← Product Code [OLO](/productcode/OLO) · K250608

# Mako Total Knee Application (3.0); Hybrid Tip Pointer (K250608)

_Mako Surgical Corp. · OLO · Apr 25, 2025 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250608

## Device Facts

- **Applicant:** Mako Surgical Corp.
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Apr 25, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Mako System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The Mako System is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include: • Total Knee Arthroplasty (TKA) The implant systems compatible with the system: • Triathlon Total Knee System (CR/CS/PS/PSR cemented and cementless primary) • Triathlon Total Knee System (TS inserts cemented primary)

## Device Story

Mako System is a stereotactic instrument for orthopedic surgery; includes robotic arm, optical detector, camera, computer, and software. Inputs: patient CT data. Operation: system uses CT data for pre-surgical planning and intraoperative navigation; robotic arm acts as intelligent tool guide; provides haptic guidance by constraining bone preparation within software-defined spatial boundaries. Output: haptic feedback and visual navigation for surgeon. Used in OR by surgeons. Benefits: assists in precise implant placement and anatomical navigation. Modifications include new battery-powered active pointer for registration/landmark selection and updated software (programming language, UI, planning tools).

## Clinical Evidence

No clinical data. Performance evaluated via non-clinical testing: instrumentation cleaning, sterility, and biocompatibility; Hybrid Tip Pointer verification; software functional and performance testing; bone registration and resection accuracy; cadaveric design validation; and EMC/electrical safety testing.

## Technological Characteristics

Stereotactic instrument; robotic arm; optical detector; camera; computer. Uses patient CT data for navigation. Connectivity: integrated system. Software: updated programming language and UI. Instrumentation: includes new battery-powered active pointer. Sterilization: reusable and disposable instruments. Standards: EMC and electrical safety testing performed.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Mako Total Knee Application ([K241011](/device/K241011.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

April 25, 2025

Mako Surgical Corp.
Shikha Khandelwal
Senior Principal Regulatory Affairs Specialist
3365 Enterprise Avenue
Weston, Florida 33331

Re: K250608
Trade/Device Name: Mako Total Knee Application (3.0); Hybrid Tip Pointer
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: February 7, 2025
Received: February 28, 2025

Dear Shikha Khandelwal:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250608 - Shikha Khandelwal
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

{2}

K250608 - Shikha Khandelwal
Page 3

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali

Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K250608  |   |
|  Device Name Mako Total Knee Application (3.0)  |   |
|  Indications for Use (Describe) The Mako System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The Mako System is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include: • Total Knee Arthroplasty (TKA) The implant systems compatible with the system: • Triathlon Total Knee System (CR/CS/PS/PSR cemented and cementless primary) • Triathlon Total Knee System (TS inserts cemented primary)  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

stryker

# 510(k) SUMMARY - K250608

Sponsor: Mako Surgical Corp.
3365 Enterprise Ave
Weston, FL 33331

Contact Person: Shikha Khandelwal, PhD
Sr. Principal Regulatory Affairs Specialist
Shikha.Khandelwal@stryker.com
510-847-9264

Date Prepared: Feb 7, 2025

Proprietary Name: Mako Total Knee Application

Common Name: Total Knee Application (TKA)

Regulation Name: Stereotaxic Instrument

Regulation Number: 21 CFR Section 882.4560

Device Classification: Class II

Product Code: OLO

# Substantial Equivalence Claimed To:

The subject device, the Mako Total Knee Application, is substantially equivalent to the predicate device, the Mako Total Knee Application, cleared in K241011.

# Device Modifications:

Instrumentation: A new battery-powered active pointer is being introduced to support the registration, landmark selection and page selection workflow.

Software Changes: The Mako Total Knee Application is being modified to update the programming language, update the look of the user interface, enable additional features, add a new optional planning tool and enhance the overall user experience.

Labeling: Labeling provided with the device has been updated to reflect the proposed modifications.

Mako Surgical Corp.
K250608

{5}

stryker

# Description:

The Mako System with the subject Mako Total Knee Application is a stereotactic instrument that includes a robotic arm, an integrated cutting system, an optical detector, a camera, a computer, dedicated instrumentation, an operating software, and tools and accessories. The system's architecture is designed to support total and partial knee procedures and total hip procedures. With application specific hardware and software, the system provides haptic guidance during orthopedic surgical procedures by using patient CT data to assist a surgeon with pre-surgical planning, implant placement, and interpretive/intraoperative navigation of the patient's anatomy.

Once configured for a specific application, the Mako robotic arm can serve as the surgeon's "intelligent" tool holder or tool guide by passively constraining the preparation of an anatomical site for an orthopedic implant with software-defined spatial boundaries.

# Summary of Technological Characteristics Compared to Predicate Device:

The technological characteristics of the Mako Total Knee Application compared to the predicate device are listed below:

|  Technological Characteristics | Mako Total Knee Application | Mako Total Knee Application – K241011  |
| --- | --- | --- |
|  Major Components | Robotic arm, Stryker Q Guidance System, cutting system. | Robotic arm, Stryker Q Guidance System, cutting system.  |
|  Tools/accessories | Various reusable and disposable instruments | Various reusable and disposable instruments  |
|  Image Use | CT | CT  |

# Intended Use

The subject device has the same intended use as that specified in the cleared 510(k) premarket notification for the predicate device listed in this 510(k) premarket notification.

# Indications for Use: Mako Total Knee Application

The Mako System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

The Mako System is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include:

Mako Surgical Corp.

K250608

{6}

stryker

- Total Knee Arthroplasty (TKA)

The implant systems compatible with the system:

- Triathlon Total Knee System (CR/CS/PS/PSR cemented and cementless primary)
- Triathlon Total Knee System (TS inserts cemented primary)

**Performance Data** – The modified Mako System with the subject Mako Total Knee Application has been evaluated through the following non-clinical performance testing:

- Instrumentation cleaning, sterility, and biocompatibility assessment
- Hybrid Tip Pointer verification
- Software Functional Testing
- Software Performance Verification
- Bone Registration and Resection Accuracy
- Cadaveric Design Validation
- EMC and electrical safety

**Conclusions:**

The hardware and software of the subject Mako System with Mako Total Knee Application is equivalent to the hardware and software of the predicate Mako System with Mako Total Knee Application. The software modifications being made do not impact the intended use or the fundamental technology of the device. Furthermore, performance testing demonstrates that the characteristics of the subject Mako Total Knee Application are equivalent to the characteristics of the predicate device. The subject device is also as safe and as effective as the predicate device and does not raise different questions of safety and effectiveness. Therefore, performance testing supports a determination of substantial equivalence.

Mako Surgical Corp.
K250608

---

**Source:** [https://fda-staging.innolitics.com/device/K250608](https://fda-staging.innolitics.com/device/K250608)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).