K250401 · Aion Biosystems, Inc. · FLL · Mar 14, 2025 · General Hospital
Device Facts
Record ID
K250401
Device Name
AION TempShield™
Applicant
Aion Biosystems, Inc.
Product Code
FLL · General Hospital
Decision Date
Mar 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The AION TempShield™ is a battery powered wearable thermometer intended for continuous measurement of human body temperature on the upper chest via wireless communication. AION TempShield™ is intended for single-use and for persons aged 1 year and older in either facilities or home envirents.
Device Story
AION TempShield™ is a wearable, single-use, battery-powered electronic thermometer; worn on the upper chest; measures skin temperature via thermistor; extrapolates to body temperature. Device pairs with smartphone/tablet app via Bluetooth LE; NFC used for power-on and pairing. App collects temperature data every 5 minutes; performs full data offload every 4 hours. Data transmitted to AION Core cloud service for visualization and fever alerts based on NIH recommendations. Used by patients or clinicians in home or clinical settings. Provides continuous monitoring; alerts users to battery status; allows historical data viewing. Benefits include long-term (up to 90 days) continuous monitoring, facilitating fever tracking and clinical decision-making.
Clinical Evidence
Clinical data included a usability study with 137 participants (ages 5+) and an addendum study with 39 participants (ages 1-4). All participants verified accurate body temperature measurement. Bench testing confirmed accuracy of +/- 0.1°C per ISO 80601-2-56.
Technological Characteristics
Wearable electronic thermometer; 42mm diameter, 3.2mm height, <8g. Materials: medical-grade silicone tape. Sensing: thermistor. Energy: 3V CR2016 Li-Mag coin cell. Connectivity: Bluetooth LE 4.2, NFC. Ingress protection: IP57. Software: firmware for data transmission and battery management. Standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2, ISO 10993 series.
Indications for Use
Indicated for continuous human body temperature measurement on the upper chest in patients aged 1 year and older in healthcare facilities or home environments.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
March 14, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" written below in a smaller font.
AION Biosystems Inc. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K250401
Trade/Device Name: AION TempShield™ Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: February 12, 2025 Received: February 12, 2025
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K250401
Device Name AION TempShield™
#### Indications for Use (Describe)
The AION TempShield™ is a battery powered wearable thermometer intended for continuous measurement of human body temperature on the upper chest via wireless communication. AION TempShield™ is intended for single-use and for persons aged 1 year and older in either facilities or home envirents.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K250401 510(k) Summary Traditional 510(k) AION TEMPSHIELD™
#### 1. SUBMITTER/510(K) HOLDER
AION Biosystems Inc. 175 Cabot street, suite 100, Lowell, MA 01854
Contact Name: Carl Shubitowski (Director of Quality & Regulatory)
Email: Carl.Shubitowski@aionbiosystems.com Telephone: (617) 279-3881
Date Prepared: February 1, 2025
Reason for 510(k): New device
#### 2. DEVICE NAME
| Proprietary / Trade Name: | AION TEMPSHIELD™ |
|----------------------------|---------------------------------|
| Common/Usual Name: | Clinical Electronic Thermometer |
| Classification Name: | Clinical Electronic Thermometer |
| Classification Regulation: | 21 CFR 880.2910 |
| Product code: | FLL |
| Classification: | Class 2 |
| Medical Specialty (Panel): | General Hospital |
## 3. PREDICATE DEVICES
| Predicate Type | Device Name | Manufacturer | 510(k) | Product Regulation | Code and |
|------------------------|-------------|-----------------|---------|-----------------------|----------|
| Primary | iTempShield | AION Biosystems | K232010 | FLL, 21 CFDR 880.2910 | |
| Reference<br>Secondary | Fever Scout | VivaLNK | K181013 | FLL, 21 CFDR 880.2910 | |
#### 4. DEVICE DESCRIPTION
The AION TempShield™ is a wearable device that senses body temperature and communicates with an app on a smartphone or tablet or certified gateway device . When the AION TempShield™ is connected to the device, it sends all temperature measurements automatically.
The AION TempShield™ can be worn for up to 90 days to check current temperature and to allow viewing of historical temperature measurements. The device has the ability to share the data with the clinician if the clinician has established an AION account.
