flow rate and pressure drop through cannulae |
| Verification/Validation | Cannula kink test |
| Verification/Validation | Label legibility |
| Verification/Validation | Drainage cannula collapse resistance |
| Verification/Validation | Cannula integrity |
| Verification/Validation | Pull strength |
| Verification/Validation | Blood trauma characterization test |
| Biocompatibility | Cytotoxicity |
| Biocompatibility | Intracutaneous reactivity |
| Biocompatibility | Sensitization |
| Biocompatibility | Systemic Toxicity |
| Biocompatibility | Pyrogenicity (material mediated) |
| Biocompatibility | Hemocompatibility (Hemolysis, Coagulation, Complement activation, Platelet and Leukocyte count assay) |
| Biocompatibility | Genotoxicity (Mutation Assay, Mouse lymphoma) |
Special 510(k) – Easyflow & Easyflow Duo cannulae
{7}
LivaNova
Health innovation that matters
## VIII. Clinical Performance Data
No clinical testing was conducted in support of the Easyflow & Easyflow Duo, as the indications for use and technical characteristics are equivalent to those of the predicate devices, which have been on the market for several years with proven safety and efficacy of use. The non-clinical testing summarized in this submission supports the substantial equivalence of the subject devices with the predicate devices when used according to their intended use.
## IX. Statement of Substantial Equivalence
Based on equivalent intended use and technological the Easyflow & Easyflow Duo can be deemed to be substantially equivalent to their predicate devices:
- the unmodified Easyflow & Easyflow Duo, cleared under K060101 and K091542 respectively;
The Easyflow & Easyflow Duo, as designed and manufactured, do not raise new questions regarding safety and effectiveness as compared to their predicate devices and are determined to be substantially equivalent to their predicate devices listed above.
Special 510(k) – Easyflow & Easyflow Duo cannulae
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.