← Product Code [FET](/productcode/FET) · K250462

# UL UHD-Clear View 4K Camera System (UL-3Chip 4K) (K250462)

_Univlabs Technologies Private Limited · FET · Aug 11, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250462

## Device Facts

- **Applicant:** Univlabs Technologies Private Limited
- **Product Code:** [FET](/productcode/FET.md)
- **Decision Date:** Aug 11, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The UL UHD-Clear View 4K Camera System (UL-3Chip 4K), consisting of a Camera Control Unit and an Endoscopic Camera Head, is indicated for use with compatible endoscopes to provide visualization during general endoscopic procedures. These procedures may involve the visual inspection of internal anatomical structures to support diagnostic or therapeutic interventions. Patient suitability for particular endoscopic procedures must be determined by the responsible physician based on the individual's clinical condition.

## Device Story

System comprises Camera Control Unit (CCU) and Endoscopic Camera Head; captures visual data via compatible endoscopes; processes electronic signals from camera head; outputs 4K resolution video to external medical-grade display monitors. Used in clinical settings for general endoscopic procedures; operated by physicians. Provides real-time visualization of internal anatomy to support diagnostic or therapeutic interventions. Benefits include high-resolution imaging for improved anatomical inspection.

## Clinical Evidence

Bench testing only. Comparative testing against predicate evaluated spatial resolution, depth of field, field of view, zoom characterization, and color enhancement. Device tested for safety/performance per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18.

## Technological Characteristics

CMOS sensor technology; 3840x2160p resolution; 12G/3G-SDI and HDMI display ports; adaptive zoom. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-18 (endoscopic equipment).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Image1 S CCU (TC200US, TC201US, TC300US, TC301US, TC302US, TC304US); Image1 S 4U Camera Head (TH120) ([K201135](/device/K201135.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

August 11, 2025

Univlabs Technologies Private Limited
Vishwanath Pratap Singh
Engineering Director
Plot No : 31, Sector -33, Near Info-City 2
Haryana
Gurugram, HR 122003
India

Re: K250462
Trade/Device Name: UL UHD-Clear View 4K Camera System (UL-3Chip 4K)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FET
Dated: June 27, 2025
Received: July 7, 2025

Dear Vishwanath Pratap Singh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250462 - Vishwanath Pratap Singh
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K250462 - Vishwanath Pratap Singh
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA
L. HITHE - S
Digitally signed by
TANISHA L. HITHE - S
Date: 2025.08.11
20:45:14 -04'00'

Tanisha Hithe, M.S., M.HS.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

K250462

Page 1 of 1

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250462 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  UL UHD-Clear View 4K Camera System (UL-3Chip 4K)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The UL UHD-Clear View 4K Camera System (UL-3Chip 4K), consisting of a Camera Control Unit and an Endoscopic Camera Head, is indicated for use with compatible endoscopes to provide visualization during general endoscopic procedures. These procedures may involve the visual inspection of internal anatomical structures to support diagnostic or therapeutic interventions. Patient suitability for particular endoscopic procedures must be determined by the responsible physician based on the individual's clinical condition.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

UL UHD-Clear View 4K Camera System

{4}

K250462
Page 1 of 2

|  510(k) #: K250462 | 510(k) Summary | Prepared on: 2025-07-04  |
| --- | --- | --- |
|  Contact Details |   | 21 CFR 807.92(a)(1)  |
|  Applicant Name | UnivLabs Technologies Private Limited |   |
|  Applicant Address | Plot No : 31, Sector -33, Near Info-City II Haryana Gurugram HR 122003 India |   |
|  Applicant Contact Telephone | +91-9560925518 |   |
|  Applicant Contact | Mr. SUNIL KUMAR SINGH |   |
|  Applicant Contact Email | sunil@univlabs.in |   |
|  Correspondent Name | UnivLabs Technologies Private Limited |   |
|  Correspondent Address | Plot No : 31, Sector - 33, Near Info-City II Haryana Gurugram HR 122003 India |   |
|  Correspondent Contact Telephone | +91-9491729764 |   |
|  Correspondent Contact | Mr. Venkataramana Anant Bhagavati |   |
|  Correspondent Contact Email | regulatory@univlabs.in |   |
|  Device Name |   | 21 CFR 807.92(a)(2)  |
|  Device Trade Name | UL UHD-Clear View 4K Camera System (UL-3Chip 4K) |   |
|  Common Name | Endoscope and accessories |   |
|  Classification Name | Endoscopic Video Imaging System/Component, Gastroenterology-Urology |   |
|  Regulation Number | 876.1500 |   |
|  Product Code(s) | FET |   |
|  Legally Marketed Predicate Devices |   | 21 CFR 807.92(a)(3)  |
|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
|  K201135 | Image1 S CCU (TC200US, TC201US, TC300US, TC301US, TC302US, TC304US); Image1 S 4U Camera Head (TH120) | FET  |
|  Device Description Summary |   | 21 CFR 807.92(a)(4)  |
|  The UL UHD-Clear View 4K Camera System (UL-3Chip-4K) consisting of a Camera Control Unit (CCU) and Endoscopic Camera Head supports video output up to 4K resolution for connection to external medical-grade display monitors. The Camera Control Unit (CCU) processes electronic signals transmitted from an endoscopic camera head and outputs the video signal to an external display monitor.  |   |   |
|  Intended Use/Indications for Use |   | 21 CFR 807.92(a)(5)  |

