P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) for daily wear are spherical lenses indicated for the correction of refractive error in not-aphakic persons with otherwise non-diseased eyes with myopia ranging from -0.00 diopters to -10.00 diopters. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are available clear or tinted and may be used to enhance or alter the apparent color of the eye. Eye care practitioners may prescribe the lenses in a frequent/planned replacement program with cleaning, disinfection, and scheduled replacement. The lenses are intended for daily wear and are to be replaced every three months (quarterly). When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Device Story
P-CON contact lenses are spherical, hydrophilic lenses fabricated from polymacon (HEMA/EGDMA copolymer). Lenses are manufactured via cast molding; tinted versions incorporate color additives via pad-printing within the lens matrix. Lenses function as a refractive medium to focus light on the retina. Prescribed by eye care practitioners for daily wear with a three-month replacement schedule; requires patient-performed cleaning and chemical disinfection. Benefits include vision correction and cosmetic eye color alteration. Used in clinical or home settings.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, acute ocular irritation, acute systemic toxicity, material-mediated pyrogenicity) and ISO 9394. Physicochemical and mechanical properties (diameter, base curve, power, water content, oxygen permeability, refractive index) verified against specifications. Shelf life validated for 5 years.
Technological Characteristics
Material: Polymacon (HEMA/EGDMA copolymer); 38% water content. FDA Group 1 (<50% H2O, non-ionic). Manufacturing: Cast molding with pad-printed colorants. Sterilization: Steam. Dimensions: Spherical, daily wear. Connectivity: None.
Indications for Use
Indicated for daily wear correction of refractive error in not-aphakic persons with non-diseased eyes; myopia range -0.00 to -10.00 diopters; refractive astigmatism ≤ 1.50 diopters. Available clear or tinted for eye color enhancement.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 30, 2025
E.O.S Co., Ltd.
% Sanglok Lee
Manager
Wise Company Inc.
#1107, #1108, 204, Gasan digital 1-ro, Geumcheon-gu
Seoul, 08502
Korea, South
Re: K242855
Trade/Device Name: P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color)
Regulation Number: 21 CFR 886.5925
Regulation Name: Soft (Hydrophilic) Contact Lens
Regulatory Class: Class II
Product Code: LPL
Dated: June 16, 2025
Received: June 16, 2025
Dear Sanglok Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242855 - Sanglok Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K242855 - Sanglok Lee
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J Angelo Green -S
J. Angelo Green, Ph.D.
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242855 | |
| Device Name | |
| P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) | |
| Indications for Use (Describe) P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) for daily wear are spherical lenses indicated for the correction of refractive error in not-aphakic persons with otherwise non-diseased eyes with myopia ranging from -0.00 diopters to -10.00 diopters. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are available clear or tinted and may be used to enhance or alter the apparent color of the eye. Eye care practitioners may prescribe the lenses in a frequent/planned replacement program with cleaning, disinfection, and scheduled replacement. The lenses are intended for daily wear and are to be replaced every three months (quarterly). When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
The assigned 510(k) Number: K242855
01. Date of Submission: 2025.07.18
02. Applicant
Company name: E.O.S Co., Ltd.
Address: 188 Yuram-ro, Dong-gu, Daegu, Korea
TEL: +82 53 962 7842
FAX: +82 53 962 7808
Email: ra@eoscl.co.kr
03. Submission Correspondent
Sanglok, Lee
Wise COMPANY Inc.
#303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea
TEL: +82 70 8812 3619 / +82 2 831 3615
FAX: +82 50 4031 3619
Email: info@wisecompany.org
04. Subject Device Identification
Trade Name: P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color)
Common Name: Contact Lens, Daily Wear
Classification Name: Soft (hydrophilic) Contact Lens (21 CFR 886.5925)
Product Code: LPL
Panel: Ophthalmic
Regulation Number: 886.5925
Device Class: Class II
05. Indication for use
P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) for daily wear are spherical lenses indicated for the correction of refractive error in not-aphakic persons with otherwise non-diseased eyes with myopia ranging from -0.00 diopters to -10.00 diopters.
