Arm-type Fully Automatic Digital Blood Pressure Monitors (DBP-6286B, DBP-6186)
Applicant
Joytech Healthcare Co. , Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jun 26, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm.
Device Story
Automatic, non-invasive blood pressure monitor for home and clinical use. Device uses oscillometric method; integral piezoelectric pump inflates arm cuff; electric valve performs rapid deflation. Microprocessor-based software analyzes pulse waveform data extracted from cuff pressure to calculate systolic/diastolic pressure and pulse rate. DBP-6286B model includes Bluetooth module for wireless data transfer to mobile app; app functions as non-device MDDS for storage/display only. Operated by patient or clinician. Output displayed on LCD; provides blood pressure, pulse rate, irregular heartbeat detection, and memory functions. Benefits include convenient, automated blood pressure monitoring for hypertension management.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018+AMD2020 using manual mercury sphygmomanometer as reference. For 22-42cm cuffs, data from 91 subjects (using representative model DBP-6279B) demonstrated accuracy within acceptable limits. For 40-56cm cuffs, data from 85 subjects (using representative model DBP-6286B) demonstrated accuracy within acceptable limits. No adverse events reported.
Technological Characteristics
Oscillometric measurement; silicon pressure transducer; piezoelectric pump; electric solenoid valve. Power: 4x1.5V AAA batteries or 5V/1A DC adapter. Connectivity: Bluetooth 5.2 (DBP-6286B). Biocompatibility per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-30, ISO 81060-2.
Indications for Use
Indicated for adults and adolescents over 12 years of age for the measurement of systolic and diastolic blood pressure and pulse rate, using arm circumferences of 22-42cm or 40-56cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 26, 2025
JOYTECH Healthcare Co., Ltd.
Cong Jing
RA Manager
No.365, Wuzhou Road
No.502, Shunda Road
Hangzhou, Zhejiang 311100
China
Re: K250548
Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitors (DBP-6286B, DBP-6186)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: February 25, 2025
Received: May 27, 2025
Dear Cong Jing:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K250548 - Cong Jing
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250548 - Cong Jing
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250548
Device Name
Arm-type Fully Automatic Digital Blood Pressure Monitors (DBP-6286B, DBP-6186)
Indications for Use (Describe)
The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
{4}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
JOYTECH 继裕
# Substantial Equivalence Comparison
## 1. Sponsor's Identification
Name: JOYTECH Healthcare Co., Ltd.
Address 1: No. 365. Wuzhou Road, Hangzhou, Zhejiang 311100, China.
Address 2: No. 502. Shunda Road, Hangzhou, Zhejiang 311100, China.
Contact Person: Cong Jing
Email address: Jingc@sejoy.com
## 2. Name of the device
Trade Name: Arm-type Fully Automatic Digital Blood Pressure Monitor
| Arm-type | DBP-6186, DBP-6286B |
| --- | --- |
Common Name: Blood Pressure Monitor
Classification name: Non-invasive blood pressure measurement System
## 3. Classification information
Regulation Number: 21 CFR 870.1130
Product Code: DXN
Device Class: II
Panel: 74 Cardiovascular
## 4. Predicate device information
The Arm-Type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following predicate devices:
| 510(k) number | Predicate device model | Product code | Manufacturer |
| --- | --- | --- | --- |
| K212115 | DBP-6286B | DXN | JOYTECH Healthcare Co., Ltd. |
Page 1 of 10
{5}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
# JOYTECH 继拓
Note: As for the subject device in this submission are derived from the DBP-6286B, DBP-6186 in cleared K212115. Compared to the previous device, the subject device has added two cuffs and also the device unit has changed the power supply voltage, air pump, bluetooth module, and appearance. Therefore we resubmitted a new traditional 510 (k).
# 5. Device description
The Arm-type Blood Pressure Monitor (BPM) series is an automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The systolic and diastolic pressures are determined using the Oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats per minute.
The DBP-6286B embed bluetooth module to transfer data to APP. With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is not be intended for active patient monitoring, therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS.
# 6. Indications for use
Page 2 of 10
{6}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
# JOYTECH 继裕
The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm.
