K250576 · Karma Mobility Co, Ltd. · IOR · Nov 4, 2025 · Physical Medicine
Device Facts
Record ID
K250576
Device Name
Flexx Junior (FLX-J00, FLX-J50)
Applicant
Karma Mobility Co, Ltd.
Product Code
IOR · Physical Medicine
Decision Date
Nov 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
Pediatric
Indications for Use
Flexx Junior manual wheelchair is intended to provide mobility to pediatrics limited to a seating position.
Device Story
Flexx Junior is a pediatric manual wheelchair providing mobility, comfort, and postural support. Available in self-propelled (FLX-J00 fixed back, FLX-J50 foldable back) and transit configurations. Input: user-applied force via handrims or attendant-applied force via push handles. Operation: lightweight aluminum folding frame with adjustable seat width, depth, height, backrest angle, and armrest/footrest positioning to accommodate pediatric growth. Output: physical mobility for the user. Used in indoor/limited outdoor environments. Healthcare providers/caregivers use the device to facilitate patient transport and positioning. Benefits: provides independent or assisted mobility, ergonomic support, and postural stability for children with mobility limitations. Includes safety features like push-to-lock brakes, anti-tippers, and crash-test compliance (ISO 7176-19) for use as a seat in motor vehicles.
Clinical Evidence
Bench testing only. Performance verified via ISO 7176 series (stability, brakes, fatigue, strength, dimensions, crash testing) and ISO 10993 biocompatibility/flammability standards. All tests passed.
Technological Characteristics
Lightweight aluminum folding frame; X-brace mechanism; manual propulsion via handrims; adjustable seat/backrest/armrest/footrest geometry; quick-release wheels. Complies with ISO 7176-1, 3, 5, 7, 8, 11, 13, 15, 19, 22. Biocompatibility per ISO 10993-1, 5, 10, 23. Flammability per EN 1021-1/2.
Indications for Use
Indicated for pediatric patients limited to a seating position requiring a manual wheelchair for mobility.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 4, 2025
Karma Mobility Co Ltd
Anoohya Pulipaka
Regulatory Affairs
30 Moo 7 Tambol Khamtaleso Amphur (KORAT)
Khamthleso, Nakhonratchasima 30280
Thailand
Re: K250576
Trade/Device Name: Flexx Junior (FLX-J00, FLX-J50)
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR, LBE
Dated: September 9, 2025
Received: September 18, 2025
Dear Anoohya Pulipaka:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250576 - Anoohya Pulipaka
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250576 - Anoohya Pulipaka
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250576
Device Name
Flexx Junior
Models: FLX-J00 and FLX-J50
Indications for Use (Describe)
Flexx Junior manual wheelchair is intended to provide mobility to pediatrics limited to a seating position.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(K) Summary - Flexx Junior (K250576)
| 510(K) Summary | |
| --- | --- |
| Applicant Contact Details | 21CFR807.92(a)(1) |
| Applicant Name | KARMA MOBILITY., CO LTD |
| Applicant Address | 30 Moo 7 Tambol Khamtaleso Amphur
Khamtaleso Nakhonratchasima (KORAT)
30280 THAILAND |
| Applicant Telephone number | +886-5-206-6688 |
| Applicant Contact Name | Pulipaka Anoohya |
| Applicant Contact Email | eva@karma.com.tw |
| Device Information | 21CFR807.92(a)(2) |
| --- | --- |
| Name of the device/ Trade Name | Flexx Junior |
| Common name | Mechanical Wheelchair |
| Classification name | Wheelchair, Mechanical |
| Device Classification | Class I |
| Classification Product Code | IOR |
| Regulation number | 890.3850 |
| Legally Marketed Predicate Device | 21CFR807.92(a)(3) |
| --- | --- |
| Predicate Device 510(K) Number | K223527 |
| Name of the Device/ Trade Name | Ki Mobility Spark |
| Common Name | Mechanical Wheelchair |
| Classification Name | Wheelchair, Mechanical |
| Device Classification | Class I |
| Classification Product Code | IOR |
| Regulation Number | 890.3850 |
Device Description: [21 CFR 807.92(a)(4)]
Designed for Comfort and Growth: The Flexx Junior is a pediatric manual wheelchair designed to provide mobility, comfort, and postural support for children with mobility limitations. Available in self-propelled (Model FLX-J00 with fixed backrest, Model FLX-J50 with foldable backrest) and transit configurations, it is primarily intended for indoor use with limited outdoor capability on smooth surfaces. The wheelchair offers both self-propulsion with 20" or 22" quick-release rear wheels and transit propulsion with 14" quick-release rear wheels. The device supports users weighing up to (165 lbs) 75 kg and is engineered for growth and incorporates adjustable features to accommodate growth.
