GBrain MRI is a post processing medical device software intended for analyzing and quantitatively reporting signal hyperintensities in the brain on FLAIR MR images in the context of diagnostic radiology. GBrain MRI is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. It is not intended for detection or specific diagnosis of any disease nor for the detection of signal hyperintensities. GBrain MRI should not be used in-lieu of a full evaluation of the patient's MRI scans. The physician retains the ultimate responsibility for making the final patient management and treatment decisions.
Device Story
GBrain MRI is a non-invasive post-processing software for adult brain MRI; operates as a computing appliance. Inputs: T2-weighted FLAIR brain MR images (DICOM). Processing: Automated pipeline performs skull stripping, signal normalization, deep learning-based segmentation of hyperintensities, volume calculation, and report generation. Outputs: Secondary capture DICOM images with segmentation overlays, DICOM Encapsulated PDF, and DICOM SR. Used in hospitals, clinics, and imaging centers by radiologists/clinicians. Software provides quantitative metrics as a support tool; does not perform diagnosis or detection. Physician reviews output to complement clinical assessment; aids in structural brain MRI evaluation. Benefits: Provides standardized, automated volumetric quantification of hyperintensities to assist clinical decision-making.
Clinical Evidence
Bench testing only. Performance evaluated by comparing software-derived segmentations to a STAPLE consensus of expert-labeled segmentations. Metrics included OLS regression and Dice similarity coefficient. Testing covered multiple scanner manufacturers and field strengths. Results met pre-defined acceptance criteria for accuracy and reliability.
Technological Characteristics
Software-only device; operates on off-the-shelf hardware. Deep learning-based automated segmentation pipeline. Inputs/outputs via DICOM (standardized). No patient-contacting components. Connectivity via DICOM file transfer from scanner or PACS.
Indications for Use
Indicated for adult patients (21+) undergoing brain MRI. Used for post-processing analysis and quantitative reporting of signal hyperintensities on FLAIR MR images. Contraindicated for pediatric populations and patients with history of brain tumor or resection surgery.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
April 11, 2025
Galileo CDS, Inc
% Mary Vater
Director of Regulatory Affairs
Innolitics LLC
1101 West 34th Street #550
Austin, Texas 78705
Re: K250416
Trade/Device Name: GBrain MRI
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: February 12, 2025
Received: February 13, 2025
Dear Mary Vater:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K250416 - Mary Vater
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250416 - Mary Vater
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250416
Device Name
GBrain MRI
Indications for Use (Describe)
GBrain MRI is a post processing medical device software intended for analyzing and quantitatively reporting signal hyperintensities in the brain on FLAIR MR images in the context of diagnostic radiology.
GBrain MRI is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. It is not intended for detection or specific diagnosis of any disease nor for the detection of signal hyperintensities.
GBrain MRI should not be used in-lieu of a full evaluation of the patient's MRI scans. The physician retains the ultimate responsibility for making the final patient management and treatment decisions.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K250416
# 510(k) Summary
## 1. CONTACT INFORMATION
| Company Name | Galileo |
| --- | --- |
| Address | 7000 N Mopac Expy, Suite 200, Austin, TX 78731 |
| Phone Number | (512) 222-9173 |
| Fax Number | {MANUFACTURER FAX NUMBER} |
| Company Representative | Abhijeet Pradhan |
| Email | ap@galileocds.com |
| Primary Contact | Mary Vater |
| Primary Contact Phone Number | +1.913.523.6988 |
| Primary Contact Email | mvater@innolitics.com |
| Date Summary Prepared | February 5, 2025 |
## 2. DEVICE INFORMATION
Trade Name: GBrain MRI
Common Name: Brain MRI Image Processing Software
Classification Name: Automated Radiological Image Processing Software
Regulation Number: 21 CFR 892.2050
Product Code(s): QIH, LLZ
## 3. DEVICE DESCRIPTION
### 3.1. Description
GBrain MRI is a non-invasive MR imaging post-processing medical device software that aids in the volumetric quantification of hyperintensities in T2-weighted Fluid Attenuated Inversion Recovery (T2w FLAIR) brain MR images. It is intended to aid the trained radiologist in quantitative measurements.
510(k) Summary
{5}
The input to the software is the T2w FLAIR brain MR images.
The outputs are volume measurements in Secondary Capture DICOM format, a DICOM Encapsulated pdf file, as well as a DICOM SR. More specifically, the total volume of hyperintensities in the input T2w FLAIR are shown in a new secondary capture image series, called GBrain T2 FLAIR with a segmentation overlay on the hyperintensities that were used to measure the total volumes. These volume measurements are summarized in the DICOM encapsulated pdf and DICOM SR files.