# 7. Comparison to the predicate device
| Comparison items | Subject devices (DBP-6186, DBP-6286B) | Predicate device DBP-6286B (K212115) | Judgement |
| --- | --- | --- | --- |
| Manufacturer | JOYTECH Healthcare Co., Ltd. | JOYTECH Healthcare Co., Ltd. | Same |
| Trade name | Arm-type Fully Automatic Digital Blood Pressure Monitor | Arm-type Fully Automatic Digital Blood Pressure Monitor | Same |
| Recommended classification regulation | 21CFR 870.1130 | 21CFR 870.1130 | Same |
| Regulatory class | Class II | Class II | Same |
| Product code | DXN | DXN | Same |
| Indications for use/ Intended use | The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm | The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age. | Same Note 1 |
Page 3 of 10
{7}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
JOYTECH 继裕
| Intended patient | Adults and adolescents over 12 years of age | Adults and adolescents over 12 years of age | Same |
| --- | --- | --- | --- |
| Intended use environment | Home and Professional healthcare facility environment. | Home and Professional healthcare facility environment. | Same |
| Working principle | Oscillometric method | Oscillometric method | Same |
| Measurement type | Determined during inflation | Determined during inflation | Same |
| Cuff location | Upper arm | Upper arm | Same |
| Cuff circumference | Fits arm circumference 22-42cm and 40-56cm | Fits arm circumference 22-36 cm | Similar Note 1 |
| Appearance | | | Similar Note 2 |
| Solenoid valve | Solenoid valve | Solenoid valve | Same |
| Sensor | Silicon pressure transducer | Silicon pressure transducer | Same |
| Pump | Air pump | Air pump | Similar Note 3 |
| Display screen | LCD | LCD | Same |
Page 4 of 10
{8}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
JOYTECH 继裕
| Power supply source | 4 x1.5V AAA batteries or Medical AC adaptor (DC 5V, 1A, recommend but not provide) | 3x1.5V AAA battery or Medical AC adaptor (DC 5V, 1A, recommend but not provide) | Similar Note 4 |
| --- | --- | --- | --- |
| Bluetooth module | DBP-6286B applicable
Frequency range: 2.4 GHz (2402-2480 MHz)
Modulation type: GFSK
Version 5.2 (BLE only) | frequency range: 2.4 GHz (2402-2480 MHz)
Modulation type: GFSK
Version 5.0 (BLE only) | Similar Note 5 |
| Measuring range | Systolic Pressure: 60mmHg ~ 260 mmHg
Diastolic Pressure:40mmHg ~ 200 mmHg
Pressure: 0mmHg~299 mmHg
Pressure: ±3mmHg
Pulse:30~180 Beats/Minute | Systolic Pressure: 60mmHg ~ 260 mmHg
Diastolic Pressure:40mmHg ~ 200 mmHg
Pressure: 0mmHg~299 mmHg
Pressure: ±3mmHg
Pulse:30~180 Beats/Minute | Same |
| Accuracy | Pressure: ±3 mmHg
Pulse: 40-180Beats/Minute, ±5%; 30-39Beats/Minute, ±5bpm | Pressure: ±3 mmHg
Pulse: 40-180Beats/Minute, ±5%; 30-39Beats/Minute, ±5bpm | Same |
| Main function | Measure blood pressure value
Measure pulse rate
Irregular Heartbeat Detection
Memory
WHO Classification Indicator
Last 3 Test Average
Low Battery Detection
Automatic Power-Off
Voice abroad
Backlight
Arm shake indicator | Measure blood pressure value
Measure pulse rate
Irregular Heartbeat Detection
Memory
WHO Classification Indicator
Last 3 Test Average
Low Battery Detection
Automatic Power-Off
Voice abroad
Backlight
Arm shake indicator | Same |
Page 5 of 10
{9}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
JOYTECH 继裕
| | Cuff loose indicator
Bluetooth function (DBP-6286B applicable) | Cuff loose indicator
Bluetooth function (DBP-6286B applicable) | |
| --- | --- | --- | --- |
| Operating condition | 10℃-40℃ (50°F~104°F)
15%-93% RH
800hPa-1060hPa | 10℃-40℃ (50°F~104°F)
15%-93% RH
800hPa-1060hPa | Same |
| Storage condition | -25℃-55℃
(-13°F~158°F)
≤93% RH | -25℃-55℃
(-13°F~158°F)
≤93% RH | Same |
| Biocompatibility | ISO 10993-1
Bio-compatible | ISO 10993-1
Bio-compatible | Same |
| Applicable standards | IEC 60601-1
IEC 60601-1-11
IEC 60601-1-2
ISO 80601-2-30
ISO 81060-2 | IEC 60601-1
IEC 60601-1-11
IEC 60601-1-2
ISO 80601-2-30
ISO 81060-2 | Same |
Note 1: The subject device has the same intended purpose and principle with cleared devices (K212115). Compare with predicate device, the modification of cuff is occur in only add two cuffs, 22-42cm and 40-56cm. Adding new cuffs generally does not result in changes in electrical safety and EMC, and its corresponding clinical accuracy has been confirmed through clinical trials according to ISO 81060-2. Thus, the difference between the subject device and the cleared predicate device do not raise different issues of safety and effectiveness.