Device Specifications:
{5}
The Flexx Junior is constructed with a lightweight aluminum folding frame, ensuring durability and ease of transport. Its compact folded dimensions facilitate storage and transport in vehicles. The device offers adjustable seat width, seat depth, and backrest height to ensure a customized fit as the child grows. The wheelchair's front caster wheel allows better maneuverability, while larger rear wheels improve stability and ease of movement, ensuring the device's structural integrity and safety. The wheelchair's compact design ensures maneuverability in tight spaces. The total width allows for easy navigation through doorways and confined spaces, and the overall height offers appropriate support and comfort. Easy to Transport and Its folding frame is built from aluminum, making it lightweight yet durable ensuring longevity. Despite its robust design, it remains lightweight and can be folded into a compact size for easy transport.
The handrim offers optimal leverage for self-propulsion. The wheelchair's turning radius ensures easy movement in confined spaces, with the pivot width supporting safe turning. Static stability defines the wheelchair's ability to maintain balance on slopes, ensuring safety on inclines and uneven terrains. The angular ranges are consistent with predicate devices and do not raise new questions of safety or effectiveness
The seat and backrest angles provide ergonomic support, promoting good posture and comfort. Adjustable seat depth and width cater to various body types, while the seat surface height facilitates ease of transfer. The wheelchair includes adjustable footrest and armrest distances, enhancing user comfort and posture.
Ideal for Indoor and Outdoor Use - Primarily designed for indoor environments but is not suitable for rough terrains.
Stability and Safety - The Flexx Junior is tested and compliant with ISO 7176-19 standards, making it suitable as a seat in a motor vehicle. Safety features include push-to-lock brakes, standard anti-tippers, and optional accessories such as a table tray, IV pole holder, and oxygen tank holder. Flip-up, height-adjustable armrests and quick-release axles enhance usability. Engineered to support children throughout their formative years, it provides the perfect balance of flexibility, safety, and comfort as they grow.
## Intended Use/Indications for Use: [21 CFR 807.92(a)(5)]
Flexx Junior manual wheelchair is intended to provide mobility to pediatrics limited to a seating position.
## Indications for Use Comparison: 21CFR807.92(a)(5)
Flexx Junior manual wheelchair is intended to provide mobility to pediatrics limited to a seating position, aligning with the intended use of the predicate device. The Flexx Junior wheelchair is substantially equivalent to the predicate devices, Ki Mobility Spark (510(k) Number: K223527). The key specifications, intended use, indications for use, weight capacity, and core design principles align with the predicate devices, ensuring comparable safety and effectiveness.
{6}
KARMA MOBILITY CO LTD
Flexx Junior -510(k) Submission
# Key Equivalences:
1. Intended Use & Indications for Use: Both devices are intended to provide mobility to pediatrics limited to a seating position. The Flexx Junior and the predicate device, the Ki Mobility Spark, share identical intended use and indications for use.
2. Weight Capacity: Both devices support a maximum user weight of (165 lbs) 75 kg, ensuring comparable structural integrity and performance for pediatric users.