The temperature measurements are in degrees Fahrenheit (°F), but can also be configured to display in degrees Celsius (°C).
{5}------------------------------------------------
Circuit board with Bluetooth and NFC (near field communication chip). A phone with NFC is used to turn on the device. The device communicates with a phone that has the AION TempShield app installed. The circuit card has two digital temperature sensors and a 32 MHz clock. The device is designed to have a minimum operating range of 10 meters and provides a 2 dBm signal to the radio. The AION TempShield™ is shipped in a shelf mode power condition (4uA) and can be on the shelf for 24 months before the battery is depleted below the 90-day threshold.
The Bluetooth transmissions are encrypted using the on-board encryption. Each device has a random pairing code that must be typed in before the sensor and the mobile device can communicate. The modulation for the low energy Bluetooth is QFSK and the device has been certified and registered with the FCC for a class II device (at home use). The hospital use is covered under this classification.
The smart phone application requires that the user allow for notifications, has access to location services, and enables both NFC and Bluetooth services to fully activate the AION TempShield™ Each AION TempShield™ has a unique serial number (UUID) and a random pin assigned. This information isaccessible via a barcode on the outside of the package and from an NFC read (Similar to apple pay or google pay type of contact with the back of the phone. The NFC read will wake up the AION TempShield™ from deep sleep and initiate the pairing. Once the initial Bluetooth contact has beenmade the mobile application will present the 6-digit pairing number that the user must type in.
Once this is done the mobile device will go into operational mode and send out temperature every five minutes. The Mobile app will look for temperature every five minutes with appropriate guard bands. Every 4hours the mobile application will request a full data offload from the AION TempShield™ to collect previous data that was not collected from the AION TempShield™.
The AION TempShield Application communicates with the AION Core (Cloud monitoring service). This service will provide NIH recommendations for fever to the shield to provide indications (alerts)in the data for visualization for the User. There is no PHI stored in the AION cloud.
The application has a home screen that displays the last read temperature. The application allows the user to put in activities, to view historical temperature history, and to select time windows to display data. For example: the application will allow the user to only view data for the past 12 hours. The total viewable data window is 48 hours.
The application in the settings window will allow the user to view the UUID and will allow the user to replace the AION TempShield™ with another one. The Application will provide the user with thepercentage of charge left in the device (battery life indication). When the battery life reaches 10% of charge left, the application will provide an alert that the user should get a new AION TempShield 100 (there will be approximately 1 week of continuous life available. At 5% life the notifications are once per two hours to replace the shield.
The AION TempShield TM requires that the user place medical grade silicone tape to hold the AION TempShield™ in place. The AION TempShield™ is provided with a patient user guide. The end user is expected to provide a mobile device (Android 10 or greater or iOS 11 or greater). The iPhone 6 and earlier will not work. The mobile device must have an NFC reader to fully operate.
{6}------------------------------------------------
## Specifications and Configurations
The AION TempShield™ is < 8 grams is approximately 42mm in diameter and 3.2 mm in height. The device is single-person use and is held in place via off the shelf silicone tape. The device is IP x57 dust and waterproof. And the operating range is from 77.0F (25C) to 109.4F (43C) degrees.
#### 5. INDICATIONS FOR USE
The AION TempShield™ is a battery powered wearable thermometer intended for continuous measurement of human body temperature on the upper chest via wireless communication to a smart device application. AION TempShield™ is intended for single-use and for persons aged 1 year and older in either healthcare facilities or home environments.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The subject device was designed to provide the benefit of continuous temperature measurement that allows wireless communication to smart devices for monitoring.
The subject device and predicate are both single use devices used for continuous temperature monitoring at healthcare facilities (hospital or hospital type facilities) and in home environments. Both devices are placed on the patient's upper chest and based on the principle that temperature measured at the skin surface can be extrapolated to the body temperature. The subject device and primary predicate are both disposable with identical accuracy of +/- 0.1°C, conforming to ISO 80601-2-56, using a thermistor sensor, supporting mobile device compatibility, utilizing Bluetooth, and powered by a CR2016 Li-Mag coin cell battery, while the secondary predicate is non-disposable and has a slightly broader accuracy range for certain temperatures.