Note 2: The subject device was added number of buttons based on predicate device. At the same time, The USB adapter jack type has no changed, the interface location of the power adapter has been changed from the right side to the top. These changes have passed electrical safety testing and will not cause new safety and effectiveness issues.
Note 3: Compared with cleared predicate device, the external dimensions and rated working voltage of the both air pump are not consistent, but the structural composition of the both air pump is the same. This subject device has passed IEC 60601-1, IEC 60601-1-2,
Page 6 of 10
{10}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
# JOYTECH 继裕
and ISO 80601-2-30 standard testing, which will not result in new safety and effectiveness issues.
Note 4: The battery type of subject devices and predicates are the same, which all of them are AAA batteries. The main difference lies in the voltage. The voltage of the subject device is 6V, and the risks caused by different voltages have been controlled through electrical safety and electromagnetic compatibility testing, without bringing new safety and effectiveness issues.
Note 5: The difference of model DBP-6286B and DBP-6186 is only bluetooth. The BLE module version of DBP-6286B was updated from V5.0 to V5.2 by adding a shield cover. The subject device DBP-6286B has been verified the safety by IEC 60601-1, IEC 60601-11, IEC 60601-1-2, FCC and Wireless coexistence test report, which those of difference between subject device and predicate device do not raise new issues of safety and effectiveness comparing with the predicated device.
# 8. Performance Data
Testing information demonstrating safety and effectiveness of the device in the intended environment of use is supported by testing that was conducted.
The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices.
The following National and International Standards were utilized for testing the subject device.
# Electrical Safety and performance requirements:
IEC 60601-1:2005+AMD1:2012+AMD2:2020 , Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.
AAMI ES60601-1:2005+AMD1:2012+AMD2:2021 , Medical Electrical Equipment Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance
Page 7 of 10
{11}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
# JOYTECH 继裕
IEC 80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.
## Home-used medical equipment requirements and environmental test:
IEC 60601-1-11:2015+AMD1:2020, General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
## Electromagnetic compatibility requirements:
IEC 60601-1-2:2014+AMD1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
## Bio-compatibility Evaluation for patient contacting components:
ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021, Biological evaluation of medical devices Part 10: Tests for skin sensitization
ISO 10993-23:2021, Biological evaluation of medical devices Part 23: Tests for irritation
## FCC Test
FCC Part15 Subpart C
RF Exposure valuation
## Guidance Document:
The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
Page 8 of 10
{12}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
JOYTECH 继裕
The test result all meet or exceed the requirement of these standards.
## 9. Discussion of Clinical Tests Performed
Clinical Validation:
- ISO81060-2:2018+AMD2020 Non-invasive sphygmomanometers Part 2: Clinical investigation of intermittent automated measurement type.
## For cuff with arm circumstance 22-42cm
Compared with Arm-type Fully Automatic Digital Blood Pressure Monitor DBP-6279B which manufactured by our company, although the subject device is different in appearance from DBP-6279B, the key factors affecting the clinical accuracy are the same. For example, both of them have the same principle and measurement type (determined by inflation), and they have the same core algorithm, same pressure sensor, same MCU, and same cuff size and material. Therefore, we believe that the clinical report of DBP-6279B can cover the subject devices.
Model DBP-6279B was selected as representative for testing. It was conducted according to ISO 81060-2: 2018+AMD2020 standard. Same arm sequential method was adopted during clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The subject demographics include 91 valid subjects data were used to analysis. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.
## For cuff with arm circumstance 40-56cm
Model DBP-6286B was selected as representative for testing. It was conducted according to ISO 81060-2: 2018+AMD2020 standard. Same arm sequential method was adopted during clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The subject demographics include 85 valid subjects data were used to analysis. The results showed
Page 9 of 10
{13}
K250548 Summary
JOYTECH HEALTHCARE CO.,LTD.
JOYTECH 继裕
the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.
# 10. Conclusion
The subject devices and predicate devices have the same intended use, measuring principle, measuring method, performance and most technical characteristic.
The difference between the subject and the predicate devices does not: (1)
affect the intended use or (2) alter the fundamental scientific technology of the device. The minor difference between the subject devices and the predicate devices have been evaluated and determined to not raise any new issues of safety or effectiveness. Therefore, the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.