3. Frame Materials: Both utilize aluminum frames with a folding X-brace mechanism that enhances rigidity and enables width adjustability.
4. Seat and Backrest Adjustability: Both devices feature adjustable seat width, depth, height, backrest angle, and armrest positioning via discrete holes, supporting pediatric growth and ergonomic positioning. The Spark offers a wider range of seat width and depth adjustments and additional backrest styles, but these differences do not affect safety or effectiveness, as the Flexx Junior's adjustment ranges meet the needs of the intended population.
5. Propulsion and Maneuverability: Both provide self-propulsion via handrims and attendant propulsion via push handles, with quick-release rear wheels and adjustable caster wheels. The Flexx Junior offers fewer rear wheel and caster options than the Spark, but both ensure equivalent maneuverability and stability for pediatric use, with adjustable center of gravity positioning.
6. Safety and Compliance: Both are crash-tested according to ISO 7176-19 standards, suitable for use as a seat in a motor vehicle and comply with equivalent performance standards ensuring comparable safety and effectiveness.
7. Accessories and Components: Both include comparable accessories, such as push-to-lock brakes, swing-away footrests, and height-adjustable armrests. The Flexx Junior offers unique accessories (e.g., IV pole holder, table tray), while the Spark includes options like spoke guards and backpacks. These differences do not introduce new risks, as both provide equivalent functionality and safety features.
The Flexx Junior complies with the same regulatory requirements and performance standards, demonstrating substantial equivalence to the legally marketed predicate devices.
## Substantial Equivalence Comparison Table
| PREDICATE COMPARISON TABLE | | |
| --- | --- | --- |
| | Subject Device | Predicate Device |
| 510(k) Number | K250576 | K223527 |
| Name of the device | Flexx Junior | Ki Mobility Spark |
{7}
| Classification name | Wheelchair, Mechanical | Wheelchair, Mechanical |
| --- | --- | --- |
| Manufacturer | Karma Mobility Co.,LTD | Ki Mobility LLC |
| Device Classification | Class 1 | Class I |
| Classification Product Code | IOR | IOR |
| Regulation Number | 890.3850 | 890.3850 |
| Intended Use | Flexx Junior manual wheelchair is intended to provide mobility to pediatrics limited to seating position. | Ki Mobility Spark manual wheelchair is intended to provide mobility to pediatrics limited to seating position. |
| Indications for use | Flexx Junior manual wheelchair is intended to provide mobility to pediatrics limited to seating position. | Ki Mobility Spark manual wheelchair is intended to provide mobility to pediatrics limited to seating position. |
| User Weight Capacity | 165lb/75kg | 165lb/75kg |
| Frame Material | Aluminum | Aluminum |
| Frame Type | Fixed backrest (FLX-J00)
Foldable backrest (FLX-J50) | Foldable backrest only |
| Transit approved | Yes | Yes |
| Seat Width | 10" - 14" | 10" - 16" |
| Seat Depth | 11" -15" | 12" - 18" |
| Seat Height | 15.5" - 18" | Front Seat-to-Floor: 13" - 21"
Rear Seat-to-Floor: 11" - 18.5" |
| Folded Width | 12.8" - 14" | 12.6" |
| Folded Length | 31.9" - 34.3" | 29.5" |
| Folded Height | 28.5" | 25.6" |
{8}
| Product Weight | 31.5-33.7lb | 28.7lb |
| --- | --- | --- |
| Turning Radius | With user - 24"-30" Without user - 20.3"-25.6" | 28.9" |
| Backrest Height | 15"-17" | Stroller Handle: 20"-24" Height Adjustable: 13"-24" |
