K250584 · Boston Scientific Corporation · KNS · Jun 11, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K250584
Device Name
Rezum System
Applicant
Boston Scientific Corporation
Product Code
KNS · Gastroenterology, Urology
Decision Date
Jun 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rezüm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia(BPH). It is indicated for men ≥ 50 years of age with a prostate volume 30 cm3 ≤ 150 cm3. The Rezüm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
Device Story
The Rezüm System treats benign prostatic hyperplasia (BPH) via convective water vapor thermal therapy. The system consists of a reusable generator and a sterile, single-use delivery device. The physician injects controlled 9-second doses of water vapor into the prostate transition zone or median lobe. The vapor condenses upon contact with tissue, releasing thermal energy that denatures cell membranes and causes cell death. This process reduces prostate volume and collapses vasculature in the treatment zone, facilitating a bloodless procedure. Denatured tissue is subsequently absorbed by the body or sloughed off and expelled via urination. The procedure is performed to relieve urinary symptoms and obstructions. The device is identical to the predicate; the update expands the indicated prostate volume range to 150 cm³.
Clinical Evidence
Evidence includes a systematic review, meta-analysis, and a manufacturer-sponsored prospective, non-randomized, single-arm study. The study compared safety and effectiveness in patients with prostate volumes >80 cm³ and ≤150 cm³ against the pivotal clinical study data for patients ≤80 cm³. Results showed comparable functional and quality of life improvements, no negative impact on sexual function, and similar adverse event rates between the two groups. No unexpected adverse events were reported.
Technological Characteristics
The system utilizes radiofrequency current to generate water vapor for thermal ablation. Components include a reusable generator and a sterile, single-use delivery device kit (delivery device, syringe, spike adaptor, water vial). The device is an endoscopic electrosurgical unit (21 CFR 876.4300). Technological characteristics are identical to the predicate device.
Indications for Use
Indicated for men ≥ 50 years of age with benign prostatic hyperplasia (BPH) and prostate volume between 30 cm³ and 150 cm³, including hyperplasia of the central zone and/or median lobe.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 11, 2025
Boston Scientific Corporation
Joseph Tabor
Regulatory Affairs Specialist
100 Boston Scientific Way
Marlborough, Massachusetts 01752
Re: K250584
Trade/Device Name: Rezum System
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic Electrosurgical Unit and accessories
Regulatory Class: II
Product Code: KNS
Dated: February 27, 2025
Received: May 6, 2025
Dear Joseph Tabor:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250584 - Joseph Tabor
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K250584 - Joseph Tabor
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark J. Antonino -S
Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250584
Device Name
Rezüm System
Indications for Use (Describe)
The Rezüm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia(BPH). It is indicated for men ≥ 50 years of age with a prostate volume 30 cm3 ≤ 150 cm3. The Rezüm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K250584
Page 1 of 3
# 510(k) Summary
## A. Date Prepared
This 510(k) summary was prepared February 27, 2025
## B. Submitter
Boston Scientific Corporation
Urology Division
100 Boston Scientific Way
Marlborough, MA 01752
## C. Contacts
Joseph Tabor
Regulatory Affairs Specialist II
Joseph.Tabor@bsci.com
(508) 382-0313
Alternate:
Cody Johnson
Regulatory Affairs Manager
Cody.Johnson@bsci.com
(952) 930-5736
## D. Subject Device
| Trade Name | Rezūm System |
| --- | --- |
| Common Name | Vapor Ablation Device |
| Regulation Name | Endoscopic electrosurgical unit and accessories |
| Regulation Number | 21 CFR 876.4300 |
| Classification | Class II |
| Product Code | KNS |
## E. Predicate Device
| Trade Name | Rezūm System |
| --- | --- |
| Common Name | Vapor Ablation Device |
| Regulation Name | Endoscopic electrosurgical unit and accessories |
| Regulation Number | 21 CFR 876.4300 |
| Classification | Class II |
| Product Code | KNS |
| 510(k) Submitter/Holder | Boston Scientific Corporation |
| 510(k) #/Clearance Date | K191505/cleared August 2, 2019 |
Page 1 of 3
{5}
K250584
Page 2 of 3
# 510(k) Summary
## F. Device Description
The Rezūm System is designed to treat patients with bothersome urinary symptoms associated with benign prostatic hyperplasia (BPH). The Rezūm System utilizes radiofrequency current to generate "wet" thermal energy in the form of water vapor, which is then injected into the transition zone and/or median lobe of the prostate tissue in controlled 9-second doses. The vapor that is injected into the prostate tissue rapidly disperses through the interstitial space between the tissue cells. As the vapor cools, it condenses immediately on contact with tissue and the stored thermal energy is released, denaturing the cell membranes and causing cell death.
