The DISPOSABLE BALLOON CATHETER B5-2Q, B7-2Q/2LA have been designed to be used with Olympus endoscopes to inject contrast medium into the biliary or pancreatic track. They can also be used for retrieval of biliary or pancreatic stones.
Device Story
Disposable balloon catheter used during ERCP procedures; inserted through endoscope instrument channel to target biliary or pancreatic ducts. Device consists of inflatable balloon, tube, branch, injection port, air feeding port, and stopcock. Operator (physician) attaches sterile syringe to air feeding port to inflate balloon with air; balloon inflation allows for stone retrieval or contrast medium injection via injection port. Device is single-use and sterilized via ETO. Benefits include facilitated stone removal and diagnostic imaging access. No electrical or software components.
Clinical Evidence
No clinical study was performed. Substantial equivalence supported by bench testing, including insertion/withdrawal, balloon diameter, infusion performance, junction strength, balloon burst, dimensional analysis, fatigue, kink stability, and radiopacity. Biocompatibility testing conducted per ISO 10993-1.
Technological Characteristics
Single-use biliary catheter; inflatable balloon, tube, branch, injection port, air feeding port, stopcock. Materials include natural rubber latex for balloon. Sterilization via ETO. No electrical components. Dimensions: B5-2Q (1.95mm diameter, 11.0mm inflated), B7-2Q/2LA (2.55mm diameter, 13.0mm inflated).
Indications for Use
Indicated for patients requiring biliary or pancreatic stone retrieval or contrast medium injection during ERCP procedures using Olympus endoscopes.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 21, 2025
Olympus Medical Systems Corporation
Eve Smith
Regulatory Affairs Specialist II
2951 Ishikawa-cho
Hachioji-shi, Tokyo 192-8507
Japan
Re: K250409
Trade/Device Name: Disposable Balloon Catheter (B5-2Q); Disposable Balloon Catheter (B7-2Q); Disposable Balloon Catheter (B7-2LA)
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary Catheter and Accessories
Regulatory Class: Class II
Product Code: FGE
Dated: September 10, 2025
Received: September 10, 2025
Dear Eve Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250409 - Eve Smith
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K250409 - Eve Smith
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANTHONY LEE -S
Anthony C. Lee Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250409
Device Name
Disposable Balloon Catheter (B5-2Q);
Disposable Balloon Catheter (B7-2Q);
Disposable Balloon Catheter (B7-2LA)
Indications for Use (Describe)
The DISPOSABLE BALLOON CATHETER B5-2Q, B7-2Q/2LA have been designed to be used with Olympus endoscopes to inject contrast medium into the biliary or pancreatic track. They can also be used for retrieval of biliary or pancreatic stones.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
OLYMPUS
K250409
Page 1 of 5
# 510(k) Summary
For
Disposable Balloon Catheter B5-2Q, B7-2Q/2LA
## General Information
Applicant: OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
Phone: (+81) 42-642-2111
Fax: (+81) 42-642-2307
Establishment Registration Number: 8010047
Manufacturer: Aomori Olympus Co., Ltd.
2-248-1 Okkonoki, Kuroishi-shi, Aomori 036-0357, Japan
510(k) Submitter: Olympus Corporation of the Americas
3500 Corporate Parkway
Center Valley, PA 18034
Contact Person: Eve Smith
Regulatory Specialist II
Mobile: (267) 373-7633
Email: eve.smith@olympus.com
Date Prepared: October 10, 2025
## Device Description
Device Name: Disposable Balloon Catheter B5-2Q, B7-2Q/2LA
Generic/Common Name: Stents, Drains and Dilators for the Biliary Ducts
Regulation Number: 21 CFR 876.5010
Regulatory Class: Class II
Classification Name: Biliary Catheter and Accessories
Product Code: FGE
Review Panel: Gastroenterology & Urology
## Predicate Device
| Device Name | 510(k) Submitter | 510(k) No. |
| --- | --- | --- |
| Olympus B5/B7 Series Balloon Catheters | Olympus America, Inc. | K962925 |
510(k) Summary
Page 1 of 5
{5}
OLYMPUS
K250409
Page 2 of 5
# Indications for Use
The DISPOSABLE BALLOON CATHETER B5-2Q, B7-2Q/2LA have been designed to be used with Olympus endoscopes to inject contrast medium into the biliary or pancreatic track. They can also be used for retrieval of biliary or pancreatic stones.
# Principle of Operation
The Balloon Catheters are constructed of an inflatable balloon, tube, branch, injection port, air feeding port, and stopcock.
Prior to use, a sterile syringe is attached to the air feeding port and injects air into the tube to inflate the balloon to the specified volume to confirm that the balloon inflates properly. The balloon is then deflated completely, and the tube is inserted into the instrument channel of the endoscope and advanced to the target area, where the balloon is inflated to allow for biliary or pancreatic stone removal or contrast medium injection through the injection port. When the procedure is completed, the balloon is deflated and the tube is removed from the patient.
# Comparison of Technological Characteristics
Table 1 compares the DISPOSABLE BALLOON CATHETER B5-2Q, B7-2Q/2LA to the predicate device with respect to intended use, and technological characteristics, providing detailed information regarding the basis for the determination of substantial equivalence.
