K250504 · Ceramoptec GmbH · GEX · Jul 14, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250504
Device Name
Leonardo Duster
Applicant
Ceramoptec GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 14, 2025
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Leonardo® Duster is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. Urology - Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH] - Laser Resection of the Prostrate (LRP) - Laser Enucleation of the Prostate (LEP) - Laser Ablation of the Prostate (LAP) - Transurethral Incision of the Prostate (TUIP) - Condylomas - Urethral/ureteral structures - Lesions of external genitalia - Bladder neck incisions (BNI) - Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors - Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi - Treatment of distal impacted fragments remaining in the ureters following lithotripsy. Lithotripsy and Percutaneous Urinary Lithotripsy Indications - Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones - Endoscopic fragmentation of renal calculi - Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed. Gastroenterology Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: - Appendectomy - Polyps - Biopsy - Gall Bladder calculi - Biliary/Bile duct calculi - Ulcers - Gastric ulcers - Duodenal ulcers - Non Bleeding Ulcers - Pancreatitis - Haemorrhoids - Cholecystectomy - Benign and Malignant Neoplasm Gynecology - Angiodysplasia - Colorectal cancer - Telangiectasias - Telangiectasias of the Osler-Weber-Renu disease - Vascular Malformation - Gastritis - Esophagitis - Esophageal ulcers - Varices - Colitis - Mallory-Weiss tear - Gastric Erosions Gynecology Open, endoscopic (including hysteroscopic) and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.
Device Story
Leonardo Duster is a Thulium-doped fiber laser system (1940nm) for soft tissue surgery and lithotripsy. Physician operates device via touchscreen to set power, pulse energy, frequency, and pulse width. System features two modes: Continuous and Pulse. RFID-encoded optical fibers ensure automatic recognition and safety. Device delivers laser energy to target tissue/calculi via fiber probes. Used in urology, gastroenterology, and gynecology settings. Output allows precise tissue ablation, coagulation, and stone fragmentation, facilitating minimally invasive surgical procedures. Benefits include reduced bleeding and effective stone removal.
Clinical Evidence
Bench testing only. No clinical data required to establish substantial equivalence. Compliance with IEC 60601-1, 60601-1-2, 60601-2-22, and ISO 10993 biocompatibility standards for fiber probes provided.
Technological Characteristics
Thulium-doped fiber laser; 1940nm wavelength; 60W max output; 525nm aiming beam. Air-cooled. Touchscreen interface. RFID-encoded fiber recognition. Reusable and single-patient-use fiber probes. Sterilization: EO (initial), autoclave (reusable). Standards: IEC 60601-1, IEC 60601-2-22, ISO 10993, ISO 14971.
Indications for Use
Indicated for patients requiring soft tissue surgery (incision, excision, resection, ablation, coagulation, hemostasis, vaporization) in urology, gastroenterology, and gynecology. Includes treatment of BPH, prostate resection/enucleation/ablation, bladder/urethral/ureteral tumors, and lithotripsy for renal/ureteral/bladder calculi.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
MegaBeam Reusable Fiber Optic Delivery System (K113688)
MegaBeam Fiber Optic Delivery System with Additional Tips & Handpiece (K113709)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 14, 2025
CeramOptec GmbH
% Daniel Kamm
Principal Engineer
Kamm & Associates
8870 Ravello Ct
Naples, Florida 34114
Re: K250504
Trade/Device Name: Leonardo Duster
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: February 20, 2025
Received: February 20, 2025
Dear Daniel Kamm:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K250504 - Daniel Kamm
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250504 - Daniel Kamm
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.07.14
14:22:56 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250504
Device Name
Leonardo Duster
Indications for Use (Describe)
The Leonardo® Duster is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
Urology
- Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- Laser Resection of the Prostrate (LRP)
- Laser Enucleation of the Prostate (LEP)
- Laser Ablation of the Prostate (LAP)
- Transurethral Incision of the Prostate (TUIP)
- Condylomas
- Urethral/ureteral structures
- Lesions of external genitalia
- Bladder neck incisions (BNI)
- Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- Treatment of distal impacted fragments remaining in the ureters following lithotripsy. Lithotripsy and
Percutaneous Urinary Lithotripsy Indications
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy
- Polyps
- Biopsy
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non Bleeding Ulcers
- Pancreatitis
- Haemorrhoids
- Cholecystectomy
- Benign and Malignant Neoplasm Gynecology
- Angiodysplasia
- Colorectal cancer
- Telangiectasias
- Telangiectasias of the Osler-Weber-Renu disease
- Vascular Malformation
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
Gynecology
Open, endoscopic (including hysteroscopic) and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
{5}
510(k) Summary: 510(k) Number K250504
CeramOptec GMBH
Siemensstrabe 44
Bonn Nordrhein-Westfalen, Germany D-53121
Telephone +49 228 979670
Date Prepared: July 10, 2025
Contact: Dr. Roland Dreschau, Director
Email Roland.Dreschau@biolitec.com
1) Identification of the Device:
Trade/Device Name: Leonardo® Duster
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II
Product Code: GEX
2) Equivalent legally marketed device: K242191 SOLTIVE™ Laser System
Trade/Device Name: SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II
Product Code: GEX
3) Reference Devices (Sterile Reusable and Single Patient Use Fiberoptic Probes)
Biolitec Medical Devices, Inc.:
MegaBeam Reusable Fiber Optic Delivery System: K113688
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II
Product Code: GEX
and
Biolitec Medical Devices, Inc.:
MegaBeam Fiber Optic Delivery System with Additional Tips & Handpiece: K113709.
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II
Product Code: GEX
4) Indications for Use: The Leonardo® Duster is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. Urology
- Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- Laser Resection of the Prostrate (LRP)
- Laser Enucleation of the Prostate (LEP)
{6}
- Laser Ablation of the Prostate (LAP)
- Transurethral Incision of the Prostate (TUIP)
- Condylomas
- Urethral/ureteral strictures
- Lesions of external genitalia
- Bladder neck incisions (BNI)
- Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- Treatment of distal impacted fragments remaining in the ureters following lithotripsy.
Lithotripsy and Percutaneous Urinary Lithotripsy Indications
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
## Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy
- Polyps
- Biopsy
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non Bleeding Ulcers
- Pancreatitis
- Haemorrhoids
- Cholecystectomy
- Benign and Malignant Neoplasm Gynecology • Angiodysplasia
- Colorectal cancer
- Telangiectasias
- Telangiectasias of the Osler-Weber-Renu disease
- Vascular Malformation
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
## Gynecology
Open, endoscopic (including hysteroscopic) and laparoscopic gynecological surgery (incision, excision, resection, ablation,
Page | 2
{7}
vaporization, coagulation and hemostasis) of soft tissue.
5) Description of the Device: All variants of the laser family LEONARDO® Duster are laser systems with functions and ergonomics specially developed for medical applications. A touchscreen is used to set treatment parameters, such as Power in Continuous mode, Pulse energy (J) and Frequency (Hz) with Pulse width options in Pulse mode. User-friendly menu navigation and microprocessor-supported control ensure reliable operation while allowing physicians to concentrate on the essential aspects of treatment. The Thulium doped fiber laser module produces a coherent laser radiation with the wavelength of $1940\mathrm{nm} \pm 3\mathrm{nm}$ (aiming beam $525\mathrm{nm} + / - 10\mathrm{nm}$ ). An optical fiber delivers this energy to affected surfaces and organs. All fields of application are listed in the indications for use statement.
The subject model LEONARDO® Duster laser has a maximum laser output power of 60W. @ 1940nm The LEONARDO® Duster is available only as a single-wavelength device with 1940nm. All LEONARDO® Duster lasers can be operated in two basic modes, CONTINUOUS or PULSE MODE. Two application modes are available which defines the intended use of the device: Lithotripsy and Other. For safety reasons, the LEONARDO® Duster laser is equipped with a system for automatic recognition of the used optical fibers. Application fibers from CeramOptec are RFID encoded for communicating with the laser device. Reusable fiber probes are available in 6 different sizes. They are initially EO sterilized. Subsequent sterilization is via autoclave. Single patient use fiber probes come in 7 different sizes. They are initially EO sterilized.