The denatured cells are absorbed by the body, which reduces the volume of prostate tissue adjacent to the urethra. The vapor condensation process also causes a rapid collapse of vasculature in the treatment zone, resulting in a bloodless procedure.
Following thermal therapy for BPH, small pieces of coagulated tissue may slough off and be expelled via urination. This sloughing process may continue for a few months post-procedure depending on the rate of healing.
The Rezūm System consists of the following:
- Rezūm Generator – reusable, non-sterile capital equipment, provided with one power cord
- Rezūm Delivery Device Kit – sterile, single-use kit containing the following disposable components:
- One sterile Delivery Device with cable and tubing
- One sterile syringe
- One sterile spike adaptor
- One sterile water vial
## G. Intended Use/Indications for Use
The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men ≥ 50 years of age with a prostate volume 30 cm³ ≤ 150 cm³. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
## H. Technological Characteristics Compared to the Predicate Device
The technological characteristics of the proposed (subject) Rezūm System are identical to those of the predicate device. There are no changes to any device features. The only change is a modification to the Indications for Use to expand the prostate volume upper limit from ≤ 80 cm³ to ≤ 150 cm³. The change in Indications for Use does not create a new intended use. The predicate and the proposed Rezūm System are both intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. The modification to the Indications for Use is supported by manufacturer-sponsored and independent clinical studies and the clinical evidence demonstrates that there are no new safety or effectiveness concerns regarding use of the Rezūm System in patients with prostates >80 cm³ and ≤ 150 cm³.
## I. Substantial Equivalence
The proposed Rezūm System is substantially equivalent to the predicate Rezūm System (K191505). The proposed and predicate Rezūm System are both intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH, and the modification to the
Page 2 of 3
{6}
K250584
Page 3 of 3
## 510(k) Summary
Indications for Use does not introduce any new safety or effectiveness concerns. The Rezūm System intended use, principle of operation and technological characteristics remain the same.
## J. Performance Data
The Indications for Use modification to the proposed Rezūm System has been evaluated in comparison to the predicate device. The expanded indication is supported by a systematic review and meta-analysis of clinical studies that examined safety and effectiveness of the Rezūm System in symptomatic BPH patients with prostate volumes >80 cm³ and compared outcomes to patients with prostate volumes ≤80 cm³. In addition, data from a manufacturer-sponsored prospective, non-randomized, single-arm study that evaluated safety and effectiveness of the Rezūm System in patients with prostate volumes >80 cm³ and ≤150 cm³ was compared to data from the Rezūm System pivotal clinical study. The clinical evidence showed overall comparable results between patients with prostate volumes >80 cm³ and ≤80 cm³. Significant and similar improvements in functional and quality of life outcomes were demonstrated in both patient populations, with no negative impact to sexual function. Patients with prostate volumes >80 cm³ had generally similar rates of adverse events, including serious complications. No unexpected adverse events were reported, and all adverse events were consistent with those reported in the Rezūm System pivotal clinical study. Use of the proposed device in the expanded patient population does not result in any new concerns of safety or effectiveness.
## K. Conclusion
Based on the intended use, technological characteristics, and the clinical evidence presented in this submission to support use of the device in the expanded patient population, the Rezūm System is substantially equivalent to the predicate device.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.