| Backrest Styles | Fixed Height, Rigid Canes Fixed Height, Foldable Canes Height-Adjustable, Rigid Canes | Stroller Handle Back Post Height Adjustable Straight with Push Handle Dynamic Rocker Back |
| Frame Options | Adjustable Seat Width Adjustable Seat Depth Adjustable Seat Height Adjustable Armrest Height Adjustable Backrest Angle Adjustable Center of Gravity Adjustable Footrest Length Adjustable Footrest Angle | Width Adjustable Folding Depth Adjustable Footrest Swing Away Footrest Super Low - Swing Away |
| Wheel Lock | Push to lock Pull to Lock Attendant Brakes | Push to lock Pull to Lock Push to Lock Flush Mount Hemi-Wheel Locks Extension handles (push and pull) Grade Aids |
| Anti-Tips | Standard anti-tippers | Removable/swing-up rear Anti-tips |
| Rear Wheels | 14", 20",22" | Stainless Steel Spoke (18Ct.): 18", 20", 22", 24" Mag: 20", 22", 24" Composite Fiber Spoke (18Ct.): 22" 24" |
| Caster Wheels | 5/6" | .75" Roller Blade (3", 4") .75" Lighted Roller Blade (3",4",5") 1" Polyurethane (4", 5",6", 7",8") 1" Polyurethane Aluminum (4", 5") 1" Pneumatic (6",8") 1.5" Polyurethane (4", 5",6") 1.5" Soft Roll Aluminum (3", 4", 5", 6") |
{9}
| | | 2" Polyurethane (6")
2" Pneumatic w/ Foam Insert (8")
Rear: 18"-24"; Casters: 3"-8", various materials |
| --- | --- | --- |
| Seat to Backrest Angle | -4° to +16° (97-117) | 85°-115°, 5° increments |
| Leg to seat angle | 94° | - |
| Footrest to seat distance | 6" - 9.8" | - |
| Armrest to seat distance | 6" - 9.3" | - |
| Seat plane angle | 2° - 5° | - |
| Static stability downhill | 15° | - |
| Static stability uphill | 14° | - |
| Static stability sideways | 14° | - |
| Push Handle | Height-Adjustable Stroller Handle Push Handle | Std Push Handle fixed on back canes
Removeable Stroller Back Handle |
| Armrest | Swing-away/ removable armrest and height-adjustable armrest | Single Post Height Adjustable Height Adjustable T-Arm (Std, Tall, Low) Angle Adjustable Locking Extendable Flip Up Pediatric Height Adjustable T-Arm Tubular Flip up Arm Pads (Std, Waterfall, Foam Grip) |
| Tires | Solid Tires (Standard) Pneumatic (optional) | Pneumatic
Pneumatic with Airless Insert
Pneumatic with puncture protection
Solid Polyurethane |
| Handrims | Ergonomic handrim Aluminum handrim | Aluminum Anodized Aluminum with Plastic Coating Aluminum with Non-Slip Tape |
{10}
| | | Projection Knob with Plastic Coating Ergonomic Thumb Grip: Flexible Polymer Ergonomic Thumb Grip: Alum with Grip Coating |
| --- | --- | --- |
| Footrest | 90° Swing away Detachable Footrests
80° Swing away Detachable Footrests Articulating Swing away Detachable Footrests | Swing Away, Footrest, Extension Mount Swing Away, Footrest, Front Mount Elevating Leg Rest (PRO, PRO Pediatric) Residual Limb Support
Hanger Release (Classic, 4 Way In) |
| Footplate | Aluminum Footplate with 3D Angle Adjustment
Aluminum Footplate with 4D Angle Adjustment | Composite
Composite Angle Adjustable
Aluminum Angle Adjustable
Aluminum Locking Multi-Angle Adjustable
One Piece Flip Up Adjustable
Swing Away Extensions (Short, Med, Long)
Front Mount Extension |
| Safety and Transportation Compliance | The Flexx Junior is crash-tested to ISO7176-19, making it suitable for use as a seat in adapted vehicles. It is recommended to use a four-point webbing system for secure restraint. | Ki Mobility Spark chair is used as seat in a motor vehicle and conforms to ISO 7176-19 Requirement. |
| Accessories | Pneumatic rear wheels
Aluminum hand rim
Attendant brake
Swing away anti-tipper
IV pole holder
Transparent Table Tray
Oxygen tank holder
Cane holder