{5}

K250462
Page 2 of 2

The UL UHD-Clear View 4K Camera System (UL-3Chip 4K), consisting of a Camera Control Unit and an Endoscopic Camera Head, is indicated for use with compatible endoscopes to provide visualization during general endoscopic procedures. These procedures may involve the visual inspection of internal anatomical structures to support diagnostic or therapeutic interventions. Patient suitability for particular endoscopic procedures must be determined by the responsible physician based on the individual's clinical condition.

## Indications for Use Comparison
**21 CFR 807.92(a)(5)**

The subject device, UL UHD-Clear View 4K Camera System, is indicated for use with compatible endoscopes to provide visualization during general endoscopic procedures. The predicate device, Karl Storz Image 1S CCU and 4U Camera Head, is indicated for visualization, image recording, and documentation during endoscopic and microscopic procedures. The subject device does not support image recording or documentation. These are minor differences, and the subject device's indications represent a subset of the predicate's. Additional labeling language in the subject device (e.g., references to physician judgment and support for diagnostic or therapeutic interventions) is intended to provide clarity and is consistent with the IMDRF definition of "Indications for Use." It does not broaden the scope of intended use. Therefore, the indications for use of the subject device do not constitute a new indication for use.

## Technological Comparison
**21 CFR 807.92(a)(6)**

The UL UHD-Clear View 4K Camera System (UL-3Chip-4K) consisting of a Camera Control Unit (CCU) and Endoscopic Camera Head supports video output up to 4K resolution for connection to external medical-grade display monitors. The Camera Control Unit (CCU) processes electronic signals transmitted from an endoscopic camera head and outputs the video signal to an external display monitor.

Both the subject and predicate device are based on CMOS Sensor technology with a sensor resolution of 3840*2160p. Both Subject and predicate device(s) have adaptive zoom and 12G/3G-SDI display port(s). Additionally, the subject device has additional two HDMI port(s).

## Non-Clinical and/or Clinical Tests Summary &amp; Conclusions
**21 CFR 807.92(b)**

Comparative Bench Testing was performed between the subject device and predicate device for the following parameters in order to ensure Proper Characterization of Image Quality :

- Spatial Resolution
- Depth of Field
- Field of View
- Zoom Characterization
- Color Enhancement

There are no performance standards or special controls developed for Camera Control Unit (CCU) and Endoscopic Camera Head(s). However, UL UHD-Clear View 4K Camera System (UL-3Chip 4K) is tested according to following FDA recognized consensus standards :

- IEC 60601-1 : 2005 + AMD1 : 2012 + AMD2 : 2020 : Medical Electrical Equipment - Part 1 : General Requirements for Basic Safety and Essential Performance.
- IEC 60601-1-2 : 2014 + AMD1 : 2020 : Medical Electrical Equipment - Part 1-2 : Collateral Standard : Electromagnetic Disturbances - Requirements and Tests.

The essential performance for the above tests confirms to IEC 60601-2-18 : 2009 : Particular requirements for the basic safety and essential performance of endoscopic equipment.

Software Verification &amp; Validation Testing was performed as per Content of Premarket Submissions for Device Software Functions (Moderate Level of Concern).

"Not Applicable"

The subject and predicate device have similar indications for use and technological characteristics as the predicate device. The minor differences between the subject and predicate devices do not raise different questions of safety and effectiveness. Non-clinical performance testing (Bench Testing), electromagnetic compatibility (EMC), electrical, mechanical, and thermal safety, demonstrates that the subject device is as safe and effective as the predicate. Therefore, the UL UHD-Clear View 4K Camera System (UL-3Chip 4K) is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K250462](https://fda-staging.innolitics.com/device/K250462)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