The subject device uses Near Field Communications (NFC) RFID for power on and pairing key, is smaller and weighs less, and has a longer battery life. These differences do not affect the safety or effectiveness of the device.
{7}------------------------------------------------
# 7. SUBSTANTIAL EQUIVALENCE COMPARISON TABLE
| Technological<br>Characteristic | Subject Device<br>AION<br>TempShield™ | Primary Predicate<br>Device<br>iTempshield<br>K232010 | Secondary<br>Predicate Device<br>Fever Scout<br>K181013 | Comparison |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The AION<br>TempShield™<br>battery powered<br>wearable thermometer<br>intended for<br>continuous<br>measurement of<br>human body<br>temperature on the<br>upper chest via<br>wireless<br>communication to a<br>smart device<br>application. AION<br>TempShield™ is<br>intended for single-<br>use and for persons<br>aged 1 year and older<br>in healthcare facilities<br>and home<br>environments. | The iTempShield<br>battery powered<br>wearable<br>thermometer intended<br>for continuous<br>measurement of<br>human body<br>temperature on the<br>upper chest via<br>wireless<br>communication to a<br>smart device<br>application.<br>iTempShield is<br>intended for single-<br>use and for persons<br>older than 5 years in<br>healthcare facilities<br>and home<br>environments. | The wireless Fever<br>Scout continuous<br>monitoring<br>thermometer is a<br>non-invasive and<br>re-usable<br>electronic device<br>for home use. This<br>product is intended<br>for non-urgent<br>ambulatory<br>continuous armpit<br>body temperature<br>monitoring from<br>ages 29 days and<br>older. | The Subject Device<br>and Primary Predicate<br>have almost identical<br>indications for patient<br>population range<br>expanded to cover 1<br>year and older<br>(compared to 5 years<br>and older).<br>The Secondary<br>Predicate Device has<br>population range that<br>covers 29 days and<br>older.<br>The indications for<br>use for all devices<br>include wireless<br>continuous<br>measurement, home<br>use for measurement<br>of body temperature. |
| Placement Location | Upper Chest | Upper Chest | Armpit | The Subject Device<br>and Primary Predicate<br>are used on the Chest<br>and the Secondary<br>Predicate is used on<br>Armpit. |
| Classification<br>Regulation / Product<br>Code | 21 CFR 880.2910 /<br>FLL | 21 CFR 880.2910 /<br>FLL | 21 CFR 880.2910 /<br>FLL | IDENTICAL |
| Regulation Description | Clinical electronic<br>thermometer | Clinical electronic<br>thermometer | Clinical electronic<br>thermometer | IDENTICAL |
| Principle of Operation | Based on the principle<br>that the temperature<br>measured at the skin<br>surface can be<br>extrapolated to the<br>body temperature.<br>The sensor within the<br>subject device detects<br>the heat and measures<br>the temperature using<br>a resistance that<br>changes with heat. | Based on the<br>principle that the<br>temperature measured<br>at the skin surface can<br>be extrapolated to the<br>body temperature.<br>The sensor within the<br>predicate device<br>detects the heat and<br>measures the<br>temperature using a<br>resistance that<br>changes with heat. | Utilizes e-skin<br>technology with<br>sensors within the<br>device to detect<br>heat and measure<br>temperature using<br>a resistance that<br>changes with heat. | IDENTICAL |
| Laboratory Accuracy | ± 0.1C in the range of<br>25C to 43C | ± 0.1C in the range of<br>25C to 43C | +/- 0.1C from 37C<br>to 39C | IDENTICAL between<br>Subject Device and<br>Primary Predicate. |
| | | | | |
| Validation Method | Conforms to ISO 80601-2-56 | Conforms to ISO 80601-2-56 | Conforms to ASTM E1112-2000 (standard specification for electronic thermometer for intermittent determination of patient temperature) | +/- 0.2C from 35C to 37C and 39C to 42C<br>The Secondary Predicate has a slightly wide range for some temperatures.<br>IDENTICAL<br>Both standards are recognized by FDA as valid standards for electronic thermometers. |
| Type of Use (sensor) | Disposable | Disposable | Reusable | Subject device is identical to primary predicate while the secondary predicate is reusable. |
| Type of Sensor | Thermistor | Thermistor | Thermistor | IDENTICAL |