The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Eye care practitioners may prescribe the lenses in a frequent/planned replacement program with cleaning, disinfection, and scheduled replacement. The lenses are intended for daily wear and are to be replaced every three months (quarterly). When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
06. Predicate devices
Predicate device
510(k) Number: K221517
Device Name: POLYVUE (polymacon) Soft (hydrophilic) Contact Lens, POLYVUE COLOR (polymacon) Soft (hydrophilic) Contact Lens
Manufacturer: Interojo Inc.
07. Device Description
The P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) is manufactured using the cast molding method. The hydrophilic characteristics allow aqueous solutions to enter the lens. The lenses are fabricated from polymacon, which is a random copolymer of 2-hydroxyethyl methacrylate (HEMA) crosslinked with ethylene glycol dimethacrylate (EGDMA). The co-polymer consists of 62% polymacon and 38% water by weight when immersed in 0.9% saline solution. The polymacon name has been adopted by the United States Adopted Names Council (USAN).
{5}
The P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) is available clear or tinted for visibility using phthalocyanine blue, tinted in unique pattern to enhance or alter the apparent color of the eye. Each unique pattern may be distributed under unique or "private label" trade names. The lenses are processed to incorporate the 'listed' color additives and contain only the amount of the additive needed to accomplish the intended coloring effect. The lenses contain one or a combination of one or more of the following 'listed' color additives:
| Color Additive | Listing |
| --- | --- |
| C.I Reactive Black 5 | 21 CFR 73.3127 |
| Titanium Dioxide (TiO2) | 21 CFR 73.3126 |
| Carbazole Violet (i.e.,C.I Pigment Violet 23) | 21 CFR 73.3107 |
| Phthalocyanine green (i.e.,C.I Pigment Green 7) | 21 CFR 73.3124 |
| D&C Yellow No. 10 | 21 CFR 74.3710 |
| D&C Red No. 17 | 21 CFR 74.3230 |
| [Phthalocyaninato (2-)] Copper (i.e.,C.I Pigment Blue 15) | 21 CFR 74.3045 |
When producing the color lenses, the manufacturing process changes the specifications of the clear lens by pad-printing the color pigment(s)—entrapping the colorants in the interpenetrating network of the contact lens material—in a location that corresponds to the iris. The color pigments used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens, the coloring process does not alter the original characteristics of the clear, pre-tinted lens.
Eye care practitioners may prescribe the lenses in a frequent/planned replacement program with cleaning, disinfection, and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
## 08. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
### Biocompatibility:
Biocompatibility testing was conducted in accordance with the ISO 10993 series and GLP (Good Laboratory Practice) regulations. Packaging material testing was conducted using the big blister configuration.
The following tests were performed:
- Cytotoxicity: Testing conducted in accordance with ISO 10993-5 confirmed that both the finished lens extracts and packaging material extracts (Big blister and Aluminum foil seal) were non-cytotoxic.
- Sensitization: Testing conducted in accordance with ISO 10993-10 confirmed that the lens extracts did not induce sensitization.
- Acute Ocular Irritation: Testing conducted in accordance with ISO 10993-23 confirmed that both the finished lens extracts and packaging material extracts (Big blister and Aluminum foil seal) were non-irritating.
- Acute Systemic Toxicity (mouse): Testing conducted in accordance with ISO 10993-11 confirmed that both the finished lens extracts and packaging material extracts (Big blister and Aluminum foil seal) were non-toxic.
- Material-Mediated Pyrogenicity (rabbit): Testing conducted in accordance with ISO 10993-11 confirmed that the finished lenses did not induce a pyrogenic response.
- Determination of Biocompatibility of Contact Lenses in Rabbit Eyes: Testing conducted in accordance with ISO 9394 confirmed that the finished lenses were non-toxic.
- Primary Packaging Biocompatibility: Cytotoxicity, acute systemic toxicity, and ocular irritation testing were conducted on the big blister and foil seal (without the lens), and all packaging components were found to be non-cytotoxic, non-irritating, and non-toxic.