The outputs are provided in standard DICOM format that can be displayed on most third-party DICOM workstations and Picture Archive and Communications Systems (PACS).
The software is suitable for use in routine patient care as a support tool for radiologists in assessment of structural adult brain MRIs, by providing them with complementary quantitative information.
The GBrain MRI processing architecture includes a proprietary automated internal pipeline that performs skull stripping, signal normalization, segmentations, volume calculations, and report generation.
From a workflow perspective, GBrain MRI is packaged as a computing appliance that is capable of supporting DICOM file transfer for input, and output of results. The software is designed without the need for a user interface after installation. Any processing errors are reported either in the output series report, or in the system log files.
GBrain MRI software is intended to be used by trained personnel only and is to be installed by trained technical personnel.
Quantitative reports and derived image data sets are intended to be used as complementary information in the review of a case.
The GBrain MRI software does not have any accessories or patient contacting components.
The GBrain MRI device is intended to be used for the adult population only.
# 4. SUBJECT DEVICE INDICATIONS FOR USE
GBrain MRI is a post processing medical device software intended for analyzing and quantitatively reporting signal hyperintensities in the brain on FLAIR MR images in the context of diagnostic radiology.
GBrain MRI is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. It is not intended for detection or specific diagnosis of any disease nor for the detection of signal hyperintensities.
GBrain MRI should not be used in-lieu of a full evaluation of the patient's MRI scans. The
510(k) Summary
{6}
physician retains the ultimate responsibility for making the final patient management and treatment decisions.
## 4.1. Subject Device Contraindications for Use
GBrain MRI is contraindicated for use in pediatric patient populations, and in patients who have undergone brain tumor or other resection surgery.
## 4.2. Subject Device Intended Use
GBrain MRI is a post processing medical device software intended for analyzing and quantitatively reporting signal hyperintensities in the brain on FLAIR MR images in the context of diagnostic radiology.
GBrain MRI is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. It is not intended for detection or specific diagnosis of any disease nor for the detection of signal hyperintensities.
GBrain MRI should not be used in lieu of a full evaluation of the patient’s MRI scans. The physician retains the ultimate responsibility for making the final patient management and treatment decisions.
## 5. SUBSTANTIAL EQUIVALENCE COMPARISON
## 5.1. Primary Predicate Device Indications for Use
OnQ Neuro is a fully automated post processing medical device software intended for analyzing and evaluating neurological MR image data.
OnQ Neuro is intended to provide automatic segmentation, quantification, and reporting of derived image metrics.
OnQ Neuro is additionally intended to provide automatic fusion of derived parametric maps with anatomical MRI data.
OnQ Neuro is intended for use on brain tumors, which are known/confirmed to be pathologically diagnosed cancer.
OnQ Neuro is intended for comparison of derived image metrics from multiple time points.
The physician retains the ultimate responsibility for making the final diagnosis and treatment decision.
510(k) Summary
{7}
5.2. Secondary Predicate Device Indications for Use
QP-Brain® is a medical imaging processing application intended for automatic labeling and volumetric quantification of segmentable brain structures and white matter hyperintensities (WMH) from a set of adults and adolescents 18 and older MR images. Volumetric measurements may be compared to reference percentile data. The application is used by clinicians with proper training, as a support tool in assessment of structural MRIs. Patient management decisions should not be based solely on the results of the device.
5.3. Subject Device Indications for Use
GBrain MRI is a post processing medical device software intended for analyzing and quantitatively reporting signal hyperintensities in the brain on FLAIR MR images in the context of diagnostic radiology.
GBrain MRI is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. It is not intended for detection or specific diagnosis of any disease nor for the detection of signal hyperintensities.
GBrain MRI should not be used in-lieu of a full evaluation of the patient's MRI scans. The physician retains the ultimate responsibility for making the final patient management and treatment decisions.
5.4. Indications for Use Equivalence Discussion
The intended use of all of the devices is the same - analyzing and evaluating neurological imaging data. The indications are equivalent to the scope of regulation 21 CFR 892.2050.
The indications for use of the Subject Device are substantially equivalent to those of the Primary and Secondary Predicate Devices.
The output of the Primary Predicate Device (OnQ Neuro, K210831) consists of fully automated segmentations, visualizations and volumetric measurements of various regions in cancerous tumors including but not limited to regions which present as hyperintense signal on T2w FLAIR MRI. These hyperintensities may be present in any part of the brain.
The output of the Secondary Predicate Device (OP-Brain, K232231) consists of segmentations, visualizations and volumetric measurements of various brain structures, and of hyperintense signal on T2w FLAIR MRI which are present in the white matter, and typically referred to as White Matter Hyperintensities. The Secondary Predicate Device has a general indication for use and is not specific to a target disease or condition.