| Temperature Measurement Intervals | Transmits Maximum once per 5 minutes and Minimum once per 240 minutes. | Transmits Maximum once per 5 minutes and Minimum once per 240 minutes. | Transmit every 15 seconds | Temperature measurement intervals are shorter for the secondary predicate device. |
| Wireless Supported Devices | Mobile (Android, Apple) | Mobile (Android, Apple) | Mobile (Android, Apple) | EQUIVALENT – both devices are supported by Android and Apple. The devices are supported by the technologies for the company (i.e. iTempShield with AION shield App, and TempShield with AION TempShield App). |
| Wireless Type | Bluetooth LE 4.2<br>NFC RFID for power ON and pairing | Bluetooth LE 4.2<br>NFC RFID for power ON and pairing | Bluetooth BLE | IDENTICAL |
| Type of Application | Wearable | Wearable | Wearable | IDENTICAL |
| Overall Dimension | Diameter 42mm<br>Thickness 3.2mm | Diameter 42mm<br>Thickness 3.2mm | Size 61mm x 41mm*<br>Thickness 5.5mm* | The subject device is identical to the primary predicate device.<br>However, it is smaller than the secondary predicate. |
| Weight | 5 grams | 5 grams | 7.3g* | The subject device is identical to the primary predicate device.<br>However, it is smaller and weighs less than the secondary predicate. |
| | 10993-1, 5, 10 | 10993-1, 5, 10 | 10993-1, 5, 10 | |
| Power Source | Internal Battery<br>(Lithium Coin Cell)<br>3V | Internal Battery<br>(Lithium Coin Cell)<br>3V | Internal Battery<br>(Lithium Rechargeable)<br>3V | IDENTICAL |
| Battery Life | 2160 hours (90 days)<br>of continuous run time | 1440 hours (60 days)<br>of continuous run time | 168 hours (7 days)<br>of continuous run time* | The battery life is<br>substantially longer<br>for the subject device. |
| Electrical Safety | Conforms to IEC<br>60601-1 | Conforms to IEC<br>60601-1 | Conforms to IEC<br>60601-1 | IDENTICAL |
| Electromagnetic<br>Compatibility | Conforms to IEC<br>60601-1-2 | Conforms to IEC<br>60601-1-2 | Conforms to IEC<br>60601-1-2 | IDENTICAL |
| Storage / Transport<br>Temperature | 32F to 86 (0C to 30C) | 32F to 86 (0C to 30C) | 50F - 104F* | The Storage<br>temperature for the<br>Subject device and<br>primary predicate are<br>identical.<br><br>The range for the<br>secondary predicate<br>includes higher<br>temperatures. |
| Storage / Transport<br>Humidity | 45% RH to 75% RH<br>(non-condensing) | 45% RH to 75% RH<br>(non-condensing) | 15-85% RH* | The storage RH range<br>is identical between<br>the subject device and<br>primary predicate.<br><br>The storage range for<br>the secondary<br>predicate is wider.<br><br>The Storage humidity<br>range for the subject<br>device was developed<br>to match the<br>instructions provided<br>by the battery<br>manufacture. This is<br>only applicable to the<br>storage conditions and<br>will not affect the<br>user. |
| Operating<br>Temperature | 10C to 40C | 10C to 40C | 50F (10C) to 104<br>(40C) | Equivalent |
| Operating Humidity | 10% RH to 95% RH<br>(non-condensing) | 10% RH to 95% RH<br>(non-condensing) | 15% to 85% RH<br>(non-condensing) | Equivalent |
| Atmospheric Pressure | 700 hPa to 1060 hPa | 700 hPa to 1060 hPa | 700 hPa to 1060<br>hPa* | IDENTICAL |
| Mode of Operation | Continuous | Continuous | Continuous | IDENTICAL |
{8}------------------------------------------------
{9}------------------------------------------------
* This information is from the Fever Scout Instructions for Use Manual and from the 510(k) Summary
K162137
{10}------------------------------------------------
## Design
The iTempShield was cleared by FDA in August 2023 under K232010. The subject device is the same device except the name was changed to AION TempShieldTM . The primary differences between the subject device and the primary device are:
- Expanded patient population age from 5 years and older to 1 year and older. -
- Expanded battery life from 60 days to 90 days. -
- Changes to firmware for battery life extension -
The iTempShield used the Masimo Radius T (K203215) as the predicate device. The Radius T used Fever Scout (K181013) as their predicate device. The Radius T was applied on the chest and used for 5 years and older whereas their predicate device was placed in the armpit and used for 29 months and older.