6) The comparison between the overall specifications of predicate device and the subject device is shown in Table 1, below.
Table 1 Substantial Equivalence Chart
| Characteristic | Olympus Surgical Technologies America K242191 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) | CeramOptec GMBH Leonardo® Duster |
| --- | --- | --- |
| Indications for Use | The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. Urology • Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH] • Laser Resection of the Prostrate (LRP) • Laser Enucleation of the Prostate (LEP) • Laser Ablation of the Prostate (LAP) • Transurethral Incision of the Prostate (TUIP) • Condylomas • Urethral/ureteral strictures • Lesions of external genitalia | The Leonardo® Duster is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. Urology • Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH] • Laser Resection of the Prostrate (LRP) • Laser Enucleation of the Prostate (LEP) • Laser Ablation of the Prostate (LAP) • Transurethral Incision of the Prostate (TUIP) • Condylomas • Urethral/ureteral strictures • Lesions of external genitalia |
{8}
| Characteristic | Olympus Surgical Technologies America K242191 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) | CeramOptec GMBH Leonardo® Duster |
| --- | --- | --- |
| | • Bladder neck incisions (BNI)
• Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
• Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
• Treatment of distal impacted fragments remaining in the ureters following lithotripsy. Lithotripsy and Percutaneous Urinary Lithotripsy Indications
• Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
• Endoscopic fragmentation of renal calculi
• Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
**Gastroenterology**
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
• Appendectomy
• Polyps
• Biopsy
• Gall Bladder calculi
• Biliary/Bile duct calculi
• Ulcers
• Gastric ulcers
• Duodenal ulcers
• Non Bleeding Ulcers
• Pancreatitis
• Haemorrhoids
• Cholecystectomy
• Benign and Malignant Neoplasm Gynecology
• Angiodysplasia
• Colorectal cancer
• Telangiectasias
• Telangiectasias of the Osler-Weber-Renu disease
• Vascular Malformation
• Gastritis
• Esophagitis
• Esophageal ulcers
• Varices | • Bladder neck incisions (BNI)
• Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
• Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
• Treatment of distal impacted fragments remaining in the ureters following lithotripsy. Lithotripsy and Percutaneous Urinary Lithotripsy Indications
• Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
• Endoscopic fragmentation of renal calculi
• Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
**Gastroenterology**
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
• Appendectomy
• Polyps
• Biopsy
• Gall Bladder calculi
• Biliary/Bile duct calculi
• Ulcers
• Gastric ulcers
• Duodenal ulcers
• Non Bleeding Ulcers
• Pancreatitis
• Haemorrhoids
• Cholecystectomy
• Benign and Malignant Neoplasm Gynecology
• Angiodysplasia
• Colorectal cancer
• Telangiectasias
• Telangiectasias of the Osler-Weber-Renu disease
• Vascular Malformation
• Gastritis
• Esophagitis
• Esophageal ulcers
• Varices |
Page | 4
{9}
| Characteristic | Olympus Surgical Technologies America K242191 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) | CeramOptec GMBH Leonardo® Duster |
| --- | --- | --- |
| | • Colitis
• Mallory-Weiss tear
• Gastric Erosions
**Gynecology**
Open, endoscopic (including hysteroscopic) and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue. | • Colitis
• Mallory-Weiss tear
• Gastric Erosions
**Gynecology**
Open, endoscopic (including hysteroscopic) and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.