Lap Belt
1" Metal Buckle Pelvic Belt. | Sideguards Composite Side guards (adult and pediatric)
Backpack/Pouch
Spoke Guards
Cane/Crutch Holder
ELR Gel Pads
Impact guards
Canopy
Tool Kit
One-Arm Drive |
| Performance compliance | ISO 7176-1: 2014
ISO 7176-3: 2012
ISO 7176-5:2008
ISO 7176-7:1998
ISO 7176-8: 2014
ISO 7176-11:2012
ISO 7176-13:1989 | ISO-7176-1:2014
ISO-7176-3:2012
ISO 7176-5:2008
ISO 7176-7:1998
ISO 7176-8:2014
ISO 7176-11:2012
ISO 7176-13:1989 |
{11}
| | ISO 7176-15: 1996
ISO 7176-19:2008
ISO 7176-22:2014
EN 1021-1 (Gap analysis)
EN 1021-2 | ISO 7176-15:1996
ISO 7176-16:2012
ISO 7176-19:2022
ISO 7176-22:2014 |
| --- | --- | --- |
# Technological Comparisons: [21CFR807.92(a)(6)]
Differences & Justification for Safety and Effectiveness: The Flexx Junior manual wheelchair and the predicate device, the Ki Mobility Spark are intended to provide mobility to pediatrics limited to a seating position. They share the same user weight capacity of (165 lbs) 75 kg.
1. Frame and Design Features: The Flexx Junior and Ki Mobility Spark share equivalent frame and design features, both using lightweight aluminum frames with an X-brace for rigidity. The Flexx Junior offers fixed or foldable backrests, while the Spark is foldable only. Both provide adjustable seat width, depth, height, backrest angle, armrests, and footrests via discrete holes, with the Spark offering more backrest styles and footrest/footplate options. Push handles and swing-away armrests are comparable. Differences in variety do not affect safety or effectiveness, as both ensure equivalent adjustability and stability, verified by ISO 7176 compliance.
2. Product Weight & Folding Dimensions: The Flexx Junior manual wheelchair and the predicate device, the Ki Mobility Spark, are equivalent in product weight and folding dimensions, ensuring ease of handling and transport for pediatric use. Both have comparable weights and folded dimensions, with the Flexx Junior's width aligning with the Ki Mobility Spark. Minor range variations in the Flexx Junior do not affect functionality, as both maintain a compact, lightweight profile suitable for pediatric use, with no implications for safety or structural integrity.
3. Turning Radius & Maneuverability: The Flexx Junior has a slightly smaller minimum turning radius compared to the predicate devices, improving maneuverability in tighter spaces without affecting overall stability. The Spark's additional wheel/tire/handrim options provide more variety, but differences do not raise safety concerns
4. Seat Height & Angle: Both devices offer similar seat height and angle adjustments for customization, with the Ki Mobility Spark wider ranges providing more flexibility. Both devices feature adjustable seat width via an X-brace mechanism, with the Flexx Junior providing a slightly narrower range than the Ki Mobility Spark, ensuring equivalent customization for growth. While the Spark offers wider adjustment ranges, these differences are minor and do not impact safety or effectiveness.
5. Accessories & Additional Features – The Flexx Junior offers comparable accessory options ensuring equivalent user positioning and support. Both provide essential safety and convenience items, with variations tailored to pediatric needs.
{12}
6. Performance Compliance: The Flexx Junior complies same standards and is substantial equivalence to the predicate device
7. Safety and Transportation Compliance: Both are crash-tested to ISO 7176-19, suitable for vehicle use with equivalent restraint recommendations.