Table 1. Subject and Predicate Device Comparison Table
| Feature / Characteristic | Subject Device (SD) | Predicate Device (PD) |
| --- | --- | --- |
| | Disposable Balloon Catheter B5-2Q and B7-2Q/2LA | Olympus B5/B7 Series Balloon Catheters (K962925) |
| Indications for Use | The DISPOSABLE BALLOON CATHETER B5-2Q, B7-2Q/2LA have been designed to be used with Olympus endoscopes to inject contrast medium into the biliary or pancreatic track. They can also be used for retrieval of biliary or pancreatic stones. | The B5-2Q and B7-2Q/B7-2LA Balloon Catheters are to be used for biliary or pancreatic stone removal and dye injection associated with ERCP. |
| Regulation Number | 876.5010 | 876.5010 |
| Regulation Name | Biliary catheter and accessories. | Biliary catheter and accessories. |
| Regulatory Class | Class II | Class II |
| Product Code | FGE
Stents, drains and dilators for the biliary ducts | FGE
Stents, drains and dilators for the biliary ducts |
| Classification Panel | Gastroenterology/Urology | Gastroenterology/Urology |
| Basic principle | The Balloon Catheter is inserted into the instrument channel of the endoscope and advanced to the target area, where the balloon | The Balloon Catheter is inserted into the instrument channel of the endoscope and advanced to the target area, where the balloon |
510(k) Summary
Page 2 of 5
{6}
OLYMPUS
K250409
Page 3 of 5
| Feature / Characteristic | Subject Device (SD) | Predicate Device (PD) |
| --- | --- | --- |
| | Disposable Balloon Catheter B5-2Q and B7-2Q/2LA | Olympus B5/B7 Series Balloon Catheters (K962925) |
| | area, where the balloon is inflated to allow for biliary or pancreatic stone removal or dye injection through the injection port. | is inflated to allow for biliary or pancreatic stone removal or dye injection through the injection port. |
| Shape of the Balloon | | |
| Maximum Insertion Portion Diameter (mm) | B5-2Q: ø 1.95
B7-2Q: ø 2.55
B7-2LA: ø 2.55 | B5-2Q: ø 1.95
B7-2Q: ø 2.55
B7-2LA: ø 2.55 |
| Working Length (mm) | B5-2Q: 1950
B7-2Q: 1950
B7-2LA: 3500 | B5-2Q: 1950
B7-2Q: 1950
B7-2LA: 3500 |
| Diameter after inflation (mm) | B5-2Q: ø 11.0
B7-2Q: ø 13.0
B7-2LA: ø 13.0 | B5-2Q: ø 11.0
B7-2Q: ø 13.0
B7-2LA: ø 13.0 |
| Maximum Air Volume (ml [cc]) | B5-2Q: 2.2
B7-2Q: 2.4
B7-2LA: 2.6 | B5-2Q: 1.6
B7-2Q: 1.8
B7-2LA: 2.0 |
| Compatible Guidewire | B5-2Q: ø 0.53mm (0.021 in)
B7-2Q: ø 0.89mm (0.035 in)
B7-2LA: ø 0.89mm (0.035 in) | B5-2Q: ø 0.53mm (0.021 in)
B7-2Q: ø 0.89mm (0.035 in)
B7-2LA: ø 0.89mm (0.035 in) |
| Compatible Endoscope | Working length less than 1400 mm; TJF Channel Inner Diameter ø4.2 | Working length less than 1400 mm; TJF Channel Inner Diameter ø4.2 |
| Reprocessing | Single Use | Single Use |
| Sterilization Method | ETO | ETO |
## Summary of Performance Testing
The following performance testing was conducted in support of substantial equivalence determination.
- Reprocessing Validation Testing
The subject device is single use, and not reprocessed.
510(k) Summary
Page 3 of 5
{7}
OLYMPUS
K250409
Page 4 of 5
- Biocompatibility Testing
Biocompatibility testing was conducted in accordance with the FDA’s Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process”. The biocompatibility testing included the following tests:
- Cytotoxicity Study using the ISO Elution method
- Intracutaneous Irritation Study
- Guinea Pig Maximization Sensitization Test
- Material-mediated Pyrogen Testing
- Acute Systemic Toxicity Study in Mice
- Electrical Safety and Electromagnetic Compatibility (EMC)
The Subject device does not contain any electrical parts.
- Software and Cybersecurity
The subject device does not include any software.
- Performance Testing – Bench (Non-Clinical)
Bench tests as listed below were conducted to ensure that the subject device performs as intended and meets design specifications.
- Insertion of the Subject Device into the Endoscope
- Withdrawal of the Subject Device from the Endoscope
- Balloon Diameter after Inflation
- Performance of Infusion
- Strength of Junction
- Safety Evaluation
- Balloon Burst
- Dimensional Analysis
- Balloon Fatigue
- Kink Stability
- Radiopacity
- Performance testing – Animal
No animal study was performed.
- Performance testing – Clinical
No clinical study was performed.
510(k) Summary
Page 4 of 5
{8}
OLYMPUS
K250409
Page 5 of 5
- Risk Management
Risk management was performed in accordance with ISO 14971:2019. The design verification tests and their acceptance criteria were performed and identified as a result of this risk management.
## Substantial Equivalence
The DISPOSABLE BALLOON CATHETER B5-2Q, B7-2Q/2LA are substantially equivalent to the legally marketed predicate device based on the claim of substantial equivalence on similarities in intended use and technological features with the predicate device:
- similar intended use
- device characteristics (design, materials, and operations) are similar or identical to the predicate device, and
- does not introduce any new or novel treatments or standard of care that differs from predicate device in commercial use.
Due to a change in the manufacturer of Natural Rubber Latex (balloon material), the subject device will hold an increased amount (volume) of air in the balloon. Although this is different between Subject Device and Predicate Device, Olympus conducted Balloon Burst, Safety Verification, and Balloon Diameter measurements/testing to confirm device safety and effectiveness.
## Conclusion
In summary, the DISPOSABLE BALLOON CATHETER B5-2Q, B7-2Q/2LA is substantially equivalent to the predicate device and raises no new questions of safety or effectiveness.
510(k) Summary
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.