IDENTICAL INDICATIONS FOR USE |
| Photo | | |
| Display | Touch Screen | Touch Screen SAME |
| Mode | Fiber laser | Fiber laser SAME |
| Laser power | 60W | 60W SAME |
| Laser type | Thulium Laser | Thulium Laser SAME |
| Wavelength | 1940 nm | 1940 nm SAME |
| Aiming beam | 500 - 550 nm, | 525 nm SAME |
| Aiming beam mW | ≤ 5 mW | ≤ 3mW +/-20% |
| Laser Class | 4 | 4 |
| Operation mode | Pulsed, Continuous Wave (CW) | Pulsed, Continuous Wave (CW) SAME |
| Pulse length/duration | 0.2 to 50ms | 0.04 to 50ms NEARLY IDENTICAL |
| Pulse frequency | 1 to 2500 Hz | 1 to 2500 Hz SAME |
| Cooling | Air cooling system | Air cooling system SAME |
Page | 5
{10}
| Characteristic | Olympus Surgical Technologies America K242191 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) | CeramOptec GMBH Leonardo® Duster |
| --- | --- | --- |
| Main power Supply | 100-240 Vac; 50/60 Hz; 1200 VA | 100-240 Vac; 50/60 Hz; 1000 VA SAME |
| Dimensions of system | 25.5 X 37.0 X 56.0 cm | 57.7 cm (W) × 36.8 cm (D) × 37.8 cm (H) SIMILAR. |
| Weight | 33kg | 41kg SLIGHTLY HEAVIER |
7) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness.
Safety and Effectiveness, comparison to predicate device: The INDICATIONS FOR USE are IDENTICAL. The results of bench and standards testing indicate that the new device is as safe and effective as the predicate device.
8) Summary of non-clinical testing:
Firmware was validated according to the FDA Guidance: Content of Premarket Submissions for Device Software Functions Guidance for Industry and Food and Drug Administration Staff (2023). Software validation documents were provided for a basic level of documentation. Reprocessing validation (for reusable fiber optic probes) was performed according to Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff
This device complies with the applicable portions of the standards listed below:
21 CFR 1040 PERFORMANCE STANDARDS FOR LIGHT-EMITTING PRODUCTS
EN 1041:2008 Information Supplied By The Manufacturer Of Medical Devices
ISO 14971: 2019-12 Medical devices — Application of risk management to medical devices FDA # 5-125
IEC ISO 15223-1: 2021-07 (Corrected Version 2017-03) Medical devices — Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements FDA # 5-134
IEC 60601-1:2005/AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) FDA # 19-49
IEC 60601-1-2:2014/AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests FDA # 19-36
IEC 60601-1-6:2010/AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability FDA # 5-132
Page | 6
{11}
IEC 60601-2-22: 2007+A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment FDA # 12-268
IEC 62366-1:2015/AMD1:2020 Medical device software - Software life cycle processes FDA # 5-129
IEC 60825-1: 2007 Safety of laser products - Part 1: Equipment classification, and requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)] FDA # 12-273
The following standards were applied to the laser fiber probes:
**Biocompatibility:**
- Cytotoxicity ISO 10993-5:2009
- Irritation ISO 10993-10:2013
- Delayed type hypersensitivity ISO 10993-10:2013
- Material mediated pyrogenicity ISO 10993-11:2018
- Acute systemic toxicity ISO 10993-11:2018
- Allowable limits for leachable substances ISO 10993-17:2009; ISO/TS 21726:2019
- Chemical characterization ISO 10993-18:2020
**Sterilization:**
- ISO 11135:2014 + AMD. 1:2019 for the sterilization process and
- EN ISO 11737-2 for the tests of sterility. T
- EN 556-1: "Sterilization of medical devices - Requirements for medical devices to be designated 'STERILE' - Part 1: Requirements for terminally sterilized medical devices"
The standards testing was conducted by accredited testing laboratories.
9) **Summary of clinical testing:** Clinical testing was not required to establish substantial equivalence.
10) **Conclusion:** After analyzing bench, software validation, and standards compliance tests, it is the conclusion of CeramOptec GMBH that the new CeramOptec GMBH Leonardo® Duster Laser is as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Page | 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.