The subject device and the primary predicate device identical in all aspects of design, technology, and principles of operations with the exception of the minor differences mentioned above. The expansion of battery life and change to firmware is supported with software validation and bench testing. The expansion of the patient population age range is supported by clinical data.
The secondary predicate is being utilized to provide a predicate that has an age group similar to the age group being pursued for the AION TempShieldTM. The subject device has a longer battery life, has different anatomical placement, has different measurement intervals and is smaller in size compared to the secondary predicate.
Based on the substantial equivalence comparison and provided testing, the above differences do not impact the safety or effectiveness of the device.
## Operational and Technological Characteristics
Both the subject and predicate devices are based on similar principles of operation, battery powered and use Bluetooth wireless technology to communicate data. The devices have the similar accuracy and support similar wireless devices.
The subject device and primary predicate are identical in characteristics except as mentioned with the expanded battery life, firmware change to support the battery life expansion and expansion in patient population age range.
The subject device and the secondary predicate have similar characteristics except for measurement intervals, anatomical placement, and size.
The difference in transmission time (measurement interval) does not impact safety or effectiveness as this device is not measuring vital physiological parameters (such as heartrate or respiration). The longer battery life offers benefit to the clinician and user to allow for longer period of time for monitoring the patient. The clinical data supports that the expanded patient population range (which is similar to the secondary predicate device).
The Subject Device has the equivalent anatomical placement, indications for use (except for patient age range), principle of operation, compliance to test standards with the updated 80601 standard-1, single use, wearable, use of thermistors, disposable, battery powered, wireless communication to smart devices using Bluetooth, and continuous mode of operation as the primary predicate device. The main differences are between patient population age range, battery life expansion and firmware change to support battery life expansion.
{11}------------------------------------------------
## Summary of Similarities and Differences
The Subject Device has similar characteristics as the secondary predicate which exception of:
- Subject device has faster measurement intervals. -
- Subiect device is smaller and weighs less than the secondary predicate. -
- -Subject device and primary predicate are placed on the chest, whereas the secondary predicate is placed in the armpit.
The minor differences in design and technology do not present an increased risk of safety and effectiveness of the device nor do they raise different questions of safety and effectiveness as supported by the compliance testing to the electrical safety IEC 60601-1 series standards, laboratory and clinical accuracy testing to the IEC 80601-2-56, biocompatibility testing to the ISO 10993 series standards as well as the performance bench testing and human factors evaluation.
# 8. PERFORMANCE TESTING
The following standards were used for testing of the subject device:
- IEC 60601-1
- IEC 60601-1-2
- IEC 60529
- IEC 60601-1-6
- IEC 62366
- IEC 60601-1-11
- ISO 80601-2-56
Biocompatibility evaluation for the following contact classification was conducted:
- For Contact Type: Surface Skin ●
- Contact Duration: Up to 90 days
- Classification: C (Intact Skin, Greater than 30 days) ●
The classification and endpoints: cytotoxicity, skin irritation, and sensitization, associated with this classification have been assessed in accordance with:
- ISO 10993-1
- ISO 10993-5
- ISO 10993-10 ●
- ISO 10993-23 .
