K250469 · Rivanna Medical, Inc. · ITX · Jun 13, 2025 · Radiology
Device Facts
Record ID
K250469
Device Name
Accuro® 3S Needle Guide Kit
Applicant
Rivanna Medical, Inc.
Product Code
ITX · Radiology
Decision Date
Jun 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
The Accuro® 3S Needle Guide Kit is intended to be used with the Accuro® 3S diagnostic ultrasound imaging system. The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide. The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminants.
Device Story
Accuro® 3S Needle Guide Kit is a sterile, single-use disposable accessory for the Accuro® 3S diagnostic ultrasound imaging system. The kit includes a needle guide, integrated probe cover, conductive ultrasound gel, and a patient drape with integrated elastic bands. The probe cover isolates the transducer from contaminants; the needle guide aligns the needle with the ultrasound imaging plane to assist clinicians in accurate needle tip placement relative to anatomical structures. The patient drape's elastic bands provide pressure against the back of the transducer to stabilize the probe and needle guide position against the patient's skin during imaging and needle advancement. Used in clinical settings by healthcare professionals. The device benefits patients by facilitating precise needle placement during ultrasound-guided procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by benchtop verification and validation testing, including needle insertion force, angulation accuracy, tip depth accuracy, sterile barrier integrity, and simulated use evaluations.
Technological Characteristics
Single-use sterile kit including needle guide, probe cover, gel, and patient drape. Sterilized via ethylene oxide (ISO 11135:2014). Biocompatibility per ISO 10993 series. Mechanical stabilization provided by elastic bands on the patient drape. No electronic or software components; purely mechanical accessory for diagnostic ultrasound transducers.
Indications for Use
Indicated for use by healthcare professionals to assist in needle placement relative to anatomical structures during ultrasound-guided procedures. Intended for use with the Accuro® 3S diagnostic ultrasound imaging system.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 13, 2025
Rivanna Medical, Inc.
Will Mauldin
Chairman, Co-founder, and CEO
2400 Hunters Way
Charlottesville, Virginia 22911
Re: K250469
Trade/Device Name: Accuro® 3S Needle Guide Kit
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic ultrasonic transducer
Regulatory Class: Class II
Product Code: ITX
Dated: May 12, 2025
Received: May 12, 2025
Dear Will Mauldin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250469 - Will Mauldin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250469 - Will Mauldin
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250469 | |
| Device Name Accuro® 3S Needle Guide Kit | |
| Indications for Use (Describe) The Accuro® 3S Needle Guide Kit is intended to be used with the Accuro® 3S diagnostic ultrasound imaging system. The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide. The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminants. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
RIVANNA
K250469
# 510(k) Summary
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
Date: June 12th, 2025
Submitter: Rivanna Medical, Inc.
2400 Hunters Way
Charlottesville, VA 22911
Primary Contact Person: F. William Mauldin, Jr.
CEO
Rivanna Medical, Inc.
T: 800-645-7508
Subject Device Trade Name: Accuro® 3S Needle Guide Kit
Common Name: Transducer, Ultrasonic, Diagnostic
Device Class: Class II
Regulation: 21 CFR 892.1570, Diagnostic Ultrasonic Transducer
Product Codes: ITX
Legally Marketed Predicate Device:
Primary Predicate Device: K171348 Pinpoint™ GT Needle Guide Kits
Common Name: Transducer, Ultrasonic, Diagnostic
Device Class: Class II
Regulation: 21 CFR 892.1570, Diagnostic Ultrasonic Transducer
Product Codes: ITX
Device Description:
The Accuro® 3S Needle Guide Kit consists of the following single use sterile disposable components: Accuro® 3S Needle Guide with integrated probe cover, conductive ultrasound gel, and a patient drape with two integrated probe elastic bands.
The kit components are assembled into a sterilized CSR Wrap then packaged in a sterile tray.
The final kit is sterilized under ethylene oxide.
RIVANNA | rivannamedical.com | info@rivannamedical.com
T: 800-645-7508 | Fax: 434-483-5720 | 2400 HUNTERS WAY, CHARLOTTESVILLE, VA 22911
{5}
2
# intended Use/Indications for Use:
The Accuro® 3S Needle Guide Kit is intended to be used with the Accuro® 3S diagnostic ultrasound imaging system.