The output of the subject device also consists of segmentations, visualizations and volumetric measurements of hyperintensities on T2w FLAIR MRI. The subject device has a general indication for use and is not specific to a target disease or condition.
510(k) Summary
{8}
To elaborate further, the subject and both the predicate devices share the intended use of automated quantification of the extent of FLAIR hyperintensity.
The Primary Predicate Device measures volumes in additional MRI series, and assigns specific clinical significance to these measurements, (enhancing tissue, whole tumor, necrosis, etc) since the measurements are made on cases with an apriori known disease. This functionality is not present in the subject device.
The Secondary Predicate Device measures brain structure volumes in additional MRI series. The subject device does not measure brain structure volumes.
The shared technical tasks performed by the subject and predicate devices, with regards to the measurement of FLAIR hyperintensities are exactly the same.
The predicate devices have other functions are outside the scope of the subject device.
# 5.5. Device Comparison Table
| Feature/Function | GBrain MRI | Primary Predicate OnQ Neuro (K210831) | Secondary Predicate QP-Brain (K232231) | Comments on Substantial Equivalence |
| --- | --- | --- | --- | --- |
| 510K Number | N/A | K210831 | K232231 | - |
| Manufacturer | Galileo CDS Inc | CorTechs Labs, Inc. | Quibim S.L. | - |
| Classification | Class II | Class II | Class II | Same |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Regulation Description | Medical image management and processing system. | Medical image management and processing system. | Medical image management and processing system. | Same |
| Classification Name | Automated Radiological Image Processing Software | Automated Radiological Image Processing Software | Automated Radiological Image Processing Software | Same |
| Product code | QIH, LLZ | QIH | QIH, LLZ | Same |
| | | | | |
| Intended Use | MR imaging data post processing software | MR imaging data post processing software | MR imaging data post processing software | Same |
| Type of Imaging Scan | MRI | MRI | MRI | Same |
| Intended Body Part | Brain | Brain | Brain | Same |
510(k) Summary
{9}
| Feature/Function | GBrain MRI | Primary Predicate OnQ Neuro (K210831) | Secondary Predicate QP-Brain (K232231) | Comments on Substantial Equivalence |
| --- | --- | --- | --- | --- |
| Indications for Use | GBrain MRI is a post processing medical device software intended for analyzing and quantitatively reporting signal hyperintensities in the brain on FLAIR MR images in the context of diagnostic radiology.
GBrain MRI is intended to provide automatic segmentation, quantification, and reporting of derived image metrics. It is not intended for the detection or specific diagnosis of any disease nor for the detection of signal hyperintensities.
GBrain MRI should not be used in-lieu of a full evaluation of the patient’s MRI scans. The physician retains the ultimate responsibility for making the final patient management and treatment decisions. | OnQ Neuro is a fully automated post processing medical device software intended for analyzing and evaluating neurological MR image data.
OnQ Neuro is intended to provide automatic segmentation, quantification, and reporting of derived image metrics.
OnQ Neuro is additionally intended to provide automatic fusion of derived parametric maps with anatomical MRI data.
OnQ Neuro is intended for use on brain tumors, which are known/confirmed to be pathologically diagnosed cancer.
OnQ Neuro is intended for comparison of derived image metrics from multiple time points.
The physician retains the ultimate responsibility for making the final diagnosis and treatment decision. | QP-Brain® is a medical imaging processing application intended for automatic labeling and volumetric quantification of segmentable brain structures and white matter hyperintensities (WMH) from a set of adults and adolescents 18 and older MR images. Volumetric measurements may be compared to reference percentile data. The application is used by clinicians with proper training, as a support tool in assessment of structural MRIs. Patient management decisions should not be based solely on the results of the device. | The indications for use are similar to the predicate devices. The subject device analyzes signal hyperintensities regardless of location similar to the primary predicate and also detects hyperintensities in white matter similar to the secondary predicate. As with the secondary predicate, the subject device has a general indication and is not specific to a particular disease or condition. |
| Environment for use | Hospital, Clinic, Imaging Center, Medical Offices | Hospital, Clinic, Imaging Center, Medical Offices | Hospital, Clinic, Imaging Center, Medical Offices | Same |
| Intended Patient Population | Adult patients aged 21 and above with a brain MRI study. | Patients with brain tumors, which are known/confirmed to be | Adult patients and adolescent patients aged 18 through 21 with brain MRI | Similar to secondary predicate and the majority of the primary predicate’s population. |
510(k) Summary
{10}
| Feature/Function | GBrain MRI | Primary Predicate OnQ Neuro (K210831) | Secondary Predicate QP-Brain (K232231) | Comments on Substantial Equivalence |
| --- | --- | --- | --- | --- |
| | | pathologically diagnosed cancer. | study. Available up to 94 years. | |