Overall Equivalence: Accessories and features are equivalent in enhancing usability and safety, with differences reflecting optional customizations rather than core functionality. Both ensure compliance with overlapping standards, confirming no new risks and substantial equivalence in performance for pediatric mobility. Differences exist in the range of seat width, seat depth, and wheel options. The Spark offers a wider range of seat width and depth, as well as more rear wheel and caster options. However, these differences do not impact safety or effectiveness, as both devices provide sufficient adjustability for pediatric users and comply with ISO 7176-19 standard for use as a seat in a motor vehicle. Since the indications for use remain unchanged compared to the predicate devices, these differences do not constitute a new intended use. Both wheelchairs have been tested to equivalent performance and biocompatibility standards, ensuring no new safety or effectiveness concerns.
## Bench testing performance tests summary and conclusions: [21 CFR 807.92(b)]
The Flexx Junior's bench testing confirms its safety, durability, and performance are equivalent to the Ki Mobility Spark. Compliance with ISO standards and biocompatibility standards verifies that the Flexx Junior performs reliably under expected use conditions, with no differences in design or adjustability raising safety or effectiveness concerns. The device is substantially equivalent to the predicate, ensuring safe and effective mobility for pediatric users.
Biocompatibility Tests and Flammability Tests: These tests ensure that the materials used in the wheelchair do not cause harm when they come into contact with the user's skin, evaluating the biological safety of medical device materials. These tests assess the fire resistance of upholstered furniture and safety-related components. Confirms components meets flammability requirements for safety aligning with the Ki Mobility Spark's compliance.
These tests collectively ensure that the materials used in the Flexx Junior wheelchair are biocompatible, safe for skin contact, and resistant to ignition, contributing to the overall safety and compliance of the device.
Standards and Test reports
| Standards/Regulations | Test Name | Result |
| --- | --- | --- |
| ISO 7176-1: 2014 | Static stability test | Pass |
| ISO 7176-3: 2012 | Determination of effectiveness of brakes | Pass |
| ISO 7176-8: 2014 | Multi drum, Drop test, Impact strength test, Manually operated parking brakes fatigue test, static strength test | Pass |
{13}
| ISO 7176-11:2012 | Test Dummies | Pass |
| --- | --- | --- |
| ISO 7176-13:1989 | Determination of coefficient of friction | Pass |
| ISO 7176-15:1996 | Determination of dimensions mass report
(Test report includes ISO 7176-7 (1998): Measurement of Seating and Wheel Dimensions and ISO 7176-5 (2008): Determination of Overall Dimensions, Mass, and Maneuvering Space) | Pass |
| ISO 7176-19:2008 | Unwin front and rear tie down, Crash test | Pass |
| ISO 7176-22:2014 | Set-up procedures | Pass |
| Test Standard | Tested Components & Test Report Number |
| --- | --- |
| ISO 10993-1,
ISO 10993-5,
ISO 10993-10
ISO 10993-23 | Fabric: EABML(E)1030421-140
GLP_SR_KL20250426-S
Handle: EABML(E)1030421-138
Armrest: EABML(E)1030421-142 |
| EN 1021-1,
EN 1021-2
Flammability test –
Upholstery, Belt | Upholstery fabric - HL 90076/2020, HL 90077/2020
Seat belt or restraint system - HL10132/2015
Gap analysis Report |
## Conclusion:
The Flexx Junior manual wheelchair demonstrates substantial equivalence to the predicate device, the Ki Mobility Spark with no new risks to safety or effectiveness. Both devices share identical intended use, indications for use, weight capacity (165 lbs /75 kg), aluminum frame, and key adjustability features such as seat width, depth, height, backrest angle. Differences, such as the Flexx Junior's optional fixed backrest (FLX-J00), narrower seat width and depth ranges, and fewer wheel options, do not affect safety or effectiveness, as both provide comparable adjustability, stability, and maneuverability for pediatric users, verified through equivalent performance testing. Testing per ISO 7176 and biocompatibility standards confirms that the Flexx Junior performs as safely and effectively as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.