← Product Code [OLO](/productcode/OLO) · K250520

# OsteoCentric Navigated Instruments System (K250520)

_Osteocentric Technologies, Inc. · OLO · Oct 27, 2025 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250520

## Device Facts

- **Applicant:** Osteocentric Technologies, Inc.
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Oct 27, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The OsteoCentric Navigated Instruments are intended to be used during the preparation and placement of OsteoCentric Pedicle Fasteners during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures. The OsteoCentric Navigated Instruments are specifically designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

## Device Story

System consists of reusable surgical instruments (navigated probes, taps, drills, drivers) used with Medtronic StealthStation S8 Navigation System; assists surgeons in spinal surgery for pedicle screw placement and Integrity SI Fusion implant placement. Instruments provide spatial tracking data to navigation system; system correlates instrument position with patient anatomy (CT/MR/fluoroscopy/landmarks). Used in OR by surgeons; provides real-time visual guidance on navigation display. Enables precise anatomical localization; facilitates minimally invasive/open procedures; improves surgical accuracy; reduces potential for malpositioned hardware.

## Clinical Evidence

Bench testing only. No clinical data presented. Equivalence established through engineering analysis, geometric comparison, and material/biocompatibility evaluation against predicate devices.

## Technological Characteristics

Reusable surgical instruments (probes, taps, drills, drivers). Materials: medical grade stainless steel per ASTM F899, ASTM F276, and ASTM F138. Designed for integration with Medtronic StealthStation navigation system. Sterilization required.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Navigated CD Horizon® Solera® Screwdrivers and Taps ([K170679](/device/K170679.md)/[K140454](/device/K140454.md))
- Navigated INFINITY™ Instruments ([K173338](/device/K173338.md))
- OsteoCentric Spine MIS Pedicle Fastener System ([K221332](/device/K221332.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 27, 2025

OsteoCentric Technologies, Inc.
% Hannah Taggart
Engineer &amp; Regulatory Specialist
Empirical Technologies
4628 Northpark Drive
Colorado Springs, Colorado 80918

Re: K250520
Trade/Device Name: OsteoCentric Navigated Instruments System
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: September 24, 2025
Received: September 25, 2025

Dear Hannah Taggart:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250520 - Hannah Taggart
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K250520 - Hannah Taggart
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250520

Device Name
OsteoCentric Navigated Instruments System

Indications for Use (Describe)
The OsteoCentric Navigated Instruments are intended to be used during the preparation and placement of OsteoCentric Pedicle Fasteners during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures. The OsteoCentric Navigated Instruments are specifically designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”

{4}

K250520 510(K) SUMMARY

|  Submitter's Name: | OsteoCentric Technologies, Inc.  |
| --- | --- |
|  Submitter's Address: | 75 West 300 North Suite 150
Logan, UT 84321  |
|  Submitter's Telephone: | 800-969-0639  |
|  Contact Person: | Hannah Taggart, MS
Empirical Technologies
719- 457-1152
htaggart@empiricaltech.com  |
|  Date Summary was Prepared: | September 24, 2025  |
|  Trade or Proprietary Name: | OsteoCentric Navigated Instruments System  |
|  Device Classification Name: | Orthopedic Stereotaxic Instruments  |
|  Classification & Regulation #: | Class II per 21 CFR §882.4560  |
|  Product Code: | OLO  |
|  Classification Panel: | Orthopedic  |

## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The OsteoCentric Navigated Instruments System is a set of reusable surgical instruments for use with the Medtronic® StealthStation™ S8 Navigation System to assist surgeons in precisely locating anatomical structures in either open or percutaneous procedures for the placement of pedicle screws. The OsteoCentric Navigated Instruments are intended for use with the OsteoCentric Pedicle Fasteners Additionally, the submission includes a hybrid technique for placement of the Integrity SI® Fusion implants.

The OsteoCentric Navigation Instrument System includes navigated probes, taps, drills, and drivers and are manufactured from medical grade stainless steels per ASTM F899, ASTM F276, and ASTM F138.

## INDICATIONS FOR USE

The OsteoCentric Navigated Instruments are intended to be used during the preparation and placement of OsteoCentric Pedicle Fasteners during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures. The OsteoCentric Navigated Instruments are specifically designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

## TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:

- Indications for Use
- Materials of manufacture
- Instrument Types
- Sizes and Functional Length of Instruments
- Manufacturing and Biocompatibility

1 | Page

{5}

2 | Page

# Predicate Devices

|  510k Number | Trade or Proprietary or Model Name | Manufacturer | Product Code | Predicate Type  |
| --- | --- | --- | --- | --- |
|  K170679/K140454 | Navigated CD Horizon® Solera® Screwdrivers and Taps | Medtronic | OLO | Primary  |
|  K173338 | Navigated INFINITY™ Instruments | Medtronic | OLO | Additional  |
|  K221332 | OsteoCentric Spine MIS Pedicle Fastener System | OsteoCentric Technologies, Inc. | NKB | Additional  |

The subject device has been evaluated through an engineering analysis and geometric comparison to predicate devices to establish the safety and efficacy for accuracy performance. The biocompatibility evaluation of the instruments leveraged equivalence in materials and manufacturing to previously cleared instruments (K221332). The results of these engineering analysis and evaluations show that the subject is substantially equivalent to the cleared predicate.

# CONCLUSION

The overall technology characteristics and geometric analysis lead to the conclusion that the OsteoCentric Navigated Instruments System is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K250520](https://fda-staging.innolitics.com/device/K250520)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