| Intended User | Radiologists, Imaging Professionals, and other Clinicians working with radiological images. The application should be used as a support tool in assessment of structural MRIs. Patient management decisions should not be based solely on the results of the device. | Radiologists, and Oncologists | The application should be used by clinicians with proper training, as a support tool in assessment of structural MRIs. Patient management decisions should not be based solely on the results of the device. | |
| | | | | |
| Design and Incorporated Technology | Automated segmentation using deep learning followed by volume calculations, and report generation. Results displayed on PACS. | Automated segmentation using deep learning followed by volume calculations. Results displayed on PACS. | Automated segmentation using deep learning followed by volume calculations, and report generation. Results displayed on PACS. | Same |
| Physical Characteristics | Software package-Operates on off-the-shelf hardware (multiple vendors). | Software package-Operates on multiple platforms | Software package-Operates on on off-the-shelf hardware (multiple vendors). | Same |
| Data Source | Supports DICOM format as input from the MRI scanner or the PACS. | Supports DICOM format as input from the MRI scanner. | Supports DICOM format as input from the MRI scanner. | Same with the addition of input from the PACS. |
| Output | Provides volumetric measurements of regions with hyperintense signal on T2w FLAIR images. Includes segmented color overlays and | Provides volumetric measurements of different regions of brain tumors including but not restricted to regions with hyperintense signal on T2w FLAIR images. Includes segmented | Provides volumetric measurements of different brain structures on T1 MR images, as well as of White Matter Hyperintensities on T2w FLAIR images. Includes segmented color overlays and volumetric | Same |
| | included in the image. Includes segmented color overlays and | included in the image. Includes segmented | included in the image. Includes segmented color | |
| | included in the image. Includes segmented | included in the image. Includes segmented | overlays and volumetric | |
510(k) Summary
{11}
| Feature/Function | GBrain MRI | Primary Predicate OnQ Neuro (K210831) | Secondary Predicate QP-Brain (K232231) | Comments on Substantial Equivalence |
| --- | --- | --- | --- | --- |
| | volumetric reports | color overlays and volumetric reports | reports | |
| Reporting | Results displayed in text and graphical formats | Results displayed in tabular and graphical formats | Results displayed in tabular and graphical formats | |
| DICOM Communication | Yes | Yes | Yes | Same |
| Safety | Automated quality control function: scan protocol verification. Results must be reviewed by a clinician with proper training. | Display/measurement data can be viewed, accepted, or rejected by a physician. | Automated quality control function: scan protocol verification. Results must be reviewed by a clinician with proper training. | Same |
## 5.6. Summary of Technological Characteristics Comparison
The subject and predicate devices have very similar technological characteristics. Both devices are software only medical devices. The inputs for both devices are Brain MRI images in DICOM format. The outputs from both devices are automatically generated in DICOM format.
All the devices use deep learning to segment hyperintensities in FLAIR MR images, and subsequently calculate the volumes of these regions. In all devices, these segmentations are overlaid as colored regions on DICOM images for users to visualize the regions used in the volume calculations.
All the devices also summarize these measurements and present them in a pdf document. In summary, both devices have extremely similar technological characteristics.
## 6. DISCUSSION OF PERFORMANCE TESTING
Performance testing included protocols demonstrating the similarity of automated segmentation compared to manual radiologist segmentations. The GBrain MRI segmentation performance was evaluated by comparing the software-derived segmentations to a Simultaneous Truth and Performance Level Estimation (STAPLE) algorithm generated consensus of expert-labeled segmentations of hyperintensities for volume measurement accuracy, and segmentation overlap agreement. Comparisons to expert segmentations were quantified using OLS Regression, and Dice similarity coefficient (extent of software-derived vs. ground truth overlap).
Acceptance criteria were set such that the GBrain MRI model performance meets clinically acceptable levels.
510(k) Summary
{12}
The results of the segmentation performance testing demonstrated that the GBrain MRI system segments hyperintensities with an accuracy that passed the planned acceptance criteria. Segmentation performance was evaluated across scanner manufacturers and field strengths. It is concluded that GBrain MRI segmentation is sufficiently accurate to be used in clinical practice in accordance with its indications for use.
## 6.1. Conclusion
The comparison between the two devices demonstrates that GBrain MRI is as safe and as effective as its predicates, the Quibim QP-Brain and the OnQ Neuro devices. The devices have substantially equivalent intended uses, intended users, and principles of operation. They have the same basic design, functions, and technological characteristics. V&V testing demonstrated that the subject device performed with similar accuracy and reliability as the predicate device to meet its intended use specifications.
Therefore, GBrain MRI is substantially equivalent to its predicate devices without raising any new concerns relating to safety or effectiveness.
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.