The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide.
The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminants.
## Comparison to Predicate Device(s):
Accuro® 3S Needle Guide Kit is substantially equivalent to the predicate device in terms of technological characteristics and safety and effectiveness.
The Accuro® 3S Needle Guide Kit indications for use is similar to the predicate device, Pinpoint™ GT Needle Guide Kits (K171348). Both are indicated to cover an ultrasound probe and isolate from microbial and other contaminants, and both have needle guides to assist with placement of a needle to a specific anatomical structure. Both subject and predicate devices are prescription use and intended for the same use environments.
The key difference between the subject and predicate device is the inclusion of a patient drape with bands that support the needle guide positioning. The elastic bands are integrated into the patient drape and do not come in direct contact with the patient. The intended function of the elastic bands is to provide pressure against the back of the transducer probe to stabilize its position, and thus the Needle Guide position, against the skin during ultrasound imaging and needle advancement. Non-clinical testing for performance, including usability, and safety has been conducted to ensure that no new risks are introduced due to the addition of elastic bands on the Patient Drape. This includes verification testing and application of ISO 14971 for risk management.
## Summary of Non-Clinical Tests:
Design functionality of the subject device was tested to ensure that it meets the requirements of the intended end users. Testing included simulated use evaluations performed by representative end users in addition to benchtop testing.
{6}
3
The following benchtop verification and validation testing was performed following internal test protocols:
- Needle guide dimensions
- Needle insertion force and damage during needle insertion
- Needle angulation accuracy
- Needle tip depth accuracy
- Needle guide attachment / removal force
- Sterile barrier integrity
- Packaging integrity
- Elastic bands holding force
- Needle gate pinch force
- Elastic bands attachment durability
- Imaging quality following administration of the needle guide and probe cover
- Probe midline needle insertion
- Material dimensions
- Packaging configuration
Assurances of quality were further established by employing the following elements of product design and development in accordance with 21 CFR 820 and ISO 14971:2019:
- Risk analysis
- Requirements development
- Design reviews
- Verification and Validation
- Safety compliance verification
The following guidance documents and standards were used to determine appropriate methods for evaluating the performance of the device:
- AAMI / ANSI / ISO 10993-5:2009/(R)2014, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- AAMI / ANSI ST72:2019, Bacterial Endotoxins - Test Methods, Routine Monitoring, And Alternatives To Batch Testing
- ASTM D4332-14, Standard Practice For Conditioning Containers, Packages, Or Packaging Components For Testing
- ASTM F1980-16, Standard Guide For Accelerated Aging Of Sterile Barrier Systems For Medical Devices
- ASTM F756, Standard Practice for Assessment of Hemolytic Properties of Materials
- ISO 10993-1:2018, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing
- ISO 10993-4:2017, Biological Evaluation of Medical Devices - Part 4: Selection of tests for interactions with blood
{7}
F
- ISO 10993-7:2008, Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals
- ISO 10993-10:2021, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- ISO 10993-11:2017, Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity
- ISO 10993-12:2021, Biological Evaluation Of Medical Devices - Part 12: Sample Preparation And Reference Materials
- ISO 11135: 2014, Sterilization of healthcare products - Ethylene Oxide- Requirements For Development, Validation And Routine Control Of A Sterilization
- ISO 11607-1:2019, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems, Amendment 1: Application of risk management 2023
- ISO 11607-2:2019, Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes, Amendment 1: Application of risk management 2023
- FDA-2013-D-0305, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff, 09/2023.
## Summary of Clinical Tests:
The subject of this premarket submission, Accuro® 3S Needle Guide Kit, did not require clinical studies to support the determination of substantial equivalence.
## Conclusion:
Accuro® 3S Needle Guide Kit exhibits similar characteristics and indications for use as legally marketed predicate and reference devices and testing demonstrates substantially equivalent performance. Therefore, Rivanna Medical considers Accuro® 3S Needle Guide Kit as substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.