The small blister configuration utilizes the same raw materials, laminate structure, and manufacturing process as the big blister. Given that the big blister has a larger surface area and is subject to a greater
{6}
amount of ink and processing stress, it is considered the worst-case configuration from a biocompatibility standpoint. Therefore, the biocompatibility testing performed on the big blister is scientifically justified as representative of the small blister configuration as well. No additional biocompatibility testing was conducted on the small blister.
## Shelf Life:
Testing was performed to evaluate the stability, sterility, and package integrity of The P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) finished contact lenses over the duration of the labeled expiration date. The data shows the 5 years validity shelf life of the subject device.
## Physicochemical & Mechanical Properties:
The following tests were completed to verify Physicochemical & Mechanical Properties: Shape and Appearance, Diameter, Basecurve, Center Thickness, Power, Visible light transmittance, Contact angle, Water contents, Oxygen permeability, Refractive Index, Specific Gravity, pH, Osmolality.
## 09. Substantially Equivalent Conclusion
Table 1: Substantial Equivalence Comparison
| Product Name | SUBJECT Device | PREDICATE Device | Equivalence Discussion |
| --- | --- | --- | --- |
| 510(k) number | Not assigned yet | K221517 | - |
| Product code | LPL | LPL | - |
| Regulatory class | Class II | Class II | - |
| Regulation Number | 21 CFR 886.5925 | 21 CFR 886.5925 | - |
| Intended use | P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color) for daily wear are indicated for the correction of refractive error in not-aphakic persons with otherwise non-diseased eyes with myopia. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.
The lenses are available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Eye care practitioners may prescribe the lenses in a frequent/planned replacement program with cleaning, disinfection, and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system. | The POLYVUE and POLYVUE COLOR (polymacon) ASPHERIC Soft (hydrophilic) Contact Lenses for daily wear are indicated for the correction of refractive error in not-aphakic persons with otherwise non-diseased eyes with myopia or hyperopia and early presbyopia up to 1.25 diopters. The lenses may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.
The lenses are available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Eye care practitioners may prescribe the lenses in a frequent/planned replacement program with cleaning, disinfection, and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system. | The intended use of the SUBJECT Device is equivalent to those of the PREDICATE because all of them correct visual acuity in aphakic and non-aphakic persons with non-diseased eyes with myopia. All the devices primarily correct myopia and handle astigmatism within set limits, making their intended uses fundamentally the same. |
| Actions | The contact lenses act as a refractive medium that focus light rays from near and distant objects on the retina | The contact lenses act as a refractive medium that focus light rays from near and distant objects on the retina | Equivalent |
| FDA Group | FDA Group 1 (<50% H 2O, non-ionic polymer) | FDA Group 1 (<50% H 2O, non-ionic polymer) | Equivalent |
| Production Method | Fully molded | Fully molded | Equivalent |
{7}
| Product Name | SUBJECT Device | PREDICATE Device | Equivalence Discussion |
| --- | --- | --- | --- |
| USAN name | polymacon | polymacon | Equivalent |
| Water Content (%) | 38±2% | 38±2% | Equivalent |
| Oxygen Permeability x 10-11 (cm2/sec)(mIO2)/(m I x mmHg @ 35°C)) (revised Fatt method) | 9.40± 20% | 10.26± 20% | Although there are differences, all values are within the criteria specified in FDA guidance, 'Soft (Hydrophilic) Daily Wear Contact Lenses – Performance Criteria for Safety and Performance Based Pathway' on March 28, 2023, making them equivalent. |
| Specific Gravity | 1.124 ± 0.0371 | Not specified | |
| Refractive Index (hydrated) | 1.437 ± 0.005 | 1.438 ± 0.005 | |
| UV Blocker | No | No | Equivalent |
| Pad-Printed Tinting | Yes | Yes | Equivalent |
| Primary Packaging | Blister pack with Aluminum foil seal (Small and Big) | blister base, foil seal | Equivalent |
| Sterilization Method | Steam sterilization | Steam sterilization | Equivalent |
The subject device, The P-CON (polymacon) Soft (hydrophilic) Contact Lenses (Tinted, Color), is determined to be Substantially Equivalent (SE) to the predicate device, K221517, in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.