The following FDA guidance documents were utilized:
- FDA Guidance for the Content of Premarket Submissions for Software Contained in ● Medical Devices – Issued May 2005
- . FDA Guidance on the content of premarket notification 510(k) submissions for clinical electronic thermometers - March 1993
{12}------------------------------------------------
| Standard | Description of Test | Results |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electromagnetic Compatibility, Electrical Safety, and Home Healthcare Safety | | |
| IEC 60601-1:2012 | Electrical Safety<br>Medical Electrical Equipment – Part 1:<br>General Requirements for Basic Safety<br>and Essential Performance | All applicable<br>requirements were tested<br>and passed. |
| IEC 60601-1-2:2014 | Electromagnetic Disturbance<br>Medical Electrical Equipment – Part 1-2:<br>General Requirements for Basic Safety<br>and Essential Performance – Collateral<br>Standard: Electromagnetic disturbances<br>Requirements and Tests | All applicable<br>requirements were tested<br>and passed. |
| IEC 60601-1-11:2015 | Safety for Home Healthcare<br>Environment<br>Medical Electrical Equipment – Part 1-<br>11: General Requirements for basic<br>safety and essential performance –<br>Collateral Standard: Requirements for<br>Medical Electrical Equipment and<br>Medical Electrical Systems Used in the<br>Home Healthcare Environment | All applicable<br>requirements were tested<br>and passed. |
| Ingress Testing | | |
| IEC 60529:2013/COR1:20<br>19 | Ingress<br>Degrees of protection provided by<br>enclosures IP57<br><br>The testing include dust penetration and<br>water ingress. | All applicable<br>requirements were tested<br>and passed. |
| Usability Testing / Human Factors | | |
| IEC 60601-1-6:2010 /<br>AMD1:2013 | Usability<br>Medical Electrical Equipment Part 1-6:<br>General Requirements for Basic Safety<br>and Essential Performance – Collateral<br>Standard: Usability | All applicable<br>requirements were tested<br>and passed. |
| IEC 62366-1:2015 | Usability<br>Medical Device – Application of<br>Usability Engineering to Medical<br>Devices | PASS – all applicable<br>requirements were tested<br>and passed. |
| No related standard | Usability | PASS – The original study<br>included 137 participants 5<br>years and old, the<br>addendum study included<br>39 participants ages 1-4<br>years. All tested passed<br>and verified accurate body<br>temperature. |
| PERFORMANCE BENCH TESTING | | |
| ISO 80601-2-56:2017 | Clinical Thermometer Performance<br>Medical Electrical Equipment – Part 2-<br>56: Particular Requirements for Basic<br>Safety and Essential Performance of<br>Clinical Thermometers for Body<br>Temperature Measurement | All applicable<br>requirements were tested<br>including laboratory<br>accuracy and passed |
{13}------------------------------------------------
| No related standard | Bluetooth Range Testing | PASS - the testing concluded the mobile app received temperature data from the AION TempShield™ at 0.1m and 10m away from the phone. | | | |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------|------------|
| No related standard | Disinfection Resistance Testing | PASS - the testing concluded that the AION TempShield™ continued to send temperature data after 90 cleanings using disinfectant wipes. There were no visible defects on the device. | | | |
| No related standard | Operational Temperature Range Testing | PASS – The AION TempShield™ continued to function and send temperature data at every temperature within the range. | | | |
| No related standard | Battery Notification Test | PASS - The AION TempShield™ transmits proper battery percentage depletion, displays raw temperaturesin DTM mode and passes encrypted data. | | | |
| No related standard | Product Life / Accelerated Aging Test | PASS – The devices were found to be fully functional at end of storage test (23 days at 70C) with no physical defects observed.<br>The devices were found to be functional and transmitted data after 15 days at 60C. | | | |
| No related standard | Software / Firmware Verification & Validation<br>Software Verification & Validation Testing<br>Mobile Application Verification and Validation | PASS - The V&V testing demonstrated that the AION TempShield Mobile App meets specified requirements and functions as intended. | | | |
| No related standard | Software Verification & Validation Testing<br>AION TempShield™ Firmware RequirementsV&V Report | PASS - The V&V testing demonstrated that the AION TempShield™ firmware meets specified requirements and functions as intended. | | | |
| ISO 10993-1:2016 | | Biocompatibility Testing<br>Biocompatibility<br>Biological Evaluation of Medical | Devices – Part 1: Evaluation and Testing<br>within a Risk